← Product Code [OLY](/productcode/OLY) · K152184

# MagView BIOMAGNETOMETER SYSTEM (K152184)

_Tristan Technologies, Inc. · OLY · Apr 8, 2016 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152184

## Device Facts

- **Applicant:** Tristan Technologies, Inc.
- **Product Code:** [OLY](/productcode/OLY.md)
- **Decision Date:** Apr 8, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1400
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Pediatric

## Indications for Use

Use of the Mag View Biomagnetometer is indicated for the patient whose physician believes that information about the magnetic fields produced by that patient's brain and information about the sources of those magnetic fields could contribute to diagnosis or therapy planning. The intended patient populations are neonates and those children with head circumferences of 50 cm or less.

## Device Story

MagView Biomagnetometer detects and amplifies magnetic fields generated by brain electrical activity. System utilizes superconducting pickup coils and Superconducting Quantum Interference Devices (SQUIDs) housed in a cryogen-cooled vacuum container. Input consists of magnetic signals from the patient's brain; signals are amplified, filtered, and digitized by an electronics subsystem. Output is digitized voltage data stored on a computer hard drive, representing magnetic field strength over time. Operated by a physician or a technologist under physician supervision. Data is analyzed and interpreted by the clinician to assist in diagnosis or therapy planning. Benefits include non-invasive mapping of brain magnetic sources in pediatric populations.

## Clinical Evidence

Bench testing only. Performance verified using external calibrated signal sources to confirm magnetic signal detection comparable to the predicate. Sensitivity measured at 10 femtoTesla/√Hz or better over a 1 Hz to 1 kHz bandwidth. Compliance with IEC 60601-1, IEC 60601-1-2, and various IEC 61000/CISPR electromagnetic compatibility and safety standards was demonstrated.

## Technological Characteristics

Superconducting pickup coils (270 primary, 114 reference) and SQUIDs. Liquid helium cooling via solid thermal conduction. Helmet-shaped vacuum housing. Sensitivity: 10 femtoTesla/√Hz; Bandwidth: 1 Hz to 1 kHz. Digitization: 24-bit at 5 kHz. Connectivity: Computer-based data storage. Standards: IEC 60601-1, IEC 60601-1-2, IEC 61000 series, CISPR 11.

## Regulatory Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

## Predicate Devices

- Magnes 2500 WH Biomagnetometer System ([K962317](/device/K962317.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 8, 2016

Tristan Technologies, Inc % Eugene Hirschkoff Consultant 3365 Calle Margarita Encinitas, California 92024

Re: K152184

Trade/Device Name: MagView Biomagnetometer Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLY Dated: August 3, 2015 Received: August 5, 2015

Dear Eugene Hirschkoff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Carlos L. Pena -S/^

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152184

Device Name MagView Biomagnetometer

#### Indications for Use (Describe)

Use of the Mag View Biomagnetometer is indicated for the patient whose physician believes that information about the magnetic fields produced by that patient's brain and information about the sources of those magnetic fields could contribute to diagnosis or therapy planning. The intended patient populations are neonates and those children with head circumferences of 50 cm or less.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10pt">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Tristan Technologies. The logo consists of a blue icon on the left and the company name on the right. The icon is a rounded square with three vertical lines inside. The company name is written in blue, with "Tristan" on the top line and "Technologies" on the bottom line.

6191 Cornerstone Court East, Suite 107 San Diego, CA 92121 USA Phone: +1 (858) 550-2700 Fax: +1 (858) 550-2799 http://www.tristantech.com

## 510(k) SUMMARY

This 510(k) Summary is being submitted pursuant to the requirements of 21 CFR 807.92(c).

| 1. | Submitted By:   | Tristan Technologies, Inc.<br>6191 Cornerstone Ct., Ste 107<br>San Diego, CA 92121<br>(858) 550-2700<br>Contact: Kevin Pratt, CEO<br>Prepared: June 12, 2015                            |
|----|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Name of Device: | Trade Name: MagView Biomagnetometer<br>Common Name: Magnetic Encephalograph<br>Classification Name: 21CFR 882.1400 (Electroencephalograph)<br>Product Code: OLY - Magnetoencephalograph |

- 3. Substantial Equivalence:
Conclusions drawn from the comparisons and testing summarized below demonstrate that the MagView Biomagnetometer is as safe, as effective, and performs as well as the Magnes 2500 WH Biomagnetometer System formerly manufactured and marketed by Biomagnetic Technologies, Inc., San Diego, CA, and is substantially equivalent to that predicate device.

{4}------------------------------------------------

- Description of Device: 4.
The Tristan Technologies MagView Biomagnetometer (hereinafter referred to as the "MagView") utilizes superconducting signal pickup coils and Superconducting Quantum Interference Devices (SQUIDs) to detect and amplify magnetic fields produced by electrical activity in brain. The MagView consists of a sensor unit, an electronics subsystem for preliminary amplification, filtering, and analog to digital conversion of the signals from each SOUID, an electronics rack containing power supplies to power the electronics subsystem, a computer to control the operation of the electronic subsystem and the SQUIDs and to acquire and store the signal values collected by the system.

- న్. Intended Uses:
The Tristan Technologies MagView Biomagnetometer is intended for use for the patient whose physician believes that information about the magnetic fields produced by that patient's brain and information about the location of the sources of those magnetic fields could contribute to diagnosis or therapy planning. The intended patient populations are neonates and infants and those children with head circumferences of 50 cm or less.

#### 6. Technological Characteristics:

The Tristan Technologies MagView utilizes superconducting signal pickup coils and Superconducting Quantum Interference Devices (SQUIDs) to detect and amplify magnetic fields produced by electrical activity in brain. This is the identical method of operation as that in the predicate Magnes 2500 WH. The MagView sensor comprises a primary array of 270 passive superconducting pickup coils, each of which is connected to a SQUID, a second array of 114 pickup coils spaced behind the primary array, each of which is also connected to a SQUID. By contrast, the predicate device employs 148 primary pickup coils and a second array of 8 pickup coils spaced behind the primary array. The pickup coils in the MagView are distributed over the same area of the head of a patient as are the coils in the predicate device. The sensitivity of each pickup coil is 10 femtoTesla//V Hz in both the MagView and the Magnes 2500 WH, 

{5}------------------------------------------------

over a bandwidth from 1 Hz to 1 kHz. In both systems, the second pickup coil array spaced behind the primary array is available for noise cancellation purposes.

In both systems, the pickup coil arrays are contained within an evacuated housing along with an insulated container for the cryogen liquid helium. The pickup coil arrays and SQUIDs are refrigerated by solid thermal conduction to the cryogen. -For both the MagView and the predicate, the vacuum container is configured to have a helmet-like external shape at the bottom. This shape is sized and oriented to accommodate the positioning of the head of a human being lying in a supine position into the helmet-like shape. The shape of the MagView "helmet" is smaller than that of the predicate device, and will be appropriate for patients with smaller head sizes. However, the sensitivity of each pickup coil is the same for both devices.

The primary array of pickup coils in the MagView is positioned within the vacuum container so as to be in close proximity to the helmet-like shape, and thus when in use, to be in close proximity to the head of the human being. This is also the identical method used in the Magnes 2500 WH although the latter was capable of being positioned to accommodate heads of human beings in the seated position as well as in a supine position.

In both the MagView and the predicate device, the output of each SQUID is a voltage the value of which is proportional to the magnetic field at the corresponding pickup coil. The voltage from each SQUID is amplified, filtered and digitized by signal processing electronics. In the MagView, the signal from each SQUID is digitized with 24 bit precision at a sample rate of 5 kHz; in the Magnes 2500, the signal from each SQUID is digitized with 16 bit precision at a sample rate of 2 kHz. The digitized signals are conveyed to a computer hard drive. In both systems, a hard drive thus contains data comprising the voltage from each SQUID recorded as a function of time. This data is available to the user of the system for analysis and interpretation. While the MagView may be operated by a physician, it may also be operated by a technologist working under the direction and supervision of a physician.

The MagView system has been tested and shown to comply with the following safety standards:

IEC 60601-1:1988 + A1:1991 + A2:1995

IEC 60601-1-2:2007 Third Edition

IEC 61000-4-2:2008

IEC 61000-4-3:2010

IEC 61000-4-4:2012

IEC 61000-4-5:2014

{6}------------------------------------------------

IEC 61000-4-6:2003

IEC 61000-4-11:2004

CISPR 11: 2010

In addition, non-clinical testing of the system was conducted on the MagView using external calibrated signal sources to demonstrate magnetic signal detection performance comparable to that of the predicate device. The sensitivity of the system, averaged over all primary channels, was measured to be 10 femtoTesla/V Hz or better, over a bandwidth of 1Hz to 1 kHz, the same sensitivity and bandwidth offered by the Magnes 2500 WH.

#### 7. Conclusions:

Measurement of magnetic fields originating from electrical activity in the brain has been undertaken for at least the past 30 years. The Magnes 2500 is a commercial device which performs these measurements and which has been legally marketed under a 510(k) procedure in 1996. See K962317. The Tristan Technologies MagView Biomagnetometer uses identical technology and methodology as used by the Magnes 2500, with a different number and geometric arrangement of the pickup coils and SQUIDs and with updated computer hardware and software. The intended use and indications for use are identical to that of the Magnes 2500.

Conclusions drawn from the non-clinical performance testing described above demonstrate that the MagView is as safe, as effective, and performs as well as the Magnes 2500 WH. Therefore, Tristan Technologies submits that the MagView is substantially equivalent to the Magnes 2500 WH

Any questions regarding the 510(k) summary may be directed to the contact person noted.

---

**Source:** [https://fda-staging.innolitics.com/device/K152184](https://fda-staging.innolitics.com/device/K152184)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).