K152004 · Nanova Biomaterials, Inc. · EBF · Nov 17, 2015 · Dental
Device Facts
Record ID
K152004
Device Name
Nanova Universal Composite
Applicant
Nanova Biomaterials, Inc.
Product Code
EBF · Dental
Decision Date
Nov 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Nanova Universal Dental Composite is indicated for use in: - Direct anterior and posterior restorations (including occlusal surfaces) - Core Build-ups - Splinting - Indirect restorations including inlays, onlays and veneers
Device Story
Nanova Universal Dental Composite is a radiopaque, methacrylate-based, visible-light-activated restorative material. Supplied in twist syringes or ampules, it contains methacrylate resins, photo-initiators, and approximately 77% inorganic filler by weight. Used by dental professionals in clinical settings for direct and indirect tooth restorations, core build-ups, and splinting. Upon irradiation with a dental curing light, the material undergoes polymerization, forming a hard composite bonded to tooth structure via dental adhesive. The device provides a restorative material for tooth shade correction and structural repair, benefiting patients by restoring tooth function and aesthetics.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and biocompatibility assessment. Bench tests included radiopacity, color stability, flexural/compressive strength, depth of cure, microtensile strength, water sorption/solubility, Knoop hardness, degree of conversion, surface gloss, and roughness. Biocompatibility testing included ISO agarose overlay, MEM elution (L-929 mouse fibroblasts), intracutaneous irritation, and guinea pig maximization sensitization tests. All tests passed.
Technological Characteristics
Methacrylate-based resin composite; 77% inorganic filler loading; visible-light-activated polymerization. Dimensions/form factor: twist syringe and ampule. Standards: ISO 4049:2009(E), ADA specification 27, ISO 10993 (biocompatibility).
Indications for Use
Indicated for direct anterior and posterior restorations (including occlusal surfaces), core build-ups, splinting, and indirect restorations (inlays, onlays, veneers).
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of a human face in profile, with three overlapping profiles to represent the department's focus on health, human services, and the well-being of the nation. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 17, 2015
Nanova Biomaterials, Inc. Mr. Andrew Ritts Senior Research Scientist 3806 Mojave Ct Columbia, Missouri, 65202
Re: K152004
Trade/Device Name: Nanova™ Universal Dental Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: II Product Code: EBF Dated: October 12, 2015 Received: October 19, 2015
Dear Mr. Ritts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Nanova Biomaterials. Inc. Nanova™ Universal Dental Composite Dental Composite 510(k) Notification
Section 4. Indications for Use Statement
(As Required by 21 CFR 807.87(e))
510(k) Number (if known): K152004
Device Name: Nanova 204 Universal Dental Composite
Indications for Use:
Nanova Universal Dental Composite is indicated for use in:
- · Direct anterior and posterior restorations (including occlusal surfaces)
- Core Build-ups
- · Splinting
- · Indirect restorations including inlays, onlays and veneers
Prescription Use X X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
:
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE F NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "NANOVA" in blue, with each letter in a bold, sans-serif font. The letters are evenly spaced and aligned horizontally. The overall impression is clean and modern.
Nanova Biomaterials, Inc. Nanova™ Universal Dental Composite Dental Composite K152004
## Section 5. 510(k) Summary
- 1) Submitted By: Nanova Biomaterials, Inc 3806 Mojave Ct Columbia, MO 65202 USA 573-875-6682
| | Contact Person: Andrew Ritts<br>Secondary Contact: Liang Chen | Phone: (573) 823-3114<br>Phone: (573) 239-8952 |
|----|---------------------------------------------------------------|---------------------------------------------------------------------|
| 2) | Establishment Registration No.: | 3011430871 |
| 3) | Date Prepared: | October 12, 2015 |
| 4) | Device Trade Name: | Nanova™ Universal Dental Composite |
| 5) | Device Common Name: | Universal Dental Composite |
| 6) | Device Classification Name: | Material, Tooth Shade, Resin<br>Product code: EBF, Reg. #: 872.3690 |
| 7) | Classification Panel: | Dental |
| 8) | Device Class: | Class II |
- 9) Predicate Devices: Nanova™ Universal Dental Composite is believed to be substantially equivalent to the following marketed products: Filtek Supreme Ultra Universal Restorative (K083610) product code EBF manufactured by 3M ESPE.
#### 10) Indication for Use: Nanova Universal Dental Composite is indicated for use in:
- Direct anterior and posterior restorations (including occlusal surfaces)
- Core Build-ups
- Splinting ●
- . Indirect restorations including inlays, onlays and veneers
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "NANOVA" in a sans-serif font. The word is written in a bright blue color. The letters are all capitalized and evenly spaced. The background is white.
Nanova Biomaterials, Inc. Nanova™ Universal Dental Composite Dental Composite K152004
## Section 5. 510(k) Summary - Cont.
- 11) Device Description:
Nanova™ Universal Dental Composite is a methacrylate based, visible-light activated, radiopaque, universal composite. This device is available in twist syringe and ampule packaging. Nanova™ Universal Dental Composite is available in a variety of tooth colored shades. Nanova"™ Universal Dental Composite contains methacrylate resins, photo-initiator, and inorganic filler material. Inorganic filler loading is approximately 77% by weight.
When irradiated by light, the methacrylate functionalities of the resins and surface-treated fillers undergo, in conjunction with the photo-initiator system, a light-induced polymerization to form a hard composite that is bonded to the tooth structure with a permanent dental adhesive.
Nanova™ Universal Dental Composite formulation is based off of several FDA approved commercially available universal dental composites. The formulation was modified to improve physical properties, such as flexural strength, and handling properties. As a result of the reformulation, a biocompatibility assessment was developed for Nanova™ Universal Dental Composite using standard risk assessment techniques and consideration of FDA & International Standards, including ISO 10993 Parts 5 and 10.
- 12) Substantial Equivalence:
The document, "Guidance on the CDRH Premarket Notification Review Program, 6/30/86 (K86-3)" was used to determine substantial equivalence:
a) The applicant device has the same intended use as the 510(k) cleared predicate listed above.
b) The technological characteristics of this product are believed to be substantially equivalent as those for the predicate device and other methacrylate based products currently on the market. Table 5.1 below shows a comparison of Nanova™ Flowable Composite and the predicate.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "NANOVA" in a sans-serif font. The word is written in blue. The letters are slightly blurred, giving the image a soft focus.
# Section 5. 510(k) Summary - Cont.
Table 5.1 Technical Comparison of NanovaTM Universal Dental Composite and Filtek Supreme Ultra Universal Restorative (K083610)
| Name | Nanova™ Universal Dental Composite | Filtek Supreme Ultra Universal Restorative (K083610) | Comparison |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Indications for use | Direct anterior and posterior restorations (including occlusal surfaces)<br>Core Build-ups<br>Splinting<br>Indirect restorations including inlays, onlays and veneers | Direct anterior and posterior restorations (including occlusal surfaces)<br>Core Build-ups<br>Splinting<br>Indirect restorations including inlays, onlays and veneers | Indicated for the same purposes |
| Composition | Methacrylate resins, photo-initiators, inorganic fillers | Methacrylate resins, photo-initiators, inorganic fillers | Chemistries are similar |
| Flexural Strength | ISO 4049:2009 (E) | ISO 4049:2009 (E) | Both above 100 MPa |
| Compressive Strength | ADA specification 27 | ADA specification 27 | Both above 300 MPa |
| Depth of Cure | Passes ISO 4049:2009(E) | Passes ISO 4049:2009(E) | Passed |
| Water Sorption | ISO 4049:2009 (E) | ISO 4049:2009 (E) | Nanova Less Adsorption |
| Water Solubility | ISO 4049:2009 (E) | ISO 4049:2009 (E) | Nanova Less Solubility |
| Packaging | Twist Syringe and Ampule | Twist Syringe and Ampule | Both use twist syringe and ampules |
Nanova™ Universal Dental Composite fulfilled ISO 4049 requirements as well as FDA 's guidance ucm071576 Dental Composite Resin Devices. The strength and physical properties show Nanova™ Universal Dental Composite performs equivalently to the predicate device. ISO 10993 biocompatibility (including cytotoxicity, irritation, and sensitization) confirms the material is biocompatible. The fulfillment of ISO 4049 and ISO 10993 show Nanova Universal Dental Composite is substantially equivalent to the predicate.
These differences do not affect the equivalence.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "NANOVA" in blue font. The letters are bold and slightly blurred. The background is white.
Nanova Biomaterials, Inc. Nanova™ Universal Dental Composite Dental Composite K152004
#### 13) Non-Clinical Performance Testing:
Non-clinical and biological testing was completed to assess its performance and biocompatibility to support substantial equivalence. The data provided in this 510(k) submission shows that the composition is biocompatible based on the biocompatibility assessment conducted based on ISO 10993 and benchtop assessment based on UCM071631, ADA specification 27, and ISO 4049. All biocompatibility tests completed were passed.
| Name | Result |
|--------------------------------------------------------|--------|
| Radiopacity Evaluation | Passed |
| Color Stability | Passed |
| Flexural Strength | Passed |
| Compressive Strength | Passed |
| Depth of Cure | Passed |
| Microtensile Strength | Passed |
| Water Sorption and Solubility | Passed |
| Knoop Hardness | Passed |
| Double Bond Conversion (Degree of Conversion) | Passed |
| ISO Agarose Overlay Using L-929 Mouse Fibroblast Cells | Passed |
| ISO Intracutaneous Irritation Test | Passed |
| ISO Guinea Pig Maximization Sensitization Test | Passed |
| ISO MEM Elution Using L-929 Mouse Fibroblast Cells | Passed |
| Shelf life | Passed |
| Surface Gloss | Passed |
| Surface Roughness | Passed |
Table 5.2 List of Tests Completed on Nanova™ Universal Composite
- 14) Clinical Performance Testing: Clinical performance data was not included.
## Conclusion:
Nanova Biomaterials Inc. believes that Nanova™ Universal Dental Composite is substantially equivalent to currently legally marketed products. It does not introduce new indications for use, has similar technological characteristics and does not introduce new potential hazards or risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.