← Product Code [LYZ](/productcode/LYZ) · K151682

# Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) (K151682)

_Daxwell, LLC · LYZ · Nov 17, 2015 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151682

## Device Facts

- **Applicant:** Daxwell, LLC
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Nov 17, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable vinyl patient examination gloves; worn on examiner's hand or finger; act as physical barrier against body fluids and bloodborne pathogens. Device conforms to hand to allow necessary medical procedure movements. Used in clinical settings by healthcare providers. Benefits include protection against contamination for both patient and examiner. Device is non-sterile.

## Clinical Evidence

No clinical data required for this submission. Substantial equivalence supported by non-clinical bench testing including dimensions, physical properties (tensile strength, elongation), freedom from pinholes (ASTM D5151), residual powder (ASTM D6124), and biocompatibility (ISO 10993-10).

## Technological Characteristics

Material: Poly (vinyl chloride) (PVC). Surface coating: Polyurethane (PU). Standards: ASTM D5250-06, ASTM D5151-06, ASTM D6124-06, ISO 10993-10. Physical properties: Tensile strength 15-20 MPa, Elongation 360-420%. Residual powder <2mg/glove. Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder-Free Vinyl Patient Examination Glove (Non-colored) ([K091663](/device/K091663.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, layered on top of each other, suggesting a sense of community and support. The profiles are rendered in a simple, elegant design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 17, 2015

Daxwell, LLC c/o Chu Xiaoan Beijing Easylink CO., LTD Rm. F302 Bldg., 41. Jing Cheng Ya Ju Courtyard 6 of Southern Dou Ge Zhang Chaoyang District Beijing 100121 CHINA

Re: K151682

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient examination glove Regulatory Class: Class I Product Code: LYZ Dated: October 4, 2015 Received: October 16, 2015

Dear Mr. Chu Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

*Tejashri Purohit-Sheth, M.D.*

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151682

Device Name

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)

Indications for Use (Describe)

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

> Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### Section C 510(k) Summary

# 510(K) Summary

"The assigned 510(k) number is: _ _ _ K151682

# Premarket Notification [510(k)] Summary

### 0.0 Application correspondent:

| Company name           | Beijing Easy-Link Company                                                                                                          |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Company address        | Rm. F302 Bldg., 41, Jing Cheng Ya Ju,<br>Courtyard 6 of Southern Dou Ge Zhuang,<br>Chaoyang District, Beijing 100121,<br>P.R.China |
| Contact Person         | Chu Xiaoan                                                                                                                         |
| Contact E-mail address | easylink_bj@aliyun.com                                                                                                             |

### 1.0 Submitter:

| Submitter's name :      | Daxwell, LLC                                                     |
|-------------------------|------------------------------------------------------------------|
| Submitter's address :   | Address: 2825 Wilcrest Dr., Suite# 500<br>Houston, TX 77042,USA. |
| Phone number :          | (281)669-0622                                                    |
| Fax number :            | (281)669-0617                                                    |
| Name of contact person: | Frank Zhang                                                      |
| Date of preparation :   | 2015-11-17                                                       |

### 2.0 Name of the Device

| Device Name:            | Powder Free Vinyl Patient Examination<br>Gloves, Clear (non-colored) |
|-------------------------|----------------------------------------------------------------------|
| Proprietary/Trade name: | Powder Free Vinyl Patient Examination<br>Gloves, Clear (non-colored) |
| Common Name:            | Exam gloves                                                          |
| Classification Name:    | Patient examination glove                                            |
| Device Classification:  | I                                                                    |
| Regulation Number:      | 21 CFR 880.6250                                                      |
| Panel:                  | General Hospital (80)                                                |
| Product Code:           | LYZ                                                                  |

## 3.0 Predicate device

| Device Name:   | Powder-Free Vinyl Patient Examination Glove<br>(Non-colored) |
|----------------|--------------------------------------------------------------|
| Company name:  | Zhang Jia Gang Fengyuan Plastic Product Co. Ltd.             |
| 510(K) Number: | K091663.                                                     |

{4}------------------------------------------------

## 4.0 Device Description:

#### 4.1 How the device functions:

PVC films form a barrier to body fluids and bloodborne Pathogens

#### 4.2 Scientific concepts that form the basis for the device

The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

- 4.3 Physical and performance characteristics such as design, materials and physical properties:
Poly (vinyl chloride) glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

## 5.0 Device Intended Use (Indication for use):

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## 6.0 Summary of the Technological Characteristics of the Device:

The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), non-sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

| Characteristics                                                    | Standard                                                                      |                                                                | Device performance  |                                            |
|--------------------------------------------------------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------|---------------------|--------------------------------------------|
| Dimension                                                          | ASTM D 5250-06(Reapproved 2011).                                              |                                                                |                     |                                            |
|                                                                    | Length                                                                        | ≥230mm                                                         | 232-240mm           |                                            |
|                                                                    | Width                                                                         | Small                                                          | 80-90 mm            | 81-90mm                                    |
|                                                                    |                                                                               | Medium                                                         | 90-100mm            | 93-98mm                                    |
|                                                                    |                                                                               | Large                                                          | 100-110mm           | 102-109mm                                  |
|                                                                    |                                                                               | X large                                                        | 110-120 mm          | 110-118mm                                  |
|                                                                    | Thickness                                                                     | Fingertip                                                      | ≥0.05mm             | 0.08-0.11 mm                               |
|                                                                    |                                                                               | Palm                                                           | ≥0.08mm             | 0.09-0.12 mm                               |
| Physical Properties                                                | ASTM D 5250-06(Reapproved 2011).                                              |                                                                |                     |                                            |
|                                                                    | Tensile strength<br>(Before & After aging)                                    | ≥11MPa                                                         | 15-20MPa            |                                            |
|                                                                    | Elongated rate<br>(Before & After aging)                                      | ≥300%                                                          | 360-420%            |                                            |
| Freedom from<br>pinholes                                           | •                                                                             | 21 CFR 800.20                                                  | Passed Standard     |                                            |
|                                                                    | •                                                                             | ASTM D5250-06(Reapproved 2011)                                 | Acceptance Criteria |                                            |
|                                                                    | •                                                                             | ASTM D5151-06(Reapproved 2011)                                 |                     |                                            |
| Powder Residual                                                    | ASTM standard D 5250-06 (Reapproved<br>2011).and D6124-06(Reaffirmation 2011) |                                                                | Meets<br><2mg/glove |                                            |
|                                                                    | Biocompatibility                                                              | Primary Skin Irritation in rabbits<br>ISO 10993-10: 2010-08-01 |                     | Passes<br>Not a Primary Skin<br>Irritation |
| Dermal sensitization in the guinea pig<br>ISO 10993-10: 2010-08-01 |                                                                               | Passes<br>Not a Dermal<br>sensitization                        |                     |                                            |

{5}------------------------------------------------

## 7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), meet requirements per ASTM D5250-06 (Reapproved 2011), per ASTM D6124-06 Reaffirmation 2011), per 21 CFR 800.20 and ISO 10993-10:2010-08-01.

The performance test data of the non clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

#### 8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data was not required for this submission.

| Features & Description                                                                              | Predicate Device                                                                                                                                                                                                          | Subject Device                                                                                                                                                                                                              | Result of Comparison        |
|-----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Company                                                                                             | Zhang Jia Gang Fengyuan Plastic Product Co.Ltd.                                                                                                                                                                           | Daxwell, LLC                                                                                                                                                                                                                | --                          |
| 510(K) Number                                                                                       | K091663                                                                                                                                                                                                                   | K151682                                                                                                                                                                                                                     |                             |
| Product name                                                                                        | Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)                                                                                                                                                         | Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)                                                                                                                                                           | same                        |
| Product Code                                                                                        | LYZ                                                                                                                                                                                                                       | LYZ                                                                                                                                                                                                                         | same                        |
| Size                                                                                                | Small/ Medium/ Large/X large                                                                                                                                                                                              | Small/ Medium/ Large/X large                                                                                                                                                                                                | same                        |
| Intend for use                                                                                      | Powder free Vinyl Patient Examination Gloves, Clear(Non-colored)is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Powder free Vinyl Patient Examination Gloves, Clear (Non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Substantially equivalent    |
| Device Description and Specifications                                                               | Meets ASTM D5250-06 (Reapproved 2011)                                                                                                                                                                                     | Meets ASTM D5250-06 (Reapproved 2011)                                                                                                                                                                                       | Substantially equivalent    |
| Dimensions --Length                                                                                 | Meets ASTM D5250 -06 (Reapproved 2011)<br>≥230mm min                                                                                                                                                                      | 230mm min for all sizes                                                                                                                                                                                                     | Substantially equivalent    |
| Dimensions -- Width                                                                                 | Meets ASTM D5250-06 (Reapproved 2011)<br>Small 80-90 mm<br>Medium 90-100mm<br>Large 100-110mm<br>X large 110-120 mm                                                                                                       | Small 81-90 mm<br>Medium 93-98mm<br>Large 102-109mm<br>X large 110-118 mm                                                                                                                                                   | Substantially equivalent    |
| Dimensions --Thickness                                                                              | Meets ASTM D5250-06 (Reapproved 2011)<br>Finger 0.05mm min.<br>Palm 0.08mm min                                                                                                                                            | Thickness (mm) min.<br>Finger 0.08-0.11<br>Palm 0.09-0.11                                                                                                                                                                   | Substantially equivalent    |
| Physical Properties                                                                                 | Meets ASTM D 5250-06<br>(Reapproved 2011)                                                                                                                                                                                 |                                                                                                                                                                                                                             | Substantially<br>equivalent |
|                                                                                                     | Before aging/after aging<br>Elongation ≥300%<br>Tensile Strength≥ 14MPa                                                                                                                                                   | Before aging/after aging<br>Elongation : 360-420%<br>Tensile Strength:15-20 MPa                                                                                                                                             |                             |
| Freedom<br>from<br>Pinholes                                                                         | Meets<br>• 21 CFR 800.20<br>• ASTM D5250-06<br>(Reapproved 2011)<br>• ASTM D 5151-06<br>(Reapproved 2011)                                                                                                                 | Meets ASTM<br>D5151-06<br>(Reapproved 2011)<br>Holes at<br>Inspection Level I AQL2.5                                                                                                                                        | Substantially<br>equivalent |
| Residual Powder                                                                                     | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)<br>below than 2 mg of<br>residual powder per glove                                                                                                                           | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)<br>below than 2 mg of<br>residual powder per glove                                                                                                                             | Substantially<br>equivalent |
| Materials used to<br>fabricate the<br>devices                                                       | PVC                                                                                                                                                                                                                       | PVC                                                                                                                                                                                                                         | Substantially<br>equivalent |
| Dusting or<br>Donning Powder:                                                                       | PU                                                                                                                                                                                                                        | PU                                                                                                                                                                                                                          | Substantially<br>equivalent |
| Dusting or<br>Donning Powder:<br>name                                                               | Surface Coating Agent                                                                                                                                                                                                     | Surface Coating Agent                                                                                                                                                                                                       | Substantially<br>equivalent |
| Compare<br>performance data<br>supporting<br>substantial<br>equivalence                             | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reaffirmation 2011)                                                                                                | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reaffirmation 2011)                                                                                                  | Substantially<br>equivalent |
| Single Patient Use                                                                                  | Single Patient Use                                                                                                                                                                                                        | Single Patient Use                                                                                                                                                                                                          | Substantially<br>equivalent |
| Biocompatibility                                                                                    | Under the conditions of this<br>study, the test article was a<br>non- irritant or non-<br>sensitizer.                                                                                                                     | Under the conditions of this<br>study, the test article was a<br>non- irritant or non-<br>sensitizer.<br><br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10 Third Edition<br>2010-08-01    | Substantially<br>equivalent |
|                                                                                                     | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION STUDIES<br>Meets ISO<br>10993-10:2002/Amd.1:2006                                                                                                                           |                                                                                                                                                                                                                             |                             |
| Labeling for the<br>legally marketed<br>device to which<br>substantial<br>equivalence is<br>claimed | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot                                                                                                                            | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot                                                                                                                              | Substantially<br>equivalent |

### 9.0 Substantial Equivalence Comparison:

{6}------------------------------------------------

#### 10.0 Substantial Equivalence Comparison:

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims.

{7}------------------------------------------------

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is as safe, as effective, and performs as well as the predicate device, Powder-Free Vinyl Patient Examination Glove (Non-colored) Zhang Jia Gang Fengyuan Plastic Product Co., Ltd. K091663.

The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.

---

**Source:** [https://fda-staging.innolitics.com/device/K151682](https://fda-staging.innolitics.com/device/K151682)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
