rainbow™ LS Pressing is used in the manufacture of Inlay, Onlay and Crown.
Device Story
rainbow™ LS Pressing is a lithium disilicate glass-ceramic ingot used in dental laboratories. It is processed via the press technique to fabricate dental restorations, specifically inlays, onlays, and crowns. The device serves as a restorative material with strength and esthetics comparable to natural teeth. It is intended for prescription use by dental professionals. The material is non-sterile and requires laboratory processing to achieve the final anatomical shape of the dental restoration. The resulting restorations are used to replace missing or damaged tooth structure, potentially benefiting the patient by restoring dental function and appearance.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 6872:2008 (dentistry - ceramic materials) including bending strength (569 MPa), linear thermal expansion coefficient, chemical solubility (35.75 µg/cm²), and glass transition temperature. Biocompatibility testing performed per ISO 10993-1:2009, including cytotoxicity (ISO 10993-5), irritation/intracutaneous reactivity (ISO 10993-10), sensitization (ISO 10993-10), acute systemic toxicity (ISO 10993-11), and genotoxicity (ISO 10993-03).
Technological Characteristics
Lithium disilicate glass-ceramic ingot. Composition: SiO2, Li2CO3, Ca3(PO4)2, P2O5, ZnO, Al2O3, La2O3, K2CO3. Bending strength: 569 MPa. Linear thermal expansion coefficient: (10.3±0.5) x 10-6K-1. Chemical solubility: 35.75 µg/cm². Glass transition temperature: (550±20) ℃. Non-sterile. Complies with ISO 6872:2008 (Type II - Class 2/3).
Indications for Use
Indicated for the manufacture of inlays, onlays, and crowns in dental patients.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
Predicate Devices
IPS e.max Press and IPS e.max Press Multi (K120134)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular pattern.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 14, 2016
GENOSS Co., Ltd. c/o Mr. Dachan Kwon iCT America, Inc. 180 Sylvan Avenue, 2nd Floor Englewood Cliffs, New Jersey 07632
Re: K151846
Trade/Device Name: rainbow™ LS Pressing Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: II Product Code: EIH Dated: December 4, 2015 Received: December 8, 2015
#### Dear Mr. Kwon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina Kiang
-s
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "GENOSYS" in a stylized font. The letters are composed of small, repeating patterns, giving them a textured appearance. The overall design is simple and clean, with a focus on the typography.
## Indication for use
510(k) Number: K151846
rainbow™ LS Pressing Device Name:
Indication for use:
rainbow™ LS Pressing is used in the manufacture of Inlay, Onlay and Crown.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Indication for use
Page 1 of 1
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for GENOSS. The word "GENOSS" is written in a sans-serif font. The letters "GEN" are in a dark green color, while the letters "OSS" are in a lighter green color. The logo is simple and modern.
# 510(k) Summary
01/05/2016
## 1. Company
| | Submitter |
|----------------|----------------------------------------------------------------------------------------------------------|
| Name | GENOSS Co., Ltd. |
| Address | 1F, Gyeonggi R&DB Center / 226, 2F, GSBC, 105<br>Gwanggyo-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea |
| Phone/Fax | +82-31-888-5100/ +82-31-888-5105 |
| Contact person | Byungsun Kim / RA<br>bskim@genoss.com |
| Summary Date | 01/05/2016 |
## 2. Device Name
| Proprietary name | : rainbow™ LS Pressing |
|---------------------|-------------------------------------|
| Regulation number | : 21 CFR 872.6660 |
| Classification name | : Porcelain powder for clinical use |
| Product code | : EIH |
| Device class | : Class II |
## 3. Predicate Device
K120134 IPS e.max Press and IPS e.max Press Multi
#### 4. Description
rainbow™ LS Pressing is a lithium disilicate glass-ceramic ingot for use with the press technique with similar strength and esthetics to natural tooth.
### 5. Indication for use
rainbow™ LS Pressing is used in the manufacture of Inlays, Onlays and Crowns.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first three letters, "GEN", are in a dark green color. The last three letters, "OSS", are in a lighter green color. The letters are connected to each other.
## 6. Technological Characteristics
The following comparison table of the technological characteristics of the subject device and the predicate devices outlines and provides the similarities and the substantial equivalency of the rainbow™ LS Pressing and the predicate.
| Device name | rainbowTM LS Pressing | IPS e.max Press<br>and IPS e.max Press Multi | Comparison |
|--------------------------------------------|----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Genoss Co., Ltd. | Ivoclar Vivadent,<br>Incorporated | N/A |
| 510(k) Number | New Device | K120134 | N/A |
| Materials | SiO2, Li2CO3, Ca3(PO4)2,<br>P2O5, ZnO, Al2O3, La2O3,<br>K2CO3. etc. | SiO2, Li2O, K2O, MgO,<br>ZnO, Al2O3, P2O5 etc. | Similar<br>The major materials of<br>lithium disilicate dental<br>ceramic(subject and<br>predicate device) are SiO2<br>and Li2O.<br>The each composition<br>of other materials is<br>less than 5%.<br>The difference of<br>composition is Ca3(PO4)2.<br>Ca3(PO4)2 is decomposed<br>into 3CaO and P2O5.<br>The CaO acts as stabilizer. |
| Form | Pre-formed Ingot | Pre-formed Ingot | Same |
| Type, class of<br>dental ceramic | Type II - Class 2 | Type II - Class 3 | Similar<br>Although the subject<br>device refers to Type II-<br>Class 2 dental ceramic, the<br>bending strength satisfies<br>requirement by<br>ISO 6872:2008 for<br>Class 3 dental ceramics<br>(> 300 MPa). |
| Sterilization | Non-sterile | Non-sterile | Same |
| | | | |
| Indication for use | rainbow™ LS Pressing is<br>used in the manufacture of<br>Inlays, Onlays and<br>Crowns. | IPS e.max Press and IPS<br>emax Press Multi is an all-<br>ceramic system for the<br>creation of Occlusal<br>veneers, Thin Veneers,<br>Veneers, Inlays, Onlays,<br>Crowns in the anterior and<br>posterior region, 3-unit<br>bridges in the anterior<br>region, 3-unit bridges in<br>the premolar region up to<br>the second premolar as the<br>terminal abutment, Crown,<br>splinted crown or 3 unit<br>bridge up to the second<br>premolar placed on top of<br>an implant abutment. | Similar<br>The subject and predicate<br>devices can be used to<br>make Inlays, Onlays and<br>Crowns, because the<br>bending strength and<br>chemical solubility of<br>devices satisfy<br>requirement by<br>ISO 6872:2008 for<br>Class 4 dental ceramics.<br>(> 300 Mpa, < 2000µg/cm²) |
| Use | Prescription | Prescription | Same |
| Technical characteristics | | | |
| Bending Strength<br>(Flexural strength) | 569 MPa | 400 Mpa | Bending strength is<br>higher than required by<br>ISO 6872:2008 for<br>Class 3 dental ceramics.<br>( > 300 Mpa ) |
| Linear thermal<br>expansion<br>coefficient | $(10.3±0.5) x 10-6K-1$ | $10.5 x 10-6K-1$ | Same |
| Chemical Solubility<br>(µg/cm²) | 35.75 | 40 | Chemical solubility<br>satisfies requirement by<br>ISO 6872:2008 for<br>Class 2 or 3 dental<br>ceramics.<br>(< 100µg/cm²) |
| Glass transition<br>temperature | (550±20) ℃ | approx. 549 ℃ | Same |
| Biocompatibility | Biocompatible | Biocompatible | The Products are<br>biocaompatible according<br>to ISO 10993-1<br>'Biological evaluation of<br>medical devices –<br>Part 1: Evaluation and<br>testing within a risk<br>management process. |
## Comparison of Characteristics
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first three letters, "GEN", are in a dark green color. The last three letters, "OSS", are in a lighter green color. The font is sans-serif and has a modern look.
# 510(K) Number: K151846
{6}------------------------------------------------
#### 510(K) Number: K151846
Image /page/6/Picture/1 description: The image shows the logo for GENOSS. The logo is in green and consists of the word "GENOSS" in a sans-serif font. The first three letters, "GEN", are in a darker green, while the last three letters, "OSS", are in a lighter green. The logo is simple and modern.
Image /page/6/Figure/2 description: The image shows a comparison of shapes and sizes of what appears to be dental materials. On the left, there is a cylinder with a diameter of 12.5mm and a height of 10mm. In the middle, there are multiple cylinders with a diameter of 12.5mm and heights of 10mm and 20mm. The right side of the image is labeled 'Similar'.
There are three minor differences that are worth discussing:
- 1) The difference of 'Indication for use' is not critical to the intended use.
Considering the bending strength of devices satisfies requirement by ISO 6872:2008 for Class 4 dental ceramics(> 300 MPa), the subject and predicate devices can be used to make Inlays, Onlays and Crowns.
- 2) The subject and predicate device have slight difference in materials. But they both are lithium disilicate dental ceramic which have SiO2 and Li2O as the major materials. The each composition of other materials is less than 5%. The difference of chemical composition is Cas(PO4)2, which is decomposed into 3CaO and P2O5. The CaO acts as stabilizer. It enhances the structure of Li2O-SiO2 system(the basic lithium disilicate system) and increases the viscosity of glass. This difference has no effect on the device. The non-clinical test and biocompatibility test) demonstrate the Technical characteristics.
- 3) The differences of technological characteristics(Bending strength, chemical solubility) are within what is expected of this type of device. The performance test results satisfy the requirement for Type II - Class 2 or 3 dental ceramic by ISO 6872:2008.
#### 7. Summary of non-clinical testing
Non-clinical device testing was conducted to confirm the performance of the subject device. Bench testing was conducted in accordance with the FDA recognized consensus standard(Recognition number : 4-178: ISO 6872 Third edition 2008-09-01, dentistry - ceramic materials) Bench tests for performance comparison of the subject device and the predicate device includes the following testing:
- Bending Strength - Linear thermal expansion coefficient
- Chemical Solubility
- Glass transition temperature
- 510(k) Summary
{7}------------------------------------------------
# GENOSS
Technical characteristics of both devices satisfy the criteria for Type II - Class 2 or 3' dental ceramic by ISO 6872:2008. Slight differences between the proposed and predicate devices do not raise any new issues of safety and effectiveness.
Biocompatibility testing was conducted on the device pursuant to the ISO 10993-1:2009 Biological evaluation of medical device - Part 1: Evaluation and testing within a risk management process.
- Cytotoxicity test(ISO 10993-5)
- Irritation or intracutaneous reactivity(ISO 10993-10)
- Sensitization(ISO 10993-10)
- Acute systemic toxicity(ISO 10993-11)
- Genotoxocity(ISO 10993-03)
The result of biocompatibility testing demonstrated that no issue of biocompatibility arises.
#### 8. Conclusion
Based on the information provided in this premarket notification of GENOSS Co., Ltd. concludes that rainbow™ LS Pressing is substantially equivalent to predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.