K151762 · Merete Medical GmbH · JDR · Dec 28, 2015 · Orthopedic
Device Facts
Record ID
K151762
Device Name
PediatrOS RigidTack/FlexTack
Applicant
Merete Medical GmbH
Product Code
JDR · Orthopedic
Decision Date
Dec 28, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The PediatrOS™ RigidTack™/ FlexTack™ bone staples are indicated for pediatric patients (children/ adolescents) with angular deformities or leg length discrepancies. Therefore PediatrOS™ RigidTack™ / FlexTack™ is intended to redirect the angle of growth of long bones and to correct leg length discrepancies by inhibition of longitudinal growth of the physis (growth plate) in growing children and adolescents. Specific pediatric conditions/diseases for which the devices will be indicated include: - Valgus, varus, or flexion, extension, deformities of the knee (femur and/or tibia) ● - Valqus or varus deformities of the elbow (humerus) ● - Leg length discrepancies
Device Story
PediatrOS™ RigidTack™/FlexTack™ are titanium alloy (TiAl6V4 ELI) bone staples for pediatric orthopedic growth modulation. RigidTack™ provides stiff mechanical fixation for epiphysiodesis; FlexTack™ features a flexible middle section allowing in vivo bending via bone growth forces for hemiepiphysiodesis. Both devices feature trapezoidal geometry, jagged legs for bone fixation, and cannulated legs for K-wire guided placement. Female threads in legs facilitate removal. Used by surgeons to correct angular deformities or leg length discrepancies in growing children. Benefit: minimally invasive growth modulation to correct skeletal alignment without major corrective osteotomy.
Clinical Evidence
Bench testing only. Mechanical strength, stiffness, mechanical behavior, and bone fixation tests were performed. Testing confirmed mechanical properties are as high or higher than predicate devices. No clinical data presented.
Technological Characteristics
Material: Titanium alloy TiAl6V4 ELI. Design: Trapezoidal bone staples with cannulated legs and jagged surfaces. FlexTack™ includes a flexible bending zone. Standards: ASTM-F564-10(2015). Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for pediatric patients (children/adolescents) with angular deformities (valgus, varus, flexion, extension of knee; valgus, varus of elbow) or leg length discrepancies requiring growth redirection or longitudinal growth inhibition of the physis.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 28, 2015
Merete Medical GmbH % Mr. Matthias Möllmann Senior Vice President Merete Medical, Incorporated 4 Crotty Lane, Suite 118 New York International Plaza New Windsor, New York 12553
Re: K151762
Trade/Device Name: PediatrOSTM RigidTack™/FlexTack™ Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDR, OBT Dated: December 8, 2015 Received: December 11, 2015
Dear Mr. Möllmann,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Ronald P. Jean - S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
- 1. Indications for Use Statement
# Indications for Use
510(k) Number (if known): K151762
Device Name: PediatrOS™ RigidTack™ / FlexTackTM
Indications for Use:
The PediatrOS™ RigidTack™/ FlexTack™ bone staples are indicated for pediatric patients (children/ adolescents) with angular deformities or leg length discrepancies. Therefore PediatrOS™ RigidTack™ / FlexTack™ is intended to redirect the angle of growth of long bones and to correct leg length discrepancies by inhibition of longitudinal growth of the physis (growth plate) in growing children and adolescents.
Specific pediatric conditions/diseases for which the devices will be indicated include:
- Valgus, varus, or flexion, extension, deformities of the knee (femur and/or tibia) ●
- Valqus or varus deformities of the elbow (humerus) ●
- Leg length discrepancies ●
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Merete. The logo consists of two interlocking circles on the left, followed by the word "merete" in a sans-serif font. A registered trademark symbol is located to the upper right of the word "merete". The logo is in a dark blue color.
# 510(k) Summary of Safety and Effectiveness Information
#### as required by 21 CFR 807.92
| Date Prepared: | 04th December 2015 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by: | Merete Medical GmbH<br>Alt-Lankwitz 102<br>12247 Berlin, Germany |
| Contact Person: | Matthias Möllmann<br>Senior Vice President<br>Merete Medical, Inc.<br>4 Crotty Lane - Suite 118<br>New York International Plaza<br>New Windsor, NY 12553<br>Phone: +49 163 06219832<br>E-Mail: mmoellmann@merete-medical.com |
| Device Name: | PediatrOS™ RigidTack™/FlexTack™ |
| Common Name: | Bone Staples |
| Classification Names: | Single/multiple component metallic bone fixation<br>appliances and accessories 21 CFR 888-3030 |
| Device Product Code: | JDR |
| Subsequent Product Code: | OBT |
| Proposed Regulatory Class: | Class II |
#### Legally marketed Devices to which substantial Equivalence is claimed:
K031493 Guided Growth Plate, UNIVERSITY OF UTAH/SCHOOL OF MEDICINE Blount Zimmalloy staple, Zimmer Inc. Warsaw, Indiana -
#### Device Description:
#### PediatrOS™ RigidTack™
The PediatrOS™ RigidTack™ bone staples are used to correct leg length discrepancies through (dual) epiphysiodesis of long bones. The PediatrOS™ RigidTack™ has a trapezoidal shape which allows the staples to fit well to the epiphyseal bone. The staples are made from titanium alloy TiAl6V4 ELI and are available in several sizes, so they can be used for children of various age groups (before reaching skeletal maturity). The jagged surfaces of the staples' legs help secure it into the bone to prevent loosening. The staples have cannulated legs to facilitate placement through K-wires. In order to achieve a stiff mechanical behavior, the staples are equipped with a support strut. Female threads, built in the cannulated legs, help to remove the staples with threaded K-wires and the appropriate explantation instruments.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Merete. The logo consists of two interlocking circles inside of a larger circle on the left, and the word "merete" in blue font on the right. There is a registered trademark symbol to the right of the word.
### PediatrOS™ FlexTack™
The PediatrOS™ FlexTack™ bone staples are used to correct axial deformities through hemiepiphysiodesis of long bones. The PediatrOS™ FlexTack™ has a trapezoidal shape which allows the staples to fit well to the epiphyseal bone. The staples are made from titanium alloy TiAl6V4 ELl and are available in several sizes, so they can be used for children of various age groups (before reaching skeletal maturity). The jagged surface of the staples' legs help secure it into the bone to prevent loosening. The staples have cannulated legs to facilitate placement through K-wires as well as a flexible middle part which can be bent in vivo by the strength of bone growth permitting successive growth delay. Female threads, built in the cannulated legs, help to remove the staples with threaded K-wires and the appropriate explantation instruments.
### Indications for Use
The PediatrOS™ RigidTack™ / FlexTack™ bone staples are indicated for pediatric patients (children/ adolescents) with angular deformities or leg length discrepancies. Therefore PediatrOS™ / FlexTack™ is intended to redirect the angle of growth of long bones and to correct leg length discrepancies by inhibition of longitudinal growth of the physis (growth plate) in growing children and adolescents.
Specific pediatric conditions/diseases for which the devices will be indicated include:
- Valgus, varus, or flexion, extension, deformities of the knee (femur and/or tibia)
- . Valgus or varus deformities of the elbow (humerus)
- Leg length discrepancies
### Comparison of technological characteristics with the predicate devices:
In order to demonstrate that the PediatrOS™ RigidTack™ / FlexTack™ bone staples have the mechanical properties necessary to perform as well as or better than the predicate devices, Merete has conducted mechanical analysis and functional worst-case tests were stipulated by the optimatized design of the staple legs compared to the Blount Zimmalloy staple and the differences in the mechanism of growth following compared to the guided growth plate. The PediatrOS™ FlexTack™ has a flexible bending zone in contrast to a rigid plate with screw holes that allow screw displacement in distal-proximal direction in the guided growth plate. Therefore the following tests have been performed:
- Mechanical strength
- Stiffness
- Mechanical behavior
- Fixation into the bone
Those tests have demonstrated, that the mechanical strength is as high or higher than the strength of the predicate device. The stiffness is comparable to both predicate devices, too. The mechanical behavior comparison illustrated that the flexible bending zone in FlexTack™ deforms under small loads as intended and allows larger opening angles than physiological possible. A secure fixation into the bone was proven. If applicable the tests were performed according to valid standards as ASTM-F564-10(2015).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Merete. The logo consists of a circular graphic on the left and the word "merete" in a dark blue sans-serif font on the right. A registered trademark symbol is located to the right of the word.
All products have successfully passed the tests. The PediatrOS™ RigidTack™ / FlexTack™ bone staples have hereby been proven to be mechanically as good as or better than the compared predicate devices.
# Substantial Equivalence:
The PediatrOS™ RigidTack™ / FlexTack™ bone staples have passed all defined criteria, have performed as well or better than the predicate devices and are therefore considered substantially equivalent to the cleared predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.