K152136 · C.R. Bard, Inc. · MMX · Feb 18, 2016 · Cardiovascular
Device Facts
Record ID
K152136
Device Name
Recovery Cone Removal System
Applicant
C.R. Bard, Inc.
Product Code
MMX · Cardiovascular
Decision Date
Feb 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
RC-15 Model: The Recovery Cone Removal System is intended for use to percutaneously remove the Recovery Filter, Recovery G2 Filter and G2 X Filter from the vena cava. FBRC Model: The Recovery Cone Removal System is intended for use to percutaneously remove the G2 X Filter, G2 Express Filter, G2 Filter and the Recovery Filter from the vena cava, or facilitate the retrieval of foreign objects from the peripheral vascular system.
Device Story
Percutaneous retrieval system for vena cava filters and foreign objects; consists of reinforced urethane cone (15-mm diameter) with wire claws, Pebax shaft, handle, and 10 French introducer sheath/dilator. Operates over 0.035" guidewire; Touhy-Borst Y-adapter allows saline flush. Used in clinical/interventional settings by physicians. Device engages target object via cone/claws, then withdraws into outer sheath for removal. Fluoroscopic visualization aided by radiopaque markers. Benefits patient by enabling minimally invasive removal of filters or foreign bodies, reducing need for open surgery.
Clinical Evidence
Bench testing included tensile, leak, torque, and simulated use testing. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemocompatibility, thromboresistance, coagulation). In-vivo animal testing and clinical references evaluated filter capture and retrieval performance.
Technological Characteristics
Reinforced urethane cone (15-mm) with wire claws; Pebax shaft; 10 French introducer sheath/dilator; radiopaque markers for fluoroscopy. Compatible with 0.035" guidewire. Sterile, single-use. Biocompatibility per ISO 10993 standards.
Indications for Use
Indicated for percutaneous removal of specific vena cava filters (Recovery, Recovery G2, G2 X, G2 Express) or retrieval of foreign objects from the peripheral vascular system.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping faces suggesting community and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 18, 2016
Bard Peripheral Vascular, Inc. Ms. Joni Creal Regulatory Affairs Program Manager 1625 West 3rd Street Tempe, Arizona 85281
Re: K152136
Trade/Device Name: Recovery Cone® Removal System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: MMX Dated: December 29, 2015 Received: January 05, 2016
Dear Ms. Creal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152136
Device Name Recovery Cone® Removal System
#### Indications for Use (Describe)
RC-15 Model: The Recovery Cone Removal System is intended for use to percutaneously remove the Recovery Filter, Recovery G2 Filter and G2 X Filter from the vena cava.
FBRC Model: The Recovery Cone Removal System is intended for use to percutaneously remove the G2 X Filter, G2 Express Filter, G2 Filter and the Recovery Filter from the vena cava, or facilitate the retrieval of foreign objects from the peripheral vascular system.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Recovery Cone® Removal System 510(k) Summary
#### Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|-------------------------------------------------------------------------------|
| Phone: | 480-638-2906 |
| Fax: | 480-449-2546 |
| Contact: | Joni Creal, Regulatory Affairs Program Manager |
| Date: | July 30, 2015 |
#### Subject Device Name:
| Device Trade Name: | Recovery Cone® Removal System |
|-----------------------|--------------------------------|
| Common or Usual Name: | Device, Percutaneous Retrieval |
| Classification: | Class II |
| Review Panel: | Cardiovascular Devices |
| Product Code: | MMX |
#### Cook FourSnare Vascular Retrieval Snare (K112185) Predicate Devices:
#### Device Description:
The Recovery Cone® Removal System is intended to percutaneously remove the G2® X Filter, G2® Express™ Filter, G2® Filter or Recovery® Filter from the vena cava or facilitate the retrieval of foreign objects from the peripheral vascular system. The device is provided sterile and is single use only. The Recovery Cone® Removal System consists of the Removal Cone and Introducer Catheter. The cone consists of a reinforced urethane cone. 15-mm in diameter. The cone is connected to a plastic (Pebax) shaft and handle. The shaft has a central lumen that accommodates a 0.035" guidewire. A Touhy-Borst Y-adapter is used to connect the cone to the Introducer Catheter and to a saline flush or drip. The Introducer Catheter consists of a 10 French
Image /page/3/Picture/12 description: The image shows the word "BARD" in a stylized, sans-serif font. The letters are bold and outlined in black, giving them a distinct and modern appearance. The font choice and design create a visually appealing and recognizable logo or wordmark.
{4}------------------------------------------------
I.D. introducer sheath and dilator. The introducer sheath has a radiopaque marker for enhanced fluoroscopic visualization.
### Indications for Use of Device:
RC-15 Model: The Recovery Cone Removal System is intended for use to percutaneously remove the Recovery Filter, Recovery G2 Filter and G2 X Filter from the vena cava.
FBRC Model: The Recovery Cone Removal System is intended for use to percutaneously remove the G2 X Filter, G2 Express Filter, G2 Filter and the Recovery Filter from the vena cava, or facilitate the retrieval of foreign objects from the peripheral vascular system.
### Technological Comparison to Predicate Devices:
The technological characteristics of the subject device, Recovery Cone® Removal System, are substantially equivalent to the legally marketed predicate device, the Cook FourSnare Retrieval Snare, in terms of intended use, indications for use, operating principle and fundamental scientific technology. Both devices are catheter based systems that have collapsible wires for filter engagement and withdraw into an outer sheath. Both systems are designed to be used over a quidewire. Both devices contain marker bands for visibility. The predicate device contains four wire loops to engage the foreign objects while the Recovery Cone® utilized a urethane cone reinforced with wire claws. The predicate wire loop system is designed to advance through a 80cm, 10 French introducer catheter, and the subject device cone is designed to advance through a 75 cm, 10 French I.D. introducer catheter.
#### Performance Testing Summary:
To demonstrate substantial equivalence of the subject device to the predicate device, the technological characteristics and performance criteria were evaluated using in-vitro testing as outlined below. In addition, the ability to capture the filter and retrieve the filter was evaluated in in-vivo animal testing and within clinical references.
- In-Vitro Filter .
- Tensile
Image /page/4/Picture/13 description: The image shows the word "BARD" in a stylized, all-caps font. The letters are bold and outlined in black, giving them a strong, graphic appearance. The font choice is geometric and modern, with sharp angles and clean lines. The overall design is simple yet impactful, making the word easily recognizable.
{5}------------------------------------------------
- | Leak Test
- Filter Retrieval ।
- -Foreign Body Retrieval
- Sheath and Dilator Tensile, Torque and Leak Integrity Test -
- Catheter Simulated Use Test -
- Marker Band Integrity —
- -Filter Removal Force
- Biocompatibility, per ISO 10993 ●
- Cytotoxicity (ISO 10993-5) -
- Sensitization (ISO 10993-10) |
- -Irritation - Intracutaneous Reactivity (ISO10993-10)
- Acute Systemic Toxicity (ISO10993-11) -
- -Pyrogenicity (ISO 10993-11)
- Hemocompatibility —
- Thromboresistance -
- Coagulation (ISO 10993-4) |
In conclusion, the results of these tests provide reasonable assurance that the device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.