Ultraviolet Lamps Intended for Use in Sunlamp Products (commonly known as metal halide sunlamps)
K152095 · Cosmedico Light, Inc. · LEJ · Oct 28, 2015 · General, Plastic Surgery
Device Facts
Record ID
K152095
Device Name
Ultraviolet Lamps Intended for Use in Sunlamp Products (commonly known as metal halide sunlamps)
Applicant
Cosmedico Light, Inc.
Product Code
LEJ · General, Plastic Surgery
Decision Date
Oct 28, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4635
Device Class
Class 2
Indications for Use
This ultraviolet lamp is intended for use in sunlamp products for tanning of the human skin.
Device Story
High-pressure mercury-metal halide discharge lamps; used as components in tanning equipment. Input: electrical power. Operation: direct excitation of mercury and metal halide vapors within a quartz envelope to generate UV radiation. Output: UV radiation for skin tanning. Operated by tanning facility staff or end-users in a salon or home setting. Healthcare provider involvement is not required. Benefits: cosmetic skin tanning. Design variables include wattage (250W-1500W), lamp dimensions, and base types (R7s, GY9.5, wire leads).
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included functionality, electrical safety, mechanical safety, dimensional integrity, and UV radiation characteristics (UVA irradiance, erythemal effective irradiance, and UVC/UVB ratio per 21 CFR 1040.20).
Technological Characteristics
High-pressure mercury-metal halide discharge lamp. Components: tubular quartz envelope, inert fill gas (argon), mercury, and metal halide salts. Tungsten electrodes, molybdenum foil, ceramic base. Power range: 250W-1500W. Base types: R7s, GY9.5, wire leads. Complies with 21 CFR 1040.20 for UV radiation limits. No software or electronic control algorithms.
Indications for Use
Indicated for tanning of human skin in sunlamp products. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.
Special Controls
*Classification.* Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
*Labeling* —(i)*Sunlamp products.* (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
*Sunlamp products and UV lamps intended for use in sunlamp products.* Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
*Performance standard.* Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.
Predicate Devices
Legally marketed Class I, 510(k) exempt sunlamp products offered for sale on or before September 2, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three heads, representing the department's commitment to health, human services, and science. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Cosmedico Light, Inc. Steven Schlitt Director Of Engineering And Quality Assurance 233 Libbey Industrial Parkway Weymouth, Massachusetts 02189
October 28, 2015
Re: K152095
Trade/Device Name: Ultraviolet Lamps Intended For Use In Sunlamp Products (commonly Known As "metal Halide Sunlamps") Regulation Number: 21 CFR 878.4635 Regulation Name: Ultraviolet Lamp For Tanning Regulatory Class: Class II Product Code: LEJ Dated: July 24, 2015 Received: July 28, 2015
Dear Steven Schlitt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Joshua C. Nipper -S
For
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K152095
#### Device Name
Ultraviolet Lamps Intended for Use in Sunlamp Products (commonly known as "metal halide sunlamps")
#### Indications for Use (Describe)
INTENDED USE: This ultraviolet lamp is intended for use in sunlamp products for tanning of the human skin.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Per Sec. 807.92(c)
## 807.92(a)(1)
| Date Prepared: | July 24, 2015 |
|-------------------------------|-----------------------------------------------------|
| 510(k) Company / Holder Name: | Cosmedico Light, Inc. |
| Contact Name: | Steven C. Schlitt |
| Title: | Director Engineering & Quality |
| Address: | 233 Libbey Industrial Parkway<br>Weymouth, MA 02189 |
| Phone Number: | 781-331-0949 (Ext. 107) |
| Fax Number: | 781-331-4766 |
# 807.92(a)(2)
| Device Proprietary Name: | Various Brands of Ultraviolet Lamps<br>See attachment F-1 for list of models<br>and part numbers. |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------|
| Device Common or Usual Name: | Ultraviolet Sunlamps (Metal Halide Sunlamps) |
| Indication for Use / Intended Use | INTENDED USE: This ultraviolet lamp is<br>intended for use in sunlamp products for<br>tanning of the human skin. |
| Classification Name: | Sunlamp products and ultraviolet lamps<br>intended for use in sunlamp products |
| Classification Code: | LEJ |
| Regulation Number: | 21 CFR Part 878.4635 |
| Device Classification | II |
{4}------------------------------------------------
#### 807.92(a)(3)
#### Predicate Devices:
Prior to September 2, 2014, the basic ultraviolet sunlamps named in this 510(k) existed and were offered for sale as legally marketed, Class I, 510(k) exempt, medical devices.
For the basis of this 510(k), Cosmedico Light claims substantial equivalence to these legally marketed devices. Since none of the ultraviolet lamps named in this 510(k) have changed in 1) intended use or 2) technological design characteristics after September 2, 2014, essentially the contemporary ultraviolet lamps are identical to the predicate devices (i.e., those basic sunlamps offered for sale prior to the FDA's cutoff date of Sept. 2, 2014).
Such an approach of using the legally marketed Class I medical devices as predicate devices is validated in the Final Reclassification Order published on June 2, 2014 in the Federal Register Vo. 79, No. 105, Page 31212 whereby it is stated that:
" FDA cleared several 510(k)s for sunlamp products prior to exempting the devices from premarket notification submission. At least one 510(k) for a sunlamp product has been cleared since then under product code LEJ. These cleared sunlamp products, as well as any 510(k)-exempt sunlamp product or UV lamp intended for use in a sunlamp product legally offered for sale on or before September 2, 2014, can serve as predicates for substantial equivalence purposes."
lt will be demonstrated in this 510(k) that the named ultraviolet lamps:
- 1) have the same intended use as the predicate devices
- 2) have the same technological characteristics as the predicate devices and
- 3) conform to the special controls required by the reclassification order
In short, the contemporary ultraviolet sunlamps of this 510(k) are as safe and as effective as the predicate devices.
#### Description of the Devices
This 510(k) applies to more 35 devices that fall into a generic class of ultraviolet lamps that have the same basic technological features and exactly the same intended use. The ultraviolet sunlamps of this 510(k) are classified scientifically as high pressure, mercury-metal halide discharge devices. The general lamp construction and technological principle of operation of all of the devices of this 510(k) is the similar to that of commonly used metal halide lamps used for
{5}------------------------------------------------
general lighting purposes. The main differentiators for the various sunlamps of this 510(k) include: lamp length, lamp diameter, lamp wattage, spectral characteristics and private labeling thereof.
The ultraviolet lamps named herein comprise a tubular quartz envelope, filled with an inert gas (such as argon) and a mixture of mercury and metal halide salts and sealed at both ends.
In principle the radiation created by the metal halide lamp is generated in the same way as in other discharge lamps. The main difference between these high pressure metal halide lamps and the low pressure discharge lamps also used as sunlamp products is in the main mechanism for production of the UV radiation. A low pressure sunlamp depends upon the conversion of short wavelength radiation inside the lamp envelope to higher wavelength radiation (UVA primarily) by a phosphor coating. The metal halide lamp (MHL) derives its radiation characteristics through the direct excitation of mercury and specific metal vapors that are created in the envelope during lamp run-up and stabilized operation. The selection of the particular metal halides, the fill gas composition, and the fill gas pressure are the main determinants of the UV output of the lamp. The physical construction: overall quartz length, diameter, and electrode spacing are variables that are design choices for the engineer in consideration of the lamp power and application requirements. However, it is the selection of the composition of the metallic salts dosed into the lamp envelope during manufacturing that is chief determinant of the spectral output of the lamp.
As with other components, the development of and final design of the metal halide sunlamp is quite dependent upon the application. Depending upon the UV requirements of the tanning equipment, a number of design components may be varied. One of the more critical application related determinants is the fixture design. Among the most critical aspects of the fixture are housing size, reflector type and shape and the cooling arrangement and capabilities. The existence of a myriad of MHL fixtures in the market has led to a wide variety of lamps with different power levels and basing options. MH lamps typically range in power ratings from about 250 W up to 1500 W. Additionally, the lamps come fitted with bases of three main types: Double Ended (clip-in or snap-in) technically named R7s, Single Ended (plug-in) with bases designated as GY9.5, and lamps with Wire Leads (also called cable terminations).
{6}------------------------------------------------
For illustrative purposes, the schematic below points out the main parts of a typical metal halide sunlamp:
Image /page/6/Figure/2 description: This image shows a metal halide lamp with labels pointing to its different parts. The lamp has a quartz tube filled with argon, mercury, and metal halide salts. The lamp also has a ceramic base, molybdenum foil, a tipping nub for exhaust and filling, and a tungsten electrode.
#### Intended Use/Indications for Use
INTENDED USE: This ultraviolet lamp is intended for use in sunlamp products for tanning of the human skin.
### Technological Characteristics
The technological characteristics of the ultraviolet sunlamps which are the subject of this 510(k) are identically the same as the technological characteristics of the predicate devices. The tables and data that are reported in Section 10 and the information summarized in Section 12, "Substantial Equivalence Discussion" of this 510(k), document that there are no significant differences between the contemporary and predicate devices. In fact, the subject/contemporary and the predicate devices are "one in the same".
{7}------------------------------------------------
## Performance Testing Summary
- . All lamps are 100% tested for functionality at the end of the manufacturing process. Lamps that do not meet critical characteristics for functionality and safety are rejected.
- According to a Quality Sampling Plan, lamps are tested in-process and / . or post-production for:
- > Electrical Characteristics and Safety
- > Mechanical Safety
- > Dimensional Integrity
- > Physical Design Attributes
- > UV radiation Characteristics, including:
- . UVA Irradiance
- . Erythemal Effective Irradiance
- · UVC/UVB ratio per 21 CFR 1040.20
As noted, the subject lamps of this 510(k) are identical to the predicate devices. Therefore, the subject lamps and the predicate devices have the same performance requirements.
| Test (Performance Criteria) | Lamps meet criteria of predicate<br>devices (Yes/No) |
|-----------------------------------|------------------------------------------------------|
| Functionality / Light up | Yes |
| Electrical Characteristics | Yes |
| Burn-in Behavior | Yes |
| Dimensional Characteristics | Yes |
| Mechanical Safety | Yes |
| Electrical Safety | Yes |
| UVA Irradiance | Yes |
| Erythemal Effective Irradiance | Yes |
| UVC/UVB Ratio (per 21 CFR1040.20) | Yes |
{8}------------------------------------------------
# TABLE F-1
#
| Item Part Number | Description / Brand |
|------------------|-------------------------------|
| 900612555 | ERGOLINE ULTRA 520W |
| 913103645 | CLEO HPA 1200 FX |
| 913132641 | CLEO HPA 250-500 / 30 SD L |
| 913134045 | CLEO HPA 250-500 SE FX |
| 913137141 | CLEO HPA 400 / 30 SD L |
| 913138841 | CLEO HPA 400 / 30 S L |
| 919220245 | CLEO HPA 400 S |
| 919300145 | CLEO HPA 400 / 30 SDC |
| 920669541 | CLEO HPA 700 S FX |
| 920677041 | CLEO HPA 250-500 / 30 SDC |
| 938670245 | CLEO HPA 1010 SE FX |
| 996000101 | ERGOLINE JK 60/80 EF |
| 21104-1 | JK-RUSA 400W R7s |
| 21112-1 | Cosmedico M 650W SE GY9.5 |
| 24003-1 | Cosmedico M 400W R7s |
| 24004-1 | Cosmedico M 400W L |
| 24006-1 | Cosmedico M 500W SE |
| 24008-1 | Cosmedico M 500W L |
| 24022-1 | Cosmedico M 1000W SBSN GY 9.5 |
| 24103-1 | Cosmedico M 1000W L |
| 24171-1 | Cosmedico M 1000W SE GY9.5 |
| 24178-1 | Cosmedico M 800W R7s |
| R02362 | Radiance HP 600-800W GY9.5 |
| R02600 | Radiance HP 1000-1400W L |
| R02606 | Radiance HP 630W P |
| R02609 | Radiance HP 300-520W GY9.5 |
| R02610 | Radiance HP 800-1000W GY9.5 |
| R02611 | Radiance HP 620W L |
| R02614 | Radiance HP 600-650W GY9.5 |
| R02615 | Radiance HP 800-1000W GY9.5c |
| R03060 | Radiance HP 400-500W R7s |
| SQ01400LWL | SUPRA Q-01400LWL |
| SQ01400WL | SUPRA Q-01400WL |
| SQ01400WL4 | SUPRA Q-01400WL4 |
| SQ03520PIJK | SUPRA Q-03520PIJK |
| SQ04500SI | SUPRA Q-04500SI |
| SQ0600PIGYL | SUPRA Q-0600PIGY-L |
| SQ06630R | SUPRA Q-06630R |
| SQ08100PIGYL | SUPRA Q-08100PIGY-L |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.