Cable-Ready Cable Grip System: Cable-Ready Bone Plate System
K151716 · Zimmer, Inc. · HRS · Jan 28, 2016 · Orthopedic
Device Facts
Record ID
K151716
Device Name
Cable-Ready Cable Grip System: Cable-Ready Bone Plate System
Applicant
Zimmer, Inc.
Product Code
HRS · Orthopedic
Decision Date
Jan 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cable-Ready® Bone Plate with Cerclage Cable is indicated for use where wire, cable, or band cerclage is used in combination with bone plates to provide fixation and/or stabilization of long bones femur, tibia and humerus. Examples include periprosthetic fractures or bone loss, comminuted shaft fractures, and nonunions of previous fractures with or without failed hardware.
Device Story
System comprises 316L stainless steel bone plates and multifilament cerclage cables; used for long bone fracture fixation (femur, tibia, humerus). Surgeon places plate on bone; inserts cable through plate and around bone; attaches and tightens cable to provide compressive force containing fracture fragments; cortical bone screws provide additional fixation. Used in clinical/surgical settings by orthopedic surgeons. Output is mechanical stabilization of bone fractures; aids healing by maintaining fragment alignment.
Clinical Evidence
No clinical data required. Bench testing included biocompatibility testing per ISO 10993-1 and Good Laboratory Practices (21 CFR § 58), which passed. Engineering analysis performed to evaluate performance differences.
Technological Characteristics
Materials: 316L stainless steel. Components: Bone plates (187mm/6-hole, 246mm/8-hole, 305mm/10-hole) and 1.8mm diameter/610mm length multifilament cerclage cable. Mechanical fixation via cable-grip and cortical bone screws.
Indications for Use
Indicated for patients requiring fixation/stabilization of long bones (femur, tibia, humerus) due to periprosthetic fractures, bone loss, comminuted shaft fractures, or nonunions of previous fractures with/without failed hardware.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Pioneer Laboratories Bone Plate with Cerclage Cable (K940729)
Pioneer Laboratories Bone Plate with Cable's Device (K972223)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping to create a sense of depth and connection. The profiles are rendered in a simple, line-art style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 28, 2016
Zimmer, Incorporated Ms. Dalene Binklev Senior Specialist, Trauma Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581
Re: K151716
Trade/Device Name: Cable-Ready® Cable Grip System: Cable-Ready® Bone Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, JDQ Dated: December 23, 2015 Received: December 28, 2015
Dear Ms. Binkley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the
{1}------------------------------------------------
quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| Form Approved: OMB No. | 0910-0120 |
|--------------------------|------------------|
| Expiration Date: | January 31, 2017 |
| See PRA Statement below. | |
| 510(k) Number ( <i>if known</i> ) | K151716 (page 1 of 1) |
|-----------------------------------|-----------------------|
|-----------------------------------|-----------------------|
| Device Name | Cable-Ready® Cable Grip System: Cable-Ready® Bone Plate System |
|-------------|----------------------------------------------------------------|
|-------------|----------------------------------------------------------------|
Indications for Use (Describe)
Cable-Ready® Bone Plate System: The Cable-Ready® Bone Plate with Cerclage Cable is indicated for use where wire, cable, or band cerclage is used in combination with bone plates to provide fixation and/or stabilization of long bones femur, tibia and humerus. Examples include periprosthetic fractures or bone loss, comminuted shaft fractures, and nonunions of previous fractures with or without failed hardware.
Type of Use (Select one or both, as applicable)
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Zimmer Biomet. The logo features a blue circle with a stylized letter "Z" inside. Below the circle, the word "zimmer" is written in lowercase, using a similar blue color as the "Z" in the circle above.
574 267-6131
## 510(k) Summary
| Sponsor: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dalene T. Binkley<br>Senior Specialist, Trauma Regulatory Affairs<br>Telephone: 574-372-4970<br>Fax: (574) 371-8760 |
| Date: | July 7, 2015 |
| Trade Name: | Cable-Ready® Cable Grip System: Cable-Ready® Bone<br>Plate System |
| Common Name: | Plate, Fixation, Bone<br>Cerclage, Bone Fixation |
| Classification Names<br>and References: | Plate, Fixation, Bone:<br>Regulation Number: CFR 888.3030<br>Classification Number: 87 HRS |
| | Cerclage, Fixation:<br>Regulation Number: CFR 888.3010<br>Classification Number: 87 JDQ |
| Classification Panel: | Orthopedics/87 |
| Predicate Device(s): | Pioneer Laboratories Bone Plate with Cerclage Cable,<br>Pioneer Laboratories, K940729, cleared December 27,<br>1994 |
| | Pioneer Laboratories Bone Plate with Cable's Device,<br>Pioneer Laboratories, K972223, cleared September 10,<br>1997 |
| Purpose and Device<br>Description: | The Cable-Ready® Bone Plate System is used to address<br>complicated fractures and reconstruction of the long bones<br>- femur, tibia and humerus. These devices are used as a<br>compressive force to aid the surgeon in containing fracture<br>fragments as they heal. The plate is placed on the bone and<br>secured with a multifilament cable that is inserted into the<br>plate and passed around the bone to the other side of the<br>plate where it is attached and tightened, securing the<br>fracture(s) in place. Cortical bone screws are used for<br>additional fixation as needed.<br><br>The bone plates are offered in 3 lengths – 187mm (6<br>holes), 246mm (8 holes) and 305mm (10 holes). The<br>cerclage cable is 1.8mm in diameter and 610mm in length.<br>Both the bone plates and cable are manufactured from<br>316L stainless steel. |
| Intended Use: | Cable-Ready® Bone Plate System:<br>The Cable-Ready® Bone Plate with Cerclage Cable is<br>indicated for use where wire, cable, or band cerclage is<br>used in combination with bone plates to provide fixation<br>and/or stabilization of long bones – femur, tibia and<br>humerus. Examples include periprosthetic fractures or<br>bone loss, comminuted shaft fractures, and nonunions of<br>previous fractures with or without failed hardware. |
| Comparison to Predicate Device: | The Cable-Ready® Bone Plate System is similar in intended<br>use, materials, basic shape and performance characteristics<br>to the predicate devices. |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>• Biocompatibility – Biocompatibility testing on the<br>Cable-Ready Bone Plates and Cerclage Cable materials<br>was conducted per ISO 10993-1 and Good Laboratory<br>Practices (21 CFR § 58). All testing passed.<br>• Performance Evaluation – The engineering analysis<br>establishes that the difference in length, number of screw<br>holes and crimp mechanism between the subject devices |
{4}------------------------------------------------
{5}------------------------------------------------
and the predicate devices do not affect the clinical strength of the subject bone plates and cable.
Conclusions: The data presented in this submission demonstrates that the subject devices are substantially equivalent to the predicate devices.
Clinical Performance and Conclusions:
• Clinical data and conclusions were not needed for these devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.