← Product Code [IOR](/productcode/IOR) · K151797

# Nuocande Plastic Manual Wheelchair Model NKD L07 (K151797)

_Ningbo Lecount Medical Technology Co., Ltd. · IOR · Aug 25, 2015 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151797

## Device Facts

- **Applicant:** Ningbo Lecount Medical Technology Co., Ltd.
- **Product Code:** [IOR](/productcode/IOR.md)
- **Decision Date:** Aug 25, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine

## Indications for Use

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. It is designed for use indoors, and outdoors, over smooth surfaces (all standard indoor flooring surfaces, concrete, asphalt and pocked dirt) that are free of large obstacles and inclines greater than 5 degrees.

## Device Story

User-propelled, manually operated folding wheelchair; provides mobility for individuals unable to walk due to illness, injury, or disability. Device features four wheels, mechanical frame, and flame-resistant upholstery. Occupant propels by turning rear wheels; handles provided for caregiver-assisted pushing. Used in indoor and outdoor environments on smooth surfaces. Output is physical mobility for the patient. Benefits include restored independence and movement for mobility-impaired users.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing per ISO 7176 standards, including static stability (ISO 7176-1:1999), brake effectiveness (ISO 7176-3:2003), static/impact/fatigue strength (ISO 7176-8:1998), and ignition resistance (ISO 7176-16:1997).

## Technological Characteristics

Mechanical wheelchair; frame constructed of hard plastics; wheel/caster rims of steel; seat fabric is flame-resistant. Features four wheels, pull-to-lock brakes, and quick-release rear axle. Dimensions: 1090x730x850 mm. Weight capacity: 100 kg. Complies with ISO 7176-1, -3, -8, -11, -13, -15, and -16.

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Manual Wheelchair Model SY100-MA02A ([K132257](/device/K132257.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a group of people. The symbol is composed of three interconnected profiles facing to the right, creating a sense of unity and collaboration.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 25, 2015

Ningbo Lecount Medical Technology Co., Ltd. c/o Charles Shen Manton Business and Technology Services 37 Winding Road Oakland, NJ 07436

Re: K151797

Trade/Device Name: Nuocande Plastic Manual Wheelchair Model NKD L07 Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: June 17, 2015 Received: July 2, 2015

Dear Mr. Shen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Michael J. Hoffmann -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K151797

Device Name

Nuocande Plastic Manual Wheelchair Model NKD L07

Indications for Use (Describe)

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. It is designed for use indoors, and outdoors, over smooth surfaces (all standard indoor flooring surfaces, concrete, asphalt and pocked dirt) that are free of large obstacles and inclines greater than 5 degrees.

Type of Use (*Select one or both, as applicable*)

| <input type="checkbox"/> Research Use (Part 21 CFR 361 Subpart D) | <input checked="" type="checkbox"/> Commercially-Oriented Use (21 CFR 361 Subpart C) |
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------|
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------|

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 5: 510(k) Summary:

This summary of 510k safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92

#### Submitter & Foreign Manufacture Identification

Ningbo Lecount Medical Technology Co., Ltd. 57 Longfei Road, Yunlong Industry Zone, Yingzhou District, Ningbo, Zhejiang, 315135 China Submitter's FDA Registration Number: N/A

#### US Agent and Contact Person

Charles Shen Manton Business and Technology Services 37 Winding Ridge Oakland, NJ 07436 Tel: 608-217-9358 Email: cyshen@aol.com

Date of Summary: June 01, 2015

| Device Name:           |                                                    |
|------------------------|----------------------------------------------------|
| Proprietary Name:      | "Nuocande Plastic Manual Wheelchair Model NKD L07" |
| Common Name:           | Mechanical Wheelchair                              |
| Classification Name:   | Wheelchair, Mechanical                             |
| Device Classification: | 1                                                  |
| Regulation Number:     | 21 CFR 890.3850                                    |
| Panel: General         | Physical Medicine                                  |
| Product Code:          | IOR                                                |

#### Predicate Device Information:

- (1) K132257, "Manual Wheelchair Model SY100-MA02A", manufactured by "Zhejiang Jiafeng Electrical & Mechanical Co., Ltd"
#### Device description:

A mechanical wheelchair is a chair with wheels, designed to be a replacement for walking, where it is propelled by the seated occupant turning the rear wheels by hand. There are also handles behind the seat for someone else to do the pushing. Wheelchairs are used by people for whom walking is difficult or impossible due to illness, injury, or disability.

{4}------------------------------------------------

"Nuocande Plastic Manual Wheelchair Model NKD L07" is user propelled, manually operated folding wheelchairs that are indicated for use indoors and outdoors, over smooth surfaces (all standard indoor flooring surfaces, concrete, asphalt and pocked dirt) that are free of large obstacles and inclines greater than 5 degrees. Each consists of four wheels, a mechanical frame and nylon upholstery that is ignition resistant.

"Nuocande Plastic Manual Wheelchair Model NKD L07" has a physical dimension of a physical dimension of 1090 (depth) x 730 (width) x 850 (height) mm, with the seat itself has a dimension of 411 (depth) x 447 (width) x 365 (height) mm. The device has a weight capacity of 100 kilograms, and weighs about 19 kilograms. The color is dark yellow.

#### Intended Use:

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. It is designed for use indoors and outdoors, over smooth surfaces (all standard indoor flooring surfaces, concrete, asphalt and pocked dirt) that are free of large obstacles and inclines greater than 5 degrees.

#### Comparison to Predicate Devices

"Nuocande Plastic Manual Wheelchair Model NKD L07" manufactured by "Ningbo Lecount Medical Technology Co., Ltd." is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.

- (1) K132257, "Manual Wheelchair Model SY100-MA02A", manufactured by "Zhejiang Jiafeng Electrical & Mechanical Co., Ltd"

{5}------------------------------------------------

The following table shows similarities and differences of use, design, and material between our devices and the predicate device.

| Description                | Subject Devices                                                                                                                                                                                                                                                                                                                                      | Predicate Device (K132257), Model<br>SY 100-MA02A                                                                                                                                                                                                                                                                                                          |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use      | The device is intended for medical<br>purposes to provide mobility to persons<br>restricted to a sitting position. It is designed<br>for use indoors and outdoors, over smooth<br>surfaces (all standard indoor flooring<br>surfaces, concrete, asphalt and pocked dirt)<br>that are free of large obstacles and inclines<br>greater than 5 degrees. | The device is intended for medical<br>purposes to provide mobility to<br>persons restricted to a sitting<br>position. It is designed for use<br>indoors and outdoors, over smooth<br>surfaces (all standard indoor<br>flooring surfaces, concrete, asphalt<br>and pocked dirt) that are free of<br>large obstacles and inclines greater<br>than 8 degrees. |
| Basic Design               | Four wheels, a plastic frame and a textilene<br>upholstery that is flame resistant.                                                                                                                                                                                                                                                                  | Four wheels, a metal frame and a<br>nylon upholstery that is flame<br>resistant.                                                                                                                                                                                                                                                                           |
| Materials                  | Steel (wheel rim and caster rim), hard plastics<br>(Frame), and flame resistant fabrics (seat)                                                                                                                                                                                                                                                       | Steel (frame) and flame resistant<br>fabrics (seat)                                                                                                                                                                                                                                                                                                        |
| Powder                     | Mechanical                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                       |
| Dimension                  | 1090 x 730 x 850 (depth x width x height)                                                                                                                                                                                                                                                                                                            | 1025 x 661 x 920 mm (depth x<br>width x height)                                                                                                                                                                                                                                                                                                            |
| Armrest                    | Non Flip Back/Non Height Adjustable                                                                                                                                                                                                                                                                                                                  | Non Flip Back/Height Adjustable                                                                                                                                                                                                                                                                                                                            |
| Rear Axle                  | Offset Axle, Quick Release Axle                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                       |
| Back Wheel                 | 58.5 cm (=23 inch)                                                                                                                                                                                                                                                                                                                                   | 61 cm (=24 inch)                                                                                                                                                                                                                                                                                                                                           |
| Casters                    | 8 inch                                                                                                                                                                                                                                                                                                                                               | 8 inch                                                                                                                                                                                                                                                                                                                                                     |
| Wheel Lock                 | Pull to Lock                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                       |
| Weight<br>Capacity         | 100 Kg                                                                                                                                                                                                                                                                                                                                               | 130 Kg                                                                                                                                                                                                                                                                                                                                                     |
| Weight                     | 19 Kg                                                                                                                                                                                                                                                                                                                                                | 16.5 Kg                                                                                                                                                                                                                                                                                                                                                    |
| Color                      | Yellow                                                                                                                                                                                                                                                                                                                                               | Dark Green                                                                                                                                                                                                                                                                                                                                                 |
| Description                | Subject Device                                                                                                                                                                                                                                                                                                                                       | Predicate Device (K132257)                                                                                                                                                                                                                                                                                                                                 |
| Static Stability           | Meets ISO 7176-1:1999                                                                                                                                                                                                                                                                                                                                | Meets ISO 7176-1:1999                                                                                                                                                                                                                                                                                                                                      |
| Effectiveness of<br>Brakes | Meets ISO 7176-3: 2003                                                                                                                                                                                                                                                                                                                               | Meets ISO 7176-3: 2003                                                                                                                                                                                                                                                                                                                                     |
| Static Strength            | Meets ISO 7176-8: 1998                                                                                                                                                                                                                                                                                                                               | Meets ISO 7176-8: 1998                                                                                                                                                                                                                                                                                                                                     |
| Impact Strength            | Meets ISO 7176-8: 1998                                                                                                                                                                                                                                                                                                                               | Meets ISO 7176-8: 1998                                                                                                                                                                                                                                                                                                                                     |
| Fatigue Strength           | Meets ISO 7176-8:1998                                                                                                                                                                                                                                                                                                                                | Meets ISO 7176-8:1998                                                                                                                                                                                                                                                                                                                                      |
| Resistance to<br>Ignition  | Meets ISO 7176-16:1997                                                                                                                                                                                                                                                                                                                               | Meets ISO 7176-16:1997                                                                                                                                                                                                                                                                                                                                     |

## Table 5.1: Comparison of Intended Use, Design, and Material

Our devices and the predicate device are almost identical in terms of all areas described in the above table (Table 5.1). There are some minor differences with the predicate device don't affect the safety or effectiveness of the subject device.

The following table shows similarities and differences of the performance between our devices and the predicate device. Tests were conducted following the recommended

{6}------------------------------------------------

procedures outlined in the respective consensus standards, and results for "Nuocande Plastic Manual Wheelchair Model NKD L07'' met all relevant requirements in the test standards, our internal specifications, and are comparable to the predicate device.

Table 5.2: Comparison of Performance Testing

The tests were performed following the general requirements outlined in ISO 7176-11: 2012, 7176-13: 1989. The product labeling conforms to requirements of 7176-15: 1996.

#### A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its user-defined acceptance criteria and intended uses:

"Nuocande Plastic Manual Wheelchair Model NKD L07" meets performance requirements per ISO 7176-1:1999, ISO 7176-3: 2003, ISO 7176-8: 1998, ISO 7176-11: 2012, 7176-13: 1989, 7176-15: 1996 and ISO 7176-16:1997. They are safe and effective, and their performances meet the requirements of their pre-defined acceptance criteria and intended uses.

#### A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence.

Clinical data is not needed for manual wheelchair cleared by the 510(k) process.

# Substantial Equivalent Conclusions

Based on the comparison of intended use, design, materials, and performance, our "Nuocande Plastic Manual Wheelchair Model NKD L07" is substantial equivalent to their predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K151797](https://fda-staging.innolitics.com/device/K151797)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).