The BOX DICOM Viewer™ is a software Teleradiology system used to receive DICOM images, scheduling information and textual reports, organize and store them in an internal formation available across a network via web and customized user interfaces. The BOX DICOM Viewer™ is used by hospitals, imaging centers, radiologist reading practices.
Device Story
BOX DICOM Viewer is a teleradiology software system for viewing DICOM-compliant medical images. Input: DICOM images, scheduling info, and reports received from third-party DICOM devices via a proxy listener. Processing: DICOM header extraction, database storage, and web-based rendering. Output: Visual display of medical images with tools for grayscale/RGB rendering, measurements (distance, area, angle, Cobb angle), annotations, and image manipulation (zoom, pan, flip, rotate). Used in hospitals, imaging centers, and reading practices by radiologists and qualified medical personnel. The system utilizes WebGL and hardware-accelerated rendering for performance. Healthcare providers use the viewer to interpret radiological data, facilitating clinical decision-making by providing convenient, network-accessible image access, replacing traditional film handling.
Clinical Evidence
Bench testing only. Verification and validation testing were conducted to evaluate input/output functions and software performance across all operational modes. Results demonstrated that the device met all predetermined acceptance criteria.
Technological Characteristics
Software-based teleradiology system; DICOM 3.x compliant; utilizes MySQL database; web-based interface (Chrome, IE, Firefox); WebGL API for 2D/3D graphics; hardware-accelerated rendering; supports high-resolution Retina displays; network connectivity via 10/100/100 Ethernet.
Indications for Use
Indicated for use by qualified medical personnel to receive, organize, store, and view DICOM images, scheduling information, and textual reports. Not intended for acquisition of mammographic image data.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 1, 2015
BOX, Inc. % Mr. Carl Alletto Consultant OTech, Inc. 8317 Belew Drive MCKINNEY TX 75071
Re: K151957
Trade/Device Name: BOX DICOM Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 18, 2015 Received: August 25, 2015
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K151957
Device Name BOX DICOM Viewer™
#### Indications for Use (Describe)
The BOX DICOM Viewer™ is a software Teleradiology system used to receive DICOM images, scheduling information and textual reports, organize and store them in an internal formation available across a network via web and customized user interfaces. The BOX DICOM Viewer™ is used by hospitals, imaging centers, radiologist reading practices.
Contraindications: The BOX DICOM Viewer™ is not intended for the acquisition of mammographic image data and is meant to be used by qualified medical personnel only who are qualified to create and diagnose radiological image data.
#### Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
#### Date Prepared:
June 16, 2015
#### Submitter's Information: 21 CFR 807.92(a)(1)
Mr. Crispen Maung, VP Compliance BOX INc. 4440 El Camino Real Los Altos, CA 94022 USA Email: compliance@ box.com
#### Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)
| Trade Name: | BOX DICOM Viewer™ |
|----------------------|--------------------------------------------|
| Common Name: | Picture, archive and communications system |
| Classification Name: | System, Image Processing, Radiological |
| Product Code: | LLZ |
### Predicate Device: 21 CFR 807. 92(a)(3)
| Device Classification Name | system, image processing, radiological23 |
|------------------------------|------------------------------------------|
| 510(k) Number | K132799 |
| Device Name | CLARISO PACS |
| Applicant | CLARISO, INC. |
| Regulation Number | 892.2050 |
| Classification Product Code | LLZ |
| Date Received | 09/06/2013 |
| Decision Date | 11/07/2013 |
| Decision | substantially equivalent (SESE) |
| Regulation Medical Specialty | Radiology |
| 510k Review Panel | Radiology |
| summary | summary |
| Type | Traditional |
| Reviewed by Third Party | No |
| Combination Product | No |
### Device Description: 21 CFR 807 92(a)(4)
The BOX DICOM Viewer™ is a software system to be used to view DICOM compliant studies, which are stored. The BOX DICOM Viewer™ is intended for professional use only as a viewing tool for medical image studies.
{4}------------------------------------------------
# 510(k) Summary
BOX Inc. has acquired Clariso Inc. which included Clariso PACS (K132799), and cleared on November 11, 2013. The full features and functions of Clariso have been imported to the BOX DICOM Viewer, which includes the CLARISO PACS Viewer technology and features;
- Grayscale Image Rendering ●
- RGB Image Rendering ●
- Localizer Lines ●
- Localizer Point ●
- Orientation Markers
- Distance Markers
- Study Data Overlays
- Stack Navigation Tool
- Window/Level Tool
- Zoom Tool
- Panning Tool
- Horizontal/Vertical Flip
- Clockwise/Counterclockwise Rotation
- Color Inversion
- Text Annotation
- Area Measurement Annotation ●
- Angle Measurement Annotation
- Cobb Angle Measurement Annotation ●
In addition, the BOX Inc. has;
- Added new styles, colors, fonts, and icons were added for "BOX" look-and-feel
- Added WebGL rendering optimizations,
- . Added Hardware accelerated rendering
- Added support for high resolution Retina displays,
- . Added keyboard shortcuts for all tools and all annotation types
The BOX DICOM VIEWER software allows for acquisition of images from DICOM devices and lets users view those images from their personal computers. A third party DICOM device sends to the Box DICOM Proxy listener, the files are then sent to the Upload Proxy, then to the DICOM processor where the DICOM header data is extracted. Finally, the BOX DICOM VIEWER will communicate with a database component to store all the information required for patients, users, studies and configuration settings.
### Indications for Use: 21 CFR 807 92(a)(5)
The BOX DICOM Viewer™ is a software Teleradiology system used to receive DICOM images, scheduling information and textual reports, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces. The BOX DICOM Viewer™ is used by hospitals, imaging centers, radiologist reading practices.
Contraindications: The BOX DICOM Viewer™ is not intended for the acquisition of mammographic image data and is meant to be used by qualified medical personnel only who are qualified to create and diagnose radiological image data.
{5}------------------------------------------------
### Technological Characteristics: 21 CFR 807 92(a)(6)
The BOX DICOM Viewer™ is a software application that handles and manipulates digital medical images. The device does not contact the patient, nor does it control any life sustaining devices.
A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed. In general, a PACS (Picture Archiving and Communication System) is a medical imaging technology which provides storage of, and convenient access to, images from multiple modalities. Electronic images and reports are transmitted digitally via the PACS; this eliminates the need to manually file, retrieve, or transport film jackets. The universal format for PACS image storage and transfer is DICOM 3.x (Digital Imaging and Communications in Medicine). Non-image data, such as scanned documents, may be incorporated using consumer industry standard formats like PDF (Portable Document Format), once encapsulated in DICOM. The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use.
The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices. Both systems have been developed to replace traditional film handling in radiology. The 2 devices are substantially equivalent in the areas of design, architecture, general function, application, and intended use. The predicate device and the new device are compared below:
The following information compares the predicate device and new device. Any differences between the predicate and the new device has no impact on safety or efficacy of the new device and does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
| Ref<br># | Functionality | Predicate:<br>CLARISO PACS<br>(K132799) | Subject Device:<br>BOX DICOM<br>Viewer | If different, Impact on<br>Safety and or Efficacy |
|----------|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Web Browser<br>Software | Google Chrome<br>for all features.<br>Microsoft Internet<br>Explorer &<br>Mozilla Firefox for<br>features except<br>the<br>DICOM Viewer | Same as<br>predicate | No differences between<br>predicate and subject<br>device. |
| 2 | Intended use | Acquiring,<br>viewing, editing<br>and storing<br>radiographs and<br>related patients<br>images | Same as<br>predicate | No difference |
| 3 | Intended user | Radiologist &<br>qualified medical<br>personnel | Same as<br>predicate | No difference |
| 4 | Network | 10/100/100<br>Ethernet | Same as<br>predicate | No difference |
| Ref<br># | Functionality | Predicate:<br>CLARISO PACS<br>(K132799) | Subject Device:<br>BOX DICOM<br>Viewer | If different, Impact on<br>Safety and or Efficacy |
| 5 | Monitor | Above 19inch<br>monitor (Using<br>1280x1024) | Same as<br>predicate | No difference |
| 6 | User interaction/input | Same, Using<br>1280x1024 | Same as<br>predicate | No difference |
| 7 | Import / export<br>images | Yes | Same as<br>predicate | No difference |
| 8 | Acquisition devices | CT, MR, US, PET | Same as<br>predicate | No difference |
| 9 | Image organization | Patient ID, Name,<br>study instance<br>UID | Same as<br>predicate | No difference |
| 10 | Image search<br>available | Same | Same as<br>predicate | No difference |
| 11 | Image storage | Yes | Same as<br>predicate | No difference |
| 12 | Database software | MySQL | Same as<br>predicate | No difference |
| 13 | Greyscale Image<br>Rendering | Yes | Same as<br>predicate | No difference |
| 14 | RGB Image<br>Rendering | Yes | Same as<br>predicate | No difference |
| 15 | Localizer Lines | Yes | Same as<br>predicate | No difference |
| 16 | Localizer Point | Yes | Same as<br>predicate | No difference |
| 17 | Orientation Markers | Yes | Same as<br>predicate | No difference |
| 18 | Distance Markers | Yes | Same as<br>predicate | No difference |
| 19 | Study Data Overlays | Yes | Same as<br>predicate | No difference |
| 20 | Stack Navigation | Yes | Same as<br>predicate | No difference |
| 21 | Window Level | Yes | Same as<br>predicate | No difference |
| 22 | Zoom in on images | Yes | Same as<br>predicate | No difference |
| 23 | Panning | Yes | Same as<br>predicate | No difference |
| 24 | Horizontal/Vertical<br>Flip | Yes | Same as<br>predicate | No difference |
| 25 | Clockwise/Countercl<br>ockwise rotate | Yes | Same as<br>predicate | No difference |
| 26 | Invert image | Yes | Same as<br>predicate | No difference |
| Ref<br># | Functionality | Predicate:<br>CLARISO PACS<br>(K132799) | Subject Device:<br>BOX DICOM<br>Viewer | If different, Impact on<br>Safety and or Efficacy |
| 27 | Text Annotation | Yes | Same as<br>predicate | No difference |
| 28 | Area measurement<br>annotation | Yes | Same as<br>predicate | No difference |
| 29 | Angle measurement<br>annotation | Yes | Same as<br>predicate | No difference |
| 30 | Cobb Angle<br>Measurement<br>Annotation | Yes | Same as<br>predicate | No difference |
| 31 | Image annotation | Yes | Same as<br>predicate | No difference |
| 32 | Security | Yes | Same as<br>predicate | No difference |
| 33 | DICOM 3.0<br>conformance | Yes | Same as<br>predicate | No difference |
| 34 | Worklist | Yes | Same as<br>predicate | No difference |
| 35 | Thumbnail viewing | Yes, thumbnails<br>on preview,<br>small, medium<br>and large | Same as<br>predicate | No difference |
| 36 | Login | Yes | Same as<br>predicate | No difference |
| 37 | Audit | Yes, a tool to<br>view access logs<br>in real time. | Same as<br>predicate | No difference |
| 38 | User Interface text<br>styles, colors, fonts,<br>and icons. | CLARISO PACS<br>styles | BOX Styles | Yes, there are differences,<br>however these changes do<br>not affect device functions<br>and does not raise new<br>potential safety risks.<br>Therefore, it is our<br>determination that there is<br>"No impact on safety or<br>efficacy" |
| 39 | WebGL rendering<br>optimizations | No hardware<br>acceleration. | Yes, Hardware<br>acceleration is<br>used. | Yes, there are differences.<br>See item 39 in section<br>7.1.1.1 below. |
| 40 | Support for high<br>resolution Retina<br>displays | Pixelated display<br>on high-DPI<br>displays only<br>(i.e., "Retina<br>Displays"). | Full pixel density<br>on all displays | Yes, there is a difference.<br>The subject device will<br>display the full pixel density<br>of the saved image where<br>the predicate device only<br>did so if it was set to<br>"Retina Display" mode. This<br>difference actually aids the<br>viewer to always see the<br>image as captured by the<br>modality. The difference |
| Ref<br># | Functionality | Predicate:<br>CLARISO PACS<br>(K132799) | Subject Device:<br>BOX DICOM<br>Viewer | If different, Impact on<br>Safety and or Efficacy |
| | | | | does not affect the device<br>IFU and does not raise new<br>potential safety risks.<br>Therefore, it is our<br>determination that there is<br>"No impact on safety or<br>efficacy" |
| 41 | keyboard shortcuts<br>for all tools and all<br>annotation types | Limited keyboard<br>shortcut support. | Keyboard<br>shortcuts allowed<br>for tools and all<br>annotation types. | Yes, there is a difference.<br>In the predicate, it was not<br>possible for the User to use<br>the inherent functions of the<br>operating system to create<br>keyboard short cuts. In the<br>subject device, keyboard<br>shortcuts allowed for tools<br>and all annotation types<br>which may help the User<br>view images. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# DIFFERENCE FOR ITEM 39 EXPLAINED
| Ref<br># | Functionality | Impact on Safety and or Efficacy |
|----------|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 39 | WebGL rendering<br>optimizations | WebGL (Web Graphics Library) is a JavaScript API for rendering<br>interactive 3D computer graphics and 2D graphics within any<br>compatible web browser without the use of plug-ins. WebGL is<br>integrated completely into all the web standards of Internet browsers,<br>allowing GPU accelerated usage of physics and image processing<br>and effects as part of the web page canvas. WebGL elements can be<br>mixed with other HTML elements and composited with other parts of<br>the page or page background. WebGL programs consist of control<br>code written in JavaScript and shader code that is executed on a<br>computer's Graphics Processing Unit (GPU). Google Chrome -<br>WebGL has been enabled on all platforms that have a capable<br>graphics card with updated drivers since version 9, released in<br>February 2011. By default, Windows Chrome uses the ANGLE<br>(Almost Native Graphics Layer Engine) renderer to translate OpenGL<br>ES to Direct X 9.0c or 11.0, which have better driver support. On<br>Linux and Mac OS X the default renderer is OpenGL however, it is<br>also possible to force OpenGL as the renderer on Windows. In<br>computing, hardware acceleration is the use of computer hardware to<br>perform some functions faster than is possible in software running on<br>a more general-purpose CPU. Examples of hardware acceleration<br>include blitting acceleration functionality in graphics processing units<br>(GPUs) and regular expression hardware acceleration for spam<br>control in the server industry. Normally, processors are sequential,<br>and instructions are executed one by one. Various techniques are<br>used to improve performance; hardware acceleration is one of them.<br>The main difference between hardware and software is concurrency,<br>allowing hardware to be much faster than software. Hardware<br>accelerators are designed for computationally intensive software<br>code. Depending upon granularity, hardware acceleration can vary |
{9}------------------------------------------------
## 510(k) Summary
| Ref<br># | Functionality | Impact on Safety and or Efficacy |
|----------|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | from a small functional unit to a large functional block (like motion<br>estimation in MPEG-2). The hardware that performs the acceleration,<br>when in a separate unit from the CPU, is referred to as a hardware<br>accelerator, or often more specifically as a 3D accelerator,<br>cryptographic accelerator, etc. Those terms, however, are older and<br>have been replaced with less descriptive terms like video card or<br>network adapter. |
| | | Yes, there are differences, between the predicate and the subject<br>device for WebGL since the subject device uses hardware<br>acceleration and the predicate does not. However this difference do<br>not affect the device IFU and does not raise new potential safety risks.<br>The device has been tested and has passed predetermined criteria<br>and therefore, it is our determination that there is "No impact on safety<br>or efficacy". |
#### Nonclinical Testing:
The BOX DICOM Viewer™ system and configuration has been assessed and tested at BOX Inc. and has passed all pre-determined testing criteria. The Verification & Validation Test Plan was designed to evaluate input functions, output functions, and actions performed by the BOX DICOM Viewer™ software in each operational mode and followed the process documented in the Validation Test Plan.
Nonclinical testing results are provided in the 510(k). Validation testing indicated, that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met. If the device is installed by BOX Inc., integration and installations verification tests are conducted aqainst acceptance criteria prior to release to the client.
#### Conclusion: 21 CFR 807 92(b)(1)
The 510(k) Pre-Market Notification for the BOX DICOM Viewer™ contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device.
The subject device will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The subject and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use. The modification to the subject device does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
Nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
Therefore, the BOX DICOM Viewer™ is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.