K151776 · Mespere Lifesciences, Inc. · PFA · Aug 12, 2015 · Cardiovascular
Device Facts
Record ID
K151776
Device Name
Central Venous Pressure System
Applicant
Mespere Lifesciences, Inc.
Product Code
PFA · Cardiovascular
Decision Date
Aug 12, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1140
Device Class
Class 2
Indications for Use
CVP-VENUS-2000 is intended to be used by healthcare professionals for assessment of central venous blood pressure (CVP) of adult individuals.
Device Story
Non-invasive system for central venous pressure (CVP) assessment; utilizes near-infrared spectroscopy (NIRS) technology to indicate right heart pressure at superior vena cava. Input: optical signals via light emitters and photo detectors. Processing: analog-to-digital conversion at sensor head; signal processing via tablet/laptop. Output: CVP measurement displayed for clinician review. Used in hospital/clinical environments by healthcare professionals. Benefits: non-invasive monitoring of venous pressure; improved sensitivity and data management compared to predicate; reduced cable noise via sensor-head digitization; enhanced usability via lighter cables and smaller sensor head.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified through safety testing (IEC 60601-1, IEC 60601-1-2), functional system testing, and bench-top comparisons against the predicate Venus 1000 system to ensure equivalent performance.
Technological Characteristics
Non-invasive NIRS technology. Core components: light emitters, photo detectors, A/D converter at sensor head. Connectivity: tablet/laptop interface. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC/ISO 62366, ISO 14971, ISO 14155, IEC/EN 62304, EN 1041, ISO 15223-1, ISO 13485, ISO 10993-1, ISO 2859-1.
Indications for Use
Indicated for assessment of central venous blood pressure (CVP) in adult individuals by healthcare professionals.
Regulatory Classification
Identification
A venous blood pressure manometer is a device attached to a venous catheter to indicate manometrically the central or peripheral venous pressure.
Predicate Devices
Central Venous Pressure System - Mespere Venus 1000 (K131085)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, suggesting a sense of community and care.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 12, 2015
Mespere Lifesciences Inc. Helen Tan Vice President 180 Frobisher Dr. Unit 1C Waterloo, Ontario, N2V2A2 CA
Re: K151776
Trade/Device Name: Central Venous Pressure System - Mespere Venus 2000 Regulation Number: 21 CFR 870.1140 Regulation Name: Venous Blood Pressure Manometer Regulatory Class: Class II Product Code: PFA Dated: July 9, 2015 Received: July 15, 2015
Dear Helen Tan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for "Mespere LIFE SCIENCES" in gray and blue. To the right of the logo, the text "Central Venous Pressure System - Special 510(k)" is written in black. The text describes the central venous pressure system.
# SECTION 006 INDICATIONS FOR USE STATEMENT
## Indications for Use
510(k) Number (if known):
Device Name: Central Venous Pressure System - Mespere Venus 2000
Indications for Use: CVP-VENUS-2000 is intended to be used by healthcare professionals for assessment of central venous blood pressure (CVP) of adult individuals.
Prescription Use _Yes_________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
No Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
SECTION 007
Image /page/3/Picture/1 description: The image shows the logo for Mespere Life Sciences. The logo features a large, bold, blue vertical line on the left side. To the right of the line is the word "Mespere" in a light gray, sans-serif font. Below "Mespere" are the words "LIFE SCIENCES" in a smaller, sans-serif font, also in light gray.
180 Frobisher Drive, Suite 1C Waterloo. ON N2V 2A2
t: 519-884-7575 £ 519-884-8810
www.mespere.com
### 510(k) SUMMARY
| Submitted by: | Mespere LifeSciences Inc.<br>180 Frobisher Drive, Unit 1C<br>Waterloo, Ontario, Canada<br>N2V 2A2<br>Phone : 1(519) 884-7575<br>Fax: 1(519) 884-8810 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company contact: | Helen Tan, Vice President |
| Date summary<br>prepared: | 11/05/2015 |
| Trade name: | Central Venous Pressure System – Mespere Venus 2000 |
| Common name: | CVP System |
| Classification<br>name: | 870.1140 (Product code PFA) |
| Classification: | Class II Medical Device |
| Predicate Device: | K131085<br>Central Venous Pressure System - Mespere Venus 1000<br>Product code: PFA<br>Regulation number: 870.1140 |
#### Device Description:
The Mespere LifeSciences Inc. CVP-VENUS-2000 Central Venous Pressure System is used to indicate right heart pressure at the superior vena cava, which is clinically referred to as central venous pressure (CVP).
The Venus 2000 is a modified version of Venus 1000 which has been cleared by FDA. The modification is to improve the detection sensitivity, data managements, and usability. The main improvements are moving the analog to digital conversion chip from the handheld to the sensor head to reduce cable noise and enhance the sensitivity; using a tablet/laptop instead of the handheld unit for faster speed of signal
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "Mespere Life Sciences". The logo features a large, bold, blue vertical line on the left side. To the right of the line is the word "Mespere" in a light gray, sans-serif font. Below "Mespere" are the words "LIFE SCIENCES" in a smaller, blue, sans-serif font.
180 Frobisher Drive, Suite 10 Waterloo ON N2V 2A7
t: 519-884-7575 £ 519-884-8810
www.mespere.com
processing and easier data management; and using lighter sensor cable and smaller sensor head to enhance usability.
The Venus 2000 system has the same fundamental scientific technology and intended use as the predicate device, the Venus 1000 System.
The device is intended for use in hospital and clinical environments by healthcare professionals.
#### Intended Use/Indications For Use:
The Venus 2000 System is intended to be used by healthcare professionals for assessment of central venous blood pressure (CVP) of adult individuals.
#### Technology Characteristics:
The Venus 2000 System and the predicate device, the Venus 1000 System, have the same intended use and technology characteristics. The Venus 2000 System and the Venus 1000 System are both based on non-invasive near infrared spectroscopy (NIRS) technology. The core hardware and software design of the Venus 2000 was kept the same as the Venus 1000, including key components such as light emitters, photo detectors, signal processing, and algorithm.
#### Complied Standards:
The design of the Mespere Venus 2000 device conforms to the following voluntary standards:
- o IEC 60601-1: 2005 + A1 2012 - edition 3.0
- IEC 60601-1-2: 2007 edition 3.0 ●
- IEC 60601-1-6: 2010 ●
- IEC/ISO 62366:2007 ●
- ISO 14971:2007 ●
- ISO 14155:2011
- IEC/EN 62304:2006 ●
- EN 1041:2008
- ISO 15223-1:2012
- ISO 13485:2003 ●
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Mespere Life Sciences. The logo features a large, blue vertical line on the left, followed by the word "Mespere" in a light gray, sans-serif font. Below "Mespere" are the words "LIFE SCIENCES" in a smaller, light blue font.
180 Fronisher Drix Waterloo ON N2V
519-884-7575 £ 519-884-8810
www.mespere.com
- ISO 10993-1:2009 ●
- ISO 2859-1:1999 ●
### Tests:
The Venus 2000 System was subjected to extensive safety testing, performance testing, verification, and validation testing.
The third party tests against IEC 60601-1: 2005 + A1 2012 and IEC 60601-1-2: 2007 of the Venus 2000 System ensured that the device met the industry and safety standards.
The system test of the Venus 2000 System ensured that the device met all its functional specifications.
The bench top test verified that the Venus 2000 System has the same performance as Venus 1000 System and the Venus 2000 System met the design requirements in an ideal non-clinical environment.
There are no further clinical tests required for the Venus 2000. The clinical performance of the Venus 1000 System has been verified and validated. The Venus 2000 system was a device modification. It continues to meet the requirements to patient safety and effectiveness.
#### Substantial Equivalence of Performance:
The Venus 2000 device is substantially equivalent to the predicate device, the Venus 1000 System, in terms of the performance and the technology safety and effectiveness.
#### Conclusions:
The conclusions drawn from the above testing is that the CVP-VENUS-2000 Central Venous Pressure System is substantially equivalent to the predicate device, the CVP-VENUS-1000 Central Venous Pressure System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.