The IMMEDIATELOAD Dental implant system is intended to be surgically inserted into the jawbone for support and retention of dental prosthesis such as artificial teeth, crowns, bridges and overdenture to restore the aesthetics of the patient and the mastication function. The IMMEDIATELOAD dental implant system is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading. The UNIVERSE implants can be used for surgery in two stages or in single stage. The POWER and SOLUTION implants can be used for surgery in single stage.
Device Story
IMMEDIATELOAD Dental Implant System consists of endosseous dental implants (UNIVERSE, POWER, SOLUTION) and prosthetic abutments; used by dental surgeons in clinical settings. Implants are surgically inserted into jawbone to support dental prostheses; UNIVERSE implants support two-stage or single-stage surgery; POWER and SOLUTION implants support single-stage surgery. Implants feature double external threads and tapered/cylindrical designs; connection to abutments via locking taper and screw. Abutments (straight or 15° inclined) engage, form, and create dentures. System restores patient aesthetics and mastication function. Immediate loading is permitted given sufficient primary stability. Healthcare providers use the system to provide stable foundations for fixed or removable dental restorations, benefiting patients by replacing missing teeth and restoring oral function.
Clinical Evidence
Bench testing only. Testing included fatigue testing per ISO 14801:2007, biocompatibility testing per ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010, and sterilization validation per ISO 17665-1:2006, 17665-2:2009, and ANSI/AAMI ST79:2010. Fatigue results were comparable to predicate devices; cytotoxicity testing demonstrated biocompatibility.
Indicated for patients requiring dental prosthesis (artificial teeth, crowns, bridges, overdentures) for jawbone support and retention to restore aesthetics and mastication. Suitable for immediate loading when primary stability and occlusal conditions are met.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 3, 2016
IMMEDIATELOAD S.A. Mr. Giovanni Canino Managing Director Via Lugano 11 Agno, 6982 SWITZERLAND
Re: K151757
Trade/Device Name: IIMMEDIATELOAD Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: June 24, 2016 Received: June 27, 2016
Dear Mr. Canino:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K151757
Device Name
IMMEDIATELOAD DENTAL IMPLANT SYSTEM
Indications for Use (Describe)
The IMMEDIATELOAD Dental Implant System is intended to be surgically inserted into the jawbone for support and retention of dental prosthesis such as artificial teeth, crowns, bridges and overdenture to restore the aesthetics of the patient and the mastication function.
The IMMEDIATELOAD dental implant system is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
The UNIVERSE implants can be used for surgery in two stages or in single stage.
The POWER and SOLUTION implants can be used for surgery in single stage.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K151757 510 (K)SUMMARY
## SUBMITTER/510(K) HOLDER:
| Company Name: | IMMEDIATELOAD S.A. |
|------------------------|------------------------------------------|
| Company Address: | Via Lugano 11<br>6982 Agno - Switzerland |
| Company Phone: | 041- 916001310 |
| Company Fax: | 041- 916001310 |
| Company e-mail: | qualityassurance@immediateload.com |
| Contact person: | Mr. Giovanni Canino<br>Managing Director |
| Date Summary Prepared: | August 2, 2016 |
## DEVICE IDENTIFICATION
| Trade/Proprietary Name: | IMMEDIATELOAD Dental Implant System |
|-------------------------|-------------------------------------|
| Classification Name: | Endosseous Dental Implant |
| Device Class: | Class II |
| Product Code: | DZE & NHA |
| Classification Panel: | Dental |
| Regulation Number: | 872.3640 |
## LEGALLY MARKETED PREDICATE DEVICE
| Predicate device | 510 (k) Holder | 510 (k) No. |
|-------------------------------------------|-----------------------|-------------|
| ANKYLOS C/X IMPLANT SYSTEM* | Dentsply Intl., Inc. | K083805 |
| NobelActive Internal Connection Implant** | Nobel Biocare USA LLC | K071370 |
| Alpha Bio Dental Implant System** | ALPHA-BIO TEC LTD | K063364 |
* - Primary Predicate
** - Reference Predicate
{4}------------------------------------------------
## DEVICE DESCRIPTION
All IMMEDIATELOAD dental implants have a double external thread that looks flattened from the proximal area to the distal area. There are three types of implants: UNIVERSE (conical biphasic), POWER (conical monophasic), and SOLUTION (cylindrical monophasic), available in different diameters and length combinations. All the implants are made of Titanium Grade 4. They are provided sterile and are ready to be implanted.
For each measure, the connection between the biphasic implant and the relative prosthetic component is secured by a locking taper connection and a preformed hexagon and by a retained screw.
The abutments are prosthetic accessories made of Titanium Grade 5; they are used to engage, form and create dentures. In particular, the abutment is designed to be applied on the biphasic implant during the surgical reopening, after osseointegration. The straight abutment IMMEDIATE has a straight axis compared to implant's axis. It has a transmucosal collar height ranging from 1 mm to 4 mm.
The inclined abutment IMMEDIATE has axis inclined of 15° compared to the implant's axis. It has a transmucosal collar height ranging from 1 mm to 3 mm. These abutments are fixed to the UNIVERSE implant via a connection screw. The OTK abutment is straight and contains an internal connection screw. It has a transmucosal collar height from 1 mm to 3 mm. The abutments are provided non-sterile and the connection screw for abutments are all provided non-sterile and are intended to be sterilized before use.
#### INDICATIONS FOR USE STATEMENT
The IMMEDIATELOAD Dental implant system is intended to be surgically inserted into the jawbone for support and retention of dental prosthesis such as artificial teeth, crowns, bridges and overdenture to restore the aesthetics of the patient and the mastication function.
The IMMEDIATELOAD dental implant system is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
The UNIVERSE implants can be used for surgery in two stages or in single stage.
The POWER and SOLUTION implants can be used for surgery in single stage.
## TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
IMMEDIATELOAD S.A. claims the substantial equivalence of the IMMEDIATELOAD Dental Implant System to the predicate devices, based on the equivalent intended use, fundamental technology, and operation characteristics. Side-by-side comparison between IMMEDIATELOAD Dental Implant System and its predicate devices is included below.
{5}------------------------------------------------
| COMPARISON CHART UNIVERSE DENTAL IMPLANTS | | | |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ATTRIBUTE /<br>CHARACTERISTICS | IMMEDIATELOAD<br>DENTAL IMPLANT<br>SYSTEM<br>(Submitted Device) | LEGALLY MARKETED<br>PREDICATE DEVICES<br>OF<br>Dentsply Intl., Inc. | LEGALLY MARKETED<br>PREDICATE DEVICES<br>OF<br>Nobel Biocare USA LLC |
| 510(k) number | K151757 | K083805 | K071370 |
| Device Name | UNIVERSE | ANKYLOS C/X IMPLANT<br>SYSTEM (Primary) | NobelActive Internal Connection<br>Implant (Reference) |
| Endosseous Dental<br>Implants | Root-form tapered | Root-form tapered | Root-form tapered |
| CFR Section | 872.3640 | 872.3640 | 872.3640 |
| Pro-code | DZE & NHA | DZE | DZE & NHA |
| Intended / Indications<br>For Use | The IMMEDIATELOAD<br>Dental implant system is<br>intended to be surgically<br>inserted into the jawbone<br>for support and retention of<br>dental prosthesis such as<br>artificial teeth, crowns,<br>bridges and overdenture to<br>restore the aesthetics of the<br>patient and the mastication<br>function.<br>The IMMEDIATELOAD<br>dental implant system is<br>indicated also for immediate<br>loading when good primary<br>stability is achieved and with<br>appropriate occlusal loading.<br>The UNIVERSE implants<br>can be used for surgery in<br>two stages or in single stage. | The ANKYLOS C/X Implant<br>System is for single-stage or two-<br>stage surgical procedures and<br>cemented or screw retained<br>restorations. The ANKYLOS<br>C/X Implant System is intended<br>for immediate placement and<br>function on single tooth and/or<br>multiple tooth applications when<br>good primary stability is<br>achieved, with appropriate<br>occlusal loading, in order to<br>restore chewing function.<br>Multiple tooth applications may<br>be splinted with a bar. | Intended for use in partially or<br>fully edentulous mandibles and<br>maxillae, in support of single or<br>multiple-unit restorations<br>including: cement-retained,<br>screw-retained, or overdenture<br>restorations and terminal or<br>intermediate abutment support<br>for fixed bridgework |
| Design-(Implants) | • Implant Type: bone Level<br>implant<br>• Connection Type: internal<br>conometric connection<br>• Neck Design: Straight<br>walled neck with<br>circumferential thread<br>provides crestal seal<br>• Body Design: Tapered<br>design enables placement<br>near impinging anatomical<br>structures while<br>maximizing prosthetic<br>table diameter for natural<br>emergence profile | • Implant Type: bone Level<br>implant<br>• Connection Type: internal<br>conometric connection<br>• Neck Design: Straight walled<br>neck with circumferential<br>thread provides crestal seal<br>• Body Design: Tapered design<br>enables placement near<br>impinging anatomical<br>structures while maximizing<br>prosthetic table diameter for<br>natural emergence profile | • Implant Type: bone Level<br>implant<br>• Connection Type: internal<br>conometric connection<br>• Neck Design: Straight walled<br>neck with circumferential<br>thread provides crestal seal<br>• Body Design: Tapered design<br>enables placement near<br>impinging anatomical<br>structures while maximizing<br>prosthetic table diameter for<br>natural emergence profile |
| Diameter (Implants) | 3,4 mm 4,0 mm 5,0 mm | 3,5 mm 4,5 mm 5,5 mm 7,0 mm | 3,0 mm 3,5 mm 4,3 mm 5,0 mm |
| Length (Implants) | from 8 to 15 mm | from 8 to 17 mm | from 8,5 to 18 mm |
| Material (Implants) | Pure titanium, ASTM F67 | Pure titanium, ASTM F67 | Pure titanium, ASTM F67 |
| Fixture Sterile -<br>(Implants) | Yes | Yes | Yes |
| Sterilization method | Beta Ray | Gamma Ray | Gamma Ray |
| Surface Finish<br>(Implants) | double acid etched | Sandblasted and etched | TiUnite Surface treatment |
| Design -(Abutments) | Healing Abutments, screws,<br>cemented and screwed fixed<br>posts; anchorage posts for<br>removable prostheses with<br>bars or ball attachments.<br>All components are<br>dedicated specifically to<br>UNIVERSE implants and fit<br>to their internal connection<br>and vertical drill hole in a<br>secure and anti-rotational<br>way. | EQUIVALENT Healing<br>Abutments, screws, cemented and<br>screwed fixed posts; anchorage<br>posts for removable prostheses<br>with bars or ball attachments. | EQUIVALENT Healing<br>Abutments, screws, cemented and<br>screwed fixed posts; anchorage<br>posts for removable prostheses<br>with bars or ball attachments. |
| Materials-(Abutments) | Titanium Ti 6Al 4V ELI,<br>ASTM F136 | Pure titanium, ASTM F67 | Titanium Ti 6Al 4V ELI,<br>ASTM F136 |
| Inclination (inclined<br>Abutments) | 15° | 7.5°, 15°, 22.5°, 30°, 37.5° | 15° , 17°, 30° |
| Fixture Sterile -<br>(Abutments) | NO, these parts of the<br>system can be sterilized by<br>the user. Labeling contains<br>instructions for<br>sterilization. | NO. These parts of the system<br>can be sterilized by the user. | NO. These parts of the system<br>can be sterilized by the user. |
{6}------------------------------------------------
| ATTRIBUTE/<br>CHARACTERISTICS | IMMEDIATELOAD DENTAL<br>IMPLANT SYSTEM (Submitted<br>Device) | LEGALLY MARKETED<br>PREDICATEDEVICES OF ALPHA-<br>BIO TEC LTD (Reference) |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K151757 | K063364 |
| Device Name | SOLUTION and POWER | Alpha Bio Dental Implant System |
| Endosseous Dental<br>Implants | Root-form tapered<br>(cylindrical and conical implant) | Root-form straight & tapered<br>(cylindrical and conical implant) |
| CFR Section | 872.3640 | 872.3640 |
| Pro-code | DZE | DZE |
| Intended / Indications<br>For Use | The IMMEDIATELOAD Dental implant<br>system is intended to be surgically inserted<br>into the jawbone for support and retention of<br>dental prosthesis such as artificial teeth,<br>crowns, bridges and overdenture to restore<br>the aesthetics of the patient and the<br>mastication function.<br><br>The IMMEDIATELOAD dental implant<br>system is indicated also for immediate<br>loading when good primary stability is<br>achieved and with appropriate occlusal<br>loading.<br><br>The POWER and SOLUTION implants can<br>be used for surgery in single stage. | The Alpha-Bio Dental Implant System is indicated<br>for use in surgical and restorative applications for<br>placement in the bone of the upper or lower jaw to<br>provide support for prosthetic devices, such as<br>artificial teeth, in order to restore the patient's<br>chewing function. The Alpha-Bio Dental Implant<br>System is indicated also for immediate loading<br>when good primary stability is achieved and<br>with appropriate occlusal loading. |
| Design-(Implants) | • Implant Type: bone Level implant<br>• Connection Type: Not Applicable<br>• Neck Design: Straight walled neck with<br>circumferential thread provides crestal seal<br>• Body Design: Cylindrical and Tapered<br>design enables placement near impinging<br>anatomical structures while maximizing<br>prosthetic table diameter for natural<br>emergence profile | • Implant Type: bone Level implant<br>• Connection Type: Not Applicable<br>• Neck Design: Straight walled neck with<br>circumferential thread provides crestal seal<br>• Body Design: Tapered design enables<br>placement near impinging anatomical<br>structures while maximizing prosthetic table<br>diameter for natural emergence profile |
{7}------------------------------------------------
| POWER diameter and<br>length combinations | | |
|----------------------------------------------------------------------------------------------|-----------------------------------|------------------------------------------------------------------|
| 3,4 mm x 8, 10, 11.5, 13, 15<br>4,0 mm x 8, 10, 11.5, 13, 15<br>5,0 mm x 8, 10, 11.5, 13, 15 | | |
| SOLUTION Diameter<br>and length combination | 3,0mm diameter x 11, 13, and 15mm | |
| Predicate Implant<br>Diameter | N/A | 3,0 mm 3,3 mm 3,6 mm 4,2 mm 5,0 mm |
| Predicate Implant<br>lengths | N/A | from 8 to 15 mm (not all lengths available for all<br>diameters) |
| Material (Implants) | Pure titanium, ASTM F67 | Titanium alloy Ti 6Al 4V ELI, ASTM F136 |
| Fixture Sterile -<br>(Implants) | Yes | Yes |
| Sterilization method | Beta Ray | Gamma Ray |
| Surface Finish<br>(Implants) | double acid etched | TiUnite Surface treatment |
## PERFORMANCE DATA
IMMEDIATELOAD S.A. has conducted laboratory testing and determined device functionality and conformance to design input requirements. The material used for the IMMEDIATELOAD dental Implant System, as well as the manufacturing methods, are similar to legally marketed devices. Surface treatment used for IMMEDIATELOAD dental Implant System is considered equivalent to the surface treatment method used for the predicate devices.
Non-clinical Testing of the subject device included the following:
- End User Steam Sterilization Test according to ISO 17665-1:2006, 17665-2:2009 and ANSI/AAMI ST79:2010.
- · Biocompatibility tests according to ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.
- · Fatigue Testing according to ISO 14801:2007
The results regarding the fatigue testing of IMMEDIATELOAD dental implants and the predicate devices were comparable. Therefore, it concluded that the IMMEDIATELOAD dental implants perform as intended. Cytotoxicity testing has demonstrated the biocompatibility of the devices. Sterility validation testing was performed and demonstrated the equivalence of the devices to their predicates.
Implants are all made of Titanium medical grade 4 (commercially named pure titanium), while the abutments are grade 5 Ti4Al6V ELI Titanium.
{8}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE DISCUSSION
The subject device and the primary predicate device have slight differences in the language of the Indications for Use Statements; however, these slight differences specifying particular models to single or two-stage surgeries do not change the intended use of the device. The subject device and the predicated devices also have similar technological characteristics, and are made of similar, if not identical materials. The subject device and predicate devices encompass a very similar or the exact same range of physical dimensions, including diameter and length of the implants and diameter. height and angle of the abutments and a comparative surface area.
## CONCLUSION
IMMEDIATELOAD S.A. has demonstrated that the IMMEDIATELOAD Dental Implants System is substantially equivalent to the predicate devices in terms of intended use, material composition, fundamental scientific technology, principles of operation, and basic design.
Based on the foregoing, the IMMEDIATELOAD Dental Implants System is substantially equivalent to the legally marketed, claimed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.