Cable-Ready® Cable Grip System: Cable-Ready Pin System, Cable-Ready Needle System
K151848 · Zimmer, Inc. · JDQ · Jan 15, 2016 · Orthopedic
Device Facts
Record ID
K151848
Device Name
Cable-Ready® Cable Grip System: Cable-Ready Pin System, Cable-Ready Needle System
Applicant
Zimmer, Inc.
Product Code
JDQ · Orthopedic
Decision Date
Jan 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cable-Ready Pin System: The Cable-Ready Pin System is indicated for use in olecranon fractures, patellar fractures, proximal humerus fractures, greater tuberosity humerus fractures, and medial malleolus tibial fractures may not be securely held by either a bone screw, bone pin, cable, or cerclage cable alone. Cable-Ready Needle System: The Cable-Ready Needle System is indicated for securing fractures of the olecranon, patella, humerus, shoulder and ankle, reducing and securing acromioclavicular dislocations as well as sternotomy surgery.
Device Story
Cable-Ready Cable Grip System (Pin and Needle variants) used in orthopedic trauma/reconstructive surgery to secure bone fractures. Pin system features 4.0mm cancellous lag screw attached to 1.3mm cable; implanted into bone. Needle system features curved surgical needle attached to stainless steel or titanium leader/cable; passed through bone; needle removed after securing. Operated by surgeons in clinical/OR settings. Instruments used to crimp and trim excess cable. Provides mechanical stabilization of bone fragments; aids healing by maintaining fracture reduction. Benefits include secure fixation in complex fractures where traditional hardware may be inadequate.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including accelerated aging (10-year shelf life), biocompatibility (ISO 10993-1), and mechanical performance testing (static, fatigue, torque, and pullout testing).
Technological Characteristics
Materials: Stainless steel, titanium. Form factor: 1.3mm cables, 4.0mm cancellous lag screws, curved surgical needles. Principle: Mechanical bone fixation/cerclage. Sterilization: Sterile. Connectivity: None. Software: None.
Indications for Use
Indicated for patients with fractures of the olecranon, patella, humerus, shoulder, ankle, or medial malleolus, and those requiring acromioclavicular dislocation reduction or sternotomy closure, where standard fixation (screws, pins, cables) is insufficient.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2016
Zimmer, Inc. Ms. Dalene Binkley Senior Specialist, Trauma Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581
Re: K151848
Trade/Device Name: Cable-Ready® Cable Grip System: Cable-Ready Pin System, Cable-Ready Needle System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II
Product Code: JDQ, HWC Dated: December 15, 2015 Received: December 17, 2015
Dear Ms. Binkley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K151848 (page 1 of 2)
Device Name
Cable-Ready® Cable Grip System: Cable-Ready® Pin System
Indications for Use (Describe)
The Cable-Ready® Pin System is indicated for use in olecranon fractures, proximal humerus fractures, greater tuberosity humerus fractures, and medial malleolus tibial fractures may not be securely held by either a bone screw, bone pin, cable, or cerclage cable alone.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
### Indications for Use
510(k) Number (if known)
K151848 (page 2 or 2)
Device Name
Cable-Ready® Cable Grip System: Cable-Ready® Needle System
Indications for Use (Describe)
The Cable-Ready® Needle System is indicated for securing fractures of the olecranon, patella, humerus, shoulder and ankle, reducing and securing acromioclavicular dislocations as well as sternotomy surgery.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Zimmer Biomet. The logo features a stylized blue "Z" inside of a blue circle. Below the circle is the word "zimmer" in lowercase blue letters.
## 510(k) Summary
| Sponsor: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dalene Binkley, MS, RAC<br>Senior Specialist, Trauma Regulatory Affairs<br>Telephone: 574-372-4970<br>Fax: (574) 371-8760 |
| Date: | January 13, 2016 |
| Trade Name: | This is a bundled traditional 510(k). The two systems<br>bundled in this submission are: |
| | Cable-Ready® Cable Grip System: |
| | 1) Cable-Ready Pin System<br>2) Cable-Ready Needle System |
| Common Name: | Cerclage Cable and Bone Screw |
| Classification Names<br>and References: | Bone Fixation Cerclage<br>21 CFR 888.3010, JDQ |
| | Smooth or threaded metallic bone fixation fastener<br>21 CFR 888.3040, HWC |
| Classification Panel: | Orthopedics/87 |
| Predicate Device(s): | Pioneer Laboratories Cable Pin (Pioneer Surgical<br>Technology, K941174, cleared 9/19/94) |
| | Zimmer® Kirschner Wires and Steinmann Pins (Zimmer,<br>Inc., K143618, cleared 2/6/15) |
| | Suture Wire and Wire Loops (Zimmer, Inc. K150889,<br>cleared 7/1/15) |
| | Songer Cable System, Modification (Pioneer Surgical<br>Technology, K935481, cleared 1/26/94) |
| Device Description: | Zimmer is requesting clearance for modifications to the<br>Cable-Ready Cable Grip System: Cable-Ready Pin<br>System and Cable-Ready Needle System. The subject<br>devices are similar in that they are both 1.3mm cables<br>used in reconstructive and trauma surgery, securing bone<br>fractures. The stainless steel Cable-Ready Pin System has<br>a partially threaded 4.0mm cancellous lag screw (pin),<br>available in differing lengths, attached to the cable which<br>is implanted into the bone. The Cable-Ready Needle<br>System has a curved stainless steel surgical needle<br>attached to either a stainless steel or titanium leader and<br>cable that is passed through the bone. The needle is cut off<br>once the fracture is secure. Both systems' cables utilize the<br>same instruments to secure the crimp and trim off excess. |
| Intended Use: | <b>Cable-Ready Pin System:</b><br>The Cable-Ready Pin System is indicated for use in<br>olecranon fractures, patellar fractures, proximal humerus<br>fractures, greater tuberosity humerus fractures, and medial<br> |
| Comparison to Predicate Device: | The subject devices incorporate similar or identical<br>materials, similar or identical indications for use, similar<br>or identical sizes of implants, and the same technological<br>characteristics as the predicate devices. |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>Shelf Life - Accelerated aging testing conducted<br>shows that the sterile devices included in this<br>submission have a shelf life of 10 years. Biocompatibility – Biocompatibility testing of the<br>subject devices was conducted per ISO 10993-1 and |
{5}------------------------------------------------
{6}------------------------------------------------
Good Laboratory Practices (21 CFR 58). All testing passed.
- . Performance Evaluation – A combination of performance testing - static, fatigue, torque and pullout testing for the cable pin and engineering analyses for the cable needle demonstrate the subject devices are safe and effective and substantially equivalent to the predicate devices.
Conclusions: The data presented in this submission demonstrates that the subject devices are substantially equivalent to their respective predicate devices.
### Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for these devices to show substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.