BRX PRO Bruxism Guard

K151933 · 12th Man Technologies, Inc. · MQC · Dec 10, 2015 · DE

Device Facts

Record IDK151933
Device NameBRX PRO Bruxism Guard
Applicant12th Man Technologies, Inc.
Product CodeMQC · DE
Decision DateDec 10, 2015
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The BRX PRO™ Bruxism Guard is indicated to protect the teeth and reduce damage caused by bruxing or nighttime grinding and reducing the noise associated with bruxing and grinding.

Device Story

BRX PRO is a prescription dual-arch bruxism guard; protects teeth; reduces damage/noise from nighttime grinding. Device consists of two polycarbonate trays; trays filled with dentist-supplied polyvinyl siloxane (PVS) impression material; covers approximately eight teeth per arch. Trays connected via stainless steel T-bar pin (upper) and slots (lower); allows four positional settings to accommodate natural bite (Class III to edge-to-edge); permits lateral mandibular movement. Creates posterior air-gap preventing tooth contact during clenching. Used in home setting; fitted by dentist/physician. Patient wears device at night. Benefits include reduced dental wear and noise reduction.

Clinical Evidence

No clinical data. Bench testing only. Testing included physical characteristics (ISO 178, ISO 179, ISO 62) and biocompatibility (ISO 10993-1, 5, 10). All tests passed.

Technological Characteristics

Materials: Polycarbonate trays, stainless steel T-bar/slots, PVS impression material. Principle: Mechanical posterior air-gap creation via custom-fitted trays. Form factor: Dual-arch, adjustable. Connectivity: None. Sterilization: Non-sterile; disinfection via denture cleaner. Standards: ISO 20795-1:2013 (denture base polymers), ISO 10993 (biocompatibility).

Indications for Use

Indicated for patients requiring protection of teeth from damage caused by bruxing or nighttime grinding and to reduce noise associated with bruxing and grinding.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 10, 2015 12th Man Technologies, Inc. Mr. Alex Stenzler President 7245 Garden Grove Blvd., Suite C Garden Grove, California 92841 Re: K151933 Trade/Device Name: BRX PROTM Regulation Number: Unclassified Regulation Name: Unclassified Regulatory Class: Unclassified Product Code: MQC Dated: October 28, 2015 Received: October 30, 2015 Dear Mr. Stenzler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Indications for Use Statement #### INDICATIONS FOR USE STATEMENT 510(k) Number (if known):K151933 Device Name: BRX PRO™ #### Indications for Use: The BRX PRO™ Bruxism Guard is indicated to protect the teeth and reduce damage caused by bruxing or nighttime grinding and reducing the noise associated with bruxing and grinding. Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ Concurrence of CDRH, Office of Device Evaluation (ODE) Page of of ___________________________________________________________________________________________________________________________________________________________________ {3}------------------------------------------------ #### 510(k) Summary | Submitter's Name | 12th Man Technologies, Inc.<br>7245 Garden Grove Blvd., Suite G<br>Garden Grove, CA 92841<br>1.714.705.4576 | |-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Registration Number | 3009108174 | | Contact Name | Alex Stenzler<br>12th Man Technologies, Inc.<br>7245 Garden Grove Blvd., Suite G<br>Garden Grove, CA 92841<br>Telephone: 1.714.705.4576<br>Fax: 1.714.373.0505<br>Email: alex.stenzler@12thmantec.com | | Date Prepared | December 10, 2015 | | Device Trade Name | BRX PRO™ | | Device Common Name | Bruxism guard | | Classification Name | None | | Product Code | MQC | | Device Classification | Unclassified | | Panel | Dental | | 510(k) Submission | Traditional | | Legally Marketed Equivalent | SmartGuard (K123161)<br>Brux-TMD QuickSplint (K111066) | ## Description The BRX PRO is a prescription dual arch bruxism guard which protects the teeth and reduces the damage caused by nighttime bruxism and reduces the noise of bruxing or nighttime grinding. The BRX PRO's two arch trays, when filled with impression material, cover approximately eight teeth on each of the upper and lower arches, thereby creating an air-gap between the posterior teeth that prevents tooth contact upon clenching. The two polycarbonate trays are connected in the front with a dual position stainless steel "T-bar" pin in the upper tray and a pair of stainless steel slots in the lower tray that can be set by the prescriber to maintain the BRX PRO trays in a neutral position by selecting one of four positions, while allowing for lateral mandibular movement. {4}------------------------------------------------ The impression materials that can be used to interface with the teeth are any of the approved for sale materials such as Mousse PVS (Parkell), individually selected by the dentist or physician fitting the patient with the BRX PRO and are typically a polyvinyl siloxane (PVS) available from many dental supply sources. #### Predicate Devices The design of the BRX PRO is substantially equivalent to the SmartGuard (K123161) and the Brux-TMD QuickSplint (K111066). The predicate devices are mouth guards designed to protect the teeth and reduce damage caused by bruxing or nighttime grinding and reduce the noise associated with bruxing and grinding. The Brux-TMD QuickSplint uses thermoplastic arches filled with impression material to interface to the teeth. The impression material is the same PVS as recommended for use in the BRX PRO and can be fitted in a prescriber's office. The BRX PRO Instructions-for-Use recommends that the dentist or prescriber uses commercially available polyvinyl siloxane impression materials to create the tray liner to obtain the ideal fit. Because these mouth guards can be used on either the upper or lower teeth, some dentists prescribe that the patient use mouth guards on both arches. The SmartGuard, Brux-TMD QuickSplint and the BRX PRO all use partial arches to create a posterior arch air gap. #### Description of Operation The BRX PRO comes with two sets of trays. A pair of impression trays (upper and lower) are sequentially filled with a PVS impression material by the prescriber to fit to the patient. Once trimmed by the dentist, the impressions are transferred to the treatment trays. The upper tray has a "T-bar" pin that extends out from the bottom of the tray. The pin has an anterior or posterior orientation so that when combined with the two slots in the lower tray, provides for a selection of four positions to fit the patient who may have a natural bite ranging from a Class III as well as an edge-to-edge bite position. The "T-bar" pin is inserted into the slot and rotated 90 degrees to connect the two trays. The slots provide for lateral movement of the mandible. #### Indications for Use: To protect the teeth and reduce damage caused by bruxing or nighttime grinding and reducing the noise associated with bruxing and grinding. {5}------------------------------------------------ | Substantial Equivalency Summary Comparison Table | | | | |--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------| | Element of<br>Comparison | BRX PRO™ | SmartGuard | Brux-TMD QuickSplint | | Indications for<br>Use | To protect the teeth and<br>reduce damage caused<br>by bruxing or nighttime<br>grinding and reducing<br>the noise associated<br>with bruxing and<br>grinding. | To reduce damage to the<br>teeth and reduce noise<br>associated with bruxing<br>and/or grinding | Prevent teeth grinding,<br>reduces jaw clenching and<br>damage to teeth | | Regulatory | Pre-Market Notification<br>510(k) | K123161 | K111066 | | Environment<br>of Use | Home Setting | Home Setting | Home Setting | | Design | Custom fitted mouth<br>guard | Custom fitted mouth<br>guard | Custom fitted mouth<br>guard | | Use | Night time | Night time | Night time | | Daily Duration<br>of Use | Up to 8 hours per night<br>for up to 30 days | Up to 12 hours per 24<br>hour period | While asleep and limited<br>daytime use | | Transitional<br>Device | Yes | NA | Yes | | Reusable | Yes, Single Patient | Yes, Single Patient | Yes, Single Patient | | Sterile | No | No | No | | Method of<br>disinfection | Denture cleaner | Denture cleaner | Soap and water | | Tray Materials | Polycarbonate and<br>stainless steel | Elvax (ethylene vinyl<br>acetate copolymer) | Polycarbonate | | Tray<br>manufacturing<br>method | Injection molded | Injection molded | Injection molded | | Impression<br>Material | Prescriber selected and<br>supplied PVS | Elvax (ethylene vinyl<br>acetate copolymer) | Prescriber selected and<br>supplied PVS | {6}------------------------------------------------ ## Non-Clinical Testing Non-clinical test results are submitted to confirm product conformance with device requirements and substantial equivalence to predicate device. These bench tests included: - 1. Packaging and Assembly / Shipping - 2. Materials Specifications - 3. Labeling Verification - 4. Physical Characteristics | Physical Parameter | Test Method | ISO 20795-1:2013<br>Requirement | Pass/Fail | |----------------------------------------------------------------------|-------------|---------------------------------|-----------| | Flexural Modulus | ISO 178 | ≥ 2000 MPa | Pass | | Flexural Stress | ISO 178 | ≥ 65 MPa | Pass | | Water Absorption | ISO 62 | ≤ 32 µg/m³ | Pass | | Water Solubility | ISO 62 | ≤ 1.6 µg/m³ | Pass | | Charpy Notched Impact<br>Strength (Linear relation<br>with Fracture) | ISO 179 | ≥ 0.9 KJ/m² | Pass | - 5. Manufacturing | Applicable Standards Met | | | | | | |------------------------------------|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|--|--| | Standard or<br>Regulation | Standard<br>Organization or<br>Regulatory<br>Body | Name of Test Performed | Test<br>Results | | | | ISO 10993-<br>1:2009 COR 1<br>2010 | International<br>Standards Office | Biological Evaluation of Medical<br>Devices - Part 1: Evaluation and<br>Testing within a Risk Management<br>Process | Pass | | | | ISO 10993-<br>5:2009 | International<br>Standards Office | Biological Evaluation of Medical<br>Devices – Part 5 Tests for in vitro<br>cytotoxicity | Pass | | | | ISO 10993-<br>10:2009 | International<br>Standards Office | Biological Evaluation of Medical<br>Devices - Part 10 Tests for irritation<br>and skin sensitization | Pass | | | | BS EN ISO<br>15223-1:2012 | International<br>Standards Office | Medical Devices - Symbols to be used<br>with Medical Devices Labels,<br>Labelling and Information to be<br>supplied- Part 1: General<br>Requirements | Pass | | | | BS EN ISO<br>20795-1:2013 | International<br>Standards Office | Dentistry - Base polymers, Part 1:<br>Denture base polymers | Pass | | | {7}------------------------------------------------ ### Substantial Equivalence Based on the design, intended use, principle of operation, technological characteristics, performance data, and attribute comparison above, the 12th Man Technologies, Inc. BRX PRO bruxism guard is substantially equivalent to the currently marketed predicate devices which have been previously reviewed for market clearance by the FDA. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ [510(k)] Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...