← Product Code [FZP](/productcode/FZP) · K151943

# Vas-Q-Clip (K151943)

_Nanova Biomaterials, Inc. · FZP · Mar 25, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151943

## Device Facts

- **Applicant:** Nanova Biomaterials, Inc.
- **Product Code:** [FZP](/productcode/FZP.md)
- **Decision Date:** Mar 25, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Vas-Q-Clip™ ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.

## Device Story

Vas-Q-Clip is a non-absorbable polyacetal ligating clip used for hemostasis; device functions by mechanically compressing vessels or tissue structures. Clips are housed in cartridges and applied by surgeons during surgical procedures. Design features include legs, bosses, hinge, and a latching mechanism with triangular teeth protrusions oriented perpendicular to the vessel to enhance grip and migration resistance. Device is supplied sterile via EtO sterilization. Output is the physical ligation of the target tissue; clinical benefit is effective hemostasis. Performance is verified through bench testing of migration resistance, vessel occlusion, application force, and axial pull-off force.

## Clinical Evidence

No clinical data provided. Evidence consists of bench-top performance testing including migration resistance, vessel occlusion, application force, axial pull-off force, and biocompatibility (ISO 10993-10, ISO 10993-6). Sterility and packaging integrity verified via USP 85, USP 151, and seal strength testing.

## Technological Characteristics

Non-absorbable polyacetal material; injection molded. Features triangular teeth protrusions oriented perpendicular to the vessel. Supplied in cartridge/blister packaging with Tyvek lidding. Sterilized via EtO (SAL 10^-6). Available in ML, L, and XL sizes.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Hem-O-lok® Ligating Clip ([K993157](/device/K993157.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and care. The logo is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2016

Nanova Biomaterials, Inc. Mr. Andrew Ritts Senior Research Scientist 3806 Mojave Ct. Columbia. MO 65202

Re: K151943

Trade/Device Name: Vas-Q-Clip Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: February 23, 2016 Received: February 26, 2016

Dear Mr. Ritts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K151943

Device Name Vas-Q-Clip^TM

Indications for Use (Describe)

Vas-Q-Clip' TM ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.

| Type of Use (Select one or both, as applicable)                                                                                            |                                                                                                          |
|--------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span> <span>   </span> Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "NANOVA" in blue, block letters. The letters are evenly spaced and appear to be a sans-serif font. The background is white, providing a clear contrast to the blue text.

#### Section 5. 510(k) Summary

- 1) Submitted By: Nanova Biomaterials, Inc 3806 Mojave Ct Columbia, MO 65202 USA 573-875-6682 Contact Person: Andrew Ritts Phone: (573) 823-3114 Secondary Contact: Richard Lebens Phone: (573) 875-6682 2) Establishment Registration No.: 3011430871 3) Date Prepared: December 14, 2015 Vas-Q-Clip™ 4) Device Trade Name: ર) Device Common Name: Ligation Clip Implantable Clip/ 878.4300/ FZP 6) Device Classification Name/ Regulation #/Product Code: Classification Panel: 7) General and Plastic Surgery Device Class: Class II 8)
- 9) Predicated Devices: Vas-Q-Clip™ is believed to be substantially equivalent to Hem-O-lok® Ligating Clip (K993157).
- 10) Indication for Use: Vas-Q-Clip™ ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "NANOVA" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The overall impression is clean and modern.

### Section 5. 510(k) Summary - Cont.

- 11) Device Description:
Nanova Biomaterials Inc.'s Vas-Q-Clip™ causes hemostasis through vessel ligation. The technological characteristics are the same as or equivalent to the predicate device. The non-absorbable polyacetal material used in the clips is the same as the predicate device and the clips are manufactured in a similar manner.

The clips are housed in a cartridge and package in a ridged plastic plaster with Tyvek coated lidding and is sold sterile. The method of sterilization is EtO with a SAL of 10-6. The blisters are fitted into a carton in which the clips are sold.

- 12) Substantial Equivalence:
The document, "Guidance on the CDRH Premarket Notification Review Program, 6/30/86 (K86-3)" was used to determine substantial equivalence:

a) The applicant device has the same intended use as the 510(k) cleared predicate listed above.

b) The technological characteristics of this product are believed to be substantially equivalent as those for the predicate device. This device and its predicates are substantially equivalent in material and manufacturing methods. Vas-Q-Clip™ is packaged in a similar manner to its predicate and many 510k cleared products already on the market. A summary of the comparison is shown in Table 5.1 below.

The Vas-O-Clip™ outer structure including legs, bosses, hook and latch mechanism, and hinge are the same as the Hem-o-lok® design (Shape and dimensions are the same for the corresponding sizes). The only difference is the inner closure feature designated as "teeth". Vas-Q-Clip"M triangular protrusions (teeth) are perpendicular to the vessel whereas the Hem-O-Lok teeth are parallel.

Vas-O-Clip™ has the same Intended Use as the predicate devices to which it was compared, and there are no differences in technological characteristics which raise new questions of safety and / or effectiveness. Therefore, Vas-Q-Clip™ is substantially equivalent.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "NANOVA" in blue, block letters. The letters are bold and sans-serif. The word is centered and takes up most of the frame. The background is white.

| Name                 | Vas-Q-ClipTM                                                                                                                                                                                                                                                                                              | Hem-O-Lok® (K993157).                                                                                                                                                                                                                                                                                   | Comparison      |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Indications for use  | Vas-Q-ClipTM ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure. | Hem-O-Lok® ligating clips are intended for use in procedures involving ligation of vessels or tissue structures. Surgeons should apply the appropriate size clip for the size of the vessel or tissue structure to be ligated such that the clip completely encompasses the vessel or tissue structure. | Same            |
| Composition          | Polyacetal                                                                                                                                                                                                                                                                                                | Polyacetal                                                                                                                                                                                                                                                                                              | Same            |
| Processing           | Injection Molding                                                                                                                                                                                                                                                                                         | Injection Molding                                                                                                                                                                                                                                                                                       | Same            |
| Sterility            | EtO at 10-6 SAL                                                                                                                                                                                                                                                                                           | EtO at 10-6 SAL                                                                                                                                                                                                                                                                                         | Same            |
| Design               | Triangular protrusions are perpendicular to the vessel                                                                                                                                                                                                                                                    | Triangular protrusion are parallel to the vessel                                                                                                                                                                                                                                                        | Different       |
| Size                 | ML, L, XL                                                                                                                                                                                                                                                                                                 | M, ML, L, XL                                                                                                                                                                                                                                                                                            | No M size       |
| Migration Resistance | Internal Testing                                                                                                                                                                                                                                                                                          | Internal Testing                                                                                                                                                                                                                                                                                        | Higher strength |
| Vessel Occlusion     | Internal Testing                                                                                                                                                                                                                                                                                          | Internal Testing                                                                                                                                                                                                                                                                                        | Higher pressure |
| Application Force    | Internal Testing                                                                                                                                                                                                                                                                                          | Internal Testing                                                                                                                                                                                                                                                                                        | Higher strength |
| Axial Pull-Off Force | Internal Testing                                                                                                                                                                                                                                                                                          | Internal Testing                                                                                                                                                                                                                                                                                        | Higher strength |
| Packaging            | The clips are housed in a cartridge and package in a ridged plastic plaster with Tyvek coated lidding.                                                                                                                                                                                                    | The clips are housed in a cartridge and package in a ridged plastic plaster with Tyvek coated lidding.                                                                                                                                                                                                  | Same            |

|  |  |  |  | Table 5.1. Technical Comparison of Vas-Q-Clip™ and Hem-O-Lok® (K993157). |
|--|--|--|--|--------------------------------------------------------------------------|
|--|--|--|--|--------------------------------------------------------------------------|

#### 13) Non-Clinical Performance Testing:

Non-clinical testing was completed to assess its performance. Table 5.2 shows the list of tests performed. The data provided in this 510(k) submission shows that the composition is safe based on the biocompatibility risk assessment conducted and a benchtop assessment.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "NANOVA" in blue, block letters. The letters are slightly rounded and have a modern, sans-serif appearance. The color of the text is a vibrant blue, which stands out against the white background.

| Table 5.2. Tests performed on Vas-Q-Clip™. |  |  |
|--------------------------------------------|--|--|
|                                            |  |  |

| Name                                         | Result |
|----------------------------------------------|--------|
| Migration Resistance                         | Passed |
| Vessel Occlusion                             | Passed |
| Application Force                            | Passed |
| Axial Pull-Off Force                         | Passed |
| USP 85 Bacterial Endoxins Test               | Passed |
| USP 151 Rabbit Pyrogenicity Test             | Passed |
| Material Comparison Test                     | Passed |
| ISO 10993-10 Skin Sensitization              | Passed |
| ISO 10993-6 Local Effects After Implantation | Passed |
| Visual Inspection of Seal                    | Passed |
| Bubble Emission                              | Passed |
| Dye Penetration                              | Passed |
| Peel-off Seal Strength                       | Passed |
| Accelerated and Real Time Aging              | Passed |
| Sterility                                    | Passed |

14) Clinical Performance Testing: Clinical performance data was not included.

#### Conclusion:

Nanova Biomaterials Inc. believes that Vas-Q-Clip™ is substantially equivalent to currently legally marketed products. It does not introduce new indications for use, has similar technological characteristics and does not introduce new potential hazards or safety risks.

---

**Source:** [https://fda-staging.innolitics.com/device/K151943](https://fda-staging.innolitics.com/device/K151943)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
