3MP (X213C3P0E, X213G3P0E) and 2MP (X213C2P0E, X213G2P0E) LCD Display
Applicant
D-View Display Technology Co., Ltd.
Product Code
LLZ · Radiology
Decision Date
Jul 28, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The D-VIEW 3MP (X213C3P0E, X213G3P0E) and 2MP (X213C2P0E, X213G2P0E) LCD Display is intended to be used in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners.
Device Story
D-VIEW LCD display system provides high-resolution visualization of digital medical images. Device accepts digital video signals via DVI-I or DP inputs; transforms signals into visual output on 21.3-inch TFT monochrome LCD screen. Features include built-in brightness stabilization circuit, front sensor, and ambient light sensor to ensure stable brightness and calibration. Anti-reflection coated protective screen facilitates cleaning and disinfection. Used in clinical environments by trained medical practitioners for image review and analysis. Output supports clinical decision-making by providing high-fidelity image display. Benefits include consistent image quality and durability in medical settings.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
21.3-inch TFT monochrome LCD panel; native resolutions 2048x1536 (3MP) or 1600x1200 (2MP). Features front sensor, ambient light sensor, and brightness stabilization circuit. Connectivity via DVI-I and DP. Power input 12V DC. Complies with IEC 60601-1 (electrical/mechanical/thermal safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for trained medical practitioners to display and view digital medical images, excluding digital mammography, for review and analysis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, composed of three overlapping shapes that resemble a person's head, torso, and legs. The figure is oriented to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 28, 2015
D-VIEW Display Technology Co., Ltd. % Mr. Jianfei Sun Regulatory Manager 3F, Building 4, No.12# Kechuang 13th Street, BDA area Beijing 100176 CHINA
Re: K151854 Trade/Device Name: 3MP (X213C3P0E, X213G3P0E) and 2MP (X213C2P0E, X213G2P0E) LCD display Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: July 3, 2015 Received: July 7, 2015
Dear Mr. Sun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D'Hara
For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
#### K151854
Device Name
3MP (X213C3P0E、X213G3P0E) and 2MP (X213C2P0E、X213G2P0E) LCD display
Indications for Use (Describe)
The D-VIEW 3MP (X213C3P0E, X213G3P0E) and 2MP (X213C2P0E) LCD Display is intended to be used in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 201 Subpart D) </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 201 Subpart C) </div> |
|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
## 1. Date Prepared [21 CFR807.92 (a) (1)]
July 3, 2015
### 2. Submitter's Information [21 CFR807.92 (a) (1)]
| Name of Sponsor: | D-VIEW Display Technology Co.,Ltd. | |
|------------------|------------------------------------------------------------------|--|
| Address: | 3F, Building 4, No.12# Kechuang 13th Street , BDA area, Beijing, | |
| P.R.China | | |
| Contact Name: | Jianfei Sun | |
| Telephone No.: | +8610 82379533 EXT. 5360 | |
| Fax No.: | +8610 82379511 | |
| Email Address: | sunjianfei@d-view.cn | |
# 3. Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)]
| Trade Name: | D-VIEW |
|-----------------------|---------------------------------------------------------------------|
| Common Name: | 3MP (X213C3P0E X213G3P0E) and 2MP (X213C2P0E X213G2P0E) LCD display |
| Classification: | 892.2050 system, image processing, radiological |
| Product code: | PGY |
| Classification Panel: | Radiology |
| Device Class: | II |
### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
The identified predicates within this submission are as follows: BARCO N.V., Coronis 3MP(MDCG-3221) has been cleared by FDA through 510(k)
{4}------------------------------------------------
No.K131246 (Decision Date -May 24, 2013), Classification: 892.2050 system, image processing, radiological Product code: LLZ Device Class: II
### 5. Description of the Device [21 CFR 807.92(a) (4)]
LCD display is a display system for medical viewing, with high resolution 1600 x 1200(X213C2P0E , X213G2P0E) and 2048 x 1536 (X213C3P0E , X213G3P0E), built-in brightness stabilization circuit, front sensor and ambient light sensor, stable brightness and persistent calibration can be guaranteed. These displays can support both landscape and portrait mode. The anti-reflection coated protective screen can prevent display from damage under hard using conditions, make the clean and disinfect easier.
### 6. Intended Use [21 CFR 807.92(a)(5)]
The D-VIEW 3MP (X213C3P0E,X213G3P0E) and 2MP (X213C2P0E,X213G2P0E) LCD Display is intended to be used in displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners.
## 7. Technological Characteristics [21 CFR 807.92(a)(6)]
| Panel | 21.3", TFT monochrome LCD screen, antiglare |
|--------------------------|--------------------------------------------------------------------------------------------------------------------|
| Brightness (typ.) | 1900cd/m2 (2MP: X213G2P0E)<br>900cd/m2 (2MP: X213C2P0E)<br>1700cd/m2 (3MP: X213G3P0E)<br>800cd/m2 (3MP: X213C3P0E) |
| CR (typ.) | 1400:1 |
| Viewing angle | R/L 176°, U/D 176° Typ. ( CR ≥ 10 ) |
| Pixel Pitch | 0.2115 × 0.0705mm ( 3MP )<br>0.270 × 0.270mm ( 2MP ) |
| Native resolution | 2048 x 1536 (3MP)<br>1600 x 1200 (2MP) |
| Display area | 433(V)mm × 324 mm(H) |
| Compatible video signals | 3MP:<br>640 x 480@60Hz(progressive)<br>2048x1536@60Hz(progressive)<br>2MP: |
{5}------------------------------------------------
| | 640 x 480@60Hz(progressive) |
|-------------------------------------------|-----------------------------------------------------|
| | 1600x1200@60Hz(progressive) |
| Horizontal resolution | 3MP: 2048x1536@60Hz(progressive) |
| | 2MP: 1600x1200@60Hz(progressive) |
| Bandwidth | 3MP: 267MHz |
| | 2MP: 162MHz |
| Aspect ratio | 4:3 |
| Screen size | 21.3" real diagonal |
| Power | 3MP, X213C3P0E: DC12V 3.5A |
| | 3MP, X213G3P0E: DC12V 2.2A |
| | 2MP, X213C2P0E: DC12V 3.5A |
| | 2MP, X213G2P0E: DC12V 2.2A |
| Input signals | DVI-I, DP |
| Digital input | TMDS ( dual ) |
| Dimension | 375 mm (W) x 434mm (H) x 70mm (D) ( without Stand ) |
| | 375 mm (W) x 589 mm (H) x 173mm (D) ( with Stand ) |
| Weight | 7.0±0.3kg ( without Stand ) |
| | 10.5kg±0.3kg ( with Stand ) |
| Operating<br>Atmospheric<br>Pressure | 70kPa ~ 106kPa |
| Operating<br>temperature and<br>humidity: | Temperature: 0℃~40℃ |
| | Humidity: 30% ~75% |
| Storage<br>temperature and<br>humidity: | Temperature: -20°C ~ 60°C |
| | Humidity: 0% ~90% |
# 8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]
#### 8.1 Intended uses:
#### Table 1 Intended Use Comparison
| ID | Comparison Item | Proposed Device | Predicate Device |
|----|-----------------|------------------------------------------------------------------------|------------------------|
| | | D-VIEW 3MP (X213C3P0E,<br>X213G3P0E) and 2MP<br>(X213C2P0E, X213G2P0E) | Coronis 3MP(MDCG-3221) |
{6}------------------------------------------------
| 1 | Intended<br>Use | The D-VIEW 3MP<br>(X213C3P0E, X213G3P0E)<br>and 2MP (X213C2P0E,<br>X213G2P0E) LCD Display is<br>intended to be used in<br>displaying and viewing digital<br>images (excluding digital<br>mammography) for review and<br>analysis by trained medical<br>practitioners. | "The Coronis 3MP<br>(MDCG-3221)" is intended to<br>be used as a tool in displaying<br>and viewing digital images<br>(excluding digital<br>mammography) for review and<br>analysis by trained medical<br>practitioners. |
|---|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### 8.2 Comparison table
| ID | Comparison<br>Item | Proposed Device<br>D-VIEW 3MP (X213C3P0E,<br>X213G3P0E) and 2MP<br>(X213C2P0E, X213G2P0E) | Predicate Device<br>Coronis 3MP(MDCG-3221) |
|-----|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| 2 | Performance | | |
| 2.1 | Panel Size and Type | 21.3", TFT LCD display | 21.3",TFT LCD display |
| 2.2 | Pixel Pitch | 0.2115 × 0.0705mm (3MP)<br>0.270 × 0.270mm(2MP) | 0.2115 mm |
| 2.3 | Available Cabinet<br>Colors | Black | Black |
| 2.4 | Native Resolutions | 1600 x 1200 (2MP)<br>2048 x 1536 (3MP) | 2048 x1536 |
| 2.5 | Brightness | 1900cd/m2 (2MP,<br>X213G2P0E)<br>900cd/m2 (2MP,<br>X213C2P0E)<br>1700cd/m2 (3MP,<br>X213G3P0E)<br>800cd/m2 (3MP,<br>X213C3P0E) | 1700 cd/m2 |
| 2.6 | Contrast Ratio | 1400:1 | 1300:1 |
| 2.7 | Network Interface | USB(1 Up, 2 Downstream) | USB(1 Up, 2 Downstream) |
| 2.8 | Active Display Size | 433mm × 324 mm | 433.2mm x 324.9mm |
| 3 | Physical Specifications | | |
| 3.1 | Dimensions<br>(Wx Hx D) | 375 mm (W) x 434mm (H) x<br>70mm (D) ( without Stand ) | 375mm x 488mm x 84mm<br>(without Stand) |
| ID | Comparison Item | Proposed Device<br>D-VIEW 3MP (X213C3P0E,<br>X213G3P0E) and 2MP<br>(X213C2P0E, X213G2P0E) | Predicate Device<br>Coronis 3MP(MDCG-3221) |
| | | 375 mm (W) x 589 mm (H) x<br>173mm (D) (with Stand) | Portrait:<br>375mm x 620~520mm<br>x 235mm<br>Landscape:<br>488mm x 563~463mm<br>x 250mm<br>(with Stand) |
| | | | |
| | | | |
| | Temperature | | |
| 3.2 | Operating | 0°C ~ 40°C | 0°C ~ 40°C |
| 3.3 | Transport/ Storage | -20°C ~ 60°C | -20°C ~ 60°C |
| | Relative humidity | | |
| 3.4 | Operating | 30% ~75%(non-condensing) | 8% ~80%(non-condensing) |
| 3.5 | Transport/ Storage | 0% ~90% | 5% ~95% |
| 4 | | Power Supply | |
| 4.1 | PowerCapacity | <50W | <40W |
| 4.2 | Input Voltage | 3MP, X213C3P0E: DC12V<br>3.5A | |
| | | 3MP, X213G3P0E: DC12V<br>2.2A | |
| | | 2MP, X213C2P0E: DC12V<br>3.5A | |
| | | 2MP, X213G2P0E: DC12V<br>2.2A | 100~240v |
| | | | |
| | | Power adapter: | |
| | | MDS-150AAS12 B | |
| | | Input: 100-240VAC 50-60Hz | |
| | | 2.5-1A Output: 12VDC 10A | |
| 5 | | Human factors (operation characteristic) | |
| 5.1 | Usability | Button operation,<br>LED indicator | Button operation,<br>LED indicator |
| 5.2 | Mode of operation | Continuous operation | Continuous operation |
| 6 | | Biocompatibility | |
| 6.1 | Evaluation | The proposed device does not<br>contain any components that | The proposed device does<br>not contain any components |
| | | come into direct or indirect | that come into direct or |
| | | contact with patients, so the | indirect contact with patients, |
| | | evaluation doesn't be needed. | so the evaluation doesn't be |
| | | | needed. |
| 7 | | Sterility | |
| ID | Comparison<br>Item | Proposed Device<br>D-VIEW 3MP (X213C3P0E,<br>X213G3P0E) and 2MP<br>(X213C2P0E, X213G2P0E) | Predicate Device<br>Coronis 3MP(MDCG-3221) |
| 7.1 | Sterilization | The proposed device does not<br>need sterilization. | The proposed device does<br>not need sterilization. |
| 8 | Electrical & Mechanical safety& Thermal safety | | |
| 8.1 | Type of protection<br>against electric<br>shock | Class I | Class I |
| 8.2 | Degree of protection<br>against harmful<br>ingress of liquid | Ordinary equipment. | Ordinary equipment. |
| 8.3 | Evaluation | The electrical, mechanical and<br>thermal safety evaluation is<br>conducted as per the<br>requirements of the standard<br>IEC 60601-1. | The electrical, mechanical<br>and thermal safety evaluation<br>is conducted as per the<br>requirements of the standard<br>IEC 60601-1. |
| 9 | Electromagnetic Compatibility | | |
| 9.1 | EMC Evaluation | Complying with<br>IEC 60601-1-2 | Complying with<br>IEC 60601-1-2 |
#### Table 2 General Comparison
{7}------------------------------------------------
{8}------------------------------------------------
#### 8.4 Discussion of Differences:
It is reasonable that there are some differences between our new system and its predicate. All of parameters comply with 21CFR1020.33 and related IEC standards. We did not use any new technology in this system, so those differences between our new system and its predicate do not affect the safety and effectiveness (SE).
Review of ID 1 - Intended use, both of them are intended to be used in displaying and viewing digital images(excluding digital mammography) for review and analysis by trained medical practitioners. So the SE is not affected.
Review of ID 2 - Performance, except the item as below, it is the same, so the SE is not affected.
- 1. Contrast Ratio, The proposed device is 1400:1 and the predicate device is 1300:1, but the 1400:1 is better than 1300:1 in terms of the image quality. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.
Review of ID 3 - Physical Specifications, Dimensions and Relative humidity are comparable, so the SE is not affected
Review of ID 4 - Power Supply, both of them comply with IEC 60601-1 and IEC 510(k) Summary 7-6
{9}------------------------------------------------
60601-1-2. Therefore, they can be considered Substantially Equivalent in safety and effectiveness. So the SE is not affected.
Review of ID 5 - Human factors, both are the same, so the SE is not affected.
Review of ID 6 - Biocompatibility, both are the same, so the SE is not affected.
Review of ID 7 - Sterility, both are the same, so the SE is not affected.
Review of ID 8 - Electrical & Mechanical safety& Thermal safety, both are the same, so the SE is not affected.
Review of ID 9 - EMC, both are the same, so the SE is not affected.
### 9. Conclusion [21 CFR 807.92(b) (3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Shenzhen Beacon Display Technology Co., Ltd concludes that D-VIEW 3MP (X213C3P0E、X213G3P0E) and 2MP (X213C2P0E、X213G2P0E) LCD display is substantially equivalent to predicate devices with regard to safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.