K152270 · Korea Engineering Plastics Co., Ltd. · EBI · Nov 13, 2015 · Dental
Device Facts
Record ID
K152270
Device Name
Denture base resin
Applicant
Korea Engineering Plastics Co., Ltd.
Product Code
EBI · Dental
Decision Date
Nov 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Indications for Use
Smiletone® (denture base resin) is a thermoplastic resin that is intended to be used in fabrication of removable full and partial dental prosthesis, as well as occlusal splints and night guards. The product is based on a compounded mixture of polyamide and pigments.
Device Story
Smiletone® is a thermoplastic resin material composed of polyamide and pigments; used by dental professionals for the fabrication of removable full and partial dental prostheses, occlusal splints, and night guards. The device is supplied as a thermoplastic blank or powder (Type 3 per EN ISO 20795-1). It serves as a structural base material for dental appliances. Clinical benefit is derived from its use as a durable, biocompatible material for restorative dental applications. The device is processed in a dental laboratory setting to form the final prosthetic shape.
Clinical Evidence
Bench testing only. Performance evaluated against EN ISO 20795-1 standards, including flexural strength, water sorption/solubility, color stability, translucency, and bonding to synthetic teeth. Biocompatibility testing conducted per ISO 10993 series, including AMES test (genotoxicity), agar diffusion (cytotoxicity), acute systemic toxicity, LLNA-BrdU (sensitization), and oral mucosa irritation. All tests passed.
Technological Characteristics
Thermoplastic resin (Polyamide 12) supplied as blank or powder. Classified as Type 3 per EN ISO 20795-1. Physical properties: flexural strength 68 ± 2 MPa, flexural modulus 1389 ± 35 MPa. Conforms to EN ISO 20795-1 and ASTM F1980 (accelerated aging).
Indications for Use
Indicated for patients requiring removable full or partial dental prostheses, occlusal splints, or night guards.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing, abstract design above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 13, 2015
Korea Engineering Plastics Co., Ltd. c/o Mr. Charlie Mack Principal Engineer International Regulatory Consultants 12226 Washington Lane Parker, Arizona 85344
Re: K152270
Trade/Device Name: Smiletone® Regulation Number: 21 CFR 872.3690 Regulation Name: Denture Relining, Repairing, Rebasing Resin Regulatory Class: II Product Code: EBI Dated: August 31, 2015 Received: September 10, 2015
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Druq Administration
# Indications for Use
510(k) Number (if known) K152270
Device Name Smiletone®
#### Indications for Use (Describe)
Smiletone® (denture base resin) is a thermoplastic resin that is intended to be used in fabrication of removable full and partial dental prosthesis, as well as occlusal splints and night guards. The product is based on a compounded mixture of polyamide and pigments.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: July 1, 2015
- 1. Company and Correspondent making the submission:
Name - KOREA ENGINEERING PLASTICS Co., Ltd. Address – 216, 217, 218, B Bldg, 17, Gosan-ro 148 beon-gil, Gunpo-si, Gyeonggi-
do, 435-833, Korea
Phone : +82-31-689-3615
Fax : +82-31-8086-8285
Contact - Mrs. Song-yi Chong
Internet — http://www.kepital.com/
Primary contact and correspondent: Name: Charles Mack Address: 12226 Washington Lane, Parker, Arizona 85344 Phone" 931-625-4938 Email: charliemack@irc-us.com
- 2. Device :
| Trade/proprietary name | : Smiletone® |
|------------------------|-------------------------------------------------|
| Common Name | : Denture base resin |
| Classification Name | : Resin, Denture, Relining, Repairing, Rebasing |
### 3. Predicate Device :
| Manufacturer | : Vertex Dental B.V. |
|---------------|------------------------------------------|
| Device | : ThermoSens Rigid |
| 510(k) Number | : K123220 (Decision Date - Dec. 3. 2010) |
{5}------------------------------------------------
| Manufacturer | : Thermoplastic Comfort Systems, Inc. |
|---------------|-------------------------------------------|
| Device | : TCS® Unbreakable |
| 510(k) Number | : K053060 (Decision Date - Jul. 20. 2006) |
- 4. Classifications Names & Citations :
21CFR872.3760, EBI, Resin, Denture, Relining, Repairing, Rebasing, Class2
- 5. Description :
Smiletone® is a thermoplastic material to be used for removable full or partial dental prosthesis. The product is based on a compounded mixture of Polyamide and pigments. Smiletone® is classified as 'Type 3-thermoplastic blank or powder' by EN ISO 20795-1[2013].
The expiry date (use-by date) is for 5 years from the date of manufacture.
- 6. Indication for use :
Smiletone® (denture base resin) is a thermoplastic resin that is intended to be used in fabrication of removable full and partial dental prosthesis, as well as occlusal splints and night guards. The product is based on a compounded mixture of Polyamide and pigments.
- 7. Comparison with predicate device :
KOREA ENGINEERING PLASTICS Co., Ltd., believes that the Smiletone® is substantially equivalent to the ThermoSens Rigid of Vertex Dental B.V. and TCS® Unbreakable of Thermoplastic Comfort Systems, Inc..
The Smiletone® described in this 510(k) has the same intended use and similar technical characteristics as the ThermoSens Rigid of Vertex Dental B.V. and TCS® Unbreakable of Thermoplastic Comfort Systems, Inc..
{6}------------------------------------------------
| | Vertex<br>(ThermoSens Rigid)<br>Primary Predicate | Thermoplastic<br>Comfort Systems,<br>Inc.<br>Reference Predicate | KOREA<br>ENGINEERING<br>PLASTICS<br>(Smiltone) |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No | K123220 | K053060 | - |
| Intended use | Vertex ThermoSens<br>Rigid is intended for<br>removable full and<br>partial dentures as<br>well as splints,<br>telescope<br>constructions and<br>temporary crown and<br>bridge constructions. | Fabrication and repair<br>of removable dental<br>prosthetic devices,<br>such as full and partial<br>dentures, orthodontic<br>devices, occlusal<br>splints, and night<br>guards. | Smiletone® is a<br>thermoplastic resin that<br>is intended to be used in<br>fabrication of removable<br>full and partial dental<br>prosthesis, as well as<br>occlusal splints and night<br>guards. |
| Biocompatibility | EN ISO 10993-3, 5,<br>10, 11 | EN ISO 10993-3, 5,<br>10, 11 | EN ISO 10993-3, 5, 10,<br>11 |
| Device<br>description | Vertex ThermoSens<br>Rigid is a<br>thermoplastic material<br>that is intended to be<br>used in the fabrication<br>of removable full and<br>partial dental<br>prostheses. The<br>product is based on a<br>compounded mixture<br>of Polyamide and<br>pigments. | TCS®<br>Unbreakable is a<br>break resistant<br>material used in the<br>fabrication and repair<br>of base plates for<br>removable dental<br>prosthetic appliances<br>where superior<br>flexibility and patient<br>comfort for the lifetime<br>of the prosthetic are<br>significant concerns.<br>This includes, but not<br>to be limited to, full<br>and partial dentures,<br>orthodontic devices,<br>occlusal splints, and<br>night guards. | Smiletone® is a<br>thermoplastic resin that<br>is intended to be used in<br>the fabrication of<br>removable full and partial<br>dental prostheses. The<br>product is based on a<br>compounded mixture of<br>Polyamide and pigments. |
| Composition of<br>Materials | Polyamide 12 | Polyamide 12 | Polyamide 12 |
| Physical<br>Properties | Flexural strength<br>: Not tested by<br>Sponsor<br>Impact Strength<br>: 8.0 ± 7.0 kJ/m²<br>Flexural modulus<br>: 1339 ± 54 MPa<br>Water absorption<br>: 31.2 ± 0.8 µg/mm²<br>Water solubility<br>: -0.20 ± 0.25 µg/mm² | Flexural strength<br>: Not tested by<br>Sponsor<br>Impact Strength<br>: 8.5 ± 1.2 kJ/m²<br>Flexural modulus<br>: 353 ± 4.24 MPa<br>Water absorption<br>: 14.6 ± 0.4 µg/mm²<br>Water solubility<br>: 2.5 ± 0.7 µg/mm² | Flexural strength<br>: 68 ± 2 MPa<br>Flexural modulus<br>: 1389 ± 35 MPa<br>Water absorption<br>: 30.7 ± 0.2 µg/mm²<br>Water solubility<br>: 0.33 ± 0.1 µg/mm² |
| Standards of<br>Conformity | ISO 20795-1 | EN ISO 20795-1 | EN ISO 20795-1 |
{7}------------------------------------------------
# 8. Biocompatibility and Performance Data :
Biocompatibility and performance testing according to standard ISO 10993 Series and ISO 20795-1 were performed. All test results were satisfactory.
| Title | Test Standard | Results |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dentistry - Medical<br>devices for dentistry -<br>Materials testing | EN 1641(2010) | Pass |
| Accelerate aging test | ASTM F1980(2002) | Pass |
| Visual Inspection | Performance test<br>Guidance of dental<br>materials (KFDA<br>2014) | Pass |
| Capacity test | Performance test<br>Guidance of dental<br>materials (KFDA<br>2014) | Pass |
| Packaging(Visual<br>Inspection) test | Performance test<br>Guidance of dental<br>materials (KFDA<br>2014) | Pass |
| Surface characteristics<br>test | EN ISO 20795-1 | Pass |
| Shape characteristics<br>test | EN ISO 20795-1 | Pass |
| Color test | EN ISO 20795-1 | Pass |
| Color Stability test | EN ISO 20795-1 | Pass |
| Translucency test | EN ISO 20795-1 | Pass |
| Freedom from porosity<br>test | EN ISO 20795-1 | Pass |
| Bonding to synthetic<br>polymer teeth test | EN ISO 20795-1 | Pass |
| Residual Monomer test | EN ISO 20795-1 | Pass |
| Sorption test | EN ISO 20795-1 | Pass |
| Solubility test | EN ISO 20795-1 | Pass |
| Ultimate flexural strength<br>test | EN ISO 20795-1 | Pass |
| Title | Test Standard | Results |
| Genotoxicity test :<br>BACTERIAL REVERSE<br>MUTATION STUDY<br>(AMES) | Under the conditions of ISO 10993-3, Tests for<br>genotoxicity carcinogenicity and reproductive<br>toxicity & OECD 471 Bacterial Reverse Mutation<br>Test, the test articles should meet the test<br>requirements. | Pass |
| Cytotoxicity Test (Agar<br>diffusion test) | Under the conditions of ISO7405:2008,<br>Evaluation of biocompatibility of medical devices<br>used in dentistry, 6.2 Agar diffusion test, the test<br>articles should meet the test requirements. | Pass |
| Acute Systemic Toxicity<br>Test | Under the conditions of ISO 10993-11, 5. Acute<br>Systemic Toxicity Test, the test articles should<br>meet the test requirements. | Pass |
| Maximization Sensitization<br>Test (LLNA-BrdU) | Under the conditions of ISO 10993-10, 7.2.<br>Murine Local Lymph Node Assay (LLNA) &<br>OECD 442B :2010, Skin Sensitization Local<br>Lymph Node Assay : BrdU-ELISA, the test<br>articles should meet the test requirements. | Pass |
| Oral mucosa irritation test | Under the conditions of ISO10993-10 : 2010,<br>Test for Irritation and skin sensitization Annex B.<br>Special irritation tests B.3 Oral mucosa irritation<br>test, the test articles should meet the test<br>requirements. | Pass |
{8}------------------------------------------------
- 9. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification KOREA ENGINEERING PLASTICS Co., Ltd. concludes that The Smiletone® is substantially equivalent to predicate devices as described herein with reference to biocompatibility, Indications for use and performance data.
- 10. KOREA ENGINEERING PLASTICS Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.
END
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.