← Product Code [NHJ](/productcode/NHJ) · K151689

# MetaNeb 4 System (K151689)

_Hill-Rom Services Pte, Ltd. · NHJ · Mar 17, 2016 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151689

## Device Facts

- **Applicant:** Hill-Rom Services Pte, Ltd.
- **Product Code:** [NHJ](/productcode/NHJ.md)
- **Decision Date:** Mar 17, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The MetaNeb® 4 System is indicated for mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and also has the ability to provide supplemental oxygen when used with compressed oxygen.

## Device Story

MetaNeb® 4 System is a pneumatically operated therapeutic device for mucus clearance and lung expansion. Inputs: 50 psi air or oxygen source (wall, cylinder, or portable compressor). Operation: Controller manages three modes—Aerosol (delivery only), CHFO (Continuous High Frequency Oscillation; pneumatic chest physiotherapy with medicated aerosol), and CPEP (Continuous Positive Expiratory Pressure; positive pressure to expand airways). Output: Medicated aerosol and pressure pulses delivered via a single-patient-use circuit (handset, mouthpiece, nebulizer). Used in hospitals, sub-acute/nursing facilities, clinics, and home settings by clinicians or patients. Healthcare providers use the manometer to monitor pressure and adjust settings to mobilize secretions and prevent atelectasis, potentially improving patient respiratory function.

## Clinical Evidence

Bench testing only. Testing included compatibility of portable air compressors, performance comparison of air sources (compressor vs. wall source), material certification, PM2.5 testing, and a usability study. The usability study involved caregivers and lay users, with 100% of participants successfully completing critical setup and treatment tasks without errors leading to unacceptable risk.

## Technological Characteristics

Pneumatically operated system; fixed venturi patient circuit. Modes: CHFO, CPEP, Aerosol. Power: 50 psi air/oxygen source (wall, cylinder, or portable compressor). Components: Controller with manometer, single-patient-use circuit (mouthpiece, handset, nebulizer). Materials: Identical to K124032; external communicating and mucosal contact. Connectivity: None (standalone).

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- MetaNeb® ([K124032](/device/K124032.md))
- PowerNeb® ([K051964](/device/K051964.md))
- Bird IPV ([K895485](/device/K895485.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of three human profiles facing to the right, arranged in a way that they appear to be interconnected. The profiles are black and are set against a white background. Encircling the profiles is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 17, 2016

Hill-Rom Services Pte Ltd c/o Paul Dryden Consultant for Hill-Rom 1 Yishun Ave 7 Singapore 768923

> K151689 Trade/Device Name: MetaNeb® 4 System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: NHJ Dated: February 12, 2016 Received: February 16, 2016

Dear Mr. Dryden,

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Expiration Date: January 31, 2017

See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

#### K151689

Device Name

### MetaNeb® 4 System

Indications for Use (Describe)

The MetaNeb® 4 System is indicated for mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and also has the ability to provide supplemental oxygen when used with compressed oxygen.

Patient Population - 5 years old and above whom can follow verbal instructions.

Environment of Use - Hospitals, sub-acute and nursing facilities, physician offices, clinics, and home settings

Type of Use (Select one or both, as applicable)

| XX Prescription Use (Part 21 CFR 801 Subpart D) |  |
|-------------------------------------------------|--|
| Over-The-Counter Use (21 CFR 801 Subpart C)     |  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary Page 1 of 9

| Date Prepared              | 15-Mar-16                                                                                                     |
|----------------------------|---------------------------------------------------------------------------------------------------------------|
|                            |                                                                                                               |
|                            | Hill-Rom Services Pte Ltd                                                                                     |
|                            | 1 Yishun Ave 7                                                                                                |
|                            | Singapore, 768923, Singapore                                                                                  |
|                            | Tel - 011 - 65 65945215                                                                                       |
|                            | Fax - 011 - 65 65945201                                                                                       |
| Official Contact:          | Huifang Zhu - RA / QA Specialist                                                                              |
| Proprietary or Trade Name: | MetaNeb® 4 System                                                                                             |
| Common/Usual Name:         | Noncontinuous ventilator (IPPB)                                                                               |
| Classification Code/Name:  | NHJ - non-continuous ventilator (IPPB)<br>21 CFR 868.5905<br>Class II                                         |
| Predicate Devices:         | K124032 – Hill-Rom - MetaNeb®<br>K051964 - CoMedica - PowerNeb®<br>K895485 - Percussionaire Corp (Bird) - IPV |

### Device Description:

The MetaNeb® 4 System is a therapeutic device that uses a systematic approach to enhance normal mucus clearance and resolve or prevent patchy atelectasis.

The system has three modes:

- Aerosol for the delivery of aerosol only. In this mode CHFO and CPEP are not available. .
- . CHFO (Continuous High Frequency Oscillation) - a pneumatic form of chest physiotherapy that delivers medicated aerosol while oscillating the airways with continuous pulses of positive pressure.
- . CPEP (Continuous Positive Expiratory Pressure) - supplies medicated aerosol combined with continuous positive pressure to help hold open and expand the airways.

There are three major components to The MetaNeb® 4 System:

- Circuit includes a mouthpiece, handset, nebulizer, tubing, and filter/tri-connector. It is a ● single patient use, assembly that is intended for multiple treatment sessions.
- . Controller - contains the controls to select the three different modes. It also has a manometer to monitor pressure. Power is supplied by a hose connected to a 50 psi air or oxygen source, e.g., wall source, cylinder or portable medical air compressor.
- . Stand - to move the MetaNeb® 4 System from room to room.

The MetaNeb® 4 System incorporates 2 changes vs. the predicate MetaNeb®, K124032. The changes are listed in Table 1.

{4}------------------------------------------------

#### 510(k) Summary Page 2 of 9 15-Mar-16

|                       | Predicate<br>MetaNeb® - K124031                                      | Difference / Modification for<br>Proposed Device                       |
|-----------------------|----------------------------------------------------------------------|------------------------------------------------------------------------|
| Environment<br>of use | Hospitals<br>sub-acute facilities<br>physician offices<br>clinics    | Expanding to include<br>nursing facility and home setting              |
| Power<br>sources      | Air source from<br>facility supply, e.g., cylinder<br>or wall source | Adding portable medical air compressor as an<br>additional air source. |

## Table 1 - Differences / Modifications between proposed MetaNeb® 4 System and Predicate MetaNeb® (K124032)

### Indications for Use:

The MetaNeb® 4 System is indicated for mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and also has the ability to provide supplemental oxygen when used with compressed oxygen.

Patient Population - 5 years old and above whom can follow verbal instructions

Environment of Use - Hospitals, sub-acute and nursing facilities, physician offices, clinics, and home settings

### Predicate Device Comparison:

The MetaNeb® 4 System is viewed as substantially equivalent to the predicate devices because:

### Indications -

The proposed indications for use for mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and also has the ability to provide supplemental oxygen when used with compressed oxygen are identical to the predicates.

Discussion: The indications for use are identical to the predicate Hill-Rom MetaNeb® system (K124032).

### Patient Population -

The patient population is similar to the predicate MetaNeb®, K124032 but the lower age has changed to 5 years and above who can follow verbal instructions.

Discussion: The difference in patient population relates to the ability to use a mouthpiece in a home setting.

#### Environment of Use -

The addition of the environments of use to include nursing care facilities and home care settings is identical to the predicate Bird IPV (K895485).

{5}------------------------------------------------

#### 510(k) Summary Page 3 of 9 15-Mar-16

Discussion: The expansion in the intended environments of use is substantially equivalent to the predicate Bird IPV (K895485). We have evaluated the risks associated with home use and performed a usability study that supports that the device may be used within this environment.

### Technology -

The design of the MetaNeb® 4 System as a pneumatically operated system with different modes of operation: CPEP, CHFO, and aerosol only are identical to the predicate in design and basic function and performance.

Discussion: The design of MetaNeb® is identical to the predicate MetaNeb® (K124032).

#### Performance -

The basic performance features and parameters are identical to the predicate MetaNeb® (K124032). Discussion: The proposed changes do not alter the performance specifications of the proposed MetaNeb® 4 System and the predicate MetaNeb® (K124032).

### Non-clinical Comparative Performance

#### Materials -

The materials in the gas and fluid pathway are considered as having 2 types of patient contact:

- External communicating, tissue contacting, permanent duration and ●
- Surface contact, mucosal contact, permanent duration for the mouthpiece ●

The materials are unchanged and identical to the predicate submission MetaNeb®, K124032. Discussion: All associated materials in the gas or fluid pathway are identical to the predicate Hill-Rom MetaNeb® (K124032). We did perform additional PM23 testing even though the controller is identical to the predicate MetaNeb® (K124032).

### Bench Testing -

We performed a series of non-clinical bench tests to demonstrate that the modified MetaNeb® 4 System is equivalent to the predicates. These tests included:

- Compatibility testing of portable air compressors to performance specifications
- Performance comparison of air source, portable compressor vs. wall source ●
- Biocompatibility Controller ●
	- o Material Certification
	- 0 PM2.5 testing for the controller reconfirmation only
- . Usability - home environment

The comparative testing demonstrates that the proposed modifications are substantially equivalent to the predicate devices.

{6}------------------------------------------------

# 510(k) Summary
 15-Mar-16

## Table 2 – Substantial Equivalence Comparative Table

|                                                    | Proposed<br>MetaNeb® 4 System                                                                                                                                                                                                                                          | MetaNeb®<br>Hill-Rom - K124032                                                                                                                                                                                                                                       | PowerNeb<br>Comedica - K051964                                                                                                                                                                                                                         | IPV<br>Bird - K895485                                                                                                                                                                                                                             |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                       | Indicated for mobilization of<br>secretions<br><br>lung expansion therapy<br><br>treatment and prevention of<br>pulmonary atelectasis<br><br>ability to provide<br>supplemental oxygen when<br>used with compressed<br>oxygen.                                         | Indicated for mobilization<br>of secretions<br><br>lung expansion therapy<br><br>treatment and prevention of<br>pulmonary atelectasis<br><br>ability to provide<br>supplemental oxygen when<br>used with compressed<br>oxygen.                                       | Indicated for mobilization<br>of secretions<br><br>lung expansion therapy<br><br>treatment and prevention of<br>pulmonary atelectasis<br><br>ability to provide<br>supplemental oxygen when<br>used with compressed<br>oxygen.                         | Indicated for the<br>mobilization and raising of<br>endobronchial secretions,<br>bronchodilation, reducing<br>mucosal edema, resolution<br>of diffuse patchy atelectasis<br>in all patient<br>populations                                         |
| Indications for Use                                | The MetaNeb® 4 System is<br>indicated for mobilization of<br>secretions, lung expansion<br>therapy, the treatment and<br>prevention of pulmonary<br>atelectasis, and also has the<br>ability to provide<br>supplemental oxygen when<br>used with compressed<br>oxygen. | The MetaNeb® System is<br>indicated for mobilization of<br>secretions, lung expansion<br>therapy, the treatment and<br>prevention of pulmonary<br>atelectasis, and also has the<br>ability to provide<br>supplemental oxygen when<br>used with compressed<br>oxygen. | PowerNeb® is indicated for<br>mobilization of secretions,<br>lung expansion therapy, the<br>treatment and prevention of<br>pulmonary atelectasis, and<br>also has the ability to<br>provide supplemental<br>oxygen when used with<br>compressed oxygen | Use of Percussaire® IPV®<br>is indicated for the<br>mobilization and raising of<br>endobronchial secretions,<br>bronchodilation, reducing<br>mucosal edema, and the<br>resolution of diffuse patchy<br>atelectasis in all patient<br>populations. |
|                                                    |                                                                                                                                                                                                                                                                        | 510(k) Summary                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                   |
|                                                    |                                                                                                                                                                                                                                                                        | Page 5 of 9                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                   |
|                                                    |                                                                                                                                                                                                                                                                        | 15-Mar-16                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                   |
|                                                    | Proposed                                                                                                                                                                                                                                                               | MetaNeb®                                                                                                                                                                                                                                                             | PowerNeb                                                                                                                                                                                                                                               | IPV                                                                                                                                                                                                                                               |
|                                                    | MetaNeb® 4 System                                                                                                                                                                                                                                                      | Hill-Rom - K124032                                                                                                                                                                                                                                                   | Comedica - K051964                                                                                                                                                                                                                                     | Bird - K895485                                                                                                                                                                                                                                    |
| Environments of Use                                | Hospital, sub-acute facilities                                                                                                                                                                                                                                         | Hospital, sub-acute facilities                                                                                                                                                                                                                                       | Hospital, sub-acute facilities                                                                                                                                                                                                                         | Hospital, sub-acute facilities                                                                                                                                                                                                                    |
|                                                    | Nursing care, Homecare                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                        | Homecare                                                                                                                                                                                                                                          |
| Patient Population                                 |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                      | All patients that require                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                   |
|                                                    | 5 years old and above whom                                                                                                                                                                                                                                             | Adult                                                                                                                                                                                                                                                                | therapy mobilization of                                                                                                                                                                                                                                | All patient populations                                                                                                                                                                                                                           |
|                                                    | can follow verbal instructions                                                                                                                                                                                                                                         | Children > 2 years old                                                                                                                                                                                                                                               | secretions, lung expansion                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                   |
|                                                    |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                      | therapy, the treatment and                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                   |
|                                                    |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                      | prevention of pulmonary                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                   |
|                                                    |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                      | atelectasis.                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                   |
| Contraindications                                  | -Untreated tension                                                                                                                                                                                                                                                     | -Untreated tension                                                                                                                                                                                                                                                   | -Untreated tension                                                                                                                                                                                                                                     | -Untreated tension                                                                                                                                                                                                                                |
|                                                    | pneumothorax                                                                                                                                                                                                                                                           | pneumothorax                                                                                                                                                                                                                                                         | pneumothorax                                                                                                                                                                                                                                           | pneumothorax                                                                                                                                                                                                                                      |
|                                                    | -Untrained or unskilled                                                                                                                                                                                                                                                | -Untrained or unskilled                                                                                                                                                                                                                                              | -Untrained or unskilled                                                                                                                                                                                                                                | -Lack of adequate, skilled                                                                                                                                                                                                                        |
|                                                    | operator                                                                                                                                                                                                                                                               | operator                                                                                                                                                                                                                                                             | operator                                                                                                                                                                                                                                               | supervision                                                                                                                                                                                                                                       |
|                                                    | -History of pneumothorax                                                                                                                                                                                                                                               | -History of pneumothorax                                                                                                                                                                                                                                             | -History of pneumothorax                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                   |
|                                                    | -Pulmonary air leak                                                                                                                                                                                                                                                    | -Pulmonary air leak                                                                                                                                                                                                                                                  | -Pulmonary air leak                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                   |
|                                                    | -Recent pneumonectomy                                                                                                                                                                                                                                                  | -Recent pneumonectomy                                                                                                                                                                                                                                                | -Recent pneumonectomy                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                   |
|                                                    | -Pulmonary hemorrhage                                                                                                                                                                                                                                                  | -Pulmonary hemorrhage                                                                                                                                                                                                                                                | -Pulmonary hemorrhage                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                   |
|                                                    | -Myocardial infarction<br>-Vomiting                                                                                                                                                                                                                                    | -Myocardial infarction<br>-Vomiting                                                                                                                                                                                                                                  | -Myocardial infarction<br>-Vomiting                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                   |
| Mode of Operation                                  | Pneumatic system using a                                                                                                                                                                                                                                               | Pneumatic system using a                                                                                                                                                                                                                                             | Pneumatic system using a                                                                                                                                                                                                                               | Pneumatic system using a                                                                                                                                                                                                                          |
|                                                    | fixed venturi patient circuit                                                                                                                                                                                                                                          | fixed venturi patient circuit                                                                                                                                                                                                                                        | fixed venturi patient circuit                                                                                                                                                                                                                          | variable (sliding) venturi                                                                                                                                                                                                                        |
|                                                    | combined with medicated                                                                                                                                                                                                                                                | combined with medicated                                                                                                                                                                                                                                              | combined with medicated                                                                                                                                                                                                                                | patient circuit combined with                                                                                                                                                                                                                     |
|                                                    | aerosol to deliver continuous                                                                                                                                                                                                                                          | aerosol to deliver continuous                                                                                                                                                                                                                                        | aerosol to deliver                                                                                                                                                                                                                                     | medicated aerosol to deliver                                                                                                                                                                                                                      |
|                                                    | high frequency oscillation                                                                                                                                                                                                                                             | high frequency oscillation                                                                                                                                                                                                                                           | continuous high frequency                                                                                                                                                                                                                              | continuous high frequency                                                                                                                                                                                                                         |
|                                                    | therapy at 2 pre-set<br>frequencies and amplitude.                                                                                                                                                                                                                     | therapy at 2 pre-set<br>frequencies and amplitude.                                                                                                                                                                                                                   | oscillation therapy at 1 pre-<br>set frequency and amplitude                                                                                                                                                                                           | oscillation therapy at multiple<br>user settable frequencies and<br>amplitudes.                                                                                                                                                                   |
|                                                    | Can deliver continuous<br>expiratory pressure (CPEP)<br>combined with medicated<br>aerosol.                                                                                                                                                                            | Can deliver continuous<br>expiratory pressure (CPEP)<br>combined with medicated<br>aerosol.                                                                                                                                                                          | Can deliver continuous<br>expiratory pressure (CPEP)<br>combined with medicated<br>aerosol                                                                                                                                                             | Equivalent CPEP function not<br>available.                                                                                                                                                                                                        |
| Page 6 of 9<br>15-Mar-16                           |                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                   |
|                                                    | Proposed<br>MetaNeb® 4 System                                                                                                                                                                                                                                          | MetaNeb®<br>Hill-Rom - K124032                                                                                                                                                                                                                                       | PowerNeb<br>Comedica - K051964                                                                                                                                                                                                                         | IPV<br>Bird - K895485                                                                                                                                                                                                                             |
| Modes                                              | CPEP<br>CHFO<br>Aerosol Only                                                                                                                                                                                                                                           | CPEP<br>CHFO<br>Aerosol Only                                                                                                                                                                                                                                         | CPEP<br>CHFO                                                                                                                                                                                                                                           | CHFO                                                                                                                                                                                                                                              |
| CPEP<br>Continuous Positive<br>Expiratory Pressure | Controlled static flow with<br>positive pressures<br>$\leq$ 30 cm H2O                                                                                                                                                                                                  | Controlled static flow with<br>positive pressures<br>$\leq$ 30 cm H2O                                                                                                                                                                                                | Controlled static flow with<br>positive pressures<br>$\leq$ 20 cm H2O                                                                                                                                                                                  | No static flow mode1,2                                                                                                                                                                                                                            |
| CHFO<br>Continuous High<br>Frequency Oscillations  | Controlled continuous flow<br>with frequencies up to 300<br>beats per minute and peak<br>positive pressures<br>$\leq$ 30 cmH2O                                                                                                                                         | Controlled continuous flow<br>with frequencies up to 300<br>beats per minute and peak<br>positive pressures<br>$\leq$ 30 cmH2O                                                                                                                                       | Controlled continuous flow<br>with frequencies up to 300<br>beats per minute and peak<br>positive pressures<br>$\leq$ 30 cmH2O                                                                                                                         | Controlled Intrapulmonary<br>Percussive Ventilation (IPV)<br>frequencies up to 300 beats per<br>minute and peak positive<br>pressures $\leq$ 80 cmH2O                                                                                             |
| Aerosol Only                                       | Controlled continuous<br>constant flow to in-line<br>nebulizer delivering<br>medicated aerosol only.                                                                                                                                                                   | Controlled continuous<br>constant flow to in-line<br>nebulizer delivering<br>medicated aerosol only.                                                                                                                                                                 | Equivalent Aerosol Only<br>function by setting CPEP to<br>minimum flow with in-line<br>nebulizer                                                                                                                                                       | Can use in-line nebulizer                                                                                                                                                                                                                         |
| Patient Circuit                                    | Disposable circuit referred to<br>as "handset" includes<br>connection for in-line<br>nebulizer                                                                                                                                                                         | Disposable circuit referred<br>to as "hand-set" includes<br>connection for in-line<br>nebulizer                                                                                                                                                                      | Disposable circuit referred<br>to as "hand-set" includes<br>connection for in-line<br>nebulizer                                                                                                                                                        | Reusable circuit includes<br>connection for in-line nebulizer                                                                                                                                                                                     |
|                                                    | Draw in room air mix with<br>medicated aerosol and gas<br>from controller                                                                                                                                                                                              | Draw in room air mix with<br>medicated aerosol and gas<br>from controller                                                                                                                                                                                            | Draw in room air mix with<br>medicated aerosol and gas<br>from controller                                                                                                                                                                              | Draw in room air mix with<br>medicated aerosol and gas from<br>controller                                                                                                                                                                         |
|                                                    | No in-line filter (Home)<br>In-line filter (Acute care)                                                                                                                                                                                                                | In-line filter (Acute care)                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                   |
| Patient Circuit settings                           | Expiratory resistance<br>adjustment $\lt$ 30 cm H2O                                                                                                                                                                                                                    | Expiratory resistance<br>adjustment $\lt$ 30 cm H2O                                                                                                                                                                                                                  | Expiratory resistance<br>adjustment $\lt$ 30 cm H2O                                                                                                                                                                                                    | Not applicable but the IPV can<br>provide pressures $\lt$ 80 cmH2O                                                                                                                                                                                |

{7}------------------------------------------------

{8}------------------------------------------------

510(k) Summary

{9}------------------------------------------------

| 510(k) Summary        |                                |                                |                                |                                |  |  |  |
|-----------------------|--------------------------------|--------------------------------|--------------------------------|--------------------------------|--|--|--|
| Page 7 of 9           |                                |                                |                                |                                |  |  |  |
| 15-Mar-16             |                                |                                |                                |                                |  |  |  |
|                       | Proposed                       | MetaNeb®                       | PowerNeb                       | IPV                            |  |  |  |
|                       | MetaNeb® 4 System              | Hill-Rom - K124032             | Comedica - K051964             | Bird - K895485                 |  |  |  |
| Patient Interface     | Mouthpiece                     | Mouthpiece                     | Mouthpiece                     | Mouthpiece                     |  |  |  |
|                       |                                | Face mask                      | Face mask                      |                                |  |  |  |
|                       | Insert into ventilator circuit | Insert into ventilator circuit |                                | Insert into ventilator circuit |  |  |  |
| Controller            | Pneumatic and air or oxygen    | Pneumatic and air or oxygen    | Pneumatic and air or oxygen    | Pneumatic and air or oxygen    |  |  |  |
| Controller settings   | On/off                         | On/off                         | On/off                         | On/off                         |  |  |  |
|                       | Mode selection                 | Mode selection                 | Mode selection                 |                                |  |  |  |
|                       |                                |                                |                                |                                |  |  |  |
|                       | Frequency selection for        | Frequency selection for CHFO   |                                | Pressure and frequency         |  |  |  |
|                       | CHFO mode                      | mode                           |                                | adjustment for CHFO mode       |  |  |  |
|                       |                                |                                |                                |                                |  |  |  |
|                       | Pressure adjustment for        | Pressure adjustment for CPEP   | Pressure adjustment for        |                                |  |  |  |
|                       | CPEP mode                      | mode                           | CPEP mode                      |                                |  |  |  |
|                       |                                |                                |                                |                                |  |  |  |
|                       | Pressure manometer             | Pressure manometer             |                                | Pressure manometer             |  |  |  |
|                       |                                |                                |                                |                                |  |  |  |
|                       | Connection for patient circuit | Connection for patient circuit | Connection for patient circuit | Connection for patient circuit |  |  |  |
| Ventilator connection | Placed in-line in the          | Placed in-line in the          | Not available                  | Placed in-line in the          |  |  |  |
|                       | inspiratory limb of the        | inspiratory limb of the        |                                | ventilator circuit with a  IPV |  |  |  |
|                       | ventilator circuit with a      | ventilator circuit with a      |                                | cone adapter                   |  |  |  |
|                       | standard "T" adapter.          | standard "T" adapter.          |                                |                                |  |  |  |
|                       |                                |                                |                                |                                |  |  |  |
|                       | Only for acute care            | Only for acute care            |                                |                                |  |  |  |
|                       | environment                    | environment                    |                                |                                |  |  |  |

{10}------------------------------------------------

## 510(k) Summary Page 8 of 9 15-Mar-16

Table 3 compares the performance specifications of the proposed MetaNeb® 4 System to the predicates for Peak Pressure under various therapy mode conditions. We should note that the test set-up parameters are identical to those used when comparing the MetaNeb® in submission K124032.

The results demonstrate substantial equivalence of the maximum peak pressure of the MetaNeb® 4 System as compared to the identified predicates.

| Therapy Mode                                           | Proposed<br>MetaNeb® 4<br>System with<br>Standard Patient<br>Circuit | MetaNeb®<br>Hill-Rom –<br>K124032<br>With original<br>Patient Circuit | PowerNeb<br>Comedica -<br>K051964 | IPV<br>Bird - K8954851,2             |
|--------------------------------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------|-----------------------------------|--------------------------------------|
| CPEP Mode @<br>High Flow<br>Peak Pressure<br>(cmH2O)   | 10.7 to 29.9                                                         | 10.7 to 29.9                                                          | 18.9 to 35.4                      | This mode is not<br>available on IPV |
| CPEP Mode @<br>Medium Flow<br>Peak Pressure<br>(cmH2O) | 7.7 to 14.6                                                          | 7.7 to 14.6                                                           | 9.4 to 18.3                       |                                      |
| CPEP Mode @<br>Low Flow<br>Peak Pressure<br>(cmH2O)    | 0.2 to 2.0                                                           | 0.2 to 2.0                                                            | 0.2 to 1.2                        |                                      |
| CHFO Mode @<br>High Flow<br>Peak Pressure<br>(cmH2O)   | 11.2 to 25.4                                                         | 11.2 to 25.4                                                          |                                   | 10.1 to 26.7                         |
| CHFO Mode @<br>Low Flow<br>Peak Pressure<br>(cmH2O)    | 8.7 to 20.3<br>From K124032<br>data                                  | 8.7 to 20.3<br>From K124032<br>data                                   |                                   | 11 to 23                             |

## Table 3 - Substantial Equivalence Comparative Performance - Expiratory Pressures

1 Note: When IPV is tested at comparable setting the differences in peak pressure is < 5 cmH2O.

2 Note: IPV is capable of a maximum peak pressure of 80 cmHzO

{11}------------------------------------------------

#### 510(k) Summary Page 9 of 9 15-Mar-16

## Usability -

We performed a usability study with 2 user groups: caregivers and lay users. The study evaluated the ability of each group to perform critical tasks related to setting the MetaNeb® 4 System up and performing simulated treatments, 100% of the participants completed all the tasks successfully without use errors that could have lead to unacceptable risk of harm.

## Substantial Equivalence Conclusion -

The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.

---

**Source:** [https://fda-staging.innolitics.com/device/K151689](https://fda-staging.innolitics.com/device/K151689)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).