To enhance fluid flow into and along the vascular and lymphatic vessels. As a treatment for venous and lymphatic disorders. As an effective in the treatment of the following clinical conditions, when combined with an individualized monitoring programme: Lymphedema. Primary and secondary (including post-surgery, radio or chemotherapy). Chronic Edema of Venous Origin Chronic venous insufficiency. Acute and chronic wounds including venous leg ulcers and post- surgical wounds. IPC may also be beneficial in the management of: Lipoedema. Varicose veins Postoperative Venous Ligation or Stripping Sports injuries Post Traumatic Edema . Lymphatic Filariasis
Device Story
Hydroven 12 system consists of air pump and inflatable limb garments; used for intermittent pneumatic compression (IPC) therapy. Pump delivers compressed air to garment chambers in sequential pattern; pressure promotes fluid/protein transport in vascular/lymphatic vessels. Garments feature overlapping air chambers; zipper closures; optional inserts for large limbs. Used in clinical settings; operated by healthcare providers. Provider selects pressure and sequence settings via front panel display. System alerts user to faults. Therapy compensates for compromised lymphatic drainage by squeezing edema fluid toward regions with normal drainage. Benefits include enhanced fluid flow and management of venous/lymphatic conditions.
System: AC-powered air pump, motor-driven rotary air valve, microcontroller-based control. Garments: 12 overlapping air bladders, RF-welded polyurethane sheet, woven nylon/PU laminate covers, zipper closure. Standards: ANSI/AAMI ES60601-1:2005/(R)2012, IEC 60601-1-2:2007. Connectivity: Standalone. Sterilization: Not specified.
Indications for Use
Indicated for patients with venous and lymphatic disorders, including lymphedema (primary/secondary), chronic edema of venous origin, chronic venous insufficiency, and acute/chronic wounds (venous leg ulcers, post-surgical). Also for management of lipoedema, varicose veins, post-operative venous ligation/stripping, sports injuries, post-traumatic edema, and lymphatic filariasis.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
Predicate Devices
Lympha Press Plus and compression Garments (K013331)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/13 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 04, 2016
Huntleigh Healthcare Ltd Greg Baily Technical Director 35 Portmanmoor Road Cardiff, CF24 5HN United Kingdom
Re: K15228
Trade/Device Name: Hydroven 12 Pump and Garments Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: February 22, 2016 Received: February 24, 2016
Dear Greg Baily:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrenner
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### 510(k) Number: K152228
Device Name: Hydroven 12 Pump and Garments.
### Indications for Use:
To enhance fluid flow into and along the vascular and lymphatic vessels. As a treatment for venous and lymphatic disorders.
As an effective in the treatment of the following clinical conditions, when combined with an individualized monitoring programme:
- Lymphedema.
- · Primary and secondary (including post-surgery, radio or chemotherapy).
- . Chronic Edema of Venous Origin
- Chronic venous insufficiency. .
- Acute and chronic wounds including venous leg ulcers and post- surgical wounds. .
IPC may also be beneficial in the management of:
- . Lipoedema.
- Varicose veins .
- Postoperative Venous Ligation or Stripping .
- . Sports injuries
- Post Traumatic Edema .
- . Lymphatic Filariasis
Prescription Use YES (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use NO (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## HUNTLEIGH
## 510K Summary Hydroven 12 Pump and Garments
| Name & Address: | Huntleigh Healthcare Ltd<br>Diagnostic Products Division<br>35 Portmanmoor Road<br>Cardiff, UK<br>CF24 5HN | | |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|---------------------------|
| Telephone: | +(44) 2920 485885 | | |
| Fax: | +(44) 2920 447142 | | |
| Prepared: | July 2015 | | |
| Contact: | Greg Baily - Technical Director | | |
| Device Name: | Hydroven 12 Pump and Garments | | |
| Common Name | Compressible Limb Sleeve | | |
| Classification | Class | Product Code | Classification Regulation |
| | II | JOW | 870.5800 |
| Classification Name: | Sleeve, Limb, Compressible | | |
| Predicate Device: | Lympha Press Plus and compression Garments (K013331) manufactured by Mego Afek AC Ltd. Kibbutz Afek, 30042, Israel | | |
| Indications for Use: | <b>Indications for Use:</b><br>To enhance fluid flow into and along the vascular and lymphatic vessels. As a treatment for venous and lymphatic disorders. As an effective in the treatment of the following clinical conditions, when combined with an individualized monitoring programme: | | |
| | Lymphedema. Primary and secondary (including post-surgery, radio or chemotherapy). Chronic Edema of Venous Origin Chronic venous insufficiency. Acute and chronic wounds including venous leg ulcers and post-surgical wounds. | | |
| | IPC may also be beneficial in the management of: Lipoedema. Varicose veins Postoperative Venous Ligation or Stripping Sports injuries Post Traumatic Edema Lymphatic Filariasis | | |
| Description : | The Hydroven 12 system comprises of a range of inflatable sleeves. | | |
{4}------------------------------------------------
inserts and a compatible air pump.
The sleeves fit around patients' upper and lower limbs. The sleeves contain separate but overlapping air chambers that intermittently compress the limbs in a specific sequence.
The resultant pressure on the patient's limb tissues promotes the transport of fluids and proteins.
The inflatable inserts can be added to allow large circumference limbs to be treated.
The pump provides the compressed air to the sleeves with user selected pressures and sequence of operation.
| Models: | |
|---------|--|
|---------|--|
| Model REF | Device | Features |
|----------------------------------------------------------------------------------------------------|-----------------------|-------------------------------------------------------|
| 316003US | Hydroven 12 | AC powered pump |
| 316A68, 316A78,<br>316AI68, 316AI78,<br>316L76S, 316L76W,<br>316L84S, 316L84W,<br>315LI76, 316LI84 | Hydroven 12 garments. | Sequential inflation providing active<br>compression. |
Comparison of Intermittent Pneumatic Compression Therapy is the technological Technological principle for both the subject and predicate devices. It is based on a Characteristics with pump and pneumatic garment combination. Whereby a garment with Predicate Device multiple internal air chambers is applied to the limb requiring therapy, the pump supplies air to inflate and deflate the garment chambers in sequence which compresses the limb. Applying a sequential compression to a limb takes over the permanently missing function of compromised lymphatics by squeezing edema tissue fluid to regions with normal lymphatic drainage. The subject and predicate are based on the following same technological elements: · An internal air source (air compressor) • An electronic pressure control system • A valve system to route air to the appropriate garment chamber • A multiway pneumatic 'garment to pump' connection system • PU Coated Nylon Garments • Limb Garments with Multiple internal chambers • Limb Garments with Zipper closures • Front panel display indicating device settings The following technological difference exist between the subject and predicate devices · Subject uses low voltage compressor - predicate uses AC line powered compressor • Subject uses motor driven rotary air valve – predicate uses a series of solenoid driven air valves
- · Subject uses multiple LED displays predicate uses a single LCD display
{5}------------------------------------------------
# HUNTLEIGH
Substantial Equivalence: The Hydroven 12 Pump and Garments is substantially equivalent to cleared device Lympha Press Plus and compression garments (K013331). The Hydroven 12 pump provides equivalent pressure profiles. The Hydroven 12 garments have similar construction and deliver substantially equivalent therapy.
| Testing to demonstrate equivalence included: | |
|----------------------------------------------|--|
| Testing to demonstrate equivalence included: | |
|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Testing conducted | Result |
| Full validation of pump software / hardware functionality, including<br>- User Interface<br>- Therapy delivery | Passed |
| Performance testing garments - Pressure cyclic test.<br>with Leg garments<br>with Arm garments | Passed |
| Electrical Testing to Standard ANSI/AAMI<br>ES60601-1:2005/(R)2012, C1:2009/(R)2012 and<br>A2:2010/(R)2012 | Complies with<br>Standard |
| EMC testing to Standard IEC 60601-1-2, 2007 | Complies with<br>Standard |
| Environmental Stability testing.<br>-Storage / Distribution Test.<br>-Operational Temperature /Humidity Test. | Passed |
| Technologies Summary: | The Hydroven 12 pump consists of an air compressor, distribution valve and a microcontroller based control and monitoring system, all housed in a portable enclosure.<br>Options of pressure settings and cycle sequences can be selected by the user.<br>The system is powered by standard 60Hz AC line supply. The monitoring system can alert the user to faults. |
| | The Hydroven 12 garments comprise of 12 separate air bladders constructed from RF welded polyurethane sheet material.<br>The cells are located in between the inner and outer covers, which are made from woven nylon/PU laminate materials.<br>The sleeve is fastened around the patient's limb by a zipper.<br>The cells overlap, in order to apply a continuous area of force along the limb. |
| Conclusion: | The data detailed within this submission demonstrates that the device is substantially equivalent to the legally marketed predicate device, identified in this summary. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.