K152470 · Great Basin Scientific, Inc. · PAM · Mar 25, 2016 · Microbiology
Device Facts
Record ID
K152470
Device Name
Great Basin Staph ID/R Blood Culture Panel
Applicant
Great Basin Scientific, Inc.
Product Code
PAM · Microbiology
Decision Date
Mar 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3365
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K152470 · Mar 25, 2016
Great Basin Staph ID/R Blood Culture Panel
Great Basin Scientific, Inc.
Prospective clinical study of patient blood culture specimens
The prospective clinical study evaluated the performance of the Great Basin Staph ID/R Blood Culture Panel using routine clinical blood culture specimens collected from patients at three geographically diverse U.S. sites.
Prospective Clinical Study of Staph ID/R Blood Culture Panel; Prospective clinical study
Individuals receiving routine care requiring blood culture testing; Sample Size: 831 clinical specimens (762 prospective, 69 archived); Number of Sites: 3
Reference Method(s)
Detection of Staphylococcus aureus, Staphylococcus lugdunensis, Staphylococcus species (genus level), and mecA gene
Indications for Use
The Great Basin Staph ID/R Blood Culture Panel is a qualitative, multiplex, nucleic acid-based in vitro diagnostic assay intended for the simultaneous identification of nucleic acid from Staphylococcus aureus, Staphylococcus lugdunensis and various Staphylococcus species to the genus level and the detection of the mecA gene for methicillin resistance directly from patient positive blood culture specimens. The test utilizes automated hot-start enabled polymerase chain reaction (PCR) for the amplification of specific DNA targets detected by hybridization probes immobilized on a silicon chip surface. The assay is performed directly on positive blood culture specimens identified as positive by continuous monitoring blood culture system that demonstrates the presence of organisms as determined by Gram stain to contain gram-positive cocci in clusters (GPCC) or gram-positive cocci in singles (GPC). The test may be performed using blood culture bottles. The Staph ID/R Blood Culture Panel identifies Staphylococcus aureus (SA), and Staphylococcus lugdunensis, and detects other Staphylococcus species without identification to species level. The Portrait Staph ID/R Blood Culture Panel is indicated for use in conjunction with other clinical or laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor these infections. Sub-culturing positive blood cultures is necessary to recover viable organisms for further identification, susceptibility testing, or epidemiological typing to identify organisms in the blood culture that are not detected by the Great Basin Staph ID/R Blood Culture Panel. If detected, mecA may or may not be associated with Staphylococcus spp. detected or the agent responsible for the disease. Negative results for mecA antimicrobial resistance gene assays do not always indicate susceptibility, as other mechanisms of resistance to methicillin exist.
Device Story
Device uses automated hot-start PCR to amplify DNA from positive blood culture bottles; targets include tuf gene (Staphylococcus genus), nuc gene (S. aureus), and mecA gene (methicillin resistance). Amplified DNA hybridizes to probes on a silicon chip; detection via HRP-conjugated antibody and TMB substrate. Portrait Optical Reader detects colored precipitate. System includes Portrait Analyzer, single-use test cartridges, and data analysis software. Used in clinical labs; provides results in ~110 minutes. Output aids clinicians in diagnosing bacterial bloodstream infections; requires sub-culturing for further identification/susceptibility testing.
Clinical Evidence
Prospective clinical study (n=831) compared device to reference methods. S. aureus PPA 98.6% (96.0-99.5%), NPA 99.5% (98.4-99.8%). S. lugdunensis PPA 100% (43.9-100%), NPA 99.9% (99.3-99.9%). mecA detection with S. aureus PPA 94.4% (86.6-97.8%), NPA 98.8% (97.7-99.4%). Simulated/frozen samples included. Bench testing confirmed LoD, inclusivity, exclusivity, and interference resistance.
Technological Characteristics
Multiplex PCR and silicon chip-based hybridization. Targets: tuf gene, nuc gene, mecA gene. Fully automated system (PA500 Portrait Analyzer). Single-use fluidic test cartridges. Optical detection of HRP-TMB precipitate. Standalone system with integrated data analysis software.
Indications for Use
Indicated for patients with positive blood culture specimens (Gram stain GPCC or GPC) to aid in diagnosis of bacterial bloodstream infections. Identifies S. aureus, S. lugdunensis, and Staphylococcus spp. (genus level), and detects mecA gene for methicillin resistance.
Regulatory Classification
Identification
A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.
Special Controls
In combination with the general controls of the FD&C Act, the Verigene® Gram Positive Blood Culture Nucleic Acid Test is subject to the following special controls: The special controls for the BC-GP Assay are contained in the guideline document entitled "Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures."
*Classification.* Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).
Predicate Devices
Verigene Gram Positive Blood Culture Nucleic Acid Test (BC-GP) (K113450)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, layered on top of each other to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 26, 2016
GREAT BASIN SCIENTIFIC, INC. CHUCK OWEN DIRECTOR, REGULATORY AFFAIRS & QUALITY ASSURANCE 2441 S. 3850 WEST SALT LAKE CITY, UT 84120
Re: K152470 Trade/Device Name: Great Basin Staph ID/R Blood Culture Panel Regulation Number: 21 CFR 866.3365 Regulation Name: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures Regulatory Class: II Product Code: PAM, OOI Dated: February 22, 2016 Received: February 25, 2016
Dear Mr. Owen:
This letter corrects our substantially equivalent letter of March 25, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
Page 2 - Mr. Owen
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Kristian M. Roth -S
For: Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152470
Device Name Great Basin Staph ID/R Blood Culture Panel
#### Indications for Use (Describe)
The Great Basin Staph ID/R Blood Culture Panel is a qualitative, multiplex, nucleic acid-based in vitro diagnostic assay intended for the simultaneous identification of nucleic acid from Staphylococcus lugdumensis and various Staphylococcus species to the genus level and the detection of the mecA gene for methicillin resistance directly from patient positive blood culture specimens. The test utilizes automated hot-start enabled polymerase chain reaction (PCR) for the amplification of specific DNA targets detected by hybridization probes immobilized on a silicon chip surface. The assay is performed directly on positive blood culture specimens identified as positive by continuous monitoring blood culture system that demonstrates the presence of organisms as determined by Gram stain to contain gram-positive cocci in clusters (GPCC) or gram-positive cocci in singles (GPC). The test may be performed using blood culture bottles. The Staph ID/R Blood Culture Panel identifies Staphylococcus aureus (SA), and Staphylococcus lugdunensis, and detects other Staphylococcus species without identification to species level.
The Portrait Staph ID/R Blood Culture Panel is indicated for use in conjunction with other clinical or laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor these infections. Subculturing positive blood cultures is necessary to recover viable organisms for further identification, susceptibility testing, or epidemiological typing to identify organisms in the blood culture that are not detected by the Great Basin Staph ID/R Blood Culture Panel. If detected, mecA may or may not be associated with Staphylococcus spp. detected or the agent responsible for the disease. Negative results for mecA antimicrobial resistance gene assays do not always indicate susceptibility, as other mechanisms of resistance to methicillin exist.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------|
| <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Great Basin Scientific. The logo consists of a series of blue oval shapes arranged in a staircase pattern above the words "GREAT BASIN" in gold, with the word "SCIENTIFIC" in smaller gold letters below. The logo is simple and modern, and the colors are eye-catching.
March 23, 2016
## 510(k) Summary: Great Basin Staph ID/R Blood Culture Panel
## A. Submitted by:
Great Basin Corporation 2441 South 3850 West Salt Lake City, Utah 84120 Phone: 801-990-1055 Fax: 801-990-1051
## Contact Information
Chuck Owen, Director of Requlatory Affairs Phone: 385-215-3313 801-990-1051 Fax: Email: cowen@gbscience.com
### B. Name of Device
| Proprietary Name: | Great Basin Staph ID/R Blood Culture Panel |
|-------------------|--------------------------------------------|
|-------------------|--------------------------------------------|
| Common or Usual Names: | Staph ID/R Blood Culture Panel<br>Staph Assay<br>SIDR |
|------------------------|-------------------------------------------------------|
|------------------------|-------------------------------------------------------|
## C. Regulatory Information:
| a. Regulation Section: | 21 CFR 866.3365 – Multiplex Nucleic Acid Assay<br>for Identification of Microorganisms and Resistance<br>Markers from Positive Blood Cultures<br>21 CFR 862.2570 – Instrumentation for clinical<br>multiplex test systems |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| b. Classification: | Class II (Staph ID/R Blood Culture Panel; non-exempt)<br>Class II (PA500 Portrait Analyzer System) |
| c. Classification panel: | Microbiology Devices, OIVD (83) Microbiology |
| d. Product Code: | PAM (Gram-positive bacteria and their resistance markers)<br>OOI (Real-time nucleic acid amplification system) |
### D. Intended use(s)/Indications for Use:
The Great Basin Staph ID/R Blood Culture Panel is a qualitative, multiplex, nucleic acid-based in vitro diagnostic assay intended for the simultaneous identification of nucleic acid from Staphylococcus aureus, Staphylococcus lugdunensis and various Staphylococcus species to the genus level and the detection of the mecA gene for methicillin resistance directly from patient positive blood culture specimens. The test utilizes automated hot-start enabled polymerase chain reaction (PCR) for the
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Great Basin Scientific. The logo features a series of blue oval shapes arranged in a step-like pattern above the words "GREAT BASIN" in a gold, sans-serif font. Below "GREAT BASIN" is the word "SCIENTIFIC" in a smaller, similar font.
amplification of specific DNA targets detected by hybridization probes immobilized on a silicon chip surface. The assay is performed directly on positive blood culture specimens identified as positive by continuous monitoring blood culture system that demonstrates the presence of organisms as determined by Gram stain to contain gram-positive cocci in clusters (GPCC) or gram-positive cocci in singles (GPC). The test may be performed using blood culture bottles. The Staph ID/R Blood Culture Panel identifies Staphylococcus aureus (SA), and Staphylococcus lugdunensis, and detects other Staphylococcus species without identification to species level.
The Portrait Staph ID/R Blood Culture Panel is indicated for use in conjunction with other clinical or laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor these infections. Sub-culturing positive blood cultures is necessary to recover viable organisms for further identification, susceptibility testing, or epidemiological typing to identify organisms in the blood culture that are not detected by the Great Basin Staph ID/R Blood Culture Panel. If detected, mecA may or may not be associated with Staphylococcus spp. detected or the agent responsible for the disease. Negative results for mecA antimicrobial resistance gene assays do not always indicate susceptibility, as other mechanisms of resistance to methicillin exist.
### E. Device Description:
#### Test Principle:
The Great Basin Staph ID/R Blood Panel on the PA500 Portrait™ Analyzer System utilizes automated hot-start enabled polymerase chain reaction (PCR) amplification technology to amplify specific nucleic acid sequences that are detected using species specific Staphylococcal DNA hybridization probes immobilized on a modified silicon chip surface.
Target genomic DNA is extracted from microbial cells and diluted to reduce potential inhibitors of the PCR reaction. During the PCR process, double-stranded DNA is separated and target nucleic acid sequences are amplified by thermal cycling. Biotin-labeled primers direct amplification of specific nucleic acid sequences within a variable region of the tuf gene for identification of coagulase-neqative Staphylococcus species, within conserved region of the thermonuclease (nuc) gene for specific identification of Staphylococcus aureus, and the mecA gene for detecting oxacillin/methicillin resistance. Following the PCR process, biotin-labeled, amplified target DNA sequences are hybridized to an array of probes immobilized on the silicon chip surface, then incubated with antibody conjugated to the horseradish peroxidase enzyme (HRP). The unbound conjugate is removed by washing and tetramethylbenzidine (TMB) is added to produce a colored precipitate at the location of the probe/target sequence complex. The resulting signal is detected by the automated Portrait Optical Reader within the Portrait System.
#### Test Device:
The Great Basin PA500 Portrait Analyzer System is a fully automated system that includes the Portrait Analyzer, single-use Staph ID/R Blood Culture Panel Test Cartridges, and the Portrait data analysis software. The PA500 Portrait Analyzer System is designed to perform automated sample preparation, PCR, and optical chip-based detection with integrated data analysis in approximately 110 minutes.
### F. Substantial Equivalence Information:
- a. Predicate Device: Verigene Gram Positive Blood Culture Nucleic Acid Test (BC-GP)
- b. Predicate 510(k) number: K113450
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Great Basin Scientific. The logo features a series of blue shapes resembling steps or a staircase above the words "GREAT BASIN" in gold lettering. Below "GREAT BASIN" is the word "SCIENTIFIC" in a smaller font, also in gold.
## c. Comparison with Predicate
| Item | Staph ID/R Blood Culture Panel | Predicate (K113450) |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Great Basin Scientific, Inc. | Nanosphere, Inc. |
| Trade Name | Portrait™ Staph ID/R Blood Culture Panel | Verigene Gram Positive Blood Culture Nucleic Acid Test (BC-GP) |
| 510(k) Number | K152470 | K113450 |
| | Similarities | |
| Classification | Class II | same |
| Intended<br>Use/Indications<br>for Use | The Great Basin Staph ID/R Blood Culture Panel is a qualitative, multiplex,<br>nucleic acid-based in vitro diagnostic assay intended for the simultaneous<br>identification of nucleic acid from Staphylococcus aureus, Staphylococcus<br>lugdunensis and various Staphylococcus species to the genus level and the<br>detection of the mecA gene for methicillin resistance directly from patient<br>positive blood culture specimens. The test utilizes automated hot-start<br>enabled polymerase chain reaction (PCR) for the amplification of specific<br>DNA targets detected by hybridization probes immobilized on a silicon chip<br>surface. The assay is performed directly on positive blood culture specimens<br>identified as positive by continuous monitoring blood culture system that<br>demonstrates the presence of organisms as determined by Gram stain to<br>contain gram-positive cocci in clusters (GPCC) or gram-positive cocci in<br>singles (GPC). The test may be performed using blood culture bottles. The<br>Staph ID/R Blood Culture Panel identifies Staphylococcus aureus (SA), and<br>Staphylococcus lugdunensis, and detects other Staphylococcus species<br>without identification to species level. The Staph ID/R Blood Culture Panel is<br>indicated for use in conjunction with other clinical or laboratory findings to<br>aid in the diagnosis of bacterial bloodstream infections; however, it is not<br>used to monitor these infections. Sub-culturing positive blood cultures is<br>necessary to recover viable organisms for further identification, susceptibility<br>testing, or epidemiological typing to identify organisms in the blood culture<br>that are not detected by the Great Basin Staph ID/R Blood Culture Panel. If<br>detected, mecA may or may not be associated with Staphylococcus spp.<br>detected or the agent responsible for the disease. Negative results for mecA<br>antimicrobial resistance gene assays do not always indicate susceptibility, as<br>other mechanisms of resistance to methicillin exist. | The Verigene® Gram Positive Blood Culture Nucleic Acid Test (BC-GP)<br>performed using the sample-to-result Verigene System is a qualitative,<br>multiplexed <i>in vitro</i> diagnostic test for the simultaneous detection and<br>identification of potentially pathogenic gram-positive bacteria which may<br>cause bloodstream infection (BSI). BC-GP is indicated for use in conjunction<br>with other clinical and laboratory findings, such as culture, to aid in the<br>diagnosis of bacterial bloodstream infections; however, it is not used to<br>monitor bloodstream infections. BC-Gp detects the following bacterial<br>genera and species: <i>Staphylococcus</i> spp., <i>Staphylococcus aureus</i> ,<br><i>Staphylococcus epidermidis</i> , and <i>Staphylococcus lugdunensis</i> , <i>Streptococcus</i><br>spp., <i>Streptococcus pneumonia</i> , <i>Streptococcus pyogenes</i> , <i>Streptococcus</i><br><i>agalactiae</i> , <i>Streptococcus anginosus</i> group, <i>Entercoccus faecalis</i> ,<br><i>Enterococcus faecium</i> , <i>Listeria</i> spp. In addition, BC-GP detects the <i>mecA</i><br>resistance marker, and the <i>vanA</i> and <i>vanB</i> resistance markers, inferring<br>vanA/vanB mediated vancomycin resistance to either <i>E. faecalis</i> or <i>E.</i><br><i>faecium</i> , or the <i>mecA</i> -mediated methicillin resistance to either <i>S. aureus</i> or<br><i>S. epidermidis</i> . BC-GP is indicated for use in conjunction with other clinical<br>or laboratory findings to aid in the diagnosis of bacterial bloodstream<br>infections; however, it is not used to monitor these infections. Sub-culturing<br>of positive blood cultures is necessary to recover organisms for susceptibility<br>testing, identification of organisms not detected by BC-GP, differentiation of<br>mixed growth, association of antimicrobial marker genes to a specific<br>organism, or for epidemiological testing. |
| Qualitative/<br>Quantitative | Qualitative | same |
| Single-Use Test<br>Cartridge | Disposable, single-use, self-contained fluidic test cartridge | same |
| Automated | Yes | same |
| Test Principle | Multiplex nucleic acid array-based detection | same |
| Sample Types | Direct blood culture positive by Gram stain for GPCC or GPC | same |
| Organism and<br>Resistance<br>Marker Detection | <i>S. aureus</i> , <i>S. lugdunensis</i> , <i>Staphylococcus</i> spp., <i>mecA</i> | <i>S. aureus</i> , <i>S. lugdunensis</i> , <i>S. epidermidis</i> , <i>Staphylococcus</i> spp., <i>mecA</i> |
| Controls | One Internal Processing Control (whole organism complete assay<br>control) | Two Internal Processing Controls (whole organism complete assay<br>control and single-stranded DNA Hybridization control) |
| Calibration | Not required | same |
| | Differences…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.