K152090 · Sterilmed, Inc. · PNE · Mar 23, 2016 · Cardiovascular
Device Facts
Record ID
K152090
Device Name
Reprocessed Steerable Introducer
Applicant
Sterilmed, Inc.
Product Code
PNE · Cardiovascular
Decision Date
Mar 23, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The reprocessed steerable introducer devices are indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
Device Story
Reprocessed steerable introducer consists of steerable sheath and plastic dilator; used to introduce cardiovascular catheters into heart, including left side via interatrial septum. Device enters through right femoral vein; guided into right atrium. Handle knob controls distal tip curl direction/amount. Shaft contains radiopaque material for fluoroscopic visualization. Guidewire (purchased off-the-shelf) used for delivery; tapered dilator enlarges septal hole. Proximal end features hemostasis valve to minimize blood loss; side port with three-way stopcock allows aspiration, fluid infusion, blood sampling, and pressure monitoring. Used in clinical settings by physicians. Reprocessing restores performance characteristics of used devices to match original predicate specifications.
Clinical Evidence
Bench testing only. No clinical or animal testing conducted. Performance testing included cap retention, shaft deflection/flexation, joint strength, leak, shaft torque, stiffness, distribution, torque response, dimensional verification, and systems use. Biocompatibility testing performed per ISO 10993-1 for short-duration blood contact (<24 hours), including cytotoxicity, sensitization, irritation, acute systemic toxicity, hemocompatibility, thrombogenicity, chemical pyrogens, hemolysis, and immune response.
Technological Characteristics
8.5F asymmetrical bi-directional steerable introducer; 61 or 71 cm length. Features hemostasis valve and side port with three-way stopcock. Shaft contains radiopaque material for fluoroscopy. Sterilized via ethylene oxide to SAL 10^-6. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
St. Jude Agilis NxT Steerable Introducers (K061363)
St. Jude Agilis NxT Steerable Introducers (K081645)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 23, 2016
Sterilmed, Inc. % Ming Chew Regulatory Consultant Libra Medical Inc. 8401 63rd Avenue North, Suite 63 Brooklyn Park, Minnesota 55428
Re: K152090
Trade/Device Name: Reprocessed Steerable Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: PNE Dated: February 25, 2016 Received: February 26, 2016
Dear Ming Chew:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Page 3 - Ming Chew
## Attachment
The following models are included in the clearance of K152090:
G408318 G408319 408309 408310 G408324
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "K152090" at the top, followed by a logo with the text "STERILMED.". Below the logo, the text "24 JULY 2015" is displayed. The logo features a circular design with green and gray colors, suggesting a medical or sterilization-related theme. The text is clear and legible, indicating a label or informational graphic.
| 5. INDICATIONS FOR USE STATEMENT | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
| Indications for Use | |
| 510(k) Number ( <i>if known</i> ) | K152090 |
| Device Name | Reprocessed Steerable Introducers |
| Indications for Use ( <i>describe</i> ) | |
| The reprocessed steerable introducer devices are indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. | |
| Type of Use ( <i>select one or both, as applicable</i> ) | |
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (Part 21 CFR 801 Subpart C</label> ) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to: | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Sterilmed logo, which includes a green and silver circular design with the word "STERILMED" in gray text. Below the logo, the text "23 MARCH 2016" is displayed in a simple, sans-serif font. The text is likely a date associated with the logo or the image itself. The overall design is clean and professional.
# 1. 510(K) SUMMARY
#### 1.1 ADMINISTRATIVE INFORMATION
### 1.1.1 Sponsor
Patricia Kaufman Sterilmed, Inc. 5010 Cheshire Parkway N, Suite 2 Plymouth, MN 55446
Tel: (763) 488-3211 Fax: (763) 488-2050 Email: tschrater@sterilmed.com
### 1.1.2 Primary Contact
Ming Cheng Chew Regulatory Consultant Libra Medical, Inc. 8401 73rd Ave North, Suite 63 Minneapolis, MN 55428 Tel: 763-232-3701 Fax: 763-477-6357 Email: mcchew@libramed.com
### 1.1.3 Secondary Contact
Julie Bodmer Regulatory Consultant Libra Medical, Inc. 8401 73rd Ave North, Suite 63 Minneapolis, MN 55428 Tel: 612-910-3412 Fax: 763-477-6357 Email: jbodmer@libramed.com
#### 1.2 DEVICE NAME
Reprocessed Steerable Introducer Trade Name Common Name Steerable Introducer Classification Name Reprocessed Catheter Introducer Classification II PNE Primary Product Code
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Sterilmed logo, which includes a stylized symbol with a green and silver color scheme. The text "STERILMED" is written in a simple, sans-serif font next to the symbol. Below the logo, the date "23 MARCH 2016" is printed in a larger, serif font.
Secondary Product Code DYB
### 1.3 DEVICE DESCRIPTION OVERVIEW
The reprocessed steerable introducer consists of a steerable sheath and plastic dilator, which is indicated for introducing various cardiovascular catheters including the left side of the heart through the interatrial septum. The device is always delivered over a guidewire. The steerable introducer device enters through the right femoral vein and is guided into the right atrium of the heart. A knob on the introducer handle dictates the amount and direction of curl at the distal tip of the sheath. The shaft of the introducer is filled with radiopaque material so the device can be seen using fluoroscopy. A guidewire pokes through the septum into the left atrium and a tapered dilator enlarges the hole. The dilator is then removed allowing the introduction of a catheter.
Note: The guidewire is not included in the scope of this submission as it is purchased off-theshelf (K935170) and packaged with the reprocessed devices. Only the steerable sheath and plastic dilator are subject in this submission. In this submission references to the "device(s)" refers to both the sheath and dilator.
### 1.4 INDICATIONS FOR USE
The Reprocessed Steerable Introducer is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
### TECHNOLOGY OVERVIEW 1.5
The device is an 8.5F asymmetrical bi-directional steerable introducer with a varying amount of curl (small, medium or large) at the distal tip and a useable length of 61 or 71 cm. The proximal end of the device sheath is fitted with a hemostasis valve to minimize blood loss during catheter insertion and/or exchange over a guidewire. A side port with three-way stopcock is provided for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. A handle is equipped with a rotating collar to deflect the large curl 90 ° in the counterclockwise direction and 180° in the clockwise direction. The sheath is filled with radiopaque material for visualization under fluoroscopy.
### 1.6 PERFORMANCE TESTING OVERVIEW
The reprocessed steerable introducer devices are substantially equivalent to the new unreprocessed predicate devices (K061363 and K081645) in that the reprocessed devices are the same as the new un-reprocessed predicate devices in terms of form, fit, function, and intended use. No animal or clinical testing was conducted for this premarket notification submission. Physical and performance testing included:
- Cap Retention
- Shaft Deflection and Flexation
- Joint Strength
- Leak ●
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Sterilmed logo, which includes a stylized green and silver arrow. The text "STERILMED" is written in a simple, sans-serif font to the right of the arrow. Below the logo, the date "23 MARCH 2016" is written in a larger, serif font.
- Shaft Torque ●
- Stiffness ●
- Distribution ●
- Torque Response ●
- Dimensional Verification ●
- Systems Use ●
In addition, the device was tested for biocompatibility per ISO 10993-1 for short duration contact with blood (<24 hours). Biocompatibility testing included:
- Cytotoxicity
- Sensitization ●
- Irritation/Intracutaneous Reactivity ●
- Acute Systemic Toxicity
- Hemocompatability ●
- Thrombogenicity
- Chemical Pyrogens
- Hemolysis ●
- Immune Response
The device is sterilized by ethylene oxide to an SAL 10-6 level. These performances are similar to that described by the predicate device.
### 1.7 SUBSTANTIAL EQUIVALENCE
Sterilmed's reprocessed steerable introducer devices are substantially equivalent to the predicate devices, St. Jude Agilis"14 NxT Steerable Introducers. There are no changes to the clinical applications, patient population, performance specifications, or method of operation. The reprocessing does not change the design or function of the device, but restores the performance characteristics of a used device. After reprocessing, the devices are identical in form, fit, functionality and intended use as the predicate devices (K061363 and K081645).
### 1.8 CONCLUSION
The reprocessed steerable introducer devices have identical indications for use and technological characteristics as the predicate devices. Functional testing has shown the reprocessing does not affect the performance of the devices, nor the safety of the devices. Therefore, the reprocessed steerable introducer devices and the predicate devices are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.