K152317 · Mederi Therapeutics, Inc. · GEI · Sep 23, 2015 · General, Plastic Surgery
Device Facts
Record ID
K152317
Device Name
Stretta Catheter
Applicant
Mederi Therapeutics, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 23, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intended for general use in the electrosurgical coagulation of tissue and intended for use specifically in the treatment of Gastroesophageal Reflux Disease (GERD).
Device Story
Stretta Catheter is a single-use monopolar electrosurgical device used for GERD treatment. Device features flexible shaft, distal multi-arm basket, and proximal control handle. Trans-oral insertion; basket expanded via internal balloon (inflated with sterile water/air). Basket arms support electrode needles that penetrate esophageal tissue to deliver radiofrequency (RF) energy from Mederi RF Generator. Thermocouples in basket arms provide temperature feedback to generator. Procedure involves sequential positioning, activation, rotation, and repositioning. Operated by physician in clinical setting. RF energy delivery induces tissue coagulation; intended to reduce GERD symptoms.
Clinical Evidence
Bench testing only. Includes in vitro functional testing (union strength, balloon/basket performance, thermocouple accuracy, system seal integrity, needle function, visual/dimensional verification) and biocompatibility testing.
Technological Characteristics
Monopolar electrosurgical catheter; RF energy source. Features flexible shaft, multi-arm basket, internal balloon, and electrode needles. Includes thermocouples for temperature monitoring. Single-use; sterile. Connects to Mederi RF Generator via custom electrical connection.
Indications for Use
Indicated for patients requiring electrosurgical coagulation of tissue, specifically for the treatment of Gastroesophageal Reflux Disease (GERD).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 23, 2015
Mederi Therapeutics, Inc. % Michele Lucey President Lakeshore Medical Device Consulting, LLC 128 Blye Hill Landing Newbury, NH 03255
Re: K152317 Trade/Device Name: Mederi Therapeutics Stretta® Catheter Regulation Number: 21 CFR§ 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 25, 2015 Received: August 28, 2015
Dear Michele Lucey,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K152317
Device Name Mederi Therapeutics Stretta® Catheter
#### Indications for Use (Describe)
Intended for general use in the electrosurgical coagulation of tissue and intended for use specifically in the treatment of Gastroesophageal Reflux Disease (GERD).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/2/Picture/14 description: The image is blurry and lacks detail, making it difficult to discern specific objects or features. The colors are muted and indistinct, with a predominance of white and light blue. Due to the lack of clarity, it is challenging to provide a more detailed description of the image's content or composition.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### MEDERI THERAPEUTICS INC
## Special
# 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 807.92(c)
Submitter Information: Submitter's Name: Mederi Therapeutics Inc. Oleg Shikhman – Chief Operating Officer Sponsor Contact : Address: 800 Connecticut Ave 1日01 Norwalk, CT 06854 Telephone: 203-9 30-9900 Fax: 203-9 39-1088 Submitter's Regulatory Contact: Michele Lucey Lakeshore Medical Device Consulting, LLC 128 Blye Hill Landing Newbury, NH 03255 603-763-3455 Date Prepared: August 15, 2015 Device Trade Name: Mederi Therapeutics Stretta® Catheter Common/Usual Name: Electrosurgical cutting and coagulation device and accessories Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulation Number: 21 CFR 878.4400 Class 2 Product Code GEI Predicate Device K000245 - Conway Stuart Stretta System Catheter
| Intended Use | |
|------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Intended for general use in the electrosurgical coagulation of<br>tissue and intended for use specifically in the treatment of<br>Gastroesophageal Reflux Disease (GERD). |
| Device Description | The Stretta® Catheter is a single use monopolar electrosurgical<br>device. The catheter connects to the Mederi RF Generator by<br>means of a custom electrical connection. The catheter consists<br>of a flexible shaft with a multi-arm basket at the distal end and<br>a control handle at the proximal end. The catheter is<br>introduced trans-orally. The catheter has depth of insertion |
| | indicators on the shaft to assist the user in inserting the device<br>to the proper location. Inside the basket is a balloon that when<br>inflated with sterile water or air to expands the basket during<br>use. The basket arms support electrode needles which contact<br>and penetrate esophageal tissue when the basket is expanded<br>and the electrode needles are advanced. The basket arms also<br>contain thermocouples to return temperature measurements to<br>the generator during activation. The balloon is inflated by<br>means of a syringe with an inline pressure relief valve. The<br>procedure, delivery of energy, occurs in several steps where<br>the catheter is positioned, activated, rotated and repositioned<br>during the course of the treatment. |
| Technological Characteristics of<br>the Device Compared to the<br>Predicate Device | The Stretta® Catheter has the same technological<br>characteristics as the predicate device, CMS Stretta Catheter.<br>Both devices are sterile, single use catheters with the same<br>intended use, principles of operation, energy type, and<br>packaging. Both devices are similar in design and<br>construction. Any differences do not introduce new questions<br>regarding safety of performance. The modifications to the<br>device include change in some materials, additional markings<br>on the catheter shaft, replacement of bonded attachments with<br>mechanical attachments, and change in location of certain<br>electrical components. |
| Principles of Operation: | The Stretta® Catheter is a single use monopolar electrosurgical<br>device. The catheter connects to the Mederi RF Generator by<br>means of a custom electrical connection. Once the catheter is<br>connected the device the user can activate the device to deliver<br>radiofrequency energy. The device may be activated multiple<br>times within the procedure to achieve the desired tissue effect. |
| Performance Data | Design verification and validation testing including <i>in vitro</i><br>functional testing (union strength, balloon/basket performance,<br>thermocouple accuracy, system seal integrity, needle function,<br>visual and dimensional verification), and biocompatibility<br>testing were conducted to confirm that the modifications do<br>not adversely affect safety and performance of the Stretta®<br>Catheter. |
| Conclusion: | Based on the performance testing and comparison to the<br>predicate device, the Stretta® Catheter is substantially<br>equivalent to the Conway Stuart Stretta System Catheter. |
{4}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.