K151754 · Medline Industries, Inc. · LYZ · Mar 22, 2016 · General Hospital
Device Facts
Record ID
K151754
Device Name
Glide-On Vinyl Examination Gloves
Applicant
Medline Industries, Inc.
Product Code
LYZ · General Hospital
Decision Date
Mar 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Glide-On Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands or fingers to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile, powder-free vinyl (PVC) examination gloves; inner polyurethane coating acts as lubricant to facilitate donning on wet, damp, or dry hands. Used in clinical settings by healthcare professionals to prevent cross-contamination between patient and examiner. Device provides physical barrier protection. Performance validated via bench testing for biocompatibility, physical properties, and frictional characteristics; user study evaluated donning ease and durability compared to similar polyurethane-coated vinyl gloves.
Clinical Evidence
No clinical testing was performed. Evidence consists of bench testing for biocompatibility (ISO 10993-10: irritation and sensitization), physical properties (ASTM D5250-06), and frictional properties (ASTM D1894). A user study (n=170) evaluated donning time and glove integrity on damp/dry hands compared to other polyurethane-coated vinyl gloves.
Technological Characteristics
Material: Polyvinyl chloride (PVC) with inner polyurethane coating (viscosity up to 18 centipoise). Dimensions and physical properties comply with ASTM D5250-06. Freedom from holes tested per ASTM D5151-06. Non-sterile. Single-use. No colorant used.
Indications for Use
Indicated for use by medical examiners to prevent cross-contamination between patient and examiner during medical procedures.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2016
Medline Industries, Inc. Jennifer Mason Senior Regulatory Affairs Specialist One Medline Place Mundelein, IL 60060
Re: K151754
Trade/Device Name: Glide-On Vinyl Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: February 22, 2016 Received: February 23, 2016
Dear Ms. Mason:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements
{1}------------------------------------------------
as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151754
Device Name Glide-On Vinyl Examination Gloves
Indications for Use (Describe)
The Glide-On Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands or fingers to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medline logo. The logo features the word "MEDLINE" in white, bold sans-serif font, positioned to the left of a white, stylized starburst symbol. The background is a solid dark blue color. The logo is simple, clean, and easily recognizable.
ledline Industries, Inc. One Medline Place Mundelein, IL 60060
# 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
Submitter / 510(k) Sponsor
Medline Industries, Inc. 1 Medline Place Mundelein, IL 60060
Registration Number: 1417592
# Contact Person
Jennifer Mason Senior Regulatory Affairs Specialist Phone: 847-643-3652 Email: jamason@medline.com
Summary Preparation Date March 21, 2016
Type of 510(k) Submission Traditional
## Device Name / Classification
Name of Device: Glide-On Vinyl Examination Gloves Proprietary Name: Glide-On Vinyl Examination Gloves Common Name: Vinyl Patient Examination Glove Classification Name: Patient Examination Glove Product Code: LYZ Product Class: Class I Regulation #: 21 CFR 880.6250 Classification Panel: General Hospital
## Predicate Device
Powder-Free PVC Vinyl Patient Examination Glove, Clear (non-colored) K142703
## Device Description
The Glide-On Vinyl Examination Gloves are single use only, disposable gloves intended for medical purposes to be worn on the hands or fingers of examiners. The gloves are powder-free and are made of poly
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is in white, sans-serif font, and to the right of the text is a white, stylized star shape that resembles a medical cross.
edline Industries. One Medline Place Mundelein II 60060
vinyl chloride (PVC) with an inner coating of polyurethane. This inner polyurethane coating acts as a lubricant to aid in the donning of the gloves when hands are wet, damp or dry. The gloves are offered nonsterile and are available in small, medium, large and extra large sizes.
## Indications for Use
The Glide-On Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands or fingers to prevent contamination between patient and examiner.
## Summary of Technological Characteristics
The Glide-On Vinyl Examination Gloves are substantially equivalent to the predicate, K142703. Both gloves have the same intended use, same material and the same device performance.
| Device Characteristic | Proposed Device | Predicate Device | Comparison<br>Analysis |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product Name | Glide-On Vinyl<br>Examination Gloves | Powder-free PVC Vinyl<br>Exam Gloves | N/A |
| 510(k) Reference | | K142703 | N/A |
| Product Owner | Medline Industries, Inc | Hebei Grandeast Plastic<br>Products Co., Ltd. | Different |
| Product Code | LYZ | LYZ | Same |
| Intended Use | The Glide-On Vinyl<br>Examination Gloves are<br>disposable device intended<br>for medical purposes that<br>are worn on the examiner's<br>hands or fingers to prevent<br>contamination between<br>patient and examiner. | The powder-free PVC vinyl<br>exam gloves are disposable<br>devices intended for<br>medical purposes that are<br>worn on the examiner's<br>hands or fingers to prevent<br>contamination between<br>patient and examiner. | Same |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Sizes | Small<br>Medium<br>Large<br>Extra-Large | Small<br>Medium<br>Large<br>Extra-Large | Same |
| Materials | PVC<br>Lubricant - Polyurethane<br>with a viscosity up to 18<br>centipoise | PVC<br>Lubricant - Polyurethane<br>with a viscosity up to 4.6<br>centipoise | Similar |
| Dimensions - Length | Complies with ASTM<br>D5250-06 (reapproved) | Complies with ASTM<br>D5250-06 (reapproved) | Same |
#### TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image features the Medline logo, which is set against a blue background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a white, stylized starburst or cross-like symbol. The symbol is oriented diagonally, with its points extending outwards from the center. The overall design is clean and corporate, conveying a sense of professionalism and reliability.
Medline Industries, Inc. One Medline Place Mundelein, IL 60060
| | 2011) ≥230mm | 2011) ≥230mm | |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------|
| Dimensions - Width | Complies with ASTM<br>D5250-06 (reapproved<br>2011)<br>Small - 85±5mm<br>Medium - 95±5mm<br>Large - 105±5mm<br>Extra Large - 115±5mm | Complies with ASTM<br>D5250-06 (reapproved<br>2011)<br>Small - 85±5mm<br>Medium - 95±5mm<br>Large - 105±5mm<br>Extra Large - 115±5mm | Same |
| Dimensions - Thickness | Complies with ASTM<br>D5250-06 (reapproved<br>2011)<br>Palm - 0.10±0.02mm<br>Finger - 0.10±0.02mm | Complies with ASTM<br>D5250-06 (reapproved<br>2011)<br>Palm - 0.10±0.02mm<br>Finger - 0.10±0.02mm | Same |
| Physical Properties | Complies with ASTM<br>D5250-06 (reapproved<br>2011)<br>Before Aging/After Aging<br>Tensile Strength ≥11MPa<br>Ultimate Elongation ≥300% | Complies with ASTM<br>D5250-06 (reapproved<br>2011)<br>Before Aging/After Aging<br>Tensile Strength ≥11MPa<br>Ultimate Elongation ≥300% | Same |
| Freedom from Holes | Complies with ASTM<br>D5250-06 (reapproved<br>2011) and ASTM D5151-<br>06 (reapproved 2011)<br>G-1, AQL 2.5 | Complies with ASTM<br>D5250-06 (reapproved<br>2011) and ASTM D5151-<br>06 (reapproved 2011)<br>G-1, AQL 2.5 | Same |
| Powder or Powder-free | Powder-free | Powder-free | Same |
| Residual Powder | Complies with ASTM<br>D5250-06 (reapproved<br>2011)<br><2mg per glove | Complies with ASTM<br>D5250-06 (reapproved<br>2011)<br><2mg per glove | Same |
| Colorant | No colorant used | No colorant used | Same |
| Sterile or Non-sterile | Non-sterile | Non-sterile | Same |
# Summary of Non-Clinical Testing
The biocompatibility evaluation for the Glide-On Vinyl Examination Gloves was conducted in accordance with ANSVAAMVISO 10993-1:2009 Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA. The Glide-On Vinyl Examination Gloves are classified as a surface contacting device with a limited contact duration of less than 24 hours.
The following tests were performed to evaluate the biocompatibility of the Glide-On Vinyl Examination Glove:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medline logo. The logo is set against a blue square background. The word "MEDLINE" is written in white, with a stylized white symbol resembling a star or a stylized cross above the text. A gray bar is visible on the right side of the blue square.
ledline Industries. Inc One Medline Place Mundelein, IL 60060
- ISO 10993-10: Irritation Intracutaneous reactivity ●
- ISO 10993-10: Delayed-Type Hypersensitivity (Sensitization) Buehler Test ●
An evaluation was performed comparing the frictional properties of the Glide-On Vinyl Examination Gloves to a plain vinyl glove. The testing was performed per ASTM D1894, Standard Test Method for Static and Kinetic Coefficients of Friction of Plastic Film and Sheeting. Both gloves were tested under wet and dry conditions.
A user study involving one hundred seventy (170) subjects was also performed to evaluate the ease of glove donning on damp hands and dry hands compared to three other vinyl gloves coated with polyurethane. The study evaluated the amount of time it took each glove and also the amount of glove tears that occurred during the donning of each glove.
## Summary of Clinical Testing
This section does not apply. No clinical testing was performed.
#### Conclusion
Based on the non-clinical performance testing Medline Industries, Inc. concludes that the Glide-On Vinyl Examination Gloves are substantially equivalent to the predicate, Powder-free PVC Vinyl Exam Gloves (K142703).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.