K152291 · Andersen Sterilizers, Inc. · FRG · Dec 17, 2015 · General Hospital
Device Facts
Record ID
K152291
Device Name
Sterisheet Sterilization Wrap
Applicant
Andersen Sterilizers, Inc.
Product Code
FRG · General Hospital
Decision Date
Dec 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Sterisheet Sterilization Wraps are single use non woven sterilization wraps intended to enclose medical devices that are to be sterilized at a healthcare facility. They are used in the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.
Device Story
Sterisheet Sterilization Wraps are single-use, non-woven wraps; constructed from cellulose, polypropylene, and synthetic binders. Used in healthcare facilities to enclose medical devices (metal instruments, plastics, fabrics, endoscopes) prior to sterilization in EOGas 4 Ethylene Oxide (EO) Gas Sterilizer. Wraps allow EO penetration while maintaining microbial barrier post-sterilization. AN85/AN86 EO indicators (yellow-green to blue color change) used to secure wrap and verify EO exposure. Healthcare providers wrap devices per instructions; process in 3-hour cycle at 50°C; maintain sterility for at least 3 months. Benefits include effective sterilization and sterile storage of surgical instruments and endoscopes.
Clinical Evidence
Bench testing only. Testing included ISO 11607-1 compliance for package integrity, material compatibility, and biocompatibility. Sterilization efficacy validated using overkill method and half-cycle/half-dose methods to achieve a sterility assurance level (SAL) of 10^-6. Shelf life studies confirmed sterility maintenance for at least 3 months post-processing.
Technological Characteristics
Materials: cellulose, polypropylene, synthetic binders. Form factor: square/rectangular non-woven sheets. Principle: porous microbial barrier allowing EO gas penetration. Connectivity: none. Sterilization: intended for use in EOGas 4 EO sterilizer. Standards: ISO 11607-1.
Indications for Use
Indicated for use in healthcare facilities to enclose medical devices for sterilization in the EOGas 4 Ethylene Oxide Gas Sterilizer (3-hour cycle at 50°C).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Public Health Service
DEPARTMENT OF HEALTH & HUMAN SERVICES • USA
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000
December 17, 2015
Andersen Sterilizers, Inc. William K. Andersen, B.E., M.D. President 3154 Caroline Dr. Haw River, NC 27258
Re: K152291
Trade/Device Name: Sterisheet® Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: November 16, 2015 Received: November 19, 2015
Dear Dr. Andersen,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152291
| K152291 | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name<br>Sterisheet® Sterilization Wrap | | | | |
| Indications for Use ( <i>Describe</i> )<br>The Sterisheet Sterilization Wraps are single use non woven sterilization wraps intended to enclose medical devices that are to be<br>sterilized at a healthcare facility. They are used in the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. | | | | |
| Critical process parameters for the cycle are summarized below in Table 1.<br>Table 1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer | | | | |
| EO Amount | Temperature | Relative Humidity | EO Exposure Time | Total Cycle Time |
| $17.6 g \pm 5%$ | $50°C \pm 3°C$ | 35-90% | 3 hours | 3.5 hours |
| The product code for the Sterisheet Sterilization Wraps is listed in Table 2.<br>Table 2. Product code for the Sterisheet Sterilization Wraps | | | | |
| Product Designation | Sterisheet S88 Blue | | | |
| Product Code | 0129 | | | |
| Maximum loads of specific materials and devices that have been validated are listed in Table 3. All validated maximum loads were<br>processed without additional devices in the sterilizer. Sterility was maintained for at least 3 month after processing in an EOGas 4<br>sterilizer. | | | | |
| Table 3. Load and material types validated in the EOGas 4 Ethylene Oxide Gas Sterilizer | | | | |
| Device Type | Maximum Load | Device Examples | | Required Aeration |
| Metal | 24 lbs (11 kg) | Surgical instruments,<br>delicate sharps, including those<br>with hinges and mated surfaces | | Metal instruments do not absorb EO;<br>Follow pouch or wrap manufacturer's<br>instructions (Example: Tyvek® pouches and<br>Sterisheet® wrap require ≥ 6 hours at 50°C). |
| Plastic | 7.0 lbs (3.2 kg) | Reusable power cords, trocars | | 24 hours at 50°C; |
| Fabric | 6.1 lbs (2.8 kg) | Reusable cloth gowns, towels | | Follow manufacturer's instructions |
| Single-lumen<br>Endoscope(s) | One (1) ≥ 2.0mm ID<br>≤ 1100mm length;<br>No additional devices | Gastrovideoscopes, gastrointestinal<br>videoscopes | | 12 hours at 50°C;<br>Follow manufacturer's instructions |
| | Four (4) ≥ 1.2 mm ID<br>≤ 700 mm length;<br>No additional devices | Bronchoscopes, bronchovideoscopes,<br>cystoscopes, ureteroscopes,<br>choledocoscopes | | Follow manufacturer's instructions |
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| 5.1 | Applicant's Name and Address | | Andersen Sterilizers, Inc.<br>3154 Caroline Drive<br>Haw River, NC 27258 | |
|------|------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------|--|
| 5.2 | Contact Person | | William K. Andersen, BE, MD, FAAOS<br>President<br>Phone: 336-376-8622, Fax: 336-376-5428 | |
| 5.3 | Date of Preparation | | November 16, 2015 | |
| 5.4 | Device | Proprietary Name<br>Common Name<br>Classification<br>Product Code | Sterisheet® Sterilization Wrap<br>Sterilization Wrap<br>Class II (21 CFR 880.6850)<br>FRG | |
| ર રે | Predicate Device | | | |
| | | Device Name<br>510(k) number<br>Manufacturer | Sterisheet® Sterilization Wrap<br>K931202<br>Arjowiggins Medical Inc. | |
This 510(k) submission modifies the indications for use of the predicate device in order to include it as a component in the EOGas 4 ethylene oxide sterilization system. No modifications were made to the technology or intended use.
5.6 Device Description
The Sterisheet Sterilization Wraps are single use, non-sterilization wraps constructed from cellulose, synthetic fibers (polypropylene), and synthetic binders, with the addition of pigmentation. They are used to enclose medical devices that are to be sterilized by a healthcare provider in the 3 hour sterilization cycle at 50°C in an EOGas 4 Ethylene Oxide Gas Sterilizer. Devices are wrapped following manufacturer's instructions. After completion of the sterilization process, the Sterisheet Sterilization Wraps maintain sterility of the enclosed medical devices for at least 3 month.
The AN85/AN86 EO Indicators, when placed on the outside of the sterilization wraps, may be used to secure the wrapping material on the devices and to indicate ethylene oxide exposure, offering a convenient way to verify processing in the sterilization cycle. The color of the AN85/AN86 EO Indicators changes from yellow-green to blue after exposure to ethylene oxide.
{5}------------------------------------------------
#### 5.7 Indications for Use
The Sterisheet Sterilization Wraps are single use non-woven sterilization wraps intended to enclose medical devices that are to be sterilized at a healthcare facility. They are used in the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Critical process parameters for the cycle are summarized in Table 5-1.
Table 5-1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer
| Ethylene Oxide | Temperature | Relative Humidity | Ethylene Oxide Exposure Time | Total Cycle Time |
|----------------|-------------|-------------------|------------------------------|------------------|
| 17.6 g ± 5% | 50°C ± 3°C | 35-90% | 3 hours | 3.5 hours |
The product code for the Sterilization Wraps is listed in Table 5-2.
| Table 5-2. Product code for the Sterisheet Sterilization Wraps | | | | |
|----------------------------------------------------------------|--|--|--|--|
|----------------------------------------------------------------|--|--|--|--|
| Product Designation | Sterisheet S88 Blue |
|---------------------|---------------------|
| Product Code | 0129 |
Maximum loads of specific materials and devices that have been validated are listed in Table 5-3. All validated maximum loads were processed without additional devices in the sterilizer. Sterility was maintained for at least 3 month after processing in an EOGas 4 sterilizer.
| Device Type | Maximum Load | Device Examples | Required Aeration |
|------------------------------|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Metal | 24 lbs (11 kg) | Surgical instruments,<br>delicate sharps, including<br>those with hinges and<br>mated surfaces | Metal instruments do not<br>absorb EO;<br>Follow pouch or wrap<br>manufacturer's instructions<br>for aeration requirements<br>(Examples: Tyvek® pouches<br>and Sterisheet® wrap require<br>$\geq$ 6 hours aeration at 50°C). |
| Plastic | 7.0 lbs (3.2 kg) | Reusable power cords,<br>trocars | 24 hours at 50°C; Follow<br>manufacturer's instructions |
| Fabric | 6.1 lbs (2.8 kg) | Reusable cloth gowns,<br>towels | manufacturer's instructions |
| Single-lumen<br>Endoscope(s) | One (1)<br>$\geq$ 2.0 mm internal<br>diameter<br>$\leq$ 1100 mm length;<br>No additional devices | Gastrovideoscopes,<br>gastrointestinal videoscopes | 12 hours at 50°C; Follow<br>manufacturer's instructions |
| | Four (4)<br>$\geq$ 1.2 mm internal<br>diameter<br>$\leq$ 700 mm length;<br>No additional devices | Bronchoscopes,<br>bronchovideoscopes,<br>cystoscopes, ureteroscopes,<br>choledocoscopes | manufacturer's instructions |
Table 5-3. Load and material types validated in the EOGas 4 Ethylene Oxide Gas Sterilizer
{6}------------------------------------------------
### 5.8 Device Comparison
The technological characteristics of the Sterilization Wraps are identical to the predicate device - both are intended for the same use, use the same technology, and are designed in the same way.
The Sterisheet Sterilization Wraps differ from the predicate device in the sterilization cycle used. The difference raises no issues related to safety or effectiveness of the subject device in the sterilization cycle. A comparison between the devices is listed in Table 5-3.
| Elements | Predicate Device: Sterisheet Sterilization<br>Wraps (K931202) | Subject Device: Sterisheet<br>Sterilization Wraps |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Manufacturer | Arjowiggins Healthcare | Arjowiggins Healthcare |
| Intended Use | To enclose medical devices, allow sterilization of<br>the enclosed devices, and maintain sterility of the<br>enclosed devices | Identical |
| Indications for Use | To be used in ethylene oxide sterilization<br>systems | To be used in the 3 hour cycle<br>at 50°C in an EOGas 4<br>Ethylene Oxide Gas Sterilizer |
| Materials | Cellulose, synthetic fibers (polypropylene), and<br>synthetic binders | Identical |
| Design | Cellulose allows EO to pass through the wrap but<br>prevents microorganisms from crossing through<br>the wrap, providing a microbial barrier for the<br>wrapped devices;<br>Synthetic fibers increase mechanical resistance;<br>Synthetic binders enhance drape ability, strength,<br>softness, and fluid repellency | Identical |
| Wrap Shape | Square or rectangular | Identical |
| Configuration in<br>Load | Double sequential envelope wrap is<br>recommended | Identical |
| Shelf Life | 5 years from date of manufacture | Identical |
Table 5-3. Device Comparison
{7}------------------------------------------------
#### 5.9 Performance Testing
The Sterisheet Sterilization Wraps conform to all applicable requirements for packaging for terminally sterilized medical devices for EO sterilization, based on ISO 11607-1. Performance testing was conducted to show that the Sterisheet Sterilization Wraps perform as intended to allow sterilization and maintain sterility of the enclosed medical device. Sterilization efficacy testing demonstrated a sterility assurance level of 10 ° using the overkill method and half cycle or half dose validation methods under worst case conditions. Shelf life studies demonstrated after completion of the sterilization process, sterility is maintained for at least 3 month. The performance of the Sterisheet Sterilization Wraps is summarized in Table 5-4.
| Table 5-4. Summary of bench tests performed to demonstrate safety and effectiveness of the | | | |
|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|--|
| Sterisheet Sterilization Wraps | | | |
| Test | Description | Results | |
| Compliance to ISO 11607-1 | | | |
| Package<br>Integrity | Porous material providing a microbial barrier;<br>Physical and chemical properties are maintained | Meet requirements | |
| Material<br>Compatibility | Compatibility with respect to forming and sealing<br>processes-suitable folding and drape ability;<br>Suitability for use in EO sterilization processes and<br>cycle parameters | Meet requirements | |
| Biocompatibility | Not direct patient-contacting devices;<br>Materials are non-toxic and meet ISO 11607-1<br>requirements;<br>Biological evaluation meets acceptable criteria;<br>Bio-burden control;<br>Provides reasonable assurance for safety | Pass | |
| Shelf Life | Physical properties and microbial barrier of the<br>processed Sterisheet Sterilization Wraps were verified<br>at the end of the clamed shelf life of 5 years;<br>Stability demonstrates reasonable assurance for<br>effectiveness | Pass | |
| Performance in the EOGas 4 Cycle | | | |
| Sterilant<br>Penetration | The sterilant penetrated the pouch under worst case<br>half-dose conditions or an extreme biological<br>challenge scenario, and inactivated 6-Log biological<br>indicators | Allow a sterility assurance<br>level of $10^{-6}$ for the<br>sterilization cycle | |
| Maintenance of<br>Package<br>Integrity | Sterility was maintained for at least 3 month after<br>processing in an EOGas 4 sterilizer. | Pass | |
Based on the intended use, technological characteristics, performance data, and nonclinical tests performed, the subject Sterilization Wrap is substantially equivalent and is as safe and as effective as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.