K151810 · Shenzhen Pump Medical System Co., Ltd. · DXQ · Dec 17, 2015 · Cardiovascular
Device Facts
Record ID
K151810
Device Name
Bpump Blood Pressure Cuff
Applicant
Shenzhen Pump Medical System Co., Ltd.
Product Code
DXQ · Cardiovascular
Decision Date
Dec 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1120
Device Class
Class 2
Indications for Use
The Bpump Blood Pressure Cuff, Model: BC1000, BC8000, are intended to be wrapped on the upper arm and used with a non-invasive blood pressure monitor system to complete the measurement of blood parameters on adults.
Device Story
Bpump Blood Pressure Cuff (Models BC1000, BC8000) is a reusable, rectangular, inelastic sleeve with an internal bladder and single-tube connection. Device wraps around the upper arm; connects to a non-invasive blood pressure monitor (NIBP) for inflation/deflation cycles. Operates as a passive accessory to an NIBP monitor; does not perform measurements independently. Used in clinical or home settings by patients or healthcare providers. Provides pneumatic interface for blood pressure monitoring; facilitates accurate cuff pressure transmission to the monitor. Benefits include reliable, non-invasive blood pressure assessment.
Clinical Evidence
Clinical validation performed in accordance with ANSI/AAMI/ISO 81060-2 and IEC 80601-2-30. Results confirmed the device met all applicable requirements for automated non-invasive sphygmomanometers.
Technological Characteristics
Reusable inelastic sleeve with bladder and single-tube pneumatic connection. Models: BC1000 (22-36cm arm range), BC8000 (32-44cm arm range). Biocompatibility per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization). Conforms to IEC 80601-2-30 for basic safety and essential performance.
Indications for Use
Indicated for adult patients requiring non-invasive blood pressure measurement using an external monitor system.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2015
Shenzhen Pump Medical System Co., Ltd. Ms. Migo Yang Clinical Registration Specialist 2/f, M-7 Sinosteel Building, Maqueling Estate Hi-tech Industrial Park Nanshan District, Shenzhen, 518057 CN
Re: K151810
Trade/Device Name: Bpump Blood Pressure Cuff, Model: BC1000, BC8000, Blood Pressure Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: June 15, 2015 Received: July 2, 2015
Dear Ms. Migo Yang,
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
Page 2 - Ms. Migo Yang
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K151810
Device Name
Bpump Blood Pressure Cuff, Model: BC1000, BC8000, Blood Pressure Cuff.
Indications for Use (Describe)
The Bpump Blood Pressure Cuff, Model: BC1000, BC8000, are intended to be wrapped on the upper arm and used with a non-invasive blood pressure monitor system to complete the measurement of blood parameters on adults.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K151810
Date Summary Prepared: Oct 16, 2015
# 1. Submitter's Identification:
SHENZHEN PUMP MEDICAL SYSTEM CO., LTD. 2/F, M-7 Sinosteel Building, Maqueling Estate, Hi-Tech Industrial Park, Nanshan District, 518057 Shenzhen, PEOPLE'S REPUBLIC OF CHINA
Contact: Xie Qiongyu Tel: 86-0755-26710795 Fax: 86-0755-26498210 E-mail:xieqy@bpump.com.cn
## 2. Device Information:
Trade Name: Bpump Blood Pressure Cuff Model No.: BC1000, BC8000 Common Name/Classification Name: Blood Pressure Cuff Product Code: DXQ Regulation Number: 21 CFR 870.1120 Regulation Class: Class II Review Panel: Cardiovascular
{4}------------------------------------------------
# 3. Predicate Device:
510(k) Number: K133117 Trade Name: Andon Blood Pressure Cuff Common Name/Classification Name: Blood Pressure Cuff Product Code: DXQ Regulation Number: 21 CFR 870.1120 Regulation Class: Class II Review Panel: Cardiovascular Manufactured By: Andon Health Co., Ltd.
# 4. Indication for Use
The Bpump Blood Pressure Cuff, Model: BC1000, BC8000, are intended to be wrapped on the upper arm and used with a non-invasive blood pressure monitor system to complete the measurement of blood parameters on adults.
#### 5. Intended Use
The Bpump Blood Pressure Cuff, Model: BC1000, BC8000, are intended to be wrapped on the upper arm and used with a non-invasive blood pressure monitor system to complete the measurement of blood parameters on adults.
# 6. Description of the device:
The proposed device, Bpump Blood Pressure Cuff, is a rectangle soft inelastic sleeve reusable with a bladder. There is a single-tube connected to the bladder and the Non Invasive Blood Pressure Monitor for inflating and deflating. The device should be connected to a Non Invasive Blood Pressure Monitor to complete the function. There are two sizes for different arm range as follows:
| Model | Arm range |
|--------|---------------------------|
| BC1000 | 22cm~36cm (8.7in~14.2in) |
| BC8000 | 32cm~44cm (12.6in~17.3in) |
{5}------------------------------------------------
Page 3 of 4
# 7. Summary Comparing Technological Characteristics with Predicate Device
The proposed changes described in this submission do not affect the intended use of the device and/or alter the fundamental scientific technology of the legally marketed predicate device described in the Kl 33177.Please refer to table below for comparing result:
| Technological Characteristics | Comparing Result |
|-------------------------------|------------------|
| Intended Use | Identical |
| Indications For Use | Identical |
| Patient Population | Identical |
| Environment of Use | Identical |
| Tube Configuration | Identical |
| Sterility | Identical |
| Applicable Arm Circumference | Similar |
| Standards Applied | Similar |
The subject device and predicate device are identical in intended use, Indications for Use, Patient Population, Environment of Use, Tube Configuration, Sterility, Standards; The subject device and predicate device are similar in Applicable Arm Circumference and Standards Applied. But the minor discrepancies mentioned above do not affect the safety or effectiveness of the subject device.
# 8. Discussion of non-clinical and clinical test performed
#### Non-clinical Tests have been done as follows:
a. Safety and performance characteristics of the test according to IEC 80601-2-30 b. Biocompatibility test performed according to ISO 10993-5 and ISO 10993-10 None of the test demonstrates that the Bpump Blood Pressure Cuff brings new questions of safety or effectiveness.
#### Clinical Test Concerning the Compliance of ANSI/AAMI/ISO 81060-2
Clinical test has been done in accordance with IEC 80601-2-30 and ANSI/AAMI/ISO 81060-2. and the test result shows, the device met all applicable requirements of the standard.
{6}------------------------------------------------
# 9. Performance:
#### Guidances
There is a guidance named "Guidance for Industry: Non-Automated Sphygmomanometer (Blood Pressure Cuff) Guidance Version I, November 19, 1998" for product code DXQ, which scope does not include any automated non-invasive blood pressure monitors whether using the oscillometric method or any other method. So, we concluded that the guidance is inapplicable to the subject device.
#### Standards
The Bpump Blood Pressure Cuff conforms to the following standards:
- · IEC 80601-2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- · ANSI/AAMI/ISO 81060-2: Non-invasive sphygmomanometers part 2: Clinical validation of automated measurement type.
- ISO 10993-1: Biological evaluation of medical devices Part I: Evaluation and testing within a risk manangement process
- · ISO 10993-5: Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity.
- · ISO 10993-10: Biological evaluation of medical devices Part I 0: Tests for irritation and skin sensitization
#### 10. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this premarket notification, SHENZHEN PUMP MEDICAL SYSTEM concludes that the subject device, the Bpump Blood Pressure Cuff, is safe, effective and substantially equivalent to Andon Blood Pressure Cuff (K133117).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.