K152219 · Gyrus Acmi, Inc. · GEI · Dec 17, 2015 · General, Plastic Surgery
Device Facts
Record ID
K152219
Device Name
PK AIM
Applicant
Gyrus Acmi, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PK AIM device is intended for monopolar cutting as well as grasping, manipulating and bipolar coagulation of selected soft tissues during electrosurgery. This device is not intended to be used for tubal ligation or female sterilization.
Device Story
The PK AIM is a 2-in-1 electrosurgical instrument featuring a pencil-style handle that integrates monopolar cutting and bipolar coagulation capabilities. It connects to the Olympus ESG-400 generator, which automatically recognizes the device and restricts settings to those specified for the PK AIM. The surgeon operates the device using handle-mounted buttons for activation or an optional foot pedal. A sliding toggle switch on the handle allows the user to switch between forceps and pencil modes. The device is used in general surgical procedures to grasp, manipulate, and treat soft tissue. By combining two modalities into one instrument, the device allows surgeons to perform multiple tasks without switching tools, potentially improving surgical efficiency and reducing procedure time. The healthcare provider uses the device output—monopolar or bipolar energy—to achieve precise tissue cutting or coagulation based on visual feedback at the surgical site.
Clinical Evidence
Bench testing only. Performance testing included dimensional measurements, mechanical force testing (closing, button, slider, grasping), load testing (tensile, compressive, vertical, lateral), cable tensile strength, and electrical functionality. Biocompatibility testing (cytotoxicity, intracutaneous injection, irritation, pyrogenicity) and packaging integrity (ISTA 2A) were performed. Ex-vivo bench testing compared cutting and coagulation performance against predicates, showing substantially equivalent results.
Technological Characteristics
Materials: plastics, metals, heatshrink, epoxy, stainless steel electrodes. Energy: Monopolar and bipolar electrosurgery. Form factor: Pencil-style handle with sliding toggle switch. Connectivity: Wired connection to Olympus ESG-400 generator. Sterilization: Ethylene oxide (EtO), 10^-6 sterility assurance level. Standards: ISO 10993-1/5/7/10, ISO 14971, ISO 15223-1, ANSI/AAMI/ISO 11607-1, ANSI/AAMI/ISO 11135-1, IEC 60601-1, IEC 60601-2-2.
Indications for Use
Indicated for monopolar cutting, grasping, manipulating, and bipolar coagulation of soft tissues during electrosurgery. Not for tubal ligation or female sterilization.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Gyrus ACMI Incorporated Mr. Neil Kelly Senior Regulatory Affairs Specialist 136 Turnpike Road Southborough, Massachusetts 01772 December 17, 2015
Re: K152219
Trade/Device Name: PK AIM Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 16, 2015 Received: November 18, 2015
Dear Mr. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
|-----------------------------------------|--|
| Food and Drug Administration | |
# Indications for Use
510(k) Number (if known) TBD
152219
Device Name PK AIM
#### Indications for Use (Describe)
The PK AIM device is intended for monopolar cutting as well as grasping, manipulating and bipolar coagulation of selected soft tissues during electrosurgery. This device is not intended to be used for tubal ligation or female sterilization.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) Notification July 2015
# 510(k) Summary of Safety and Effectiveness
Gyrus ACMI, Inc.
# Olympus PK AIM
## General Information
Manufacturer:
Gyrus ACMI, Inc. 9600 Louisiana Ave North Brooklyn Park, MN 55445 Phone: 508-804-2690
Establishment Registration Number:
510(k) Submitter:
3011050570
Gyrus ACMI, Inc. 136 Turnpike Rd. Southborough, MA 01772-2104
Establishment Registration Number:
Contact Person:
Date Prepared:
# Device Description
Classification Name:
Regulation number Product code Regulatory class Review Panel
Trade Name:
Generic/Common Name:
3003790304
Neil Kelly Senior Regulatory Affairs Specialist 508-804-2690 Neil.kelly@olympus-osta.com
July 2015
Electrosurgical cutting and coagulation device and accessories 21 CFR 878.4400 GEI Class II General and Plastic Surgery
Olympus - PK AIM
Electrosurgical cutting and coagulation device
K152219 Page 1 of 5
{4}------------------------------------------------
Traditional 510(k) Notification July 2015
## Predicate Devices
| K813071 | ValleyLab/Covidien Electrosurgical Pencil (monopolar pencil) |
|---------|--------------------------------------------------------------|
| K093108 | Stryker Bipolar Forceps (bipolar forceps) |
| K063195 | ValleyLab LigaSure Advance (dual plugs) |
#### Comparison to Predicate Devices:
The Olympus - PK AIM has been compared to the above mentioned predicates with respect to intended use, design, and fundamental scientific technology. The comparisons and testing results presented in this 510(k) notification to the FDA show this device to be substantially equivalent to predicate devices and raises no new concerns or safety or effectiveness.
#### Product Description
This 2 in 1 device has a pencil type handle and combines the technologies of a monopolar pencil and a bipolar forceps. The device has buttons that allow hand activation and a sliding toggle switch to allow the surgeon to switch between a forceps, and a pencil device. Foot pedals connected to the generator are also available to allow for foot pedal activation of the device. The proposed device plugs into the Olympus ESG-400 generator (K141225). The generator and device make up a medical electrical system. The instrument is to be used only with the Olympus ESG-400 Generator.
The proposed device is comprised of a mixture of plastics, metals, heatshrink and epoxy. The device has passed all applicable biocompatibility testing and additional information is provided within this submission for any patient contacting materials that contain colorants.
#### Technological Characteristics
The proposed PK AIM uses monopolar energy in order to cut and bipolar energy in order to coagulate soft tissue in general surgical procedures. For safety and convenience the ESG-400 generator recognizes the proposed PK AIM when it is connected and limits generator settings to those dedicated for use with the proposed PK AIM device, which are called out in the IFU.
The PK AIM can be activated using buttons located on the device handle, or via a foot pedals which are sold separately. The hand activation allows the physician to activate either cut or coagulation (coag) mode without taking their eyes off the surgical site. A sliding toggle switch located on the handle allows the user to switch between using the device as a pencil or forceps.
{5}------------------------------------------------
Traditional 510(k) Notification July 2015
### Indications for Use
The PK AIM device is intended for monopolar cutting as well as grasping, manipulating and bipolar coagulation of selected soft tissues during electrosurgery. This device is not intended to be used for tubal ligation or female sterilization.
#### Compliance to Voluntary Standards
The design of the proposed device complies with the following standards:
ISO 10993-1, 2009 ISO 10993-5. 2009 ISO 10993-7 2008 ISO 10993-10, 2010 ANSI/AAMI/ISO 11607-1, 2006 ANSI/AAMI/ISO 11135-1, 2007 ISO 14971, 2007 ISO 15223-1; 2012 IEC 60601-1: 2005 IEC 60601-2-2: 2009
#### Summary of Sterilization and Shelf Life Discussion
The PK AIM is delivered in a sterile state and is intended for single patient use only. The sterilization method used is ethylene oxide and has a shelf life of 1 year.
#### Summary of Performance Testing
The following performance tests were conducted on sterilized units. Items marked with an * were also performed post accelerated aging:
- . Dimensional Measurements
- o Cord Length
- Tip Width, Length, Thickness o
All devices tested met specification and were graded as Pass.
- Expected Forces on the Device ●
- o Bipolar Closing Force
- o Button Force
- Slider Force o
- Grasping Force o
All devices tested met specification and were graded as Pass
- . Overall Design Testing
- O Load Testing (Monopolar and Bipolar)
- Tensile*, Compressive, Vertical, Lateral .
{6}------------------------------------------------
Traditional 510(k) Notification July 2015
- o Cable Tensile
- Device Symmetry o
- o Hand Switching
- Tip Orientation o
- o Grasping Force*
- Slider Force and Distance* ্ত
- o Bipolar Gap*
- Slider Endurance o
- O Electrical Functionality*(device performs as intended)
All devices met specification and were graded as Pass
- Biocompatibility Testing .
- o Cytotoxicity
- Intracutaneous Injection o
- o Irritation
- Pyrogenicity o
All devices passed ISO 10993 acceptance criteria and were graded as Pass
- Packaging Testing (ISTA 2A) .
- · Label/package visual inspection *
- Bubble Leak* o
- Peel Strength* o
- Electrical Functionality*(device performs as intended) o
All devices met specification and were graded as Pass.
In addition ex-vivo bench top testing was performed comparing the proposed PK AIM device to the predicates. All ex-vivo results showed substantially equivalent results to the predicate devices in terms of cutting and coagulation performance.
* indicates test was also performed on 1 year accelerated aged units.
#### Substantial Equivalence
The proposed PK AIM has the same intended use, design, and scientific technology as the predicate devices combined into one device. The proposed device is both a pencil and a forceps device with a dual cord. Each of the predicate devices includes these very same features. In addition the proposed and predicate devices were shown to perform substantially equivalent in bench testing. There were no new issues of safety or effectiveness with the proposed device. Please see Table 1 - Substantial Equivalence Table on the following page.
{7}------------------------------------------------
Traditional 510(k) Notification July 2015
# Table 1 – Substantial Equivalence Table
| Design Feature | Proposed<br>Olympus PK AIM | Predicate<br>ValleyLab/Covi<br>dien<br>Electrosurgical<br>Pencil<br>(K813071) | Predicate<br>Stryker SilverGlide<br>Bipolar Forceps<br>(K093108) | Predicate<br>ValleyLab<br>LigaSure Advance<br>(K063195)* | Comments |
|------------------------|-----------------------------------------------|-------------------------------------------------------------------------------|------------------------------------------------------------------|----------------------------------------------------------|-------------|
| Function of<br>Device | Forceps/Pencil | Pencil | Forceps | N/A | Same |
| Tip Size | Forceps<br>1mm tip, 2mm base<br>Pencil<br>2mm | 2mm | 2mm | N/A | Same |
| Energy | Monopolar/Bipolar | Monopolar | Bipolar | Monopolar/<br>Bipolar | Same |
| Activation | Hand / Footpedal | Hand | Foot Pedal | N/A | Same |
| Plug | Dual (mono/bi) | Single | Single | Dual (mono/bi) | Same |
| Electrode<br>Materials | Stainless Steel | Stainless<br>Steel | Steel with<br>Silver coating | N/A | Same |
| Sterility | 10-6<br>EtO | 10-6<br>Irradiation | 10-6<br>Gravity,<br>Autoclave | N/A | Same (10-6) |
| Single Use | Yes | Yes | No | N/A | Same |
# Conclusion:
In summary, the Olympus – PK AIM is substantially equivalent to the predicate devices and presents no new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.