← Product Code [MNI](/productcode/MNI) · K152174

# Sniper Spine System (K152174)

_Spine Wave, Inc. · MNI · Aug 31, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152174

## Device Facts

- **Applicant:** Spine Wave, Inc.
- **Product Code:** [MNI](/productcode/MNI.md)
- **Decision Date:** Aug 31, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

When used as a pedicle screw fixation system of the noncervical spine in skeletally mature patients, the Sniper® Spine System is indicated for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The Sniper® Spine System is also indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3-S1), and for whom the device is intended to be removed after solid fusion is attained. The Sniper® Spine System percutaneous instruments are intended to be used with the Sniper® Spine System implants. The percutaneous instruments when used with the percutaneous cannulated screws and percutaneous rods, are intended to provide the surgeon with a percutaneous approach for posterior spinal surgery for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). As well as, for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3-S1), and for whom the device is intended to be removed after solid fusion is attained.

## Device Story

The Sniper® Spine System is a pedicle screw fixation system consisting of titanium alloy rods and screws; used to stabilize the noncervical spine during fusion of vertebral bodies. The system includes percutaneous instruments to facilitate a percutaneous approach for posterior spinal surgery. The device is intended for use by surgeons in an operating room setting. The system provides rigid spinal construct stabilization; intended for removal after solid fusion is achieved. The device benefits patients by providing spinal stabilization for various degenerative and traumatic conditions.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including dynamic axial compression bending testing per ASTM F1717, and cadaveric validation to demonstrate equivalence to the predicate device.

## Technological Characteristics

Constructed of titanium alloy. Components include rods and screws (including extended tab screws). System is non-sterile and single-use. Mechanical performance validated via ASTM F1717 dynamic axial compression bending testing.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Sniper® Spine System ([K151813](/device/K151813.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked one behind the other, representing the department's mission to protect the health of all Americans and provide essential human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 31, 2015

Spine Wave, Incorporated Ms. Sanja Jahr Regulatory Affairs Specialist Three Enterprise Drive. Suite 210 Shelton, Connecticut 06484

Re: K152174

Trade/Device Name: Sniper® Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH Dated: August 3, 2015 Received: August 4, 2015

Dear Ms. Jahr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Sanja Jahr

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv vours.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

#### K152174

Device Name

Sniper® Spine System

#### Indications for Use (Describe)

When used as a pedicle screw fixation system of the noncervical spine in skeletally mature patients, the Sniper® Spine System is indicated for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

The Sniper® Spine System is also indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3-S1), and for whom the device is intended to be removed after solid fusion is attained.

The Sniper® Spine System percutaneous instruments are intended to be used with the Sniper® Spine System implants. The percutaneous instruments when used with the percutaneous cannulated screws and percutaneous rods, are intended to provide the surgeon with a percutaneous approach for posterior spinal surgery for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis). As well as, for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3-S1), and for whom the device is intended to be removed after solid fusion is attained.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                               |                                                                                                                                                                                |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="12" width="12"> <rect fill="lightgray" height="12" width="12"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black"></path> </svg> </span>  Prescription Use (Part 21 CFR 801 Subpart D)  </div> | <div> <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" width="12"></rect> </svg> </span>  Over-The-Counter Use (21 CFR 801 Subpart C)  </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SPINEWAVE" in a bold, sans-serif font. The word is all in uppercase letters. A blue wave design is located under the word, adding a visual element to the logo.

## 510(k) Summary Sniper® Spine System

#### 1. Submitter Information

| Submitter: | Spine Wave, Inc.                                         |
|------------|----------------------------------------------------------|
| Address:   | Three Enterprise Drive<br>Suite 210<br>Shelton, CT 06484 |
| Telephone: | 203-712-1870                                             |
| Telefax:   | 203-944-9493                                             |
| Contact:   | Sanja Jahr                                               |

Date Prepared: August 3, 2015

#### 2. Device Information

| Trade Name:          | Sniper® Spine System                   |
|----------------------|----------------------------------------|
| Common Name:         | Pedicle Screw Spinal System            |
| Classification Name: | Pedicle Screw Spinal System            |
| Classification/Code: | Class II per 21 CFR 888.3070; MNI, MNH |

#### 3. Purpose of Submission

The purpose of this submission is to gain clearance for the addition of extended tab screws to the cleared Sniper® Spine System.

#### 4. Predicate Device Information

The Sniper® Spine System described in this submission is substantially equivalent to the following primary predicate:

| Predicate Device     | Manufacturer     | 510(k) No. |
|----------------------|------------------|------------|
| Sniper® Spine System | Spine Wave, Inc. | K151813    |

## 5. Device Description

The Sniper® Spine System consists of a selection of non-sterile, single use titanium alloy rod and screw components that are assembled to create a rigid spinal construct. The rod and screw components of the Sniper® Spine System are attached to the non-cervical spine in order to stabilize the spine during fusion of the vertebral bodies, and are intended to be removed after spinal fusion is achieved.

{4}------------------------------------------------

## 6. Indications for Use

When used as a pedicle screw fixation system of the noncervical spine in skeletally mature patients, the Sniper® Spine System is indicated for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

The Sniper® Spine System is also indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3-S1), and for whom the device is intended to be removed after solid fusion is attained.

The Sniper® Spine System percutaneous instruments are intended to be used with the Sniper® Spine System implants. The percutaneous instruments when used with the percutaneous cannulated screws and percutaneous rods, are intended to provide the surgeon with a percutaneous approach for posterior spinal surgery for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). As well as, for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral ioint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3-S1), and for whom the device is intended to be removed after solid fusion is attained.

## 7. Comparison of Technological Characteristics

The substantial equivalence of the subject Sniper® Spine System is shown by similarity in intended use, indications for use, materials and performance to the cited predicate device.

## 8. Performance Data

Engineering analyses, dynamic axial compression bending testing according to ASTM F1717, as well as cadaveric validation was performed to demonstrate that the modified Sniper Spine System is substantially equivalent to the predicate device.

## 9. Conclusion

Based on the indications for use, technological characteristics, performance testing and comparison to the predicate, the subject Sniper® Spine System has been shown to be substantially equivalent to the predicate device identified in this submission.

---

**Source:** [https://fda-staging.innolitics.com/device/K152174](https://fda-staging.innolitics.com/device/K152174)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).