Ebony PTA .014 RX Pheripheral Dilatation Catheter

K152230 · Natec Medical · LIT · Sep 3, 2015 · Cardiovascular

Device Facts

Record IDK152230
Device NameEbony PTA .014 RX Pheripheral Dilatation Catheter
ApplicantNatec Medical
Product CodeLIT · Cardiovascular
Decision DateSep 3, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Ebony® PTA 0.014" RX Peripheral Dilatation Catheter is intended for dilatation of lesions in the iliac, femoral, popliteal, infra popliteal, and renal arteries. This catheter is not intended for use in coronary or neuro-vasculature.

Device Story

Rapid exchange (RX) peripheral transluminal angioplasty (PTA) catheter; features semi-compliant inflatable balloon and tapered distal tip. Includes two radiopaque markers for positioning and hydrophilic coating for navigation. Used by physicians in clinical settings to treat arterial stenosis via balloon dilatation. Device is advanced through vasculature to lesion site; balloon is inflated to dilate vessel. Benefits patient by restoring blood flow in peripheral arteries. Modification involves change in hydrophilic coating material from Slipskin to Proglide.

Clinical Evidence

Bench testing only. Testing included coating integrity, particulate evaluation, and biocompatibility assessments to verify performance of the modified hydrophilic coating.

Technological Characteristics

Rapid exchange (RX) PTA catheter; semi-compliant inflatable balloon; tapered distal tip; two radiopaque markers; hydrophilic coating (Proglide).

Indications for Use

Indicated for dilatation of lesions in iliac, femoral, popliteal, infra popliteal, and renal arteries. Contraindicated for coronary or neuro-vasculature use.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center of the logo is an abstract symbol of a human figure, represented by three overlapping profiles. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 3, 2015 Natec Medical, Ltd Xavier de Buchere Regulatory Affairs and Quality Manager Maeva Centre Building Silicon Avenue Ebene Business Park Reduit 72201 - Mauritius Re: K152230 Trade/Device Name: Ebony PTA .014" RX Peripheral Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: July 25, 2015 Received: August 7, 2015 Dear Mr. de Buchere: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {1}------------------------------------------------ CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K152230 Device Name Ebony® PTA 0.014" RX Peripheral Dilatation Catheter Indications for Use (Describe) The Ebony® PTA 0.014" RX Peripheral Dilatation Catheter is intended for dilatation of lesions in the iliac, femoral, popliteal, infra popliteal, and renal arteries. This catheter is not intended for use in coronary or neuro-vasculature. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ ## Ebony® PTA 0.014'' RX Peripheral Dilatation Catheter 510(k) SUMMARY 21 CFR 807.92 ### Applicant: | Company Name: | Natec Medical Ltd. | |------------------|-----------------------------------------------------------------------------------------------| | Company Address: | Maeva Centre Building<br>Silicon Avenue,<br>Ebene Business Park<br>Reduit 72201,<br>Mauritius | | Telephone: | +230 466 30 54 | | Fax: | +230 466 67 70 | | Contact Person: | Xavier De Buchere<br>Regulatory Affairs & Quality Manager<br>xdbuchere@natec-medical.com | # Summary Preparation Date: August 25th, 2015 ### Device Name: | Trade Name: | Ebony® PTA 0.014" RX Peripheral Dilatation Catheter | |----------------------|-----------------------------------------------------| | Common/Usual Name: | PTA Catheter | | Classification Name: | Percutaneous Transluminal Angioplasty Catheter | | Regulation Number: | 21 CFR 870.1250 | | Product Code: | LIT | | Device Class: | Class II | ### Predicate Devices: - Ebony® PTA 0.014" RX Peripheral Dilatation Catheter (K112513) - Ebony® PTA 0.014" RX Peripheral Dilatation Catheter (K142459) ## Device Description: The EBONY® PTA 0.014" RX Dilatation Catheter is a standard rapid exchange (RX) PTA catheter with a semi-compliant inflatable balloon and tapered tip on the distal end. Two radiopaque markers facilitate proper balloon positioning across the stenosis, and a hydrophilic coating facilitates advancement of the catheter through the vasculature. The modified EBONY® PTA 0.014" RX Dilatation Catheter has a new hydrophilic coating, applicable to all lengths to the currently available sizes. {4}------------------------------------------------ ## Indication for Use: The Ebony® PTA 0.014" RX Catheter is intended for dilatation of lesions in the iliac, femoral, popliteal, infra popliteal, and renal arteries. This catheter is not for use in coronary or neuro-vasculature. ## Technological Characteristics: The modified Ebony® PTA 0.014" RX Peripheral Dilatation Catheter has the same indication for use and is manufactured using the same rapid exchange design and materials as the Ebony® PTA 0.014" RX Peripheral Dilatation Catheter cleared under K112513 and K142459. The hydrophilic coating on the Ebony® PTA 0.014" RX Peripheral Dilatation Catheter was modified from Slipskin to Proglide. ## Performance Data: Based on a risk analysis, the following bench testing was performed on the modified Ebony PTA .014" RX Catheter or leveraged from similar catheters: - I Coating Integrity test - 트 Particulate evaluation - I Biocompatibility ## Conclusion The modified Ebony® 0.014" RX Peripheral Dilatation Catheters and the predicate Ebony 0.014" RX Peripheral Dilatation Catheter (K112513) & (K142459) have the same intended use, same materials and the same RX design. The modified catheters met all of the verification/validation acceptance criteria and demonstrated that the modified Ebony® 0.014" RX Peripheral Dilatation Catheters are substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...