Value Care Cuff Large Adult, Value Care Cuff Adult XL (extra long), Value Care Cuff Adult, Value Care Cuff Small Adult, Value Care Cuff Pediatric, Value Care Cuff Infant, Efficia Adult NIBP Airhose, 3 M, Efficia Adult NIBP Air Hose, 1.5 M
K152363 · Philips Medical Systems · DXQ · Oct 27, 2015 · Cardiovascular
Device Facts
Record ID
K152363
Device Name
Value Care Cuff Large Adult, Value Care Cuff Adult XL (extra long), Value Care Cuff Adult, Value Care Cuff Small Adult, Value Care Cuff Pediatric, Value Care Cuff Infant, Efficia Adult NIBP Airhose, 3 M, Efficia Adult NIBP Air Hose, 1.5 M
Applicant
Philips Medical Systems
Product Code
DXQ · Cardiovascular
Decision Date
Oct 27, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Philips multi-patient and single-patient cuffs are intended for use by, or under the supervision of, a licensed physician or other healthcare provider for the non-invasive measurement of infant, pediatric, and adult human blood pressure. Philips air hoses are intended for use by, or under the supervision of, a licensed physician or other healthcare provider for the noninvasive measurement of neonatal, pediatric, and adult human blood pressure.
Device Story
Value Care Cuffs are reusable, single-hose NBP cuffs with integrated, non-removable bladders; Efficia Air Hoses provide pneumatic connection between cuffs and automated NBP monitoring devices. Used in clinical settings by healthcare providers to facilitate non-invasive blood pressure measurement. Cuffs feature color-coded tubing for size identification; hook-and-loop closure mechanism. Device functions as a passive pneumatic interface; no internal electronics or active sensing components. Output is the transmission of pressure signals to an automated monitor for blood pressure calculation. Benefits include standardized, reliable pneumatic coupling for patient monitoring.
Clinical Evidence
Bench testing only. Performance testing demonstrated compliance with ANSI/AAMI/EC 80601-2-30 and ANSI/AAMI/EC 81060-1 standards, confirming safety and effectiveness equivalent to the predicate device.
Indicated for non-invasive blood pressure measurement in neonatal, infant, pediatric, and adult patient populations by or under the supervision of a licensed healthcare provider.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
Predicate Devices
Philips Series of Multi-Patient cuffs and Single Patient Cuffs (K071885)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, with three curved lines above them, possibly representing the flow of information or services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2015
Philips Medical Systems Ms. Mary Kruitwagen QA and Regulatory Affairs 3000 Minuteman Rd Andover, Massachusetts 03104-1099
Re: K152363
> Trade/Device Name: Value Care Cuff Large Adult, Value Care Cuff Adult Xl (extra Long), Value Care Cuff Adult, Value Care Cuff Small Adult, Value Care Cuff Pediatric, Value Care Cuff Infant, Efficia Adult NIBP Airhose, 3 M, Efficia Adult NIBP Air Hose, 1.5 M Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: August 14, 2015 Received: August 20, 2015
Dear Ms. Mary Kruitwagen,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D.Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152363
#### Device Name
Value Care Cuff Large Adult, Value Care Cuff Adult, Value Care Cuff Adult, Value Care Cuff Small Adult, Value Care Cuff Pediatric, Value Care Cuff Infant, Efficia Adult NIBP Airhose, 3 M, Efficia Adult NIBP Air Hose, 1.5 M
Indications for Use (Describe)
Philips multi-patient and single-patient cuffs are intended for use by, or under the supervision of, a licensed physician or other healthcare provider for the non-invasive measurement of infant, pediatric, and adult human blood pressure.
Philips air hoses are intended for use by, or under the supervision of, a licensed physician or other healthcare provider for the noninvasive measurement of neonatal, pediatric, and adult human blood pressure.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K152363
# Page 1 of 3
# Philips Value Care NBP Cuffs and Philips Efficia Air Hoses 510(k) Summary
| Prepared | September 18, 2015 | | |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Submission Type: | Special 510(k) | | |
| Regulatory Information<br>Classification | Blood Pressure Cuff | | |
| Common/Usual name | Blood Pressure Cuff | | |
| Device trade name | Philips Value Care<br>Philips Efficia | | |
| Proprietary Name | Value Care Cuff Large Adult<br>Value Care Cuff Adult XL (extra long)<br>Value Care Cuff Adult<br>Value Care Cuff Small Adult<br>Value Care Cuff Pediatric<br>Value Care Cuff Infant<br>Efficia Adult NIBP Air Hose, 3 M<br>Efficia Adult NIBP Air Hose, 1.5 M | | |
| Model (same as reference number)<br>(respectively to the above models) | 989803160861<br>989803160851<br>989803160841<br>989803160831<br>989803160821<br>989803160811<br>989803160891<br>989803160881 | | |
| Device Sponsor | Philips Medical Systems,<br>3000 Minuteman Rd,<br>Andover, MA, USA 01810-1099 | | |
| Device Owner | Philips Medical Systems,<br>3000 Minuteman Rd,<br>Andover, MA, USA 01810-1099 | | |
| Establishment Registration number:<br>Sponsor/Manufacturer/ Owner/Operator: | 1218950<br>1217116 | | |
{4}------------------------------------------------
K152363 Page 2 of 3
| Contact | Mary Kruitwagen,<br>Philips Medical Systems,<br>3000 Minuteman Rd,<br>Andover, MA 01810-1099 USA<br>Email: mary.kruitwagen@philips.com<br>Phone: (978) 659-4932 / Fax: 978-659-7323 |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Alternate contact | Peter Schipelliti,<br>Philips Medical Systems, 3000 Minuteman Rd,<br>Andover, MA 01810-1099 USA<br>Email: peter.schipelliti@philips.com<br>Phone: (978) 659-4744/ Fax: 978-659-7323 |
#### Device Classification:
The product code and individual classification monograph applicable to the subject devices is listed below
| Device Panel | Classification | ProCode | Description |
|----------------|----------------|---------|---------------------|
| Cardiovascular | 870.1120, II | DXQ | Blood Pressure Cuff |
### Predicate Device
| 510(k) | Date | Device | Owner/<br>Manufacturer | Trade<br>Name | Models |
|---------|------------|-----------------------------------------------------------------------|----------------------------|---------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| K071885 | 12/20/2007 | Philips Series of Multi-<br>Patient cuffs and<br>Single Patient Cuffs | Philips Medical<br>Systems | Easy<br>Care | M4552B, M4553B, M4554B, M4555B,<br>M4556B, M4557B, M4558B, M4559B,<br>M4562B, M4563B, M4564B, M4565B,<br>M4566B, M4567B, M4568B, M4569B |
## Device Description
The Value Care Cuffs are a reusable single hose NBP cuff with a non-removable bladder. The Efficia Air hoses are standard tubing with compatible connector for the Value Care cuffs. The initial release encompasses six cuffs (Large Adult, Adult Long, Adult, Small Adult, Pediatric, Infant) and two air hoses (1.5 m and 3.0 m). They have color-coded tubings to differentiate them from other brands of cuffs and allow for quick cuff size identification. The patient contacting cuffs material is polyurethane coated polyester. The air hose tubing is made of PVC.
## Fundamental Scientific Technology:
The design of the subject cuffs is similar to the predicate devices, with the integrated bladder, single hose, hook and loop closure. There are differences in material but testing of the subject devices demonstrates the same level of performance. The subject cuffs with subject air hoses demonstrated compliance with the requirements of the standards ANS/AAM//EC 80601-2-30 and ANSI/AAMI/EC 81060-1 as applicable to cuffs. The cuffs are for automated NBP devices. The fundamental scientific technology employed for the subject Value Care NBP Cuffs and Efficia Air Hoses is the same as the predicate NBP cuffs and air hoses.
The performance testing completed demonstrates that the subject devices have level of safety and effectiveness as their predicate devices. The Value Care cuffs with Efficia air hoses demonstrate substantial equivalence as the Philips Easy Care family of cuffs cleared with K071885.
#### Indications for Use
Philips multi-patient and single-patient cuffs are intended for use by, or under the supervision of, a licensed physician or
{5}------------------------------------------------
other healthcare provider for the non-invasive measurement of infant, pediatric, and adult human blood pressure.
Philips air hoses are intended for use by, or under the supervision of, a licensed physician or other healthcare provider for the noninvasive measurement of neonatal, pediatric, and adult human blood pressure.
### Conclusion
The subject Value Care Cuffs with Efficia Air hoses demonstrate the same level of performance as the predicate devices as determined from the results of the performance testing. The results support that the subject devices should perform to the same level of safety as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.