K152266 · Stryker Medical · DWJ · Dec 2, 2016 · Cardiovascular
Device Facts
Record ID
K152266
Device Name
Altrix Precision Temperature Management System
Applicant
Stryker Medical
Product Code
DWJ · Cardiovascular
Decision Date
Dec 2, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Altrix system is intended for circulating temperature controlled warm or cold water via patient contact thermal transfer devices for the application of regulating human body temperature in situations where a physician with prescription privileges determines that temperature therapy is necessary or desirable. Indications for use for the Altrix system include: a. Maintain pre-set body temperature as determined by the physician b. Maintain normal body temperature during surgical procedures c. For use in all clinical settings including coronary care units, operating, recovery and emergency departments, burn units, and medical/surgical units d. Adult and pediatric patients e. Monitoring and controlling patient temperature f. Temperature reduction in patients where clinically indicated, e.g. in hyperthermic patients
Device Story
Altrix Precision Temperature Management System regulates human body temperature by circulating temperature-controlled water through patient-contact thermal transfer devices (blankets/wraps). System inputs include user-selected temperature settings via graphical interface and real-time patient temperature data from 400-series probes. Controller uses a microprocessor to manage a 500W cartridge heater and refrigeration system, providing closed-loop feedback for automatic temperature control. Output includes regulated water flow to single or multiple circuits. Used in clinical settings (OR, ICU, etc.) by physicians/clinicians. System provides visual/audible alarms for safety limit breaches, system irregularities, or temperature deviations. Benefits include precise maintenance of normothermia or therapeutic temperature reduction in hyperthermic patients. Includes patient temperature output reference signal for third-party system integration.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including warming/cooling performance, mechanical/thermal overload, environmental testing, fluid intrusion, packaging, and disinfection validation (M. Mucogenicum). Compliance with recognized safety and performance standards (IEC/ISO/ANSI) verified.
Technological Characteristics
Microprocessor-controlled thermal regulating system. Materials: Vinyl hoses with Clik-tite/Colder connectors. Energy: 120V/60Hz AC, 500W cartridge heater, refrigeration system. Dimensions: 42.5"h x 23"d x 15"w. Connectivity: Patient temperature output reference signal for third-party systems. Sterilization: Sterile patient probes. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-10, IEC 80601-2-35, ISO 80601-2-56, ISO 10993.
Indications for Use
Indicated for adult and pediatric patients requiring body temperature regulation (warming or cooling) in clinical settings including coronary care, operating, recovery, emergency, burn, and medical/surgical units. Contraindications not specified.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 02, 2016
Stryker Medical Brian Orwat Sr. Staff Regulatory Affairs Specialist 3800 East Centre Avenue Portage, Michigan 49002
Re: K152266
Trade/Device Name: Altrix Precision Temperature Management System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ, FLL Dated: November 18, 2016 Received: November 21, 2016
Dear Brian Orwat:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrand
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152266
Device Name
Altrix Precision Temperature Management System
#### Indications for Use (Describe)
The Altrix system is intended for circulating temperature controlled warm or cold water via patient contact thermal transfer devices for the application of regulating human body temperature in situations where a physician with prescription privileges determines that temperature therapy is necessary or desirable.
Indications for use for the Altrix system include:
- a. Maintain pre-set body temperature as determined by the physician
- b. Maintain normal body temperature during surgical procedures
- c. For use in all clinical settings including coronary care units, operating, recovery and emergency departments, burn units, and medical/surgical units
- d. Adult and pediatric patients
- e. Monitoring and controlling patient temperature
- f. Temperature reduction in patients where clinically indicated, e.g. in hyperthermic patients
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: bold;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> |
escription OSC (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is in black and is set against a white background. A small registered trademark symbol is located to the upper right of the word.
# Section 5
# 510(k) Summary
# Altrix Precision Temperature Management System
# Submitter / 510(k) Holder
| Name: | Stryker Medical |
|-----------------|-----------------------------------------|
| Address: | 3800 E. Centre Ave<br>Portage, MI 49002 |
| Contact Person: | Brian L. Orwat |
| Telephone: | 269 389 6817 |
| Date Prepared: | August 10, 2015 |
# Device Name
| Proprietary Name: | Altrix Precision Temperature Management System |
|----------------------|---------------------------------------------------|
| Catalog Number: | 8001 |
| Common/Usual Name: | Hyper/Hypothermia System |
| Classification Name: | Thermal Regulating System (21 CFR 870.5900) |
| Product Code: | DWJ |
| Regulation Class: | Class II |
| Review Panel: | Cardiovascular |
| Classification Name: | Clinical Electronic Thermometer (21 CFR 880.2910) |
| Product Code: | FLL |
| Regulation Class: | Class II |
| Review Panel: | General Hospital |
# Predicate Device
Stryker Medical claims substantial equivalence to Stryker Medical (previously Gaymar Industries Inc.) Medi-Therm Hyper/Hypothermia System cleared per K100585.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "stryker" in bold, black font. The letters are all lowercase, except for the first letter, which is uppercase. There is a registered trademark symbol to the right of the word.
## Device Description
The Altrix Precision Temperature Management System components include the controller, reusable hose set(s), thermal transfer devices, patient temperature probes and reusable adaptor cable(s). The controller regulates water temperatures between 4.0° C (39.2° F) and 40.0° C (104.0° F) and circulates the heated or cooled water via hose set(s) through the thermal transfer device(s). A graphical display provides the user an interface for selecting desired water or patient temperature settings, operating modes, help menus and other key parameters. Visual indicators are displayed to inform the user of system status or when the user must confirm a setting selection. The system's water temperature and flow outputs are monitored to ensure optimal system operation.
The controller can supply water to an individual or multiple thermal transfer devices simultaneously with each of these circuits monitored separately. Three operating modes are available to facilitate patient care: Automatic, Manual and Monitor. The patient temperature probe is used with the controller to provide closed loop feedback for automatic patient temperature management and monitoring. The controller alarms activate visual and audible indications for when safety parameters are exceeded or it detects system function or performance irregularities. The Altrix system also includes the ability to provide a patient temperature output reference signal to be connected to a non-specific third party device/system.
### Indications for Use
The Altrix system is intended for circulating temperature controlled warm or cold water via patient contact thermal transfer devices for the application of regulating human body temperature in situations where a physician or clinician with prescription privileges determines that temperature therapy is necessary or desirable.
Indications for use for the Altrix system include:
- a. Maintain pre-set body temperature as determined by the physician
- b. Maintain normal body temperature during surgical procedures
- c. For use in all clinical settings including coronary care units, operating, recovery and emergency departments, burn units, and medical/surgical units
- d. Adult and pediatric patients
- e. Monitoring and controlling patient temperature
- f. Temperature reduction in patients where clinically indicated, e.g. in hyperthermic patients
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all connected, and the word is in black. There is a small circle with an "R" inside it in the upper right corner of the image. The background is white.
## Technological Characteristics and Substantial Equivalence Summary
Both Altrix and Medi-Therm share the same basic principles of operation. Altrix and Medi-Therm are temperature management systems that circulate temperature controlled water via centrifugal pump through a configurable combination of hoses connected to single or multiple thermal transfer devices which can be applied around, placed over or under a patient. Both systems use a heater to increase water temperature and a refrigeration system to decrease water temperature. Delivery and removal of thermal energy to and from the patient occurs at the contact point of the patient skin and thermal transfer device to raise or lower patient temperature. A patient temperature probe can also be used with each device to provide closed-loop feedback for patient temperature control capability.
| Category | Subject Device: Altrix Precision<br>Temperature Management System | Predicate Device: Medi-Therm<br>Hyper/Hypothermia System | Comparison |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k) | | K100585 | |
| | General | | |
| Indications for<br>Use | The Altrix system is intended for<br>circulating temperature controlled warm<br>or cold water via patient contact thermal<br>transfer devices for the application of<br>regulating human body temperature in<br>situations where a physician or clinician<br>with prescription privileges determines<br>that temperature therapy is necessary or<br>desirable.<br>a. Maintain pre-set body temperature<br>as determined by the physician<br>b. Maintain normal body temperature<br>during surgical procedures<br>c. For use in all clinical settings<br>including coronary care units,<br>operating, recovery and emergency<br>departments, burn units, and<br>medical/surgical units<br>d. Adult and pediatric patients<br>e. Monitoring and controlling patient<br>temperature<br>f. Temperature reduction in patients<br>where clinically indicated, e.g. in<br>hyperthermic patients | The Medi-Therm is intended for use in<br>supplying warm or cold water at<br>controlled temperatures via water<br>circulating blankets or body wraps for<br>the application of regulating patient<br>temperature in situations where a<br>physician determines that temperature<br>therapy is necessary and desirable.<br>a. To maintain pre-set body temperature<br>as determined by the physician<br>b. To maintain normal body temperature<br>during surgical procedures<br>c. For use in all hospital areas including<br>invasive and coronary care units, in<br>operating, recovery and emergency<br>rooms, in burn units and on<br>medical/surgical floors<br>d. This system can be used with adult<br>and pediatric patients<br>e. Monitoring and controlling patient<br>temperature<br>f. Temperature reduction in patients<br>where clinically indicated, e.g. in<br>hyperthermic patients | Same |
| Anatomical<br>Site | Trunk; upper and lower extremities | Trunk; upper and lower extremities | Same |
| Sterility | Accessory patient probes only | Non-sterile accessories | Different |
| Rx Only | Yes | Yes | Same |
| Voltage -<br>AC/hertz | 120/60 | 120/60 | Same |
| Current (A) | 12 | 11.5 | Similar |
| Heat Source | 500W Cartridge heater | 500W Cartridge heater | Same |
| Circulating<br>Fluid | Water | Water | Same |
| Flow Rate | 1.2 L/Min (minimum) | 1 L/Min (minimum) | Similar |
| Reservoir<br>Volume | 5.0 L | 9.5 L | Different |
| Outlet Ports | 3 | 2 | Different |
| Control | Microprocessor | Microprocessor | Same |
| Modes | Automatic, Manual, Monitor | Automatic, Manual, Monitor | Same |
| Automatic<br>Mode<br>Warming<br>Rates | Min, Med, Max, Custom | Gradual, Moderate, Rapid | Similar |
| Automatic<br>Mode Cooling<br>Rates | Min, Med, Max | Gradual, Moderate, Rapid | Same |
| Setting Range<br>- Water<br>Temperature | 4.0 - 40.0° C | 4 - 42° C | Different |
| Controller<br>Accuracy -<br>Water<br>Temperature | $\pm$ 0.3° C | $\pm$ 0.8° C | Different |
| Display<br>Accuracy -<br>Water<br>Temperature | $\pm$ 0.2° C | $\pm$ 0.3° C | Different |
| Settings Range<br>-Patient<br>Temperature | 32.0 - 38.0° C | 30 - 41° C | Similar |
| Controller<br>Accuracy -<br>Patient<br>Temperature | $\pm$ 0.3°C (25.0° - 45.0° C) | $\pm$ 0.5° C | Different |
| Patient<br>Temperature<br>Display<br>Accuracy | $\pm$ 0.3° ( 25.0° - 45° C) | $\pm$ 0.3° C | Same |
| Patient<br>Temperature<br>Output<br>Capability | Yes | No | Different |
| Controller<br>Dimensions | 42.5" h x 23" d x 15" w | 37"h x18.75"d x14"w | Similar |
| Weight<br>(without<br>water) | 150 lbs | 141 lbs | Similar |
| <b>SAFETY</b> | | | |
| Patient Probe<br>Alarm | Yes | Yes | Same |
| Patient<br>Temperature<br>Deviation<br>Alarm | Yes | No | Different |
| Normothermia<br>Deviation<br>Alarm | Yes | No | Different |
| High Water<br>Temp Limits<br>(software) | $>42.5° C$ | $43° C$ | Similar |
| Water<br>Temperature<br>Deviation<br>Alarm | Yes; ±0.8° C | No | Different |
| System Self-<br>Test at Power-<br>up | Yes | Yes | Same |
| Water Level<br>Alarm | Yes | Yes | Same |
| Water Flow<br>Alarm | Yes (for each water circuit) | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.