rainbow™ Shade is a pre-sintered zirconia(Y-TZP) block for fabrication of single-tooth and bridgework restoration.
Device Story
rainbow™ Shade is a pre-sintered zirconia (Y-TZP) ceramic block used in dental laboratories. The device is milled into tooth cores using CAD/CAM or manual milling machines; subsequently fired in a furnace to achieve final hardness. The hardened core is then layered with porcelain to create a finished dental restoration. Used by dental professionals to fabricate single-tooth or bridgework prostheses. The device provides a durable, biocompatible structural foundation for dental crowns and bridges, restoring patient oral function and aesthetics.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included biocompatibility (ISO 10993-5, -10, -11, -3) confirming non-toxicity and non-sensitization. Bench testing per ISO 6872 evaluated visual, size, uniformity, radioactivity, chemical solubility (0µg/cm²), flexural strength (913MPa), linear thermal expansion, and low-temperature degradation. All results met established criteria.
Technological Characteristics
Pre-sintered zirconia (Y-TZP) ceramic block. Non-sterile. Flexural strength: 913MPa. Sintering density: 6g/cm³. Radioactivity: 0.01Bq/kg. Chemical solubility: 22.65µg/cm². Complies with ISO 6872 (Dentistry - Ceramic materials) and ISO 10993 (Biocompatibility).
Indications for Use
Indicated for fabrication of single-tooth and bridgework dental restorations in patients requiring prosthetic treatment.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The caduceus is rendered in black, and the text is also in black.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 16, 2015
Genoss Co., Ltd. c/o Dachan Kwon iCT America. Inc. 180 Sylvan Avenue, 2nd Floor Englewood Cliffs, NJ 07632
Re: K151844
Trade/Device Name: Rainbow™ Shade Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: July 15, 2015 Received: July 22, 2015
Dear Dachan Kwon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Tina Kiang 5
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indication for use
510(k) Number: K151844
rainbow™ Shade Device Name:
Indication for use:
rainbow™ Shade is a pre-sintered zirconia(Y-TZP) block for fabrication of single-tooth and bridgework restoration.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Indication for use
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for GENOSS. The word "GENOSS" is written in a sans-serif font. The letters "GEN" are in a dark green color, while the letters "OSS" are in a lighter green color. The logo is simple and modern.
# 510(k) Summary
10/06/2015
### 1. Company
| | Submitter |
|----------------|----------------------------------------------------------------------------------------------------------|
| Name | GENOSS Co., Ltd. |
| Address | 1F, Gyeonggi R&DB Center / 226, 2F, GSBC, 105<br>Gwanggyo-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea |
| Phone/Fax | +82-31-888-5100/ +82-31-888-5105 |
| Contact person | Byungsun Kim / RA<br>bskim@genoss.com |
| Summary Date | 10/06/2015 |
### 2. Device Name
Proprietary name: Regulation description : Porcelain powder for clinical use Classification name: Powder, porcelain
### 3. Predicated Device
K092513 RAINBOW BLOCK
### 4. Description
rainbow™ Shade is a dental ceramic made out of colored ZrO2(Y-TZP). rainbow™ Shade is milled into cores of teeth and then is fired in a furnace to harden the ZrO2. Then, the core is layered with porcelain to make a finished tooth.
#### 5. Indication for use
rainbow™ Shade is a pre-sintered zirconia(Y-TZP) block for fabrication of single-tooth and bridgework restoration.
#### 510(k) Summary
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first three letters, "GEN", are in a dark green color, while the last three letters, "OSS", are in a lighter green color. The font is bold and slightly italicized, giving the word a dynamic appearance. The overall design is simple and modern.
### 6. Technological Characteristics
rainbow™ Shade has the similar technological characteristics as the predicate device; main material, indication for use and design. Technological characteristics of rainbow™ Shade and RAINBOW BLOCK are as following
| Device name | rainbow™ Shade | RAINBOW BLOCK |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Genoss Co., Ltd. | Genoss Co., Ltd. |
| 510(k) Number | New Device | K092513 |
| Materials | ZrO2(Y-TZP) | ZrO2(Y-TZP) |
| Form | Preformed block | Preformed block |
| Sterilization | Non-sterile | Non-sterile |
| Indication for use | rainbow™ Shade is a pre-sintered<br>zirconia(Y-TZP) block for<br>fabrication of single-tooth and<br>bridgework restoration. | Rainbow Block is used in the<br>manufacture of a dental core through<br>milling by machine (MAD/MAM or<br>CAD/CAM) followed by sintering. |
| Use | Prescription | Prescription |
| Properties | | |
| Bending Strength<br>(Flexural strength) | 913MPa | 1144MPa |
| Sintering Density<br>(g/cm³) | 6 | 6.00~6.04 |
| Radioactivity<br>(Bq/kg) | 0.01 | 0.01 |
| Chemical Solubility<br>(µg/cm²) | 22.65 | 14.47 |
| Biocompatibility | Biocompatible | Biocompatible |
The differences of 'Indication for use' are not critical to the intended use. The general intended use of zirconia blank is to manufacture a dental prosthesis using milling machine. The dental core means single or bridgework restoration.
And the differences of technological characteristics(bending strength, chemical solubility) are within what is expected of this type of device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first four letters, "GENO", are in a dark green color. The last three letters, "OSS", are in a lighter green color.
### 7. Performance Data
Biocompatibility testing on the proposed rainbow™ Shade has been completed. Requirements for biological evaluation of the proposed device were based on FDA recognized concensus standard of ISO10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological bone and tissues with its intended use. The following biocompatibility tests were completed:
| No. | Items | Criteria | Result | Standard |
|-----|-------------------------------------------|---------------------------------|---------------------------------|--------------|
| 1 | Cytotoxicity | None cytotoxicity | None cytotoxicity | ISO 10993-5 |
| 2 | Sensitization<br>(Local Lymph Node Assay) | None sensitization | None sensitization | ISO 10993-10 |
| 3 | Oral mucosa irritation | Less than stimulus<br>threshold | Minimal irritation | ISO 10993-10 |
| 4 | Acute systemic toxicity | None acute Systemic<br>toxicity | None acute Systemic<br>toxicity | ISO 10993-11 |
| 5 | Genotoxicity | None genotoxicity | None genotoxicity | ISO 10993-3 |
| | | | | ISO 10993-3 |
The proposed rainbow™ Shade was evaluated using the following performance bench testing to confirm the performance characteristics:
| No. | Items | Criteria | Result | Standard |
|-----|--------------------------------------|------------------------------------------|-----------------------------------------------------------|-------------------------------|
| 1 | Visual | No impurities<br>and No specific changes | No impurities<br>and No specific changes | ISO 6872 |
| 2 | Size | Size error of;<br>Standard Size < ±5% | Size error of:<br>Standard Size < ±5%<br>H 0.43%, Ø 0.03% | ISO 6872 |
| 3 | Package | No damage | No damage | ISO 6872 |
| 4 | Uniformity | Uniform | Uniform | ISO 6872 |
| 5 | Freedom from<br>extraneous materials | Not Freedom from<br>extraneous materials | Not Freedom from<br>extraneous materials | ISO 6872 |
| 6 | Radioactivity | 238U Less than 1.0 Bq/g | 238U: < 0.000124Bq/g<br>226Ra: < 0.010Bq/g | ISO 6872 |
| 7 | Chemical solubility | Less than 100 µg/cm² | Solubility: 0µg/cm² | ISO 6872 |
| 8 | Flexural strength | More than 800MPa | Average: 913MPa | ISO 6872 |
| 9 | Linear thermal<br>expansion | 10.8(±0.5) X 10⁻⁶K⁻¹ | Average: 10.44 X 10⁻⁶K⁻¹ | ISO 6872 |
| 10 | Shipping Test | Result: Pass | Pass | ISTA2011<br>Integrity Test 3A |
{6}------------------------------------------------
| | Low Temperature degradation test | | | |
|---|----------------------------------|------------------------------------------------------------------------------|------------------------------------------|----------|
| 1 | Monoclinic phase fraction | Normal: Less than 20% after low temperature degradation: Less than 25% | Normal: 1.26%<br>134°C, 5 hours: 1.69% | ISO 6872 |
| 2 | Flexural strength | Less than 20% and more than 800MPa before/ after low temperature degradation | Normal: 850MPa<br>134°C, 5 hours: 832MPa | ISO 6872 |
All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the rainbow™ Shade met the established specifications necessary for consistent performance according to its intended use.
## 7. Conclusion
Based on the information provided in this premarket notification of GENOSS Co., Ltd. Concludes that rainbow™ Shade is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.