K152508 · Dongguan Ageless Health Industrial Co., Ltd. · FLL · Dec 31, 2015 · General Hospital
Device Facts
Record ID
K152508
Device Name
Ageless Health Medical Digital Thermometer
Applicant
Dongguan Ageless Health Industrial Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Dec 31, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
Ageless Health Medical Digital Thermometers are intended for use in measuring human body temperature (Armpit or Oral).
Device Story
Digital thermometer for human body temperature measurement (oral/armpit). Input: resistance change from thermistor sensor; processed by ASIC; output: digital temperature reading on LCD. Operated by patient or clinician in home or clinical settings. Provides temperature data to assist in health monitoring.
Clinical Evidence
Bench testing only. Performance evaluated against IEC 60601-1, IEC 80601-2-56, IEC 60601-1-2, IEC 60601-1-11, ASTM E1112-00, ISO 10993-5, and ISO 10993-10. No clinical data provided.
Technological Characteristics
Thermistor-based sensing; ASIC signal processing; 1.5V button battery power; ABS plastic and stainless steel patient-contact materials; IP22 ingress protection. Standards: IEC 60601-1, IEC 80601-2-56, IEC 60601-1-2, IEC 60601-1-11, ASTM E1112-00, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for human patients requiring body temperature measurement via armpit or oral routes.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 31, 2015
Dongguan Ageless Health Industrial Co., Ltd c/o Ms. Cecilia Ceng Regulatory Manager Guangzhou Glomed Biological Technology Co., Ltd Suit 306, Kecheng Mansion. No.121 Science Road, Guangzhou Science Park Guangzhou. 510663 CHINA
Re: K152508
Trade/Device Name: Ageless Health Medical Digital Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: November 17, 2015 Received: November 23, 2015
Dear Ms. Ceng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152508
### Device Name
Device Name: Ageless Health Medical Digital Thermometer
Indications for Use (Describe)
Ageless Health Medical Digital Thermometers are intended for use in measuring human body temperature (Armpit or Oral).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## K152508
This summary of 510(K) information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
#### 1. Submitter Information
Sponsor Name: Dongguan Ageless Health Industrial Co.,Ltd. Address: 3/F, A1 Bldg, Dongshen Sima Industrial Area, No.33 Shenbei Road, Sima Village, Changping Town, Dongguan City, Guangdong Province, China Contact Person: Victor Wan (Vice-president) Phone: +86-769-81158038 Fax: +86-769-82289331 E-mail: victor@agelh.com
## Application Correspondent:
Contact Person: Ms. Cecilia Ceng / Mr. Tim Wong Guangzhou GLOMED Biological Technology Co., Ltd. Tel: +86-20-61099984 Email: regulatory@glomed-info.com
#### 2. Subject Device Information
| Type of 510(k): | Traditional |
|----------------------|--------------------------------------------|
| Common Name: | Thermometer, electronic |
| Trade Name: | Ageless Health Medical Digital Thermometer |
| Classification Name: | Thermometer, electronic, clinical |
| Review Panel: | General Hospital |
| Product Code: | FLL |
{4}------------------------------------------------
Requlation Number: 880.2910
Regulation Class: 2
#### 3. Predicate Device Information
Sponsor: SHENZHEN PUMP MEDICAL SYSTEM CO., LTD. Classification Name: Thermometer, electronic, clinical Trade Name: Electronic thermometer 510(k) number: K131210 Review Panel: General Hospital Product Code: FLL Regulation Number: 880.2910 Regulation Class: 2
#### 4. Device Description
Ageless Health Medical Digital Thermometers comprises a thermistor for getting temperature signal, a reference resistor for comparing the resistance of the thermistor, a buzzer for sounding effect, an ASIC for processing the target temperature digitally, and a LCD for displaying the temperature result. The design principle of thermometer is based on thermosensor and ASIC technology. A thermistor is used as thermosensor. The ASIC gets the sensor's signal from human body, then processes the signal and calculates the result, after that displays the temperature result by a LCD.
#### 5. Intended Use
Ageless Health Medical Digital Thermometers are intended for use in measuring human body temperature (Armpit or Oral).
#### 6. Test Summary
Ageless Health Medical Digital Thermometer has been evaluated the performance by lab bench testing according to the following standards:
{5}------------------------------------------------
- IEC 60601-1 (2005 + CORR.1:2006 + CORR.2:2007 + AM1:2012) Medical I Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- . IEC 80601-2-56 (2009) Medical Electrical Equipment - Part 2-56: Particular Requirements for Basic Safety and Essential Performance of Clinical Thermometers for Body Temperature Measurement
- " IEC 60601-1-2 (2007) Medical Electrical Equipment - Part 1-2: Collateral Standard: Electromagnetic Compatibility
- .. IEC 60601-1-11 (2010), Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used
- ' ASTM E1112-00 (2006) Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
- . ISO 10993-5 (2009) Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro cytotoxicity
- l ISO 10993-10 (2010) Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization
The same tests on Ageless Health Medical Digital Thermometer and the predicate device demonstrate the substantial equivalence of the two.
#### 7. Comparison to Predicate Device
Compare with predicate device, the subject device is very similar in design principle, intended
| Elements of<br>Comparison | Subject Device | Predicate Device | Conclusion |
|-------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------|
| Company | Ageless Health Industrial Ltd. | SHENZHEN PUMP MEDICAL<br>SYSTEM CO., LTD. | -- |
| Name and<br>Model | Ageless Health Medical Digital<br>Thermometer, models: BT-301, | Electronic thermometer, models:<br>TF3100, TF3101, TF3102 | -- |
| Elements of<br>Comparison | Subject Device | Predicate Device | Conclusion |
| | BT-302, BT-303, BT-305, BT-306,<br>BT-308, BT-311, BT-318 | | |
| 510(k) Number | Applying | K131210 | -- |
| Thermometer<br>Type | Digital Thermometer | Digital Clinical Thermometer | SE |
| Intended Use | Ageless Health Medical Digital<br>Thermometers are intended for use<br>in measuring human body<br>temperature (Armpit or Oral). | It is intended for use in measuring<br>temperature in human body<br>(Armpit or Oral). | SE |
| Indication for<br>Use | Ageless Health Medical Digital<br>Thermometers are intended for use<br>in measuring human body<br>temperature (Armpit or Oral). | It is intended for use in measuring<br>temperature in human body<br>(Armpit or Oral). | SE |
| Sensor | Thermistor | Thermistor | SE |
| Signal<br>Processing and<br>Display | Using the resistance change of<br>thermal resistor to detect body<br>temperature and displayed through<br>the LCD. | Using the resistance change of<br>thermal resistor to detect body<br>temperature and displayed<br>through the LCD. | SE |
| Power<br>Requirements | 1.5V button battery | 1.5V button battery | SE |
| Measurement<br>Temperature<br>Range | 32.0°C ~ 42.9°C | 32.0°C ~ 42.0°C | SE<br>Note 1 |
| Accuracy | 35.0 ~ 39.0°C : +/- 0.1°C<br>The rest: +/- 0.2°C | 0.05°C (35.30 ~ 39.00°C )<br>0.1°C (< 35.30°C or > 39.0°C ) | SE<br>Note 1 |
| Ambient<br>temperature | 10~35°C | 5~40°C | SE<br>Note 1 |
| Response Time | 60s | 5min | SE<br>Note 1 |
| Protection<br>Degree against<br>Ingress of Water<br>and Particulate<br>Matter | IP22 | IP22 | SE |
| Patient<br>Contacting | ABS plastic and stainless steel | ABS plastic and stainless steel | SE |
| Elements of<br>Comparison | Subject Device | Predicate Device | Conclusion |
| Material | | | |
| Complied<br>Standard | IEC 60601-1, IEC 80601-2-56, IEC<br>60601-1-2, IEC 60601-1-11, ASTM<br>E1112, ISO 10993-5, ISO 10993-10 | IEC 60601-1, IEC 60601-1-2, IEC<br>60601-1-11, ASTM E1112, ISO<br>10993-5, ISO 10993-10 | SE |
use, indications for use, functions, material and the applicable standards.
{6}------------------------------------------------
{7}------------------------------------------------
## Note 1
Although some specifications of Measurement Temperature Range, Accuracy, and Response Time are different for subject device and predicate device, they are both complied with related performance standards.
## Conclusion
The subject device Ageless Health Medical Digital Thermometer has all features of the predicate device,only with few differences(accuracy, Ambient temperature, Response Time). The technological principle is also the same with the predicate device's. Thus, the subject device is substantially equivalent to the predicate device.
## 8. Summary Prepared Date
30 December 2015
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.