K151698 · Becton Dickinson Infusion Therapy Systems, Inc. · FOZ · Jul 27, 2015 · General Hospital
Device Facts
Record ID
K151698
Device Name
BD Angiocath IV Catheter, BD Insyte IV Catheter
Applicant
Becton Dickinson Infusion Therapy Systems, Inc.
Product Code
FOZ · General Hospital
Decision Date
Jul 27, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
An intravascular catheter is a device that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.
Device Story
Peripheral IV catheters (Angiocath and Insyte) designed for short-term vascular access; inserted into patient's vascular system by healthcare professionals. Used for blood sampling, blood pressure monitoring, or fluid administration. Subject devices feature modified needle lubricant compared to predicate. Design verification confirms substantial equivalence. No electronic or software components; purely mechanical device.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through design verification testing per ISO 594-1, ISO 594-2, ISO 10993-1, and ISO 10555-1/5, covering biocompatibility, leakage, tensile force, flow rate, and penetration force.
Technological Characteristics
Peripheral IV catheters. Materials: FEP or BD Vialon (polyurethane) tubing, polypropylene hubs/covers, stainless steel needles/wedges, polyethylene vent plugs. Lubricants: Polydimethylsiloxane-based. Dimensions: 14G-24G. Sterilization: Not specified. No software or electronic components.
Indications for Use
Indicated for any patient population requiring short-term (less than 30 days) vascular access for blood sampling, blood pressure monitoring, or intravenous fluid administration, provided vascular anatomy is adequate.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 27, 2015
Becton Dickinson Infusion Therapy Systems, Incorporated c/o Mr. Mark Job Regulatory Technology Services, Inc. 1394 25th Street, NW Buffalo, MN 55313
Re: K151698
Trade/Device Name: BD Angiocath™ and Insyte™ Catheters Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: July 15, 2015 Received: July 16, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
**Indications for Use**
| 510(k) Number ( <i>if known</i> ) | K151698 |
|-----------------------------------|-------------------------------------------------|
| Device Name | BD Angiocath IV Catheter and Insyte IV Catheter |
Indications for Use (Describe)
An intravascular catheter is a device that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">Over-The-Counter Use</span> (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5
# 510(k) Summary
# 510(k) Summary
## 21 CFR §807.92
## BD Angiocath™ and Insyte™ IV Catheters
| Submitter<br>Information | Submitter Name: | Becton Dickinson Infusion Therapy Systems Inc. |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| | Submitter Address: | 9450 South State Street<br>Sandy, UT 84070 |
| | Contact Person: | Henry Boland<br>Senior Regulatory Affairs Specialist<br>Phone: 801.565.2550<br>Fax: 801.304.3963<br>Email: henry_boland@bd.com |
| | Date of Preparation: | 14 May 2015 |
| Subject<br>Devices | Trade Name: | BD Angiocath™ IV Catheter |
| | Common Name: | Peripheral Intravascular or IV Catheter |
| | Classification Name: | FOZ - Intravascular Catheter |
| | CFR Reference: | 21 CFR 880.5200 - Class II |
| | Classification Panel: | General Hospital |
| | Trade Name: | BD Insyte™ IV Catheter |
| | Common Name: | Peripheral Intravascular or IV Catheter |
| | Classification Name: | FOZ - Intravascular Catheter |
| | CFR Reference: | 21 CFR 880.5200 - Class II |
| | Classification Panel: | General Hospital |
| Predicate<br>Devices | Trade Name: | BD Angiocath™ and Insyte™ IV Catheter |
| | 510(k) Reference: | K013800 |
| | Common Name: | Peripheral Intravascular or IV Catheter |
| | Classification Name: | FOZ - Intravascular Catheter |
| | CFR Reference: | 21 CFR 880.5200 - Class II |
| | Classification Panel: | General Hospital |
| Device<br>Description | Insyte™, Insyte-N™, and Insyte-W™ are available in 14-24 gauge and are made<br>of BD Vialon™ Material. Angiocath™ is available in 14-24 gauge and is made of<br>FFP Polymer material | |
{4}------------------------------------------------
#### Indications for An intravascular catheter is a device that is inserted into the patient's vascular Use system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.
Technological characteristics of the subject devices are equivalent to the Technological predicates. The BD Angiocath™ and Insyte™ IV Catheters achieve their intended Characteristics use based on the same technology and principles of operation. The subject devices have been modified from the predicate devices as listed below. The results of design verification demonstrate that these changes are substantially equivalent to the predicate devices. All other aspects of the subject device are identical to those of the predicate device.
· Change to lubricants used in the assembly on the device
| Attribute | SUBJECT DEVICE | PREDICATE DEVICE | | |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | An intravascular catheter is a device<br>that is inserted into the patient's<br>vascular system for short term use<br>(less than 30 days) to sample blood,<br>monitor blood pressure, or administer<br>fluids intravenously. These catheters<br>may be used for any patient population<br>with consideration given to adequacy<br>of vascular anatomy and<br>appropriateness of procedure. | An intravascular catheter is a device<br>that is inserted into the patient's<br>vascular system for short term use<br>(less than 30 days) to sample blood,<br>monitor blood pressure, or administer<br>fluids intravenously. These catheters<br>may be used for any patient population<br>with consideration given to adequacy<br>of vascular anatomy and<br>appropriateness of procedure. | | |
| Design | A fluorinated ethylene propylene (FEP)<br>or polyurethane peripheral IV catheter | A fluorinated ethylene propylene (FEP)<br>or polyurethane peripheral IV catheter | | |
| | Component | Subject Device | Component | Predicate Device |
| Materials | Catheter Tubing | Angiocath™ IV Catheter<br>Fluorinated ethylene<br>propylene (FEP)<br><br>Insyte™ IV Catheter<br>BD Vialon™ polyurethane | Catheter Tubing | Angiocath™ IV Catheter<br>Fluorinated ethylene<br>propylene (FEP)<br><br>Insyte™ IV Catheter<br>BD Vialon™ polyurethane |
| | Catheter Hub | Polypropylene | Catheter Hub | Polypropylene |
| | Metal Wedge | Stainless Steel | Metal Wedge | Stainless Steel |
| | Needle | Stainless Steel | Needle | Stainless Steel |
| | Needle Hub | Angiocath™ IV Catheter<br>Propionate<br><br>Insyte™ IV Catheter<br>Propionate | Needle Hub | Angiocath™ IV Catheter<br>Propionate<br><br>Insyte™ IV Catheter<br>Propionate |
| | Needle Cover | Polypropylene | Needle Cover | Polypropylene |
{5}------------------------------------------------
| Materials<br>(cont.) | Vent Plug | Angiocath™ IV Catheter<br>Polyethylene | Angiocath™ IV Catheter<br>Polyethylene | |
|--------------------------------------------|---------------------------------------------------------------------------------------|-----------------------------------------------------|---------------------------------------------------------------------------------------|--|
| | | Insyte™ IV Catheter<br>Polypropylene with filter | Insyte™ IV Catheter<br>Polypropylene with filter | |
| | Catheter Tipping<br>Lubricant | Polydimethylsiloxane-based<br>Lubricant | Polydimethylsiloxane-based<br>Lubricant | |
| | Catheter Lubricant | Polydimethylsiloxane-based<br>Lubricant | Polydimethylsiloxane-based<br>Lubricant | |
| | Wedge Lubricant | Polydimethylsiloxane-based<br>Lubricant | Polydimethylsiloxane-based<br>Lubricant | |
| | Needle (Cannula)<br>Lubricant | Two-Part<br>Polydimethylsiloxane-based<br>Lubricant | Single-Part<br>Polydimethylsiloxane-based<br>Lubricant | |
| Physical /<br>Mechanical<br>Specifications | SUBJECT DEVICES | | PREDICATE DEVICES | |
| | Angiocath™ IV Catheter | | Angiocath™ IV Catheter | |
| | Catheter Diameters:<br>14G, 16G, 18G, 20G, 22G, 24G | | Catheter Diameters:<br>14G, 16G, 18G, 20G, 22G, 24G | |
| | Catheter Lengths:<br>0.75", 1.00", 1.16", 1.88"<br>Non-winged only | | Catheter Lengths:<br>0.75", 1.00", 1.16", 1.88"<br>Non-winged only | |
| | Insyte™ IV Catheter | | Insyte™ IV Catheter | |
| | Catheter Diameters:<br>14G, 16G, 18G, 20G, 22G, 24G | | Catheter Diameters:<br>14G, 16 G, 18G, 20G, 22G, 24G | |
| | Catheter Lengths:<br>0.56", 0.75", 1.00", 1.16", 1.77", 1.88"<br>Winged or Non-winged | | Catheter Lengths:<br>0.56", 0.75", 1.00", 1.16", 1.77", 1.88"<br>Winged or Non-winged | |
#### Summary of Performance Tests
Pursuant to 21 CFR 820.30, Design Controls, design verification tests were performed based on the risk analysis, and the results of these tests demonstrate that the BD Angiocath™ and Insyte™IV Catheters are substantially equivalent to the predicate devices. Design verification testing included the following:
| Characteristic / Test Performed | Test Method | Results | |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|------|
| Biocompatibility | Per ISO 10993-1 | Pass | |
| Liquid leakage | Per ISO 594-1, -2 | Pass | |
| Air leakage | Per ISO 594-1, -2 | Pass | |
| Separation force | Per ISO 594-1, -2 | Pass | |
| Stress cracking | Per ISO 594-1, -2 | Pass | |
| Unscrewing torque | Per ISO 594-2 | Pass | |
| Ease of assembly | Per ISO 594-2 | Pass | |
| Resistance to overriding | Per ISO 594-2 | Pass | |
| General | Per ISO 10555-1 | Pass | |
| Radio-detectability | Per ISO 10555-1 | Pass | |
| Biocompatiblity | Per ISO 10555-1 | Pass | |
| Surface | Per ISO 10555-1 | Pass | |
| Corrosion resistance | Per ISO 10555-1 | Pass | |
| Peak tensile force | Per ISO 10555-1 | Pass | |
| Freedom from leakage | Per ISO 10555-1 | Pass | |
| Hubs | Per ISO 10555-1 | Pass | |
| Flowrate | Per ISO 10555-1 | Pass | |
| Power injection | Per ISO 10555-1 | Pass | |
| Side holes | Per ISO 10555-1 | NA | |
| Distal tip | Per ISO 10555-1 | Pass | |
| Color code | Per ISO 10555-5 | Pass | |
| Catheter unit | Per ISO 10555-5 | Pass | |
| Needle point | Per ISO 10555-5 | Pass | |
| Needle hub | Per ISO 10555-5 | Pass | |
| Strength of union between needle hub and needle tube | Per ISO 10555-5 | Pass | |
| Vent fitting | Per ISO 10555-5 | Pass | |
| Needle (Cannula) penetration force | Internal design input, identical to predicate | Pass | |
| Catheter penetration force | Internal design input, identical to predicate | Pass | |
| Catheter average drag force | Internal design input, identical to predicate | Pass | |
| | Minimal catheter and cannula tip adhesion | Internal design input, identical to predicate | Pass |
| Summary of Substantial Equivalence | Based on the indications for use, technological characteristics, and performance testing, the subject BD AngiocathTM and InsyteTM IV Catheters meet all predetermined requirements in accordance with 21 CFR 820.30, Design Controls, and demonstrates that the subject devices are substantially equivalent to the predicate devices. | | |
{6}------------------------------------------------
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.