K152473 · Cascination AG · JAK · Jan 20, 2016 · Radiology
Device Facts
Record ID
K152473
Device Name
CAS-One IR
Applicant
Cascination AG
Product Code
JAK · Radiology
Decision Date
Jan 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K152473 · Jan 20, 2016
CAS-One IR
Cascination AG
Post-clinical evaluation of patient interventions
The sponsor conducted a post-clinical evaluation of interventions performed with the CAS-One IR system to demonstrate that the device is accurate, safe, and effective in a clinical setting compared to predicate devices.
CAS-One IR is a user controlled, stereotactic accessory intended to assist in planning, navigation and manual advancement of one or more instruments, as well as in verification and performance during Computed Tomography (CT) guided procedures. In planning, the desired needle configuration and performance is defined relative to the target anatomy. In navigation, the instrument position is displayed relative to the patient and guidance for needle alignment is provided while respiratory levels are monitored. In verification, the achieved instrument contiguration and performance are displayed relative to the previously defined plan through an overlay of the pre- and post- treatment image data. CAS-One IR is indicated for use with rigid straight instruments such as needles and probes used in CT guided interventional procedures performed by physicians trained for CT procedures.
Device Story
CAS-One IR is a mobile, stereotactic navigation system for CT-guided interventional procedures. It consists of a mobile platform with touch screens, a camera, and a computer, plus an aiming device or instrument adapter clamp. The system tracks rigid instruments (needles/probes) relative to patient anatomy. During planning, users define needle trajectories on CT images. During navigation, the system displays instrument position relative to the patient, provides alignment guidance, and monitors respiratory levels. During verification, it overlays pre- and post-treatment images to confirm instrument placement. Used in radiology suites by physicians trained in CT procedures, the device assists in precise needle advancement. Benefits include improved accuracy and workflow efficiency for interventional procedures, reducing potential for misplacement.
Clinical Evidence
Evidence includes bench testing, phantom studies, and clinical evaluation. Positional accuracy was benchmarked against predicate technology. Patient registration methods and integrated clinical workflows were validated. Accuracy tests across all needle configurations were performed on phantoms. Post-clinical evaluation of patient interventions confirmed safety and effectiveness comparable to predicates. Usability studies involving novice and experienced users demonstrated ease of use and accuracy.
Technological Characteristics
Mobile navigation platform with dual touch screens, camera, and computer. Features multi-axis mechanical arm for aiming or instrument adapter clamp for direct needle attachment. Uses optical tracking. Conforms to IEC 60601-1:2005/A1 2012 and IEC 60601-1-2:2007. Patient-contacting components are surface devices (skin, <24h) compliant with ISO 10993. Software-based workflow assistance.
Indications for Use
Indicated for use with rigid straight instruments (needles/probes) in CT-guided interventional procedures performed by physicians trained in CT procedures.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a symbol that appears to be three stylized human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 20, 2016
CAScination AG % Mr. Matthias Peterehans CEO Steigerhubelstrasse 3 CH-3008 Bern SWITZERLAND
Re: K152473
Trade/Device Name: CAS-One IR Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: November 20, 2015 Received: November 27, 2015
Dear Mr. Peterehans:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152473
Device Name CAS-One IR
#### Indications for Use (Describe)
CAS-One IR is a user controlled, stereotactic accessory intended to assist in planning, navigation and manual advancement of one or more instruments, as well as in verification of instrument position and performance during Computed Tomography (CT) guided procedures.
In planning, the desired needle configuration and performance is defined relative to the target anatomy.
In navigation, the instrument position is displayed telative to the patient and guidance for needle alignment is provided while respiratory levels are monitored.
In verification, the achieved instrument configuration and performance are displayed relative to the previously defined plan through an overlay of the pre- and post- treatment image data.
CAS-One IR is indicated for use with rigid straight instruments such as needles and probes used in CT guided interventional procedures performed by physicians trained for CT procedures.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "CAScination" in green font. The letters are connected and have a modern, sans-serif style. The word appears to be a logo or brand name, and the green color suggests a connection to nature or sustainability.
Steigerhubelstrasse 3 CH-3008 Bern Switzerland
# 5: 510(k) Summary
ID: CASC-11-228 Version: 3.0 Valid from: Tuesday January 12, 2015
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "CAScination" in green font. The letters are all capitalized except for the "c" in the middle of the word. The font is sans-serif and the letters are bolded.
# Contents
| Intended Use: | 3 |
|--------------------------------------|---|
| Description: | 3 |
| Substantial Equivalence: | 4 |
| Biocompatibility | 4 |
| Electrical Safety | 4 |
| Software Verification and Validation | 5 |
| Performance Data | 5 |
| Conclusion | 5 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "CAScination" in a sans-serif font. The word is written in green color. The letters are connected to each other.
# 510(k) Summary
| Manufacturer: | CAScination AG |
|------------------------|-----------------------------------------------------------------------------------------------------------|
| Contact: | Matthias Peterhans<br>CEO<br>Steigerhubelstrasse 3<br>CH-3008 Bern<br>Switzerland<br>Tel: +41 31 552 0440 |
| Date Prepared: | 12 January 2016 |
| Trade Name: | CAS-One IR |
| Common Name: | Navigated system for CT guided interventions |
| Classification Number: | 21 CFR 892.1750, Computed Tomography X-Ray system |
| Product Code: | JAK, System, X-ray, Tomography, Computed |
| Predicate device(s): | ig4 Image Guided System [K060903] (JAK, 21 CFR 892.1750)<br>MAXIO [K132108] (JAK, 21 CFR 892.1750) |
#### Intended Use:
CAS-One IR is a user controlled, stereotactic accessory intended to assist in planning, navigation and manual advancement of one or more instruments, as well as in verification and performance during Computed Tomography (CT) guided procedures.
In planning, the desired needle configuration and performance is defined relative to the target anatomy.
In navigation, the instrument position is displayed relative to the patient and guidance for needle alignment is provided while respiratory levels are monitored.
In verification, the achieved instrument contiguration and performance are displayed relative to the previously defined plan through an overlay of the pre- and post- treatment image data.
CAS-One IR is indicated for use with rigid straight instruments such as needles and probes used in CT guided interventional procedures performed by physicians trained for CT procedures.
### Description:
The system consists of three main components:
- A mobile navigation platform can be moved in and out of radiology rooms and is positioned next to the patient in front of the CT scanner. The platform includes two touch screens, a camera and a computer.
- Aiming device with trackable aiming insert: To aim the needles to their correct locations, the system uses an aiming device. The aiming device is attached to a multi-axis mechanical arm that can align the position of the aiming device around the expected needle entry position. The aiming device is first aligned to the desired entry point (translational alignment) and then alignment to the desired needle insertion angle is performed using a remote center of rotation principle (rotational alignment). There are two possible configurations of the aiming device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "CAScination" in green font. The letters are all capitalized, and the font is sans-serif. The word is written in a single line, and the letters are evenly spaced. The image is simple and clean, with no other elements present.
- Instrument adapter clamp with trackable marker shield: As an alternative to the aiming device, trackable markershields can be attached directly to rigid needles by means of an instrument adapter. Calibration of the needle geometry is performed with a calibration unit supplied by CAScination.
- CAS-One IR software: The software provides the step-by-step workflow assistance for needle navigation. It provides a means for users to precisely plan a single or multiple needle trajectories, navigate a needle to this exact position and validate the inserted needle's position to the planned position.
#### Substantial Equivalence:
CAS-One IR has been shown to be equivalent to:
- MAXIO (Perfint Healthcare, Pvt. Ltd.) [K132108] |
- ig4 Image Guided System (Veran Medical Technologies) [K060903]
CAS-One IR was compared and evaluated to the predicate devices with respect to the following characteristics:
- Planning
- Needle configuration and performance -
- Navigation ー
- Respiratory motion control -
- -Intra-interventional verification
- Post treatment verification —
- -Interventional instruments
- -User
- lmaging modality |
The above characteristics were evaluated and found to have the same intended use as the predicate devices. Where technological differences were identified, performance data was supplied to support the conclusion that CAS-One IR does not raise different questions of safety and effectiveness.
CAS-One IR is shown to be substantially equivalent to the predicate devices.
### Biocompatibility
All devices of CAS-One IR have the following patient contacting characteristics:
- Surface Devices
- -Skin
- -Limited (<24h)
These components were assessed for their biocompatibility through their adherence to material standards and appropriate testing according to ISO 10993.
#### Electrical Safety
Electrical safety tests according to IEC 60601-1:2005/A1 2012 and IEC 60601-1-2:2007 were conducted on CAS-One IR and the system was found to conform.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "CAScination" in a sans-serif font. The word is written in green color. The letters are connected to each other, giving the impression of a single, continuous word. The image is simple and clean, with a focus on the text.
#### Software Verification and Validation
Through risk management, the software was classified as having a moderate level of concern as a latent flaw in the software could result in minor injury to the patient. Verification testing appropriate to the software classification was carried out.
#### Performance Data
A positional accuracy bench test was conducted between the real-time tracking system used in the predicate device and CAS-One IR to show that the two technologies were substantially equivalence.
Performance of the patient registration method was evaluated and shown to be a safe and effective method of registering.
The integrated clinical workflow was benchmarked against the predicate devices and CAS-One IR was shown to be safe and effective.
An accuracy test that evaluated all needle insertion configurations of CAS-One IR was conducted on a phantom within the context of clinical use and each method was shown to be accurate and as safe and effective as the predicate devices.
A post-clinical evaluation of interventions conducted with CAS-One IR showed the system to be accurate and as safe and effective on patients within a clinical context as the predicate devices.
Usability studies with qualified users of varying degrees of experience was conducted in terms of both risk management and human factors. CAS-One IR was shown to be easy and accurate to use for both novice and experienced users.
### Conclusion
The above performance testing and substantial equivalence to predicate devices shows that CAS-One IR is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.