Zimmer NexGen Complete Knee Solution Tibial and Femoral Augments
K152494 · Zimmer, Inc. · JWH · Nov 25, 2015 · Orthopedic
Device Facts
Record ID
K152494
Device Name
Zimmer NexGen Complete Knee Solution Tibial and Femoral Augments
Applicant
Zimmer, Inc.
Product Code
JWH · Orthopedic
Decision Date
Nov 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For NexGen Augments used with NexGen Knee Systems - Cruciate Retaining (CR), Cruciate Retaining Augmentable (CRA), Posterior Stabilized (PS), Legacy Posterior Stabilized (LPS), and Legacy Constrained Condylar Knee (LCCK): This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Collagen disorders, and/or avascular necrosis of the femoral condyle. - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. - Moderate valgus, varus, or flexion deformities. - The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. · CR and LPS porous coated femoral and tibial baseplate components may be used cemented biological fixation). The CR Hydroxyapatite/tricalcium phosphate [HA/TCP] coated femoral or tibial baseplate components may only be used uncemented. All other femoral, tibial baseplate and all-polyethylene patella components are indicated for cemented use only. · NexGen Augments that may be used with cemented CR, CRA, PS, and LCCK femoral and tibial baseplate components are indicated for cemented use only. For NexGen Augments used with NexGen Rotating Hinge Knee System: This device is indicated for patients with: - Moderate to severe knee instability. - Significant bone loss and/or ligament deficiencies caused by neoplasms, traumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, and/or avascular necrosis of the femoral condyle. - Valgus, varus, or flexion deformities. - The salvage of previously failed surgical attempts. This device is intended for cemented use only.
Device Story
Zimmer NexGen Tibial and Femoral Augments are orthopedic prosthetic accessories used to address bone loss in total knee arthroplasty. Augments are provided in various shapes and sizes to supplement existing NexGen Knee Systems (CR, CRA, PS, LPS, LCCK, and Rotating Hinge). Components are attached to tibial baseplates or femoral components using bone cement or screws. Used by orthopedic surgeons in clinical/OR settings to restore joint stability and configuration in patients with significant bone defects. Output is a physically augmented prosthetic construct that provides structural support, enabling successful joint reconstruction and improving patient mobility/pain outcomes.
Clinical Evidence
No clinical data required. Bench testing only. Accelerated aging confirmed 10-year shelf life. Mechanical testing demonstrated adequate attachment strength to resist joint reaction forces and showed increased shear load for PMMA-coated titanium substrates compared to uncoated titanium.
Technological Characteristics
Prosthetic augments made of metal/polymer. Attachment via bone cement or screws. PMMA coating used on titanium substrates to enhance cement bonding. Designed for use with NexGen knee systems. No software or electronic components.
Indications for Use
Indicated for patients with severe knee pain, disability, or instability due to rheumatoid/osteo/traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, post-traumatic joint loss, or neoplasms. Also indicated for moderate valgus/varus/flexion deformities, significant bone loss, ligament deficiency, or salvage of failed surgical attempts.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
NexGen Complete Knee Solution Rotating Hinge Knee (K013385)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing to the right, arranged in a way that they appear to be interconnected. The profiles are rendered in a simple, line-art style. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2015
Zimmer, Incorporated Mr. Mark D. Warner Sr. Specialist. Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708
Re: K152494
Trade/Device Name: Zimmer® NexGen® Complete Knee Solution Tibial and Femoral Augments Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, KRO Dated: August 31, 2015 Received: September 2, 2015
Dear Mr. Warner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152494
#### Device Name
Zimmer® NexGen® Complete Knee Solution Tibial and Femoral Augments
Indications for Use (Describe)
For NexGen Augments used with NexGen Knee Systems - Cruciate Retaining (CR), Cruciate Retaining Augmentable (CRA), Posterior Stabilized (PS), Legacy Posterior Stabilized (LPS), and Legacy Constrained Condylar Knee (LCCK): This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Collagen disorders, and/or avascular necrosis of the femoral condyle.
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities.
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
· CR and LPS porous coated femoral and tibial baseplate components may be used cemented biological fixation). The CR Hydroxyapatite/tricalcium phosphate [HA/TCP] coated femoral or tibial baseplate components may only be used uncemented. All other femoral, tibial baseplate and all-polyethylene patella components are indicated for cemented use only.
· NexGen Augments that may be used with cemented CR, CRA, PS, and LCCK femoral and tibial baseplate components are indicated for cemented use only.
For NexGen Augments used with NexGen Rotating Hinge Knee System:
This device is indicated for patients with:
- Moderate to severe knee instability.
- Significant bone loss and/or ligament deficiencies caused by neoplasms, traumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, and/or avascular necrosis of the femoral condyle.
- Valgus, varus, or flexion deformities.
- The salvage of previously failed surgical attempts.
This device is intended for cemented use only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized letter "Z" inside it. Below the circle is the word "zimmer" in lowercase blue letters.
| | 510(k) Summary |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
| Contact Person: | Mark Warner<br>Sr. Specialist, Regulatory Affairs<br>Telephone: (574) 372-4150<br>Fax: (574) 372-4605 |
| Date: | August 31, 2015 |
| Trade Name: | Zimmer® NexGen® Complete Knee Solution Tibial and<br>Femoral Augments |
| Common Name: | Prosthesis, Knee, Patellofemorotibial, Semi-constrained,<br>Cemented, Polymer/Metal/Polymer |
| | Prosthesis, Knee, Femorotibial, Constrained, Cemented,<br>Metal/Polymer |
| Classification Names<br>and References: | Knee joint patellofemorotibial polymer/metal/polymer<br>semi-constrained cemented prosthesis, 21 CFR 888.3560,<br>JWH |
| | Knee joint femorotibial metal/polymer constrained<br>cemented prosthesis, 21 CFR 888.3510, KRO |
| Classification Panel: | Orthopedics/87 |
| Predicate Device(s): | NexGen Complete Knee Solution Cruciate Retaining<br>(Augmentable) and Legacy Constrained Condylar Knee,<br>manufactured by Zimmer, Inc. (K946150) |
| | NexGen Complete Knee Solution Rotating Hinge Knee,<br>manufactured by Zimmer, Inc. (K013385) |
| Purpose and Device<br>Description: | The NexGen augments are used with the following<br>NexGen Knee Systems: |
{4}------------------------------------------------
NexGen Rotating Hinge Knee NexGen Complete Knee Solution:
- · Cruciate Retaining Augmentable (CRA)
- Legacy Constrained Condylar (LCCK)
- · Cruciate Retaining (CR)
- · Posterior Stabilized (PS)
- · Legacy Posterior Stabilized (LPS)
This submission includes tibial and femoral augments of various shapes and sizes. Augments are attached via bone cement or screw to tibial baseplate and femoral components indicated for cemented use only. Use of these augments is optional.
For NexGen augments used with NexGen Knee Systems -Cruciate Retaining (CR), Cruciate Retaining augmentable (CRA), Posterior Stabilized (PS), Legacy Posterior Stabilized (LPS), and Legacy Constrained Condylar Knee (LCCK):
This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Collagen disorders, and/or avascular necrosis of the femoral condyle.
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities.
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
CR and LPS porous coated femoral and tibial baseplate components may be used cemented or uncemented biological fixation). The CR Hydroxyapatite/tricalcium phosphate [HA/TCP] coated femoral or tibial baseplate components may only be used uncemented. All other femoral, tibial baseplate and all-polyethylene patella components are indicated for cemented use only.
NexGen augments that may be used with cemented CR, CRA, PS, LPS, and LCCK femoral and tibial baseplate components are indicated for cemented use only.
### Intended Use:
{5}------------------------------------------------
| | For NexGen augments used with NexGen Rotating Hinge<br>Knee System: |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | This device is indicated for patients with:<br>- Moderate to severe knee instability.<br>- Significant bone loss and/or ligament deficiencies caused<br>by neoplasms, trauma, rheumatoid arthritis, osteoarthritis,<br>traumatic arthritis, polyarthritis, collagen disorders,<br>and/or avascular necrosis of the femoral condyle.<br>- Valgus, varus, or flexion deformities.<br>- The salvage of previously failed surgical attempts.<br>This device is intended for cemented use only. |
| Comparison to Predicate Device: | NexGen tibial and femoral augments were initially cleared<br>in K946150, K013385. The subject devices include<br>additional shapes and sizes of the predicate devices. They<br>have the same intended use, function, and fundamental<br>technology as the predicate devices. The difference<br>between the subject and predicate devices do not raise new<br>issues of safety and effectiveness. |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>Shelf Life Testing:<br>Accelerated aging testing conducted demonstrated that the<br>subject NexGen augments have a shelf life of 10 years. |
| | Performance Testing:<br>Non-clinical testing demonstrated that the subject NexGen<br>augments have adequate attachment strength to resist joint<br>reaction forces generated during weight-bearing activities<br>and that PMMA coated titanium substrate to bone cement<br>demonstrated an increased maximum shear load compared<br>to uncoated titanium. |
| | Clinical Performance and Conclusions:<br>Clinical data and conclusions were not needed for this<br>device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.