K151745 · Shaoxing Fuqing Health Products Co., Ltd. · EZD · Mar 3, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K151745
Device Name
FULL CARE Disposable PVC Nelaton Catheter
Applicant
Shaoxing Fuqing Health Products Co., Ltd.
Product Code
EZD · Gastroenterology, Urology
Decision Date
Mar 3, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The FULL CARE Disposable PVC Nelaton Catheter is indicated for intermittent catheterization of the urethra for those individuals who are unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder-voiding episode. The catheter is inserted into urethra to reach the bladder allowing urine to drain. The urinary catheter comes in a variety of sizes packaged sterile for single-use.
Device Story
Disposable PVC Nelaton Catheter; sterile, single-use device for intermittent bladder drainage. Input: manual insertion into urethra. Transformation: provides physical pathway for urine to drain from bladder. Output: urine drainage. Used in clinical or home settings; operated by patients or healthcare providers. Device consists of PVC catheter, available in uncoated or hydrophilic-coated variants, with optional accessories (collection bag, lubricant, povidone-iodine swabs, gloves, underpad). Benefits: enables bladder emptying for patients with urinary retention.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by biocompatibility testing (cytotoxicity, sensitization, irritation per ISO 10993) and bench testing (EN 1616:1997+A1:1999, EN 1618:1997).
Technological Characteristics
Materials: Polyvinyl chloride (PVC). Design: Intermittent urinary catheter, straight or rounded tip, two drainage eyelets. Sizes: 6Fr to 20Fr. Variants: Uncoated or hydrophilic-coated (P41001 coat). Sterilization: Ethylene Oxide (EO) for catheter; radiation for water sachets. Connectivity: None (standalone).
Indications for Use
Indicated for intermittent catheterization of the urethra in male, female, and pediatric patients unable to promote natural urine flow or those with significant residual urine volume following bladder-voiding.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with a flowing design beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 3, 2016
Shaoxing Fuqing Health Products Co., Ltd. % Mike Gu RA Manager Osmunda Medical Device Technology Service Co., Ltd Floor 7, Jingui Business Building 982 Congyun Rd., Baiyum District Guangzhou 510420 China
Re: K151745
Trade/Device Name: FULL CARE Disposable PVC Nelaton Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: Class II Product Code: EZD Dated: January 28, 2016 Received: February 2, 2016
Dear Mike Gu,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K151745
### Device Name
FULL CARE Disposable PVC Nelaton Catheter
### Indications for Use (Describe)
The FULL CARE Disposable PVC Nelaton Catheter is indicated for intermittent of the urethra for those individuals who are unable to promote a natural urine for those individuals who have a significant volume of residual urine following a natural bladder-voiding episode. The catheter is inserted into urethra to reach the bladder allowing watheter comes in a variety of sizes packaged sterile for single-use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
#### l. SUBMITTER
Shaoxing Fuqing Health Products Co., Ltd.
| No. 599 Wuxie Road, Shaoxing High-Tech Industrial Development Zone, | |
|---------------------------------------------------------------------|--------------------------------------------|
| Shaoxing Zhejiang, CHINA 312000 | |
| Phone: | +86-575-88675000-830 |
| Fax: | +86-575-88676000 |
| Primary Contact Person: | Mike Gu |
| | Regulatory Affairs Manager |
| | OSMUNDA Medical Device Consulting Co., Ltd |
| | Tel: (+86) 20-6232 1333 |
| | Fax: (+86) 20-8633 0253 |
| Secondary Contact Person: | Mr. Dean Wu |
| | Shaoxing Fuqing Health Products Co., Ltd. |
| Date Prepared: | April 14, 2015 |
II. DEVICE
| Name of Device: | FULL CARE Disposable PVC Nelaton Catheter |
|-----------------------|-------------------------------------------|
| Common/Usual Name: | Urethral Catheter |
| Classification Names: | Catheter, Urethral (21 CFR876.5130) |
| Regulation Class: | II |
| Product Code: | GBM, EZD, FCM |
{5}------------------------------------------------
#### lll. PREDICATE DEVICE
Cure Catheter™, K072539
Cure Pediatric Catheter, K110653
Cure Catheter™ Hydrophilic Coated, K132500
Cure Catheter™ Closed System, K080881
These predicates have not been subject to a design-related recall.
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The FULL CARE Disposable PVC Nelaton Catheter is a sterile, single use device designed as an intermittent pathway for drainage of the bladder. It is available for male, female and pediatric, in uncoated and coated variants and in two different tip configurations (straight and rounded). There are two drainage eyelets on the catheter.
The PVC catheter is designed as an intermittent pathway for drainage of the bladder. The devices consist of the following parts(with different combination):
- -Intermittent Urinary Catheter;
- hydrophilic coated Nelaton Catheter;
- water sachet; -
- collection bag;
- gauze tissue;
- povidone iodine swabs ; -
- two gloves;
- underpad;
- BZK; -
- -Lubricant;
- -Interface tube;
- pouch
The device is made of polyvinyl chloride. The catheter is provided sterile in a variety of lengths and sizes.
{6}------------------------------------------------
#### > INDICATIONS FOR USE
The FULL CARE Disposable PVC Nelaton Catheter is indicated for intermittent catheterization of the urethra for those individuals who are unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder-voiding episode. The catheter is inserted into urethra to reach the bladder allowing urine to drain. The urinary catheter comes in a variety of sizes packaged sterile for single-use.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The sponsor identified the similarities and differences of the proposed FULL CARE Disposable PVC Nelaton Catheter device to the legally marketed predicates K072539, K132500 and K080881 to which substantial equivalency is claimed.
| Specification | Predicate Device | Predicate Device | Proposed |
|---------------|-------------------|-------------------------|------------------------------------------------|
| Device name | Cure Catheter™ | Cure Pediatric Catheter | Nelaton Catheter<br>(NC) |
| 510k number | K072539 | K110653 | NA |
| Manufacturer | Cure Medical, LLC | Cure Medical, LLC | Shaoxing Fuqing<br>Health Products<br>Co., Ltd |
{7}------------------------------------------------
| | | An intermittent urinary<br>catheter that is inserted<br>through the urethra and<br>indicated for the purpose<br>of bladder drainage for<br>pediatric males and<br>females. The urinary<br>catheter comes in a variety<br>of sizes packaged sterile<br>for single-use. | The FULL CARE<br>Disposable PVC<br>Nelaton Catheter<br>is indicated for<br>intermittent<br>catheterization of<br>the urethra for<br>those individuals<br>who are unable to<br>promote a natural<br>urine flow or for<br>those individuals<br>who have a<br>significant volume<br>of residual urine<br>following a natural<br>bladder-voiding<br>episode. The<br>catheter is<br>inserted into<br>urethra to reach<br>the bladder<br>allowing urine to<br>drain. The urinary<br>catheter comes in<br>a variety of sizes<br>packaged sterile<br>for single-use. |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <i>Indications<br/>for Use</i> | The FULL CARE Disposable PVC Nelaton<br>Catheter is indicated for intermittent<br>catheterization of the urethra for those<br>individuals who are unable to promote<br>a natural urine flow or for those<br>individuals who have a significant<br>volume of residual urine following a<br>natural bladder-voiding episode. The<br>catheter is inserted into urethra to<br>reach the bladder allowing urine to<br>drain. The urinary catheter comes in a<br>variety of sizes packaged sterile for<br>single-use. | | |
| <i>Patient<br/>Population</i> | Available for male and female | Available for pediatric male<br>and female | Available for male<br>and female, and<br>pediatric male and<br>female. |
{8}------------------------------------------------
| Dimension | Female straight tip 6 inch from 8Fr to<br>18Fr with 2Fr increments;<br>Male straight tip 16 inch from 8Fr to<br>18Fr with 2Fr increments. | Pediatric straight tip 10<br>inch from 8Fr to 14Fr with<br>2Fr increments; | Male NC without<br>lubricant is<br>available from 8Fr<br>to 20Fr, female NC<br>without lubricant<br>is available from<br>8Fr to 16Fr with<br>2Fr increments;<br>Both male and<br>female NC with<br>lubricant is<br>available from 8Fr<br>to 18Fr with 2Fr<br>increments;<br>Pediatric is<br>available from 6Fr<br>to 10Fr with 2Fr<br>increments. |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>method | EO | EO | EO |
| Material | PVC | PVC | PVC |
| Specification | Predicate Device | Proposed |
|---------------|------------------------------------|-------------------------------------------|
| Device name | Cure CatheterTM Hydrophilic Coated | Hydrophilic Coated Nelaton Catheter (HNC) |
| 510k number | K132500 | NA |
| Manufacturer | Cure Medical, LLC | Shaoxing Fuqing Health Products Co., Ltd |
{9}------------------------------------------------
| Indications<br>for Use | The Cure Catheter™ Hydrophilic<br>Coated is an intermittent urinary<br>catheter that is inserted through the<br>urethra and indicated for the purpose<br>of bladder drainage for males and<br>females. The urinary catheter comes in<br>a variety of sizes packaged sterile for<br>single-use. | The FULL CARE Disposable PVC Nelaton Catheter<br>is indicated for intermittent catheterization of<br>the urethra for those individuals who are unable<br>to promote a natural urine flow or for those<br>individuals who have a significant volume of<br>residual urine following a natural bladder-<br>voiding episode. The catheter is inserted into<br>urethra to reach the bladder allowing urine to<br>drain. The urinary catheter comes in a variety of<br>sizes packaged sterile for single-use. |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient<br>Population | Available for male and female | Available for male and female |
| Dimension | Female 6 inch straight tip in size from<br>6Fr to 18Fr with 2Fr increments;<br>Male 16 inch straight and coude tip in<br>size from 8Fr to 18Fr with 2Fr<br>increments;<br>Pediatric 10 inch straight tip in size<br>from 6Fr to 14Fr with 2Fr increments; | The CB is available in size from 8Fr to 16Fr with<br>2Fr increments;<br>The Smile Cath is available in size from 8Fr to<br>18Fr with 2Fr increments;<br>The CS1 and CS2 are available in size from 8Fr to<br>16Fr with 2Fr increments; |
| Sterilization<br>method | EO for catheter and Radiation for water<br>sachet | EO for catheter and Radiation for water sachet |
| Material | PVC catheter with DSM Comfort Coat-<br>polyvinylpyrrolidone-Hydrophilic<br>Material Coating-and sterile water<br>sachet | PVC catheter with P41001 coat, and purified<br>water. |
| Specification | Predicate Device | Proposed |
|---------------|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Cure Catheter® Closed Systems | Intermittent Catheterization Bag (CB)+<br>Intermittent Catheterization Kit (Smile- Cath)+<br>Intermittent Catheterization Kit (CS1)+<br>Intermittent Catheterization Kit (CS2) |
| 510k number | K080881 | NA |
{10}------------------------------------------------
| Manufacturer | Cure Medical, LLC | Shaoxing Fuqing Health Products Co., Ltd |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Cure Catheter Closed System is an<br>intermittent urinary catheter attached<br>to a collection bag that is inserted<br>through the urethra and indicated for<br>the purpose of bladder drainage for<br>males and<br>females. The urinary<br>catheter comes in a variety of sizes<br>packaged sterile for single-use. | The FULL CARE Disposable PVC Nelaton Catheter<br>is indicated for intermittent catheterization of<br>the urethra for those individuals who are unable<br>to promote a natural urine flow or for those<br>individuals who have a significant volume of<br>residual urine following a natural bladder-<br>voiding episode. The catheter is inserted into<br>urethra to reach the bladder allowing urine to<br>drain. The urinary catheter comes in a variety of<br>sizes packaged sterile for single-use. |
| Patient<br>Population | Available for male and female | Available for male and female |
| Dimension | The CB is available in size from 8Fr to<br>16Fr with 2Fr increments;<br>The CS is available in size from 8Fr to<br>16Fr with 2Fr increments; | The CB is available in size from 8Fr to 16Fr with<br>2Fr increments;<br>The Smile Cath is available in size from 8Fr to<br>18Fr with 2Fr increments;<br>The CS1 and CS2 are available in size from 8Fr to<br>16Fr with 2Fr increments; |
| Anatomical<br>Sites | Urethra, bladder | Urethra, bladder |
| Sterilization<br>method | EO | EO |
| Material | PVC | PVC |
| Kit | Cure Catheter, collection bag, a BZK<br>wipe, three Povidone-iodine<br>swabsticks, one pair of gloves, a sterile<br>wipe, and an underpad | CS1:CB, underpad, gauze tissue, povidone iodine<br>swab, a BZK wipe, one pair of gloves.<br>CS2: NC, collection bag, a BZK wipe, underpad,<br>one pair of gloves, lubricant, |
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility testing:
{11}------------------------------------------------
The biocompatibility evaluation for the Disposable PVC Nelaton Catheter was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The catheter testing included the following tests:
- Cytotoxicity
- Sensitization
- . Irritation
The catheter is considered surface contacting for duration of less than 24 hours.
### Bench testing:
- . EN 1616:1997+A1:1999
- EN 1618:1997
It's found from the test results that the proposed devices meet the requirements.
### Animal and clinical study:
The subject of this premarket submission, Disposable PVC Nelaton Catheter, does not require clinical studies to support substantial equivalence.
#### VIII. CONCLUSIONS
Shaoxing Fuqing Health Products Co., Ltd. considers the Disposable PVC Nelaton Catheter does not raise any new issues of safety or effectiveness. And is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.