The SleepWeaver Advance Pediatric Nasal Mask is intended to provide an interface for noninvasive Positive Airway Pressure (PAP) ventilation therapy. This mask is intended for single-patient use in the home and single-patient use in the hospital / institutional environment. This mask is to be used on patients between 2 and 7 years of age.
Device Story
SleepWeaver Advance Pediatric Nasal CPAP Mask provides interface for noninvasive PAP ventilation therapy. Device consists of cloth cushion, adjustable elastic headgear, and 360-degree rotating 22-mm tubing swivel connector. Used in home or hospital/institutional environments by pediatric patients (ages 2-7). Air supplied by external therapy device; patient inhales air from mask; exhales into mask; continuous airflow from therapy device purges exhaled CO2 through mask exhalation holes. Cloth cushion contacts patient face; headgear secures mask. Optional Feather Weight tube extension available. Healthcare provider prescribes therapy; mask facilitates delivery of prescribed pressure (4-20 cm H2O) to maintain open airway. Benefits include noninvasive airway support for pediatric patients.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing per ISO 17510-2, including CO2 rebreathing, flow resistance, leak rate, and physical dead space measurements. Biocompatibility testing performed per ISO 10993-1, 10993-5, and 10993-10.
Technological Characteristics
Materials: Blue polyester fabric, nylon, elastane, polyurethane, polycarbonate swivel. Sensing/Actuation: Passive interface for PAP therapy. Dimensions: 1 size. Connectivity: Standard 22mm tubing interface. Standards: ISO 14971, ISO 10993-1/5/10, ISO 5356-1, ISO 5367, ISO 17510-2, IEC 62366.
Indications for Use
Indicated for pediatric patients aged 2 to 7 years requiring noninvasive Positive Airway Pressure (PAP) ventilation therapy. No contraindications reported.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 5, 2016
Circadiance, LLC % Pooja Roychoudhury Operation Manager Regulatory & Quality Solutions, LLC 2790 Mosside Blvd, Suite 800 Monroeville, Pennsylvania 15146
Re: K151683
Trade/Device Name: SleepWeaver Advance Pediatric Nasal CPAP Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: March 1, 2016 Received: March 3, 2016
Dear Ms. Roychoudhury:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K151683
#### Device Name
SleepWeaver Advance Pediatric Nasal CPAP Mask
#### Indications for Use (Describe)
The SleepWeaver Advance Pediatric Nasal Mask is intended to provide an interface for noninvasive Positive Airway Pressure (PAP) ventilation therapy. This mask is intended for single-patient use in the home and single-patient use in the hospital / institutional environment. This mask is to be used on patients between 2 and 7 years of age.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
Circadiance, LLC 1300 Rodi Road, Turtle Creek, PA 15145 Phone: (724) 858-2837 Facsimile: (412) 202-4583 Contact Person: David Groll, CEO, David.Groll@Circadiance.com
Date Originally Prepared: June 19, 2015 Date Revised: April 5, 2016
#### Name of Device and Name/Address of Sponsor
SleepWeaver Advance Pediatric Nasal CPAP Mask
## Common or Usual Name
Vented Nasal Mask
## Classification Name
Accessory to Non-Continuous Ventilator, Class II, 21 CFR 868.5905, Product Code: BZD
#### Predicate Devices
ResMed Ltd., Pixi Pediatric Mask (K102224)
#### Indications for Use
The SleepWeaver Advance Pediatric Nasal Mask is intended to provide an interface for noninvasive Positive Airway Pressure (PAP) ventilation therapy. This mask is intended for single-patient use in the home and single-patient use in the hospital / institutional environment. This mask is to be used on patients between 2 and 7 years of age.
#### Description of Device
The SleepWeaver Advance Pediatric Nasal CPAP Mask serves as a mechanism for connecting a pediatric patient between 2 and 7 years of age for whom PAP therapy has been prescribed to maintain an open airway. The nasal mask is placed over a patient's nose and held in place by use of an adjustable elastic headgear. A cloth cushion contacts the patient's face. The mask assembly has a tubing swivel connector which is compatible with the industry standard 22 mm air tubing.
Air is supplied to the mask by a therapy device. The patient inhales air from the mask and exhales into the mask where continuous airflow from the therapy device purges the exhaled carbon dioxide from the mask through the mask exhalation holes.
{5}------------------------------------------------
The SleepWeaver Advance Pediatric Nasal CPAP Mask can be used in the home environment (single-patient use) or the hospital/institutional environment (single-patient use).
The mask features a cloth cushion made of a blue polyester fabric with a polyester / nylon / lycra / polyurethane interface material. The mask exhalation feature is incorporated into the cloth cushion. Attached to the cushion is a 22-mm tubing swivel connector that can rotate 360°. The tubing connector can be removed by unwrapping the white medical tape for cleaning and cushion replacement. The cloth headgear is connected to the mask through slots in the cushion wings. The headgear is removable for cleaning and replacement.
The SleepWeaver Advance Pediatric Nasal CPAP Mask product also offers an optional accessory: the Feather Weight tube. The optional SleepWeaver FeatherWeight tube may be used with this mask, but it remains outside the scope of this submission as it has been previously cleared by its inclusion in K120757. The Feather Weight tube is an 18" plastic extension tube compliant to ISO 5356-1 and ISO 5367 that connects between the SleepWeaver Advance Pediatric Nasal CPAP Mask and standard CPAP tubing.
# Technological Characteristics
The SleepWeaver Advance Pediatric Nasal CPAP Mask consists of a cloth cushion, headgear, and a swivel connector. Table 1 is provided below which lists additional technological characteristics for the SleepWeaver Advance Pediatric Nasal CPAP Mask as well as a comparison of the described feature to the predicate device.
| Feature | Proposed Device:<br>SleepWeaver Advance<br>Pediatric Nasal CPAP Mask | Predicate Device: ResMed Pixi<br>Pediatric Mask (K102224) | Equivalence Narrative |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use | The SleepWeaver Advance<br>Pediatric Nasal Mask is intended<br>to provide an interface for<br>noninvasive Positive Airway<br>Pressure (PAP) ventilation<br>therapy. This mask is intended<br>for single-patient use in the home<br>and single-patient use in the<br>hospital / institutional<br>environment. This mask is to be<br>used on patients between 2 and<br>7 years of age. | The Pixi Pediatric Mask channels<br>airflow noninvasively to a patient from a<br>continuous positive airway pressure<br>(CPAP) or bilevel device.<br>The Pixi Pediatric Mask is:<br>To be used by children aged<br>●<br>between 2 and 7 for whom<br>continuous positive airway<br>pressure (CPAP) or bilevel<br>therapy has been prescribed.<br>Intended for single-patient re-use<br>●<br>in the home environment and<br>single-patient re-use in the<br>hospital/ institutional environment. | Same |
| Contra-<br>indications | None | None | Same |
| Product<br>Code | BZD | BZD | Same |
| Environment<br>of Use | Hospital, Home | Hospital, Home | Same |
| Patient<br>Population | Pediatric patients between 2 and<br>7 years of age | Pediatric patients between 2 and 7<br>years of age | Same |
| Use Case(s) | Single Patient Use | Single Patient Use | Same |
| | | | |
| Components<br>/<br>Accessories | Headgear, Mask, Interface,<br>Swivel Connector, Feather<br>Weight Tube (Optional) | Headgear, Mask frame, Frame nodes,<br>Mask socket, Short tube assembly,<br>Elbow, Plug tether, Plug, Cushion, Vent,<br>Cushion shoulder, Pressure port, Top<br>headgear strap and buckle, Middle<br>headgear strap and buckle, Lower<br>headgear strap and buckle, Quick<br>release clip, Lower headgear arms, 22<br>mm Swivel | The SleepWeaver Pediatric and<br>predicate ResMed Pixi both have<br>a nasal mask, adjustable<br>headgear, an interface swivel<br>connection, and a tube assembly.<br>The SleepWeaver Pediatric mask<br>is manufactured from cloth<br>whereas the ResMed Pixi mask<br>has a silicone cushion and a<br>plastic stability feature. The<br>SleepWeaver Pediatric Mask is<br>supported by performance testing;<br>thus it can be concluded that the<br>SleepWeaver Advance Pediatric<br>Nasal CPAP Mask, is substantially<br>equivalent to the predicate device,<br>the ResMed Pixi Pediatric Mask, in<br>regards to this comparison. |
| Design | Image: SleepWeaver Pediatric Mask | Image: ResMed Pixi Pediatric Mask | The SleepWeaver Pediatric Mask<br>and predicate ResMed Pixi are<br>both nasal CPAP masks with<br>adjustable headgear. The<br>SleepWeaver Pediatric mask is<br>manufactured from cloth whereas<br>the ResMed Pixi mask has a<br>silicone cushion and a plastic<br>stability feature. The<br>SleepWeaver Pediatric Mask is<br>supported by performance testing;<br>thus it can be concluded that the<br>SleepWeaver Advance Pediatric<br>Nasal CPAP Mask, is substantially<br>equivalent to the predicate device,<br>the ResMed Pixi Pediatric Mask, in<br>regards to this comparison. |
| Materials | Headgear: Nylon,<br>Elastane, Polyurethane<br>Mask: Blue Polyester<br>Interface: Polyester, Nylon,<br>Lycra, Polyurethane<br>Swivel Connector: Polycarbonate | Information not available | The SleepWeeaver Pediatric Mask<br>and headgear are made of<br>materials used in legally marketed<br>devices (SleepWeaver Advance<br>Nasal CPAP Mask (K092362),<br>SleepWeaver Elan Nasal CPAP<br>Mask (K120757), and<br>SleepWeaver Anew Full Face<br>Mask CPAP (K130481)) and the<br>predicate ResMed Pixi is made of<br>plastic and silicone. The<br>SleepWeaver Pediatric Mask uses<br>biocompatible materials; thus it<br>can be concluded that the<br>SleepWeaver Advance Pediatric<br>Nasal CPAP Mask, is substantially<br>equivalent to the predicate device,<br>the ResMed Pixi Pediatric Mask, in<br>regards to this comparison. |
| Therapy<br>Pressure<br>Range | $4 - 20 cm H_2O$ | $3 to 20 cm H_2O$ | The SleepWeaver Pediatric has a<br>therapy range from $4 to 20 cm$<br>$H_2O$ whereas he ResMed Pixi has<br>a therapy range from $3 to 20 cm$<br>$H_2O$ . Since the SleepWeaver<br>Pediatric therapy range is within<br>the therapy range for the ResMed |
# Table 1: Technological Characteristics and Comparison to the Predicate
{6}------------------------------------------------
{7}------------------------------------------------
| | | | Pixi, it can be concluded that the<br>SleepWeaver Advance Pediatric<br>Nasal CPAP Mask, is substantially<br>equivalent to the predicate device,<br>the ResMed Pixi Pediatric Mask, in<br>regards to this comparison. | | | | | | | | | | | | | | | | | | | | | | | | |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|----------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sizing | 1 size | 1 size | SleepWeaver Pediatric and<br>ResMed Pixi are both sized to<br>accommodate the intended patient<br>population. | | | | | | | | | | | | | | | | | | | | | | | | |
| Physical<br>Dead Space | 130 mL | 101.7 mL | The dead space for the<br>SleepWeaver Pediatric is larger<br>than the predicate ResMed Pixi;<br>however, since the SleepWeaver<br>Pediatric mask has exhalation<br>holes and seams that enable<br>continuous purging of the volume<br>of air around the periphery of the<br>mask, it prevents CO2 build-up.<br>This is supported by the<br>performance testing since the<br>mask has passed the rebreathing<br>tests under ISO 17510-2. Thus it<br>can be concluded that the<br>SleepWeaver Advance Pediatric<br>Nasal CPAP Mask, is substantially<br>equivalent to the predicate device,<br>the ResMed Pixi Pediatric Mask, in<br>regards to this comparison. | | | | | | | | | | | | | | | | | | | | | | | | |
| A-weighted<br>Sound<br>levels | Sound Power Level:<br>26 dBA;<br>Sound Pressure Level at 1m:<br>18 dBA | Not available | Not applicable | | | | | | | | | | | | | | | | | | | | | | | | |
| Resistance<br>(pressure<br>drop) | 0.4 cm H2O at 50 L/min;<br>1.0 cm H2O at 100 L/min | 0.8 cm H2O at 50 L/min;<br>3.1 cm H2O at 100 L/min | The resistance for the<br>SleepWeaver Pediatric is lower<br>than the predicate ResMed Pixi;<br>however, a lower pressure drop<br>may be considered preferential as<br>the patient will receive therapy<br>closer to the intended pressure.<br>Thus it can be concluded that the<br>SleepWeaver Advance Pediatric<br>Nasal CPAP Mask, is substantially<br>equivalent to the predicate device,<br>the ResMed Pixi Pediatric Mask, in<br>regards to this comparison. | | | | | | | | | | | | | | | | | | | | | | | | |
| Carbon<br>Dioxide<br>Washout<br>Profile | The amount of CO2 being<br>rebreathed by the patient in all<br>scenarios listed in ISO 17510-<br>2:2009 is less than the maximum<br>allowable percentage increase<br>under normal conditions (20% for<br>4, 5, and 10 cm H2O). | Not available | The SleepWeaver Pediatric meets<br>ISO 17510-2. | | | | | | | | | | | | | | | | | | | | | | | | |
| Leak Rate | Pressure<br>(cm H2O) Flow<br>(L/min) 4 14 5 16 10 25 15 33 20 40 | | | | | | | | | | | | | Pressure<br>(cm H2O) Flow<br>(L/min) 3 18 7 28 12 36 16 43 20 49 | | | | | | | | | | | | | The leak rate is lower than the<br>predicate ResMed Pixi. Since the<br>leak rate of the SleepWeaver<br>Pediatric mask is greater than a<br>patient would exhale per minute<br>and the SleepWeaver Pediatric<br>mask passed the rebreathing tests<br>under ISO 17510-2, it can be<br>concluded that the SleepWeaver<br>Advance Pediatric Nasal CPAP<br>Mask, is substantially equivalent to |
{8}------------------------------------------------
| | | | the predicate device, the ResMed<br>Pixi Pediatric Mask, in regards to<br>this comparison. |
|----------------------|-------------------------------------------------------------------------------------------------------|-------------|--------------------------------------------------------------------------------------------|
| Applied<br>Standards | ISO 14971, ISO 10993-1, 10993-<br>5, ISO 10993-10, ISO 5356-1,<br>IEC 62366, ISO 5367, ISO<br>17510-2 | ISO 10993-1 | Not applicable |
# Non-Clinical Testing Summary:
Bench top testing for the SleepWeaver Advance Pediatric Nasal CPAP Mask included the following tests or evaluations:
- . Product performance;
- Fixed leak rate per ISO 17510-2; O
- Resistance to flow per ISO 17510-2; O
- o CO2 rebreathing testing per ISO 17510-2:
- Physical dead space; and O
- O Headgear attachment;
- Reliability/Cleaning:
- Fit evaluation and anthropometric analysis; and ●
- Simulated use case;
- . Biocompatibility:
- Patient contact Duration: Permanent o
- Gas Pathway (External Communicating) Components: Swivel connector, O mask body material, nasal interface, and internal surface of the oral interface. The materials in these components have been used in legally marketed devices (SleepWeaver Advance Nasal CPAP Mask (K092362), SleepWeaver Elan Nasal CPAP Mask (K120757), and SleepWeaver Anew Full Face Mask CPAP (K130481)).
- Surface Contact (Skin) Components: Headgear/mask wings and external o surface of the oral interface. The materials in these components have been used in legally marketed devices (SleepWeaver Advance Nasal CPAP Mask (K092362), SleepWeaver Elan Nasal CPAP Mask (K120757), and SleepWeaver Anew Full Face Mask CPAP (K130481)).
The packaging of the SleepWeaver Advance Pediatric Nasal CPAP mask uses the same packaging and is stored in the same way as other legally marketed devices (SleepWeaver Advance Nasal CPAP Mask (K092362)).
{9}------------------------------------------------
# Clinical Data
Use of the SleepWeaver Advance Pediatric Nasal CPAP Mask is supported by bench testing and is sufficient to demonstrate substantial equivalence to the ResMed Pixi Pediatric Mask.
## Substantial Equivalence (Conclusion)
The new device, the SleepWeaver Advance Pediatric Nasal CPAP Mask, is substantially equivalent to the predicate device, the ResMed Pixi Pediatric Mask (K102224):
- . it has the same intended use;
- it has the same patient population;
- . it has similar technological characteristics;
- it has similar performance characteristics; and
- it does not raise new questions of safety and effectiveness ●
Thus it can be concluded that the SleepWeaver Advance Pediatric Nasal CPAP Mask, is substantially equivalent to the predicate device, the ResMed Pixi Pediatric Mask.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.