K152261 · Nostix, LLC · LJS · Oct 21, 2015 · General Hospital
Device Facts
Record ID
K152261
Device Name
PICC Tip Positioning Aid
Applicant
Nostix, LLC
Product Code
LJS · General Hospital
Decision Date
Oct 21, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
The PICC Tip Positioning Aid is intended to provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava.
Device Story
Device provides real-time PICC tip location during insertion. Inputs: patient cardiac electrical activity via ECG cables/electrodes. Operation: monitors P-wave changes as catheter approaches right atrium via superior vena cava. Output: real-time tip location information displayed on monitor. Used in clinical settings by clinicians. Replaces traditional chest x-ray or fluoroscopy for tip confirmation. Benefits: reduces radiation exposure and procedural time. System includes monitor, software, battery, ECG patient cable, remote control cable, probe cover, and ECG clip cable.
Clinical Evidence
Bench testing only. Device relies on simulated use/human factors testing previously conducted for the predicate Celerity System to validate the alternative method to chest x-ray/fluoroscopy. Compliance with IEC 60601-1 and IEC 60601-1-2 standards confirmed.
Technological Characteristics
Standalone monitor with software, battery, and power cord. Includes ECG patient cable, remote control cable, probe cover, and ECG clip cable. Complies with IEC 60601-1 (General safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for positioning Peripherally Inserted Central Catheters (PICC) in adult patients. Provides real-time tip location via cardiac electrical activity as an alternative to chest x-ray or fluoroscopy. Contraindicated/limited for patients with atrial fibrillation, atrial flutter, severe tachycardia, pacemaker-driven rhythm, or COPD; these patients require additional confirmation methods.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and clean, and it is easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 21, 2015
Nostix LLC c/o Mr. Jim Lewis Salus Ventures LLC 5335 Holmes Place Boulder, Colorado, 80303
Re: K152261
Trade/Device Name: PICC Tip Positioning Aid Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, implanted, long-term intravascular catheter Regulatory Class: II Product Code: LJS Dated: August 7, 2015 Received: August 11, 2015
Dear Mr. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K152261
Device Name
PICC Tip Positioning Aid
Indications for Use (Describe)
The PICC Tip Positioning Aid is indicated for the postioning of Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location by using the patient's cardiac electrical activity. The PICC Tip Positioning Aid is indicated for use as an alternative method to chest x-ray or fluoroscopy confirmation of PICC tip placement in adult patients.
Note: In general, devices that utilize ECG technique to observe P-Wave arc limited, but not contraindicated, for patients where cardiac rhythms may change presentation of the P-Wave; including
- Atrial fibrillation
- Atrial flutter
- Severe tachycardia
- Pacemaker-driven rhythm
- Chronic obstructive pulmonary disease (COPD)
Such patients are easily identified prior to PICC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☑</span> |
|----------------------------------------------|--------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span>☐</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Nostix" in blue font with a gray shadow. The word is written in a serif font. The shadow is slightly offset from the word, giving it a 3D effect.
# 510(k) Summary
Prepared: 25 September 2015
## Submitter
| Company | Nostix LLC<br>5541 Central Av, Suite 170<br>Boulder, CO 80301 |
|---------|-----------------------------------------------------------------|
| Tel | 303 245 8895 |
| Fax | 303 245 8909 |
| Contact | Pete Nelson<br>Director of Engineering<br>petenelson@nostix.com |
| Device | |
| Trade Name | PICC Tip Positioning Aid |
|--------------|-----------------------------------------------------------|
| Common Name | PICC placement accessory |
| Class Name | Percutaneous, implanted, long-term intravascular catheter |
| Product Code | LJS |
| Regulation | 21 CFR 880.5970 |
| Class | 2 |
# Predicate
| Trade Name | Celerity System |
|--------------|-----------------------------------------------------------|
| Clearance | K142889, 27 January 2015 |
| Common Name | PICC placement accessory |
| Class Name | Percutaneous, implanted, long-term intravascular catheter |
| Product Code | LJS |
| Regulation | 21 CFR 880.5970 |
| Class | 2 |
{4}------------------------------------------------
### Device Description
The PICC Tip Positioning Aid includes a standalone Monitor containing software, battery and power cord accompanied by an ECG Patient Cable, a Remote Control Cable, probe cover and ECG Clip Cable.
Other procedural accessories; including ECG Snap Leads, ECG Surface Electrodes, Cable Cover, Gloves and Prep Pads; may be provided as a convenience for the clinician but are not in the scope of this submission.
#### Intended Use
The PICC Tip Positioning Aid is intended to provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava.
#### Indications for Use
The PICC Tip Positioning Aid is indicated for the positioning of Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location information by using the patient's cardiac electrical activity. The PICC Tip Positioning Aid is indicated for use as an alternative method to chest x-ray or fluoroscopy confirmation of PICC tip placement in adult patients.
Note: In general, devices that utilize ECG technique to observe P-Wave arc limited, but not contraindicated, for patients where cardiac rhythms may change presentation of the P-Wave; including
- Atrial fibrillation 트
- Atrial flutter
- I Severe tachycardia
- Pacemaker-driven rhythm
- I Chronic obstructive pulmonary disease (COPD)
Such patients are easily identified prior to PICC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location.
#### Technological Characteristics
Technological Characteristics of the subject PICC Tip Positioning Aid are identical to the predicate device. The name and address changes on the labels between the predicate and proposed devices do not raise new technological questions.
#### Performance Data
As the only differences between the device and its predicate are names, logos, and addresses in the labeling, the following recognized standards from the IEC 60601 (3ra Edition) series continue to be satisfied.
- IEC 60601-1-1 Medical electrical equipment—Part 1-1: General requirements for safety-Collateral standard: Safety requirements for medical electrical systems
{5}------------------------------------------------
- IEC 60601-1-2 Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility - Requirements and tests
Human Factors Evaluation - Simulated Use Testing alternate to chest x-ray and fluoroscopy:
Simulated Use / Human Factors Testing has been conducted to evaluate the application of the PICC Tip Placement Aid as embodied in the predicate Celerity System when used to confirm PICC tip position as an alternative method to chest X-ray or fluoroscopy confirmation. The use related events noted in the studies have been adequately reviewed and addressed in order to ensure the appropriate use of the device as an alternate to x-ray techniques for confirmation of tip location of PICC.
Based on the content of the proposed PICC Tip Positioning Aid's Risk Analysis / Use and Design FMEAs, and the content of the Instructions for Use, the PICC Tip Positioning Aid has demonstrated its suitability for its intended purpose.
## Substantial Equivalence Conclusion
As this device design and manufacturing are the same as the predicate except for name changes to the device and the manufacturer, the device is clearly equivalent to its predicate. The proposed device is clearly substantially equivalent to the predicate device based on identical:
- 트 Intended Use
- 트 Indications for Use
- ' Design
- 트 Production
- I Operating principles, characteristics, and user interface
- 트 Technology and specifications
- I Labeling
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.