← Product Code [GCJ](/productcode/GCJ) · K151990

# Tissue Removal System (K151990)

_Boehringer Laboratories, LLC · GCJ · Sep 11, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151990

## Device Facts

- **Applicant:** Boehringer Laboratories, LLC
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Sep 11, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Tissue Removal System is intended for the containment and removal of tissue during minimally invasive surgical procedures.

## Device Story

Tissue Removal System facilitates contained tissue removal during minimally invasive surgery. System consists of a tissue retrieval pouch (previously cleared) and a manual sectioning device. During surgery, pouch is placed in peritoneal space via port; resected tissue is placed inside for isolation. Pouch opening is exteriorized. Sectioning device, comprising a blunt wire, flexible passer, and handle with blunt hook, is used to divide tissue manually within the contained environment. Passer encircles specimen with wire; back-and-forth motion of wire divides tissue. System allows specimen removal en bloc or in smaller pieces without sharp instruments. Used by surgeons in clinical settings. Benefits include maintaining containment integrity during tissue division, reducing risk of tissue dissemination compared to uncontained methods.

## Clinical Evidence

No clinical data. Bench testing only. Testing included simulated use, extreme use, bubble leak, dye penetration, viral barrier, puncture resistance, tear resistance, and film cutting tests to verify containment integrity and mechanical performance.

## Technological Characteristics

System includes a polymer-based tissue retrieval pouch and a manual sectioning device (blunt wire, flexible passer, handle with blunt hook). Operates via manual mechanical force. Single-patient use; provided sterile. No energy source or software components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Tissue Removal Pouch ([K150781](/device/K150781.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other, resembling a stylized bird.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 11, 2015

Boehringer Laboratories, LLC Mr. Christopher Radl Engineering, Product Development Manager 300 Thoms Drive Phoenixville, Pennsylvania 19460

Re: K151990

Trade/Device Name: Tissue Removal System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: July 17, 2015 Received: July 20, 2015

Dear Mr. Radl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{1}------------------------------------------------

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Binita S. Ashar -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K151990

Device Name

Tissue Removal System

Indications for Use (Describe)

The Tissue Removal System is indicated for the contained removal of tissue, including contained manual tissue sectioning, during minimally invasive surgical procedures.

Contraindications:

The Tissue Removal System is contraindicated for laparoscopic power morcellation during gynecologic procedures.

The Tissue Removal System is contraindicated for use with powered cutting devices (e.g., power morcellators, electrosurgical and laser instruments), and when, in the judgment of the physician, use of such a device would be contrary to the best interest of the patient.

Type of Use (Select one or both, as applicable)

| <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Prepared: July 17, 2015

#### APPLICANT INFORMATION:

Name: Boehringer Laboratories. LLC Address: 300 Thoms Drive Phoenixville, PA 19460 Phone: 610-278-0900 Fax: 610-278-0907 Contact: Christopher Radl; Engineering, Product Development Manager

Signature: Date:

July 17, 2015

## TRADE NAME:

Tissue Removal System

#### COMMON NAME:

Laparoscopic Retrieval System

#### Device Classification:

Accessory to Endoscope, Class II Product Code: GCJ Regulation: 876.1500 Endoscope and Accessories Classification Panel: General & Plastic Surgery

#### PREDICATE DEVICES:

Predicate: Tissue Removal Pouch K150781

#### Device Description:

The Tissue Removal System is comprised of the Tissue Removal Pouch cleared under K150781 and an additional accessory Sectioning Device. It is provided sterile, and is for single patient use.

The Tissue Removal Pouch of the Tissue Removal System includes a bag with a single large opening held open by a thin, opening ring. During laparoscopic surgery, the bag is placed within the peritoneal space via an existing port site. Tissue that has been resected is placed within the bag for isolation, containment and removal from the patient. The bag opening is exteriorized. Two halves of the Tissue Removal Pouch flip ring are attached to each other and to the bag is rolled around the flip ring by inverting the flip ring, in order to draw the resected tissue toward the port site.

The resected tissue is then removed en bloc (for smaller specimens), or by first dividing the specimen using the accessory Sectioning Device. The Sectioning Device is comprised of a blunt wire, a flexible passer and a handle with a blunt hook. The passer is used to encircle the specimen with the wire. The handle includes a blunt hook that can be used to aid in the advancement of the passer. The specimen is divided by the manual application of a back and forth motion to the wire. Use of the accessory Sectioning Device permits division of the specimen while maintaining the containment effectiveness of the Tissue Removal Pouch

{4}------------------------------------------------

### INTENDED USE:

The Tissue Removal System is intended for the containment and removal of tissue during minimally invasive surgical procedures.

#### COMPARISON SUMMARY:

The Tissue Removal System is identical to the predicate Tissue Removal Pouch with the exception of the accessory Sectioning Device. With the predicate Tissue Removal Pouch, the only way to divide large specimens for removal was with the use of sharp instruments such as scalpels. In comparison, the accessory Sectioning Device of the Tissue Removal System allows for specimen division with the use of a blunt wire.

#### Non-Clinical Testing:

The following non-clinical tests have been included with this submission:

- Simulated Use Testing followed by Bubble Leak, Dye Penetration and Viral Barrier Testing, ● demonstrating that the Tissue Removal System performs as intended and that the Tissue Removal Pouch maintains containment integrity after use with the accessory Sectioning Device.
- Extreme Use Testing followed by Bubble Leak. Dve Penetration and Viral Barrier Testing. . demonstrating that the Tissue Removal System maintains its integrity and containment effectiveness after sectioning of multiple tissue specimens, far in excess of actual clinical conditions.
- . Sectioning Device Puncture Resistance Testing demonstrating that the flexible passer of the Sectioning Device will not puncture the Tissue Removal Pouch of the Tissue Removal System.
- Sectioning Device Hook Tear Resistance Testing demonstrating that the blunt hook of the hooked handle will not tear the Tissue Removal Pouch of the Tissue Removal System.
- o Film Cutting Testing demonstrating that forcing the wire of the Sectioning Device into the film of the Tissue Removal Pouch does not result in a cut to the Tissue Removal Pouch.

## CONCLUSIONS:

As evidenced by the provided information, the Tissue Removal System is as safe and effective as, and performs as well as, or better than, the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K151990](https://fda-staging.innolitics.com/device/K151990)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).