← Product Code [IYN](/productcode/IYN) · K152467

# Touch Ultrasound (K152467)

_Carestream Health, Inc. · IYN · Dec 9, 2015 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152467

## Device Facts

- **Applicant:** Carestream Health, Inc.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Dec 9, 2015
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal (incl Obstetrics) Abdominal Intraoperative Neurosurgery Pediatrics Small Parts (organs) Adult Cephalic Neonatal Cephalic Transrectal Transvaginal Transurethral Muskulo-skeletal (conventional and superficial) Cardiac Peripheral Vessel

## Device Story

Touch Ultrasound is a diagnostic ultrasound system using transducers to acquire B-mode, M-mode, PWD, CWD, Color Doppler, Tissue/Contrast Harmonic Imaging, and Elastography data. It features Vector Flow Imaging (VFI) for axial/transverse velocity visualization and 3D reconstruction of 2D images. Operated by clinicians in clinical settings, the system processes acoustic signals to display images and perform geometrical measurements/calculations for biopsy needle guidance and parameter estimation. Output is viewed on-screen to assist in diagnostic decision-making. The device maintains acoustic output within Track 3 limits (Ispta 720 mW/cm², MI 1.9).

## Clinical Evidence

Bench testing only. The device was evaluated for acoustic output, thermal, electrical, electromagnetic, and mechanical safety, and biocompatibility. It conforms to standards including AIUM/NEMA UD-2, AIUM/NEMA UD-3, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 14971, EN ISO 10993-1, IEC 62304, and IEC 62359.

## Technological Characteristics

Diagnostic ultrasound system utilizing pulsed echo and Doppler imaging. Features include B-mode, M-mode, PWD, CWD, Color Doppler, Tissue/Contrast Harmonic Imaging, Elastography, and Vector Flow Imaging. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, and IEC 62304. Acoustic output follows Track 3 (Ispta < 720 mW/cm², MI < 1.9). Biocompatible patient-contact materials. No changes to physical transducer design.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Touch Ultrasound ([K150342](/device/K150342.md))
- Ultrasound scanner bk2300 ([K140428](/device/K140428.md))
- Ultrasound scanner Pro Focus 2202 ([K043524](/device/K043524.md))
- Ultrasound scanner Pro Focus 2202 ([K100919](/device/K100919.md))
- Ultrasound scanner system Acuson S1000 ([K130619](/device/K130619.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, with a series of curved lines above them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 9, 2015

Carestream Health, Inc. % Ms. Carolyn Wagner Supervising Sr. Manager, Regulatory Affairs & Quality Systems 150 Verona Street ROCHESTER NY 14608

Re: K152467

Trade/Device Name: Touch Ultrasound Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: November 5, 2015 Received: November 9, 2015

Dear Ms. Wagner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Oakes

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

CEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Annroved: OMB No. 0910-0120

510(k) Number (if known)

K152467

Device Name

#### Touch Ultrasound

#### Indications for Use (Describe)

Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD, CWD, Color Doppler, Vector Flow Imaging and combined mode imaging and Elastography.

Signal Analysis.

Guidance of biopsy needles, geometrical measurements and calculation of parameters. And optimal 3-D unit can reconstruct a series of 2-D images into a single 3-D volume and display this on the screen.

An optional Vector Flow Imaging (VFI) module: Color Flow Mapping (CFM) imaging mode with the ability to visualize both the axial and the transverse velocity.

Clinical Applications:

- Fetal (Sometimes called Obstetrics) ●
- Abdominal ●
- Intra-operative
- Intra-operative (Neuro) (sometimes called Neuro Surgical) ●
- Pediatrics ●
- Small Organ (Sometimes called Small Parts) ●
- . Neonatal Cephalic (Cephalic is sometimes called trans-cranial)
- Adult Cephalic (Cephalic is sometimes called trans-cranial)
- Trans-rectal
- Trans-vaqinal
- Trans-urethral
- Musculo-skeletal (Conventional)
- Musculo-skeletal (Superficial)
- . Cardiac Adult
- . Peripheral Vessel (Sometimes called Peripheral Vascular)

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The Counter Use (21 CFR 801 Suppail C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED. PLEASE DO NOT WRITE BELOW THIS LINE

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

#### This section applies only to requirements of the Paperwork Recuction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response including the time to review instructions, search existing data sources, gather and maintain the dista needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burder, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Intornation Officer Paperwork Reduction Act (PRA) Staff PRAStaff@ida.hhs.gov

Ar agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number

{4}------------------------------------------------

#### Fill out one form for each ultrasound system and each transducer.

#### System: Touch Ultrasound

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            |   | Modes |     |            |                               |                                        |                  |                      |                              |                |              |
|---------------------------------|---|-------|-----|------------|-------------------------------|----------------------------------------|------------------|----------------------|------------------------------|----------------|--------------|
| Specific<br>(Tracks I & III)    | B | M     | PWD | CWD<br>25) | Tissue<br>Harmonic<br>Imaging | Contrast<br>Harmonic<br>Imaging<br>26) | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>1) | Other        |
| Ophthalmic                      |   |       |     |            |                               |                                        |                  |                      |                              |                |              |
| Fetal 2)                        | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Abdominal 3)                    | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Intra-operative (Specify 4)     | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Intra-operative (Neuro 5)       | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Laparoscopic 6)                 |   |       |     |            |                               |                                        |                  |                      |                              |                |              |
| Pediatric 7)                    | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Small Organ (Specify 8)         | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Neonatal Cephalic 9)            | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Adult Cephalic 10)              | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Trans-rectal 11)                | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Trans-vaginal 12)               | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Trans-urethral 13)              | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Trans-esoph. (non-Card.) 14)    |   |       |     |            |                               |                                        |                  |                      |                              |                |              |
| Musculo-skel. (Conventional 15) | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Musculo-skel. (Superficial 16)  | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Intra-luminal 17)               |   |       |     |            |                               |                                        |                  |                      |                              |                |              |
| Other (Specify 18)              |   |       |     |            |                               |                                        |                  |                      |                              |                |              |
| Cardiac Adult 19)               | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N24)         |
| Cardiac Pediatric 20)           |   |       |     |            |                               |                                        |                  |                      |                              |                |              |
| Trans-esoph. (Cardiac 21)       |   |       |     |            |                               |                                        |                  |                      |                              |                |              |
| Other (Specify 22)              |   |       |     |            |                               |                                        |                  |                      |                              |                |              |
| Peripheral vessel 23)           | P | P     | P   | N          | P                             | N                                      | P                | P                    | P                            | P              | N23)<br>N24) |
| Other (Specify 24)              |   |       |     |            |                               |                                        |                  |                      |                              |                | N            |

N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.

The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006 Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.

Additional Comments:

| 1. | Mode combinations: | B+M,<br>B+D,<br>B+C,<br>B+D+C.<br>B includes Tissue Harmonic Imaging<br>D:PWD.<br>C: Color Flow mapping Doppler incl.<br>Amplitude(power)Doppler |
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|

Vector Flow Imaging (VFI) 23)

24) Elastography.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)

{5}------------------------------------------------

### Fill out one form for each ultrasound system and each transducer.

#### System: Touch Ultrasound

#### Transducer: 9018

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            | B | M | PWD | CWD | Tissue<br>Harmonic<br>Imaging | Modes | Contrast<br>Harmonic<br>Imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined 1) | Other |
|---------------------------------|---|---|-----|-----|-------------------------------|-------|---------------------------------|------------------|----------------------|------------------------------|-------------|-------|
| Specific<br>(Tracks I & III)    |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Ophthalmic                      |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Fetal 2)                        |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Abdominal 3)                    |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Intra-operative (Specify 4)     |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Intra-operative (Neuro 5)       |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Laparoscopic 6)                 |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Pediatric 7)                    |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Small Organ (Specify 8)         |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Neonatal Cephalic 9)            |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Adult Cephalic 10)              |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Trans-rectal 11)                | P | P | P   | P   | P                             |       | P                               | P                | P                    | P                            | P           | N24)  |
| Trans-vaginal 12)               | P | P | P   | P   | P                             |       | P                               | P                | P                    | P                            | P           | N24)  |
| Trans-urethral 13)              |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Trans-esoph. (non-Card.) 14)    |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Musculo-skel. (Conventional 15) |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Musculo-skel. (Superficial 16)  |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Intra-luminal 17)               |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Other (Specify 18)              |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Cardiac Adult 19)               |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Cardiac Pediatric 20)           |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Trans-esoph. (Cardiac 21)       |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Other (Specify 22)              |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Peripheral vessel 23)           |   |   |     |     |                               |       |                                 |                  |                      |                              |             |       |
| Other (Specify 24)              |   |   |     |     |                               |       |                                 |                  |                      |                              |             | N     |

N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.

The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.

Additional Comments:

| 1. Mode combinations: | B+M,<br>B+D,<br>B+C,<br>B+D+C.<br>B includes Tissue Harmonic Imaging<br>D:PWD.<br>C: Color Flow mapping Doppler incl.<br>Amplitude(power)Doppler |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|

24) Elastography.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)

{6}------------------------------------------------

### Fill out one form for each ultrasound system and each transducer.

#### System: Touch Ultrasound

#### Transducer: 9051

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            |   |   |     |     |                               | Modes                           |                  |                      |                              |             |              |  |
|---------------------------------|---|---|-----|-----|-------------------------------|---------------------------------|------------------|----------------------|------------------------------|-------------|--------------|--|
| Specific<br>(Tracks I & III)    | B | M | PWD | CWD | Tissue<br>Harmonic<br>Imaging | Contrast<br>Harmonic<br>Imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined 1) | Other        |  |
| Ophthalmic                      |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Fetal 2)                        |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Abdominal 3)                    |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Intra-operative (Specify 4)     |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Intra-operative (Neuro 5)       |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Laparoscopic 6)                 |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Pediatric 7)                    | P | P | P   | P   | P                             | P                               | P                | P                    |                              | P           | N24)         |  |
| Small Organ (Specify 8)         | P | P | P   | P   | P                             | P                               | P                | P                    |                              | P           | N24)         |  |
| Neonatal Cephalic 9)            |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Adult Cephalic 10)              |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Trans-rectal 11)                |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Trans-vaginal 12)               |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Trans-urethral 13)              |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Trans-esoph. (non-Card.) 14)    |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Musculo-skel. (Conventional 15) | P | P | P   | P   | P                             | P                               | P                | P                    |                              | P           | N24)         |  |
| Musculo-skel. (Superficial 16)  | P | P | P   | P   | P                             | P                               | P                | P                    |                              | P           | N24)         |  |
| Intra-luminal 17)               |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Other (Specify 18)              |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Cardiac Adult 19)               |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Cardiac Pediatric 20)           |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Trans-esoph. (Cardiac 21)       |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Other (Specify 22)              |   |   |     |     |                               |                                 |                  |                      |                              |             |              |  |
| Peripheral vessel 23)           | P | P | P   | P   | P                             | P                               | P                | P                    |                              | P           | N23)<br>N24) |  |
| Other (Specify 24)              |   |   |     |     |                               |                                 |                  |                      |                              |             | N            |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.

The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.

Additional Comments:

| 1. | Mode combinations: | B+M,<br>B+D,<br>B+C,<br>B+D+C.<br>B includes Tissue Harmonic Imaging<br>D:PWD.<br>C: Color Flow mapping Doppler incl.<br>Amplitude(power)Doppler |
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|

Vector Flow Imaging (VFI) 23)

24) Elastography.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)

{7}------------------------------------------------

### Fill out one form for each ultrasound system and each transducer.

#### System: Touch Ultrasound

#### Transducer: 9062

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            | Modes |   |     |            |                               |                                        |                  |                      |                              |             |       |
|---------------------------------|-------|---|-----|------------|-------------------------------|----------------------------------------|------------------|----------------------|------------------------------|-------------|-------|
| Specific<br>(Tracks I & III)    | B     | M | PWD | CWD<br>25) | Tissue<br>Harmonic<br>Imaging | Contrast<br>Harmonic<br>Imaging<br>26) | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined 1) | Other |
| Ophthalmic                      |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Fetal 2)                        |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Abdominal 3)                    |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Intra-operative (Specify 4)     | P     | P | P   |            | P                             |                                        | P                | P                    |                              | P           | N24)  |
| Intra-operative (Neuro 5)       | P     | P | P   |            | P                             |                                        | P                | P                    |                              | P           | N24)  |
| Laparoscopic 6)                 |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Pediatric 7)                    | P     | P | P   |            | P                             |                                        | P                | P                    |                              | P           | N24)  |
| Small Organ (Specify 8)         |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Neonatal Cephalic 9)            | P     | P | P   |            | P                             |                                        | P                | P                    |                              | P           | N24)  |
| Adult Cephalic 10)              |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Trans-rectal 11)                |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Trans-vaginal 12)               |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Trans-urethral 13)              |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Trans-esoph. (non-Card.) 14)    |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Musculo-skel. (Conventional 15) |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Musculo-skel. (Superficial 16)  |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Intra-luminal 17)               |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Other (Specify 18)              |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Cardiac Adult 19)               |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Cardiac Pediatric 20)           |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Trans-esoph. (Cardiac 21)       |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Other (Specify 22)              |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Peripheral vessel 23)           |       |   |     |            |                               |                                        |                  |                      |                              |             |       |
| Other (Specify 24)              |       |   |     |            |                               |                                        |                  |                      |                              |             | N     |

N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.

The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.

Additional Comments:

| 1. | Mode combinations: | B+M,<br>B+D,<br>B+C,<br>B+D+C.<br>B includes Tissue Harmonic Imaging<br>D:PWD.<br>C: Color Flow mapping Doppler incl.<br>Amplitude(power)Doppler |
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|

24) Elastography.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)

{8}------------------------------------------------

### Fill out one form for each ultrasound system and each transducer.

#### System: Touch Ultrasound

#### Transducer: 9077

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            | Modes |   |     |     |                    |                      |                  |                      |                              |             |       |
|---------------------------------|-------|---|-----|-----|--------------------|----------------------|------------------|----------------------|------------------------------|-------------|-------|
| Specific                        | B     | M | PWD | CWD | Tissue<br>Harmonic | Contrast<br>Harmonic | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined 1) | Other |
| (Tracks I & III)                |       |   |     | 25) | Imaging            | Imaging<br>26)       |                  |                      |                              |             |       |
| Ophthalmic                      |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Fetal 2)                        |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Abdominal 3)                    | P     | P | P   | N   | P                  | P                    | P                | P                    | P                            | P           |       |
| Intra-operative (Specify 4)     |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Intra-operative (Neuro 5)       |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Laparoscopic 6)                 |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Pediatric 7)                    |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Small Organ (Specify 8)         |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Neonatal Cephalic 9)            |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Adult Cephalic 10)              | P     | P | P   | N   | P                  | P                    | P                | P                    | P                            | P           |       |
| Trans-rectal 11)                |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Trans-vaginal 12)               |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Trans-urethral 13)              |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Trans-esoph. (non-Card.) 14)    |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Musculo-skel. (Conventional 15) |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Musculo-skel. (Superficial 16)  |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Intra-luminal 17)               |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Other (Specify 18)              |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Cardiac Adult 19)               | P     | P | P   | N   | P                  | P                    | P                | P                    | P                            | P           |       |
| Cardiac Pediatric 20)           |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Trans-esoph. (Cardiac 21)       |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Other (Specify 22)              |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Peripheral vessel 23)           |       |   |     |     |                    |                      |                  |                      |                              |             |       |
| Other (Specify 24)              |       |   |     |     |                    |                      |                  |                      |                              |             |       |

N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.

The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.

#### Additional Comments:

| 1. | Mode combinations: | B+M,<br>B+D,<br>B+C,<br>B+D+C.<br>B includes Tissue Harmonic Imaging<br>D:PWD.<br>C: Color Flow mapping Doppler incl.<br>Amplitude(power)Doppler |
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|

25) Continuous Wave Doppler (CWD/CW)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)

{9}------------------------------------------------

#### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: bk2300_ Transducer: 9022 (Product Name 10L2w)

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Specific | B | M | PWD | CW | Tissue<br>Harmonic<br>Imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify 1) | Other |
|----------------------------------|---|---|-----|----|-------------------------------|------------------|----------------------|------------------------------|-------------------------|-------|
| Ophthalmic                       |   |   |     |    |                               |                  |                      |                              |                         |       |
| Fetal                            |   |   |     |    |                               |                  |                      |                              |                         |       |
| Abdominal                        |   |   |     |    |                               |                  |                      |                              |                         |       |
| Intra-operative (Specify)        |   |   |     |    |                               |                  |                      |                              |                         |       |
| Intra-operative (Neuro)          |   |   |     |    |                               |                  |                      |                              |                         |       |
| Laparoscopic                     |   |   |     |    |                               |                  |                      |                              |                         |       |
| Pediatric                        |   |   |     |    |                               |                  |                      |                              |                         |       |
| Small Organ (Specify)            |   |   |     |    |                               |                  |                      |                              |                         |       |
| Neonatal Cephalic                |   |   |     |    |                               |                  |                      |                              |                         |       |
| Adult Cephalic                   |   |   |     |    |                               |                  |                      |                              |                         |       |
| Trans-rectal                     |   |   |     |    |                               |                  |                      |                              |                         |       |
| Trans-vaginal                    |   |   |     |    |                               |                  |                      |                              |                         |       |
| Trans-urethral                   |   |   |     |    |                               |                  |                      |                              |                         |       |
| Trans-esoph. (non-Card.)         |   |   |     |    |                               |                  |                      |                              |                         |       |
| Musculo-skel. (Conventional)     |   |   |     |    |                               |                  |                      |                              |                         |       |
| Musculo-skel. (Superficial)      |   |   |     |    |                               |                  |                      |                              |                         |       |
| Intra-luminal                    |   |   |     |    |                               |                  |                      |                              |                         |       |
| Other (Specify)                  |   |   |     |    |                               |                  |                      |                              |                         |       |
| Cardiac Adult                    |   |   |     |    |                               |                  |                      |                              |                         |       |
| Cardiac Pediatric                |   |   |     |    |                               |                  |                      |                              |                         |       |
| Trans-esoph. (Cardiac)           |   |   |     |    |                               |                  |                      |                              |                         |       |
| Other (Specify)                  |   |   |     |    |                               |                  |                      |                              |                         |       |
| Peripheral vessel                | N | N | N   | N  | N                             | N                | N                    | N                            | N                       |       |
| Other (Specify) 2)               |   |   |     |    |                               |                  |                      |                              |                         | N     |

N= new indication; P= previously approved.

*Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.

Additional Comments:

1) Mode combinations: B+M, B+D, B+C, B+D+C. (D is PWD, C is Color Flow mapping Doppler including Amplitude(power)Doppler)

2) Elastography

{10}------------------------------------------------

# 510(k) Summary:

This summary is provided as part of this Premarket Notification in compliance with 21CRF, Section 807.92.

Submitters name: Carestream Health, Inc. Address: 150 Verona St., Rochester, NY 14608 Phone: 585-627-6588 Fax: 585-323-7643 Contact person: Carolyn Wagner, Sr. Supervising Manager, Regulatory Affairs & Quality Systems Date prepared: August 25, 2015

Trade name: Touch Ultrasound Common name: Diagnostic Ultrasound System Classification names: Ultrasonic Pulsed Echo Imaging System (90 IYO, CFR 892.1560) Ultrasonic Pulsed Doppler Imaging System (90 IYN, CFR 892.1550) Diagnostic Ultrasonic Transducer (90 ITX, CFR 892.1570)

Identification of predicate, legally marketed devices: B-K Medical Ultrasound Scanner Pro Focus 2202, K043524 B-K Medical Ultrasound Scanner Pro Focus 2202, K100919 BK Medical Ultrasound Scanner bk2300, K140428 Carestream Health Touch Ultrasound, K150342 Siemens Ultrasound Scanner System Acuson S1000, K130619

## Device description:

Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD, CWD, Color Doppler, Smart Flow Imaging, combined mode imaging, and Elastography.

Signal Analysis.

Guidance of biopsy needles, geometrical measurements and calculation of parameters. A 3-D unit that can reconstruct a series of 2-D images into a single 3-D volume and display this on the screen.

An optional Smart Flow Imaqinq module: Color Flow Mapping (CFM) imaqinq mode with the ability to visualize both the axial and the transverse velocity.

{11}------------------------------------------------

## Transducers

All the transducers in this submission have previously been cleared by FDA.

In this 510(k) application the new modes:

- 1) Elastography
- 2) Smart Flow Imaging (referred to as Vector Flow Imaging [VFI] in K100919)
- 3) Continuous Wave Doppler (CWD/CW)
- 4) Contrast Harmonic Imaging (CHI or Ci)

have been added to the functionality of relevant transducers as indicated in the Instructions For Use Section in VOL 004.

There are no changes to the physical design of the transducers.

There are no changes to the patient contact materials. All patient contact materials are biocompatible.

All transducers used together withTouch Ultrasound are Track 3 transducers.

## Acoustic output

The system controlling the Acoustic Output in the modified Touch US is the same as the system in the predicate device Pro Focus 2202. The system will assure that the acoustic output always will stay below the pre-amendments upper limits i.e. Ispta 720 mW/cm² and MI 1.9 (Track 3, non ophthalmic).

The Thermal Index values are maximum 6.0, i.e. TI [ 6.0

## Clinical measurement accuracy

Clinical measurements and calculations are described and accuracies are provided in the User Information.

## Thermal, mechanical and electrical safety.

The ultrasound scanner system Touch Ultrasound has been tested by a recognized Certified Body.

## Acoustic Output Reporting

The Acoustic Output Reporting is made according to the standards required by "Information for Manufacturers Seeking Clearance of Diagnostic Ultrasound Systems and Transducers, FDA, CDRH, September 9, 2008".

Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body.

{12}------------------------------------------------

## Summary of Technological Characteristics - Predicate Devices Compared to Modified Device

| Devices and<br>Predicates | Predicate Device<br>K150342, Touch<br>Ultrasound                                                  | Predicate Device<br>K140428,<br>Ultrasound scanner<br>bk2300                                      | Predicate Device<br>K043524,<br>Ultrasound scanner<br>Pro Focus 2202                              | Predicate Device<br>K100919,<br>Ultrasound scanner<br>Pro Focus 2202                              | Predicate Device<br>K130619,<br>Ultrasound scanner<br>system Acuson<br>S1000                      | Modified Device<br>(this application),<br>Touch Ultrasound                                        |
|---------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|                           |                                                                                                   |                                                                                                   |                                                                                                   |                                                                                                   |                                                                                                   |                                                                                                   |
| Ref: [1] Appendix G       | B                                                                                                 | B                                                                                                 | B                                                                                                 | B                                                                                                 | B                                                                                                 |                                                                                                   |
|                           | THI                                                                                               | THI                                                                                               | THI                                                                                               | THI                                                                                               | THI                                                                                               |                                                                                                   |
|                           | M,                                                                                                | M,                                                                                                | M,                                                                                                | M,                                                                                                | M,                                                                                                |                                                                                                   |
|                           | PWD,                                                                                              | PWD,                                                                                              | PWD,                                                                                              | PWD,                                                                                              | PWD,                                                                                              |                                                                                                   |
|                           | CFM1)                                                                                             | CFM1)                                                                                             | CFM1)                                                                                             | CFM1)                                                                                             | CFM1)                                                                                             |                                                                                                   |
|                           | and combinations                                                                                  | and combinations                                                                                  | and combinations                                                                                  | and combinations                                                                                  | and combinations                                                                                  |                                                                                                   |
|                           |                                                                                                   |                                                                                                   | CHI                                                                                               |                                                                                                   |                                                                                                   |                                                                                                   |
|                           |                                                                                                   |                                                                                                   | CWD,                                                                                              |                                                                                                   |                                                                                                   |                                                                                                   |
|                           |                                                                                                   |                                                                                                   |                                                                                                   | VFI                                                                                               |                                                                                                   |                                                                                                   |
|                           |                                                                                                   |                                                                                                   |                                                                                                   |                                                                                                   | Elastography                                                                                      |                                                                                                   |
| Intended Use:             | Diagnostic<br>Ultrasound<br>imaging or fluid<br>flow analysis of<br>the human body<br>as follows: | Diagnostic<br>Ultrasound imaging<br>or fluid flow<br>analysis of the<br>human body as<br>follows: | Diagnostic<br>Ultrasound imaging<br>or fluid flow<br>analysis of the<br>human body as<br>follows: | Diagnostic<br>Ultrasound imaging<br>or fluid flow<br>analysis of the<br>human body as<br>follows: | Diagnostic<br>Ultrasound imaging<br>or fluid flow<br>analysis of the<br>human body as<br>follows: | Diagnostic<br>Ultrasound<br>imaging or fluid<br>flow analysis of<br>the human body<br>as follows: |
|                           | Fetal (incl                                                                                       | Fetal (incl                                                                                       | Fetal (incl                                                                                       | Fetal (incl                                                                                       | Fetal (incl                                                                                       | Fetal (incl                                                                                       |
| Indications for Use       | Obstetrics)                                                                                       | Obstetrics)                                                                                       | Obstetrics)                                                                                       | Obstetrics)                                                                                       | Obstetrics)                                                                                       | Obstetrics)                                                                                       |
|                           | Abdominal                                                                                         | Abdominal                                                                                         | Abdominal                                                                                         | Abdominal                                                                                         | Abdominal                                                                                         | Abdominal                                                                                         |
|                           | Intraoperative                                                                                    | Intraoperative                                                                                    | Intraoperative                                                                                    | Intraoperative                                                                                    | Intraoperative                                                                                    | Intraoperative                                                                                    |
|                           | Neurosurgery                                                                                      | Neurosurgery                                                                                      | Neurosurgery                                                                                      | Neurosurgery                                                                                      | Neurosurgery                                                                                      | Neurosurgery                                                                                      |
|                           | Pediatrics                                                                                        | Pediatrics                                                                                        | Pediatrics                                                                                        | Pediatrics                                                                                        | Pediatrics                                                                                        | Pediatrics                                                                                        |
|                           | Small Parts<br>(organs)                                                                           | Small Parts (organs)                                                                              | Small Parts (organs)                                                                              | Small Parts (organs)                                                                              | Small Parts (organs)                                                                              | Small Parts<br>(organs)                                                                           |
|                           | Adult Cephalic                                                                                    | Adult Cephalic                                                                                    | Adult Cephalic                                                                                    | Adult Cephalic                                                                                    | Adult Cephalic                                                                                    | Adult Cephalic                                                                                    |
|                           | Neonatal Cephalic                                                                                 | Neonatal Cephalic                                                                                 |                                                                                                   |                                                                                                   | Neonatal Cephalic                                                                                 | Neonatal Cephalic                                                                                 |
|                           | Transrectal                                                                                       | Transrectal                                                                                       | Transrectal                                                                                       | Transrectal                                                                                       | Transrectal                                                                                       | Transrectal                                                                                       |
|                           | Transvaginal                                                                                      | Transvaginal                                                                                      | Transvaginal                                                                                      | Transvaginal                                                                                      | Transvaginal                                                                                      | Transvaginal                                                                                      |
|                           | Transurethral                                                                                     | Transurethral                                                                                     | Transurethral                                                                                     | Transurethral                                                                                     | Transurethral                                                                                     | Transurethral                                                                                     |
|                           | Muskulo-skeletal<br>(conventional and<br>superficial)                                             | Muskulo-skeletal<br>(conventional and<br>superficial)                                             | Muskulo-skeletal<br>(conventional and<br>superficial)                                             | Muskulo-skeletal<br>(conventional and<br>superficial)                                             | Muskulo-skeletal<br>(conventional and<br>superficial)                                             | Muskulo-skeletal<br>(conventional and<br>superficial)                                             |
|                           | Cardiac                                                                                           | Cardiac                                                                                           | Cardiac                                                                                           | Cardiac                                                                                           | Cardiac                                                                                           | Cardiac                                                                                           |
|                           | Peripheral Vessel                                                                                 | Peripheral Vessel                                                                                 | Peripheral Vessel                                                                                 | Peripheral Vessel                                                                                 | Peripheral Vessel                                                                                 | Peripheral Vessel                                                                                 |
| Features                  |                                                                                                   |                                                                                                   | ECG (not<br>monitoring)                                                                           | ECG (not<br>monitoring)                                                                           |                                                                                                   |                                                                                                   |

{13}------------------------------------------------

1) CFM - Color Flow Mapping = Color Doppler and Amplitude (Power) Doppler.

{14}------------------------------------------------

## A brief discussion of non-clinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence.

The device has been evaluated for acoustic output, thermal, electrical, electromagnetic and mechanical safety, biocompatibility and has been found to conform with applicable medical device safety standards. The system complies with the following voluntary standards:

- Σ AIUM/NEMA UD-2, Acoustic Output Measurement Standard For Diagnostic Ultrasound Equipment
- 2 AIUM/NEMA UD-3, Real-time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- ル IEC 60601-1, Medical Electrical Equipment, Part 1: General requirements for safety
- IEC 60601-1-2, General requirements for safety, Collateral Standard, Electromagnetic ど Compatibility - Requirements and tests
- 人 IEC 60601-2-37, Particular requirements for the safety of ultrasonic diagnostic medical and monitoring equipment
- Z ISO 14971, Application of Risk Management of Medical Devices
- EN ISO 10993-1. Biocompatibility
- IEC 62304, Medical Device Software Software lifecycle processes
- IEC 62359, Ultrasonics Field characterization Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields

Thermal, electrical, electromagnetic, mechanical and biocompatibility safety is unchanged.

## Technological characteristics compared to the predicate device

The predicate devices have the same major technological characteristics as the subject device described above.

Minor differences with respect to the Ultrasound Scanner System consist of: Addition of new modes.

---

**Source:** [https://fda-staging.innolitics.com/device/K152467](https://fda-staging.innolitics.com/device/K152467)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com