uwe sun tanning beds/booths P90, iBED EVG, iSUN HP, SILVERBULLET SE, LOTUS
K152364 · Pc Marketing, Inc. · LEJ · Apr 14, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152364
Device Name
uwe sun tanning beds/booths P90, iBED EVG, iSUN HP, SILVERBULLET SE, LOTUS
Applicant
Pc Marketing, Inc.
Product Code
LEJ · General, Plastic Surgery
Decision Date
Apr 14, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4635
Device Class
Class 2
Indications for Use
Intended to provide ultraviolet light to tan the skin.
Device Story
Device consists of tanning beds (lying area and canopy) or booths (cabin and door) providing artificial UV radiation. Primary components include UV lamps, filters, reflectors, electronic control boards, and fans. Operation involves user positioning within the device for a factory-preset duration to achieve tanning via UV-A (superficial, rapid tan) and UV-B (long-term tan) radiation. Devices are used in tanning salons; operated by users or salon staff. Output is controlled UV exposure; maximum irradiance is factory-preset to comply with 21 CFR 1040.20 to prevent overexposure. Benefits include cosmetic skin tanning.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via bench testing, including electrical/thermal safety, electromagnetic compatibility, mechanical/environmental tolerance, and functional verification/validation.
Technological Characteristics
UV radiation source (UV-A/UV-B lamps), filters, reflectors, mechanical structure. Electronic components include main/relay boards and fans. Complies with IEC 60601-1:2005, IEC 60601-1-2:2007, and ISO 17491:2007. Maximum irradiance limited by factory-preset timers per 21 CFR 1040.20.
Indications for Use
Indicated for individuals seeking to tan the skin via ultraviolet light exposure.
Regulatory Classification
Identification
A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.
Special Controls
*Classification.* Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
*Labeling* —(i)*Sunlamp products.* (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
*Sunlamp products and UV lamps intended for use in sunlamp products.* Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
*Performance standard.* Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.
Predicate Devices
uwe sun tanning beds/booths (legally marketed Class I sunlamp products offered for sale on or before September 2, 2014)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of three human profiles facing right, with a design that resembles a bird or wing above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2016
PC Marketing, Inc. c/o Shepard Bentley, RAC Bentley Biomedical Consulting, LLC 28241 Crown Valley Parkway, Suite 510(k) Laguna Niguel, CA 92677
Re: K153479
Trade/Device Name: uwe sun tanning beds/booths P90, iBED EVG, iSUN HP, SILVERBULLET SE, LOTUS Regulation Number: 21 CFR 878.4635 Regulation Name: Ultraviolet lamp for tanning Regulatory Class: Class II Product Code: LEJ Dated: March 13, 2016 Received: March 17, 2016
Dear Mr. Bentley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K152364
Device Name
uwe sun tanning beds/booths P90, iBED EVG, iSUN HP, SILVERBULLET SE, LOTUS
Indications for Use (Describe) Intended to provide ultraviolet light to tan the skin.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5
## 510(k) Summary
| Name and address<br>of submitter: | PC Marketing, Inc.<br>1040 Wilt Avenue<br>Ridgefield, NJ 07657<br>Phone: (201) 943 – 6100<br>Fax: (201) 943 - 4234<br>www.pctan.com |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Shepard Bentley, RAC<br>Bentley Biomedical Consulting, LLC<br>28241 Crown Valley Parkway, Suite 510(k)<br>Laguna Niguel, CA 92677<br>Phone: (949) 374 - 9187<br>Fax: (949) 606 - 8191<br>Email: sbentley@bentleybiomed.com |
| Date of submission: | 20 August, 2015 |
| Trade name of the device: | The bundle of uwe sun tanning beds/booths consists of:<br>• P90<br>• iBED EVG<br>• iSUN HP<br>• SILVERBULLET SE<br>• LOTUS |
| Common name of the device: | uwe sun tanning beds/booths (bundled) |
| Classification reference: | 21 CFR 878.4635 |
| Product code: | LEJ |
| Predicate Devices: | uwe sun tanning beds/booths |
{4}------------------------------------------------
#### 5.1 DEVICE DESCRIPTION
#### 5.2 Intended Use
Intended to provide ultraviolet light to tan the skin.
#### 5.2.1 Device Components
The uwe sun tanning beds/booths consist of several essential parts. These parts differ between the bed and the booth models.
The beds consist of a lying area and a canopy. The booth consists of a cabin and a door. Both sunlamp products contain the same electronic components like the main board and the relay board, fans, lamps etc.
The most significant differences are the decorating design elements and the implemented features like "Body Mist" and "Aroma". These constitute the several looks of the beds and booths.
The primary technical components of a tanning device are an artificial source of UV radiation, a variety of filters and reflectors as well as a mechanical structure with a defined active surface. Different tanning results can be achieved in varying strength tanning devices. This is due to the different strength UV lamps and the different UV-A and UV-B proportions of the UV radiation.
The UV-A proportion primarily generates a superficial tan, which appears rapidly and is intensive but also fades more rapidly, the UV-B radiation is primarily responsible for more long-term tanning results.
The disadvantage of the UV-B radiation is that the tan does not become visible until one or two days after visiting the tanning salon.
Therefore, the appropriate tanning device is to be selected depending on the desired tanning goal.
{5}------------------------------------------------
#### 5.2.2 Comparison to Predicate Devices and Summary of Differences
Prior to September 2, 2014, the basic ultraviolet sunlamps named in this 510(k) existed and were offered for sales as legally marketed Class I 510(k)-exempt medical devices.
For the basis of this 510(k), PC Marketing, Inc. claims substantial equivalence to these legally marketed devices. Since none of the sun tanning booths named in this 510(k) have changed in 1) intended use or 2) technological design characteristics after September 2, 2014, essentially the contemporary sun tanning booths are identical to the predicate devices (i.e., those basic sun tanning booths offered for sale prior to the FDA's cutoff date of Sept. 2, 2014).
Such an approach of using the legally marketed Class I medical devices as predicated in the Final Reclassification Order published on June 2, 2014 in the Federal Register Vol. 79, No. 105, Page 31212, whereby it stated that:
"FDA cleared several 510(k)s for sunlamp products prior to exempting the devices from premarket notification submission. At least one 510(k) for a sunlamp product has cleared since then under product code LEJ. These cleared sunlamp products, as well as any 510(k)-exempt sunlamp product or UV lamp intended for use in a sunlamp product legally offered for sale on or before September 2, 2014, can serve as predicates for substantial equivalence purposes."
This 510(k) provides ample evidence that the subject devices all have the same intended uses and substantially equivalent technological characteristics with each other, and conform to the special controls required by the reclassification order.
In short, the sun tanning booths described within this 510(k) submission are as safe and as effective as the predicate devices, as they are unchanged from their Class I configurations.
The devices differ in position, intensity, type and number of lamps. Related to these dimensions the duration until the maximum energy intensity and hence the maximum exposure time is reached, differs. The maximum exposure time factory preset. This set time cannot be exceeded by any other component. The maximum irradiance of all devices fulfill the requirements of FDA 21 CFR 1040.20.
The devices are equivalent to the predicate devices within the intended uses. which are identical, and the technological characteristics, which are nearly identical.
#### 5.3 Discussion of Non-clinical Tests
The bundled uwe sun tanning beds/booths comply with the applicable safety and performance standards and therefore are safe and effective for their intended use. The devices have been thoroughly tested including electrical and thermal safety, electromagnetic compatibility, mechanical and environmental tolerance, environmental conditions, and functional verification and validation of specifications. The bundled uwe sun tanning beds/booths meet the requirements of the applicable FDA-recognized standards, specifically the IEC 60601-1:2005 and IEC 60601-1-2:2007 standards, and hazard analyses conforming to ISO 17491:2007 have been conducted to assure proper mitigations of risk of hazards leading to harm.
{6}------------------------------------------------
The following quality assurance measures were applied to the bundled uwe sun tanning beds/booths:
- Risk Analysis ●
- . Requirements Review
- . Design Review
- Testing on unit level (Module verification) ●
- . Integration testing (System verification)
- . Final acceptance testing (Validation)
- . Performance testing
- . Safety testing
- Environmental testing
- . Usability testing
#### Discussion of Clinical Tests 5.4
Clinical studies on the bundled uwe sun tanning beds/booths were not required in order to demonstrate the safety, effectiveness and performance of the device.
{7}------------------------------------------------
#### 5.5 Conclusions demonstrating Safety, Effectiveness and Performance:
Based on its years of experience in the marketplace of the United States of America and the information provided in this submission, PC Marketing, Inc. concludes that the bundled uwe sun tanning beds/booths are safe, effective and substantially equivalent to the predicate devices.
End of 510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.