← Product Code [EIH](/productcode/EIH) · K152118

# Zenostar MT, Zenostar Color Liquids (K152118)

_Wieland Dental + Technik GmbH & Co. KG · EIH · Dec 16, 2015 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152118

## Device Facts

- **Applicant:** Wieland Dental + Technik GmbH & Co. KG
- **Product Code:** [EIH](/productcode/EIH.md)
- **Decision Date:** Dec 16, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Zenostar MT (medium translucency) are machinable zirconium oxide discs for the production of all-ceramic crowns, partial crowns and bridges in the anterior and posterior region. Zenostar® MT Color Liquids are ready-made, water-based solutions for shading full contour restorations made from Zenostar MT. They are applied prior to sintering using the brush technique.

## Device Story

Zenostar MT consists of partially sintered zirconium oxide discs (98.5mm diameter; 14mm or 18mm thickness) used by dental technicians to fabricate all-ceramic dental restorations (crowns, partial crowns, bridges) via CAD/CAM milling. Zenostar MT Color Liquids are water-based shading solutions applied to the milled restoration via brush technique prior to final sintering in a compatible furnace. The device is used in dental laboratories. The output is a custom-fitted dental prosthesis. The healthcare provider uses the final sintered restoration for permanent dental replacement, benefiting the patient by restoring dental function and aesthetics. The device is not sterile and is for single-use fabrication.

## Clinical Evidence

Bench testing only. Device tested for flexural strength, chemical solubility, coefficient of thermal expansion, and radioactivity per ISO 6872:2008. Biocompatibility assessed per EN ISO 10993-1:2009 and EN ISO 7405:2008. Results confirm compliance with Class 5 material requirements (minimum mean flexural strength 500 MPa).

## Technological Characteristics

Material: Yttria-stabilized zirconium dioxide (Y-TZP). Form factor: 98.5mm discs (14mm/18mm thickness). Processing: Dry CAD/CAM milling followed by sintering. Connectivity: None (standalone material). Sterilization: Non-sterile. Standards: ISO 6872:2008 (mechanical), ISO 10993/7405 (biocompatibility).

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- IPS e.max CAD/IPS e.max ZirCAD ([K051705](/device/K051705.md))

## Reference Devices

- Zenotec Zr Bridge ([K112710](/device/K112710.md))
- Zenostar Zr Translucent ([K112710](/device/K112710.md))
- Zenotec Color Zr ([K112710](/device/K112710.md))
- Zenostar Color Zr ([K112710](/device/K112710.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 16, 2015

Wieland Dental + Technik Gmbh & Co. Kg % Donna Hartnett Director QA/Regulatory Affairs Ivoclar Vivadent. Inc. 175 Pineview Drive Amherst. New York 14228

Re: K152118

Trade/Device Name: Zenostar MT, Zenostar MT Color Liquids Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: November 13, 2015 Received: November 16, 2015

Dear Ms. Donna Hartnett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply 

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Susan Runno DDS, MA

Erin Keith Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K152118

Device Name ZENOSTAR MT

Indications for Use (Describe)

Zenostar MT (medium translucency) are machinable zirconium oxide discs for the production of all-ceramic crowns, partial crowns and bridges in the anterior and posterior region.

Zenostar® MT Color Liquids are ready-made, water-based solutions for shading full contour restorations made from Zenostar MT. They are applied prior to sintering using the brush technique.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Wieland Dental. The word "WIELAND" is written in large, bold, blue letters. Below that, the word "DENTAL" is written in smaller, blue letters.

EMBER OF THE IVOCLAR VIVADENT GR

Contact: Donna Marie Hartnett

Wieland Dental + Technik GMBH & Co KG Company: Lindenstrasse 2 71175 Pforzheim, Germany

December 15, 2015 Date Prepared:

Zenostar MT Proprietary Name:

Classification Name: Powder, Porcelain (872.6660) (Classification Code EIH)

- Primary predicate: IPS e.max CAD/IPS e.max ZirCAD (K051705) Predicate Device: by Ivoclar Vivadent, AG (Liechtenstein) Reference device: Zenotec Zr Bridge, Zenostar Zr Translucent, Zenotec Color Zr, Zenostar Color Zr, by Wieland Dental (K112710).
Device Description: Zenostar MT are zirconia discs with 98.5mm width and 2 thicknesses, 14 mm and 18 mm. for use in the fabrication of dental prosthesis through the CAD/CAM milling technology. In addition, coloring liquids for shading full-contour restorations made from Zenostar MT.

Indications for Use:

Zenostar MT (medium translucency) are machinable zirconium oxide discs for the production of all-ceramic crowns, partial crowns and bridges in the anterior and posterior region.

Zenostar MT Color Liquids are ready-made, water-based solutions for shading full-contour restorations made from Zenostar MT. They are applied prior to sintering using the brush technique.

Comparison to Predicate: The predicate device to which Zenostar MT has been compared is IPS e.max ZirCAD (K051705) and Zenostar MT Coloring Liquids have been compared to those in the predicate Zenotec Zr Bridge, Zenostar Zr Translucent, Zenotec Color Zr. Zenostar Color Zr, by Wieland Dental (K112710). All the ingredients except three nitrates in the coloring liquids have been used in the predicates. The biocompatibility of the new coloring liquid formulation was fully assessed and found to be equivalent to that of the predicate. For this application, Zenostar MT has been compared to its predicate with regard to chemical composition, performance data and indications for use. The comparison shows that Zenostar MT is substantially equivalent to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Wieland Dental. The word "WIELAND" is in large, bold, blue letters. Below that, in smaller blue letters, is the word "DENTAL" with a trademark symbol.

a member of the ivoclar vivadent group

| Device Name                      | Zenostar MT<br>K152118                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Primary Predicate<br>IPS e.max CAD/IPS<br>e.max ZirCAD<br>K051705                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Reference Device<br>Zenotec Zr Bridge,<br>Zenostar Zr Translucent,<br>Zenotec Color Zr,<br>Zenostar Color Zr<br>K112710                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                     | Weiland Dental + Technik<br>GmbH & Co. KG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Ivoclar Vivadent, AG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Weiland Dental + Technik<br>GmbH & Co. KG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| 510K Number                      | K152118                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | K051705                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | K112710                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code                     | EIH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | EIH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | EIH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for Use              | Zenostar MT (medium<br>translucency) are<br>machinable zirconium<br>oxide discs for the<br>production of all- ceramic<br>crowns, partial crowns<br>and bridges in the anterior<br>and posterior region.<br><br>Zenostar MT Color<br>Liquids are ready-<br>made, water-based<br>solutions for shading<br>full-contour restorations<br>made from Zenostar MT.<br>They are applied prior to<br>sintering using the brush<br>technique.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | IPS e.max CAD is a<br>CAD/CAM machinable<br>glass ceramic based on<br>lithium disilicate for the<br>preparation of full ceramic<br>crowns, inlays, onlays and<br>full ceramic 3-unit anterior<br>bridges.<br><br>IPS e.max ZirCAD<br>consists of machinable<br>zirconia blocks for the<br>preparation of full ceramic<br>crowns, onlays and 3- and<br>4-unit bridges and inlay<br>bridges (anterior and<br>molar).                                                                                                                                                                                                                                      | Zeno Zr Disc is a group of<br>medical devices, which is<br>intended to fabricate all-<br>ceramic dental<br>restorations like crowns<br>and bridges. It consists of<br>ceramic blanks and<br>coloring liquids. Zenotec<br>Zr Bridge and Zenostar Zr<br>Translucent are yttria<br>stabilized zirconium<br>dioxide (Y-TZP) ceramic<br>(zirconia) blanks for the<br>CAD/CAM production of<br>dental restorations.<br>Zenotec Color Zr and<br>Zenostar Color Zr are<br>coloring liquids for the<br>shading of white zirconia<br>ceramic materials. The<br>products are intended to<br>be used by dental<br>technicians for fabrication<br>of all-ceramic single-tooth<br>and bridgework<br>restorations, with one or<br>two pontics, in the anterior<br>as well as in the posterior<br>tooth region |
| Contraindicated Use<br>of Device | Zenostar MT  Bridge restorations with more than 3 units Very deep sub-gingival preparations Patients with substantially reduced residual dentition Bruxism Extended bridges Any other use no listed in the indications Temporary insertion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | IPS e.max CAD  Full veneers on molar crowns Very deep sub gingival preparations Patients with substantially reduced residual dentition bruxism any other use not listed in the indications              IPS e.max ZirCAD                                                                                                                                                                                                                                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                  | Zenostar MT Color Liquids                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                  | -             All applications which are not listed in the labeling are considered as contraindications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | -             Very deep sub gingival preparations             Patients with substantially reduced residual dentition             Bruxism                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Delivery Forms                   | 98.5mm discs<br>Coloring Liquids packaged in bottles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Any other use not listed in the labeling<br>Machinable blocks in various sizes with attached mandrel to fit milling machine                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 98.5 mm discs<br>Zenostar Color Zr liquids packaged in bottles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Design Parameters                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Minimum Wall<br>Thickness        | Zenostar MT for 3 unit anterior bridges:<br>Minimum wall thickness 1.2mm<br>Zenostar MT for 3 unit posterior bridges:<br>Minimum wall thickness 1.5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | IPS e.max CAD for 3 unit anterior bridges:<br>Minimum wall thickness circular 1.2mm<br>Minimum wall thickness incisal/occlusal 1.5mm staining technique, 0.8mm cut-back technique<br>IPS e.max ZirCAD for 3 unit posterior bridges:<br>Minimum framework dimension circular 0.5mm<br>Minimum wall thickness occlusal 0.7mm                                                                                                                                                                                                                                                                                                                              | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Minimum Connector<br>Dimension   | Anterior Bridges:<br>Minimum Connector dimension 12mm²<br>Posterior Bridges<br>Minimum Connector dimension 16mm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | IPS e.max CAD:<br>Minimum Connector dimension 16mm²<br>IPS e.max ZirCAD:<br>Minimum Connector dimension 9mm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification                   | Type II Class 5                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | IPS e.max CAD:<br>Type II Class 3/3<br>IPS e.max ZirCAD:<br>Type II Class 6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Biocompatibility                 | ISO 10993 and ISO 7405                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | ISO 10993 and ISO 7405                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| CAD/CAM<br>Processing            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Physical Properties              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Flexural Strength                | >550 MPa                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | IPS e.max CAD:<br>>360 MPa<br>IPS e.max ZirCAD<br>>900 MPa                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Mechanical Testing               | ISO 6872:2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | ISO 6872:2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | ISO 6872:2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Sterile                          | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Reusable                         | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Principles of<br>Operation       | Zenostar MT milling discs<br>are supplied in a partly<br>sintered state. They are<br>processed in the Wieland<br>Zenotec milling machine<br>using dry machining. Final<br>sintering takes place after<br>milling in a compatible<br>sintering furnace. The<br>Ivoclar Vivadent<br>Programat S1 sintering<br>furnace or the Wieland<br>Zenotec System sintering<br>furnace are<br>recommended.<br>Prior to sintering the<br>Zenostar MT restorations<br>can be infiltrated with<br>Zenostar MT Color Liquids<br>using the brush technique<br>to achieve individual<br>shades.<br>The dental restoration can<br>be conventionally<br>cemented using either<br>phosphate cement or<br>glass ionomer cement or<br>adhesively using a<br>composite based<br>adhesive. Ivoclar Vivadent<br>recommends using<br>Multilink® Automix for<br>adhesive cementation,<br>SpeedCEM® for self-<br>adhesive cementation and<br>Vivaglass®CEM PL for<br>conventional cementation. | IPS e.max CAD:<br>Pre-crystallized<br>blocks<br>ready to be processed in a<br>compatible<br>CAD/CAM<br>milling machine.<br>IPS e.max ZirCAD<br>Blocks are supplied in a<br>partly-sintered state. They<br>are processed in a<br>compatible milling machine<br>using wet machining. Final<br>sintering takes place after<br>milling in a compatible<br>sintering<br>furnace.<br>The<br>Ivoclar Vivadent Programat<br>S1 sintering furnace is<br>recommended.<br>The dental restoration can<br>be<br>conventionally<br>cemented using either<br>phosphate c e m e n t or<br>glass ionomer cement or<br>adhesively using a<br>composite-based<br>adhesive. | Zenostar Zr Translucent<br>are<br>partly-sintered<br>zirconium milling discs<br>processed using dry<br>machining. Final sintering<br>takes place after milling.<br><br>Zenostar Color Zr liquids<br>are used prior to sintering<br>to infiltrate the zirconia<br>restorations using the<br>brush technique to<br>achieve individual shades.                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Wieland Dental. The word "WIELAND" is in large, bold, blue letters. Below that, the word "DENTAL" is in smaller, blue letters.

a member of the ivoclar vivadent group

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Wieland Dental. The logo is in blue and consists of the word "WIELAND" in large, bold letters, with the word "DENTAL" in smaller letters below it. The logo is simple and modern, and the blue color gives it a professional look.

A MEMBER OF THE IVOCLAR VIVADENT GROUP

The predicate includes IPS e.max CAD which is a silicate-based dental ceramic and IPS e.max ZirCAD which is zirconium oxide. Zenostar MT is zirconium oxide. The composition of the zirconium materials is very similar. IPS e.max ZirCAD and Zenostar MT do not contain pigments. It is possible to achieve esthetic restorations with IPS e.max ZirCAD by veneering them using IPS e.max Ceram. To achieve aesthetic restorations with Zenostar MT the milled restoration is infiltrated with Zenostar MT Color Liquids before sintering. Zenostar MT Color liquids are ready-made, water-based solutions for shading full-contour restorations made from Zenostar MT. The shades are as follows: A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4,

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Wieland Dental. The logo is in blue and consists of the word "WIELAND" in large, bold letters. Below the word "WIELAND" is the word "DENTAL" in smaller letters. There is a trademark symbol to the right of the word "DENTAL".

EMBER OF THE IVOCLAR VIVADENT GRO

Testing Summary: The device has been designed and tested in accordance with ISO 6872:2008 Dentistry: Ceramic Materials for Flexural Strength, Chemical Solubility, Co-efficient of thermal expansion and Radioactivity. The product-specific standard ISO 6872:2008 specifies for a Class 5 material for three-unit prostheses involving molar restoration a minimum mean flexural strength of 500 MPa. Zenostar MT fulfills these criteria. The test results are equivalent to those of the predicate device. Zenostar

MT. Zenostar MT Color Liquids have been thoroughly assessed for biocompatibility in accordance with EN ISO 10993-1:2009 Biological assessment of medical devices, EN ISO 7405:2008 Dentistry – Evaluation of biocompatibility of medical devices used in dentistry, and EN ISO 14971:2012 "Medical devices – Application of risk management to medical devices The biocompatibility is substantially equivalent to the predicate devices.

CONCLUSION: The indications, design parameters and physical properties of Zenostar MT lie between those of IPS e. max CAD and IPS e. max ZirCAD. Zenostar MT also includes Zenostar MT Color liquids which are applied to the milled restoration prior to sintering to achieve individual shades. This technique is already used as shown in the cleared Zenostar Zr Translucent predicate.

The above data and analysis demonstrates that Zenostar MT is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K152118](https://fda-staging.innolitics.com/device/K152118)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com