The intended use of the Belmont® Hyperthermia Pump is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.
Device Story
Device provides rapid perfusion of warmed fluid to thoracic or peritoneal cavities; used for hyperthermia therapy. System inputs: fluid temperature, patient temperature, line pressure, air-in-line detection. Operation: microprocessor-controlled mechanical pump circulates sterile solution; warms fluid to target temperature. Output: continuous lavage; touch screen displays flow rate, fluid volume, temperatures, pressure, and alarm status. Used in clinical settings; operated by physicians/clinical staff. Safety features: hardware override circuit for computer failure; battery backup for transport (warming disabled, monitoring active). Benefits: precise thermal control for cavity lavage; automated safety monitoring/alarms for over-temperature, high pressure, or air-in-line conditions.
Clinical Evidence
Bench testing only. Validation included: fluid pumping accuracy across flow rates/pressures/temperatures; warming performance; alarm verification for over-temperature, high pressure, out-of-fluid, and air-in-line conditions; risk assessment; software algorithm validation; and compliance with medical electrical equipment safety (IEC 60601-1) and electromagnetic compatibility standards.
Technological Characteristics
Microprocessor-controlled mechanical pump; high-speed fluid perfusion. Features touch screen interface, monochromatic high-contrast LCD. Monitors fluid/patient temperature, line pressure, air-in-line. Includes hardware override circuit for safety. Battery backup for mobile transport. Connectivity: standalone system. Sterilization: compatible with sterile solutions.
Indications for Use
Indicated for patients requiring thoracic or peritoneal cavity lavage with warmed sterile solution to reach a physician-selected target temperature.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a way that they appear to be interconnected. The profiles are black against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 2, 2015
Belmont Instrument Corporation Uraiwan P. Labadini 780 Boston Road Billerica, Massachusetts 01821
Re: K152208
Trade/Device Name: Belmont® Hyperthermia Pump Regulation Number: N/A Regulation Name: N/A Regulatory Class: Unclassified Product Code: LGZ Dated: August 6, 2015 Received: August 7 2015
Dear Mr. Labadini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin |. Keith -S
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152208
Device Name The Belmont Hyperthermia Pump
Indications for Use (Describe)
The intended use of the Belmont Hyperthermia Pump is to raise the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Together. Saving lives.
Image /page/3/Picture/1 description: The image shows the logo for Belmont Instrument Corporation. The logo features the word "Belmont" in a stylized, blue font, with the "B" extending into a curved shape. Below "Belmont" is the text "INSTRUMENT CORPORATION" in a smaller, gray font.
# PREMARKET NOTIFICATION 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS [As Required By 21 CFR 807.92(c)]
| Special 510(k) No. | K152208 |
|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name and Address: | Belmont Instrument Corporation<br>780 Boston Road<br>Billerica, MA 01821 |
| Contact Person: | Uraiwan P. Labadini<br>Quality Assurance/Regulatory Affairs Manager<br>Telephone: (508) 663-0212 Ext. 128<br>Fax: (508) 663-0214<br>Email address: ulabadini@belmontinstrument.com |
| Date Summary Prepared: | September 2, 2015 |
| Trade/Proprietary Name of Device: | The Belmont® Hyperthermia Pump |
| Regulatory Class: | Unclassified |
| Product Code: | LGZ (Warmer, Thermal, Infusion Fluid) |
| Legally Marketed Device Under<br>Which Substantial Equivalence is<br>Claimed: | The Belmont® Hyperthermia Pump cleared for<br>market entry originally under 510(k) K070654,<br>1st modification cleared under 510(k) K090089. |
| Device Description: | The Belmont® Hyperthermia Pump combines<br>advanced microprocessor technology with an<br>efficient mechanical system to provide a high<br>speed, simple and safe system for rapid perfusion<br>of warmed fluid. It can raise the temperature of the<br>thoracic or peritoneal cavity to the desired target<br>temperature by continuously lavaging the cavity<br>with circulating warmed sterile solution, according<br>to a protocol to be selected by the physician.<br><br>The system monitors fluid temperature, patient<br>temperature, line pressure, and air in the fluid path<br>to ensure safe operation and alarms at all unsafe<br>conditions. A hardware override circuit prevents<br>unsafe operation in case of system computer<br>failure. A touch screen displays flow rate, total fluid<br>pumped, output fluid temperature, target<br>temperature, patient temperature, line pressure,<br>alarm and status messages and proper procedures |
{4}------------------------------------------------
to proceed safely after an alarm situation.
A battery backup allows for mobile transport of the patient and system. During battery operation, fluid warming is disabled while pump operation and safety monitoring remain active.
Intended Use: The intended use of the Belmont® Hyperthermia Pump is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.
Comparison of Technological Belmont Instrument Corporation is claiming Characteristics of the Device vs. substantial equivalence of of the Belmont® Hyperthermia Pump, to our Belmont® Hyperthermia the Legally Marketed Device: Pump which was originally cleared to market under 510(k) K070654 on June 8, 2007, and to the Belmont® Hyperthermia Pump 1000 ml/min flow rate was cleared to market under 510(k) K090089 on February 12, 2009.
> Belmont® The Hyperthermia Pump was a modification of the Rapid Infuser which has been in clinical use for more than 15 years; consequently a component used in the system had become obsolete and hard to obtain. Therefore, the system hardware (chip replacement), and software (to correspond with the chip changes) have been upgraded. At the same time, we are using a brighter monochromatic display with a higher contrast LCD to improve visibility. We have performed a rigorous validation to ensure that these updates are backward compatible with our existing system and no major functions have been altered.
Brief Discussion of Nonclinical Tests and their Results Submitted in the Application:
Specific Modifications being
reviewed in this submission:
In order to verify performance of the Belmont® Hyperthermia Pump in support of substantial equivalence, the following tests were carried out:
- a. The ability of the system to pump fluids accurately over the full range of flow rate and operating conditions including different input fluid temperatures, different back pressure levels, and change in ambient temperature.
{5}------------------------------------------------
Brief Discussion of Nonclinical Tests and their Results Submitted in the Application: Cont'd
- b. The ability of the system to warm cold fluids to the user set temperature over the full range of flow rate and operating conditions.
- c. The ability of the system to detect and alarm at unsafe or ineffective operating conditions including over temperature, the unsafe high line pressure condition, out of fluid, air in the line and at any of several internal fault conditions.
- d. We performed risk assessment and performed the verification and validation tests to ensure that these risks were mitigated and method of controls were implemented correctly.
- e. We validated each algorithm for each function element in the software specification using an approved procedure. We also performed extensive testing for the entire system with software in place to test both the hardware and software functions to ensure that all system requirements are met.
- f. Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.
- g. Medical Electrical Equipment Collateral Standard: Electromagnetic Compatibility
The Belmont® Hyperthermia Pump performed within specifications in all of the above tests.
The Belmont® Hyperthermia Pump is substantially equivalent to the Belmont® current product which received 510(k) approval at various times. Both systems are capable of high infusion flow rates, while maintaining infusate at user-set temperature, and both systems are suitable for use with sterile solutions. Both systems monitor flow rate. temperature, line pressure, and the presence of air at the input line or within the system, and alarm and stop the system at all unsafe conditions.
Belmont® The tests demonstrate that the Hyperthermia Pump is substantial equivalent and performs as well as the legally marketed device.
Conclusion:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.