← Product Code [LZA](/productcode/LZA) · K151680

# Powder Free Nitrile Patient Examination Gloves, Blue Color (K151680)

_Suqian Xingye Glove Co,Ltd · LZA · Feb 12, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151680

## Device Facts

- **Applicant:** Suqian Xingye Glove Co,Ltd
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Feb 12, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Nitrile patient examination glove; provides barrier against body fluids and bloodborne pathogens. Device worn on examiner's hand/finger during medical procedures. Nitrile material provides tensile strength and flexibility to conform to hand. Non-sterile; disposable; single-patient use. Bench testing confirms barrier integrity and physical properties. Used in clinical settings to protect both patient and examiner from contamination.

## Clinical Evidence

No clinical data. Substantial equivalence is based on non-clinical bench testing, including dimension, tensile strength, elongation, water leak (pinhole) testing per ASTM D5151-06, powder residual testing per ASTM D6124-06, and biocompatibility (skin irritation and dermal sensitization) per ISO 10993-10.

## Technological Characteristics

Nitrile rubber material; non-sterile; powder-free with PU surface coating agent. Dimensions and physical properties conform to ASTM D6319-10. Biocompatibility tested per ISO 10993-10. Barrier integrity tested per ASTM D5151-06 and 21 CFR 800.20.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Nitrile Powder Free Patient Examination Gloves, Cloured (Blue) ([K120970](/device/K120970.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 12, 2016

Suqian Xingye Glove Co., Ltd. c/o Mr. Chu Xiaoan Beijing Easylink Co., Ltd. Room F302 Bldg. 41. Jing Cheng Ya Ju Courtyard 6 of Southern Dou Ge Zhuang Beijing 100121 CHINA

Re: K151680

Trade/Device Name: Powder Free Nitrile Patient Examination Gloves, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: January 3, 2016 Received: January 11, 2016

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151680

### Device Name

Powder Free Nitrile Patient Examination Gloves, Blue Color

Indications for Use (Describe)

Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### Section 6 510(k) Summary

# 510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

" "The assigned 510(k) number is: K 151680

## Premarket Notification [510(k)] Summary

#### 1.0 Submitter:

| Submitter's name :      | Suqian Xingye Glove Co,Ltd                                                            |
|-------------------------|---------------------------------------------------------------------------------------|
| Submitter's address :   | Dongwu Road,Economic Development<br>Zone,Suqian City,Jiangsu<br>Province,223800,China |
| Phone number :          | 0086-527-82860533                                                                     |
| Fax number :            | 0086-527-82860080                                                                     |
| Name of contact person: | Jian Zhong Deng                                                                       |
| Date of preparation :   | 2015-11-07                                                                            |

#### 2.0 Name of the Device

| Device Name:            | Powder Free Nitrile Patient Examination<br>Gloves, Blue Color |
|-------------------------|---------------------------------------------------------------|
| Proprietary/Trade name: | Powder Free Nitrile Patient Examination<br>Gloves, Blue Color |
| Common Name:            | Exam gloves                                                   |
| Classification Name:    | Patient examination glove                                     |
| Device Classification:  | I                                                             |
| Regulation Number:      | 21 CFR 880.6250                                               |
| Panel:                  | General Hospital (80)                                         |
| Product Code:           | LZA                                                           |

## 3.0 Predicate device

| Device Name:   | Nitrile Powder Free Patient Examination<br>Gloves, Cloured (Blue) |
|----------------|-------------------------------------------------------------------|
| Company name:  | Tangshan Zhonghong Pulin Plastic Co.,Ltd.                         |
| 510(K) Number: | K120970                                                           |

### 4.0 Device Description:

- 4.1 How the device functions: Nitrile films form a barrier to body fluids and bloodborne Pathogens
#### 4.2 Scientific concepts that form the basis for the device

{4}------------------------------------------------

The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

### Physical and performance characteristics such as design, 4.3 materials and physical properties:

Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

# 5.0 Device Intended Use (Indication for use):

Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

# 6.0 Summary of the Technological Characteristics of the Device:

The Powder Free Nitrile Patient Examination Gloves, Blue Color, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

| Characteristics          | Standard                                                         | Device performance                        |                                                                             |
|--------------------------|------------------------------------------------------------------|-------------------------------------------|-----------------------------------------------------------------------------|
| Dimension                | ASTM standard D 6319-10.                                         |                                           |                                                                             |
|                          | Length                                                           | ≥230mm                                    | 234-243mm                                                                   |
|                          | Width                                                            | Small 70-90 mm                            | 78-85mm                                                                     |
|                          |                                                                  | Medium 85-105mm                           | 95-99mm                                                                     |
|                          |                                                                  | Large 100-120mm                           | 108-113mm                                                                   |
|                          |                                                                  | X large 110-130 mm                        | 117-126mm                                                                   |
|                          | Thickness                                                        | Fingertip ≥0.05mm                         | 0.08-0.11 mm                                                                |
|                          |                                                                  | Palm ≥0.05mm                              | 0.08-0.12 mm                                                                |
| Physical<br>Properties   | ASTM standard D 6319-10.                                         |                                           |                                                                             |
|                          | Tensile strength<br>(Before aging)                               |                                           | 20-23MPa                                                                    |
|                          | Tensile strength<br>(After aging)                                | ≥14MPa                                    | 18-21 MPa                                                                   |
|                          | Elongated rate<br>(Before aging)                                 | ≥500%                                     | 560-600%                                                                    |
|                          | Elongated rate<br>(After aging)                                  | ≥400%                                     | 530-580%                                                                    |
| Freedom from<br>pinholes | • 21 CFR 800.20                                                  |                                           | Passed Standard<br>Acceptance Criteria                                      |
|                          | • ASTM standard D 6319-10.                                       |                                           |                                                                             |
|                          | Test method in accordance with<br>ASTM D5151-06(Reapproved 2011) |                                           |                                                                             |
| Powder Residual          | ASTM standard D 6319-10.                                         |                                           | 0.1mg/glove                                                                 |
|                          | Test method in accordance with<br>D6124-06(Reaffirmation 2011)   |                                           |                                                                             |
| Biocompatibility         | Primary Skin Irritation in rabbits                               |                                           | Under the conditions<br>of the study, the<br>device was a<br>non-irritant.  |
|                          |                                                                  | ISO 10993-10: Third Edition<br>2010-08-01 |                                                                             |
|                          |                                                                  |                                           |                                                                             |
|                          | Dermal sensitization in the guinea pig                           |                                           | Under the conditions<br>of the study, the<br>device was a<br>non-sensitizer |
|                          | ISO 10993-10: Third Edition<br>2010-08-01                        |                                           |                                                                             |

{5}------------------------------------------------

# 7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:

The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. The Powder Free Nitrile Patient Examination Gloves, Blue Color made by Sugian Xingye Glove Co,Ltd met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing.

- . Dimension per ASTM D6319-10,
- Tensile strength (Before aging/ After aging) and Elongated rate (Before aging/ . After aging) per ASTM D6319-10,
- Water leak test on pinhole per ASTM D6319-10 and per 21 CFR 800.20, ●
- . Powder Residual tests per ASTM D6319-10(Reapproved 2011)
- Biocompatibility test per ISO 10993-10: Third Edition 2010-08-01.

### Substantial Equivalent Based on Assessment of Clinical Performance Data: 8.0 -

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

So determination of substantial equivalence is not based on an assessment of clinical performance data.

| Features & Description                      | Predicate Device                                                                                                                                                                                                                                    | Subject Device                                                                                                                                                                                                                                     | Result of Comparison        |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Company                                     | Tangshan Zhonghong<br>Pulin Plastic Co., Ltd.                                                                                                                                                                                                       | Suqian Xingye Glove<br>Co,Ltd                                                                                                                                                                                                                      | --                          |
| 510(K)<br>Number                            | K120970                                                                                                                                                                                                                                             | K151680                                                                                                                                                                                                                                            |                             |
| Product name                                | Nitrile Powder Free<br>Patient Examination<br>Gloves, Cloured (Blue)                                                                                                                                                                                | Powder Free Nitrile<br>Patient Examination<br>Gloves, Blue Color                                                                                                                                                                                   | same                        |
| Product Code                                | LZA                                                                                                                                                                                                                                                 | LZA                                                                                                                                                                                                                                                | same                        |
| Size                                        | Small/ Medium/<br>Large/X large                                                                                                                                                                                                                     | Small/ Medium/<br>Large/X large                                                                                                                                                                                                                    | same                        |
| Intend for use                              | Nitrile Powder Free<br>Patient Examination<br>Gloves, Cloured (Blue) is<br>a disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand or<br>finger to prevent<br>contamination between<br>patient and examiner. | Powder Free Nitrile<br>Patient Examination<br>Gloves, Blue Color is<br>a disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's<br>hand or finger to<br>prevent contamination<br>between patient and<br>examiner. | Substantially<br>equivalent |
| Device Description<br>and<br>Specifications | Meets ASTM D6319-10                                                                                                                                                                                                                                 | Meets ASTM<br>D6319-10                                                                                                                                                                                                                             | Substantially<br>equivalent |

### 9.0 Substantial Equivalence Comparison:

{6}------------------------------------------------

| Dimensions                                                              |         | Meets ASTM<br>D6319-10                                                                               | 234 mm min for all<br>sizes                                                                        | Substantially<br>equivalent |         |  |
|-------------------------------------------------------------------------|---------|------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------|---------|--|
| Dimensions<br>--Length                                                  |         | ≥230mm min                                                                                           |                                                                                                    |                             |         |  |
| Dimensions<br>-- Width                                                  |         | Meets ASTM D6319-10                                                                                  |                                                                                                    | Substantially<br>equivalent |         |  |
|                                                                         | Small   | 70-90 mm                                                                                             | Small 78-85 mm                                                                                     |                             |         |  |
|                                                                         | Medium  | 85-105mm                                                                                             | Medium 95-99 mm                                                                                    |                             |         |  |
|                                                                         | Large   | 100-120mm                                                                                            | Large 108-113mm                                                                                    |                             |         |  |
|                                                                         | X large | 110-130 mm                                                                                           | X large 117-126 mm                                                                                 |                             |         |  |
| Dimensions<br>--Thickness                                               |         | Meets ASTM D6319-10                                                                                  |                                                                                                    | Substantially<br>equivalent |         |  |
|                                                                         |         | Finger 0.05mm min.<br>Palm 0.05mm min.                                                               | Thickness (mm) min.<br>Finger 0.08<br>Palm 0.08                                                    |                             |         |  |
| Physical<br>Properties                                                  |         | Meets ASTM D D6319-10                                                                                | Aging Before After                                                                                 | Substantially<br>equivalent |         |  |
|                                                                         |         | Before aging/after aging<br>Elongation ≥500%<br>Tensile Strength≥ 14MPa                              | Elongation<br>(%)                                                                                  | 560-600                     | 530-580 |  |
|                                                                         |         |                                                                                                      | Tensile<br>Strength<br>(MPa)                                                                       | 20-23                       | 18-21   |  |
| Freedom from<br>Pinholes                                                |         | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10                                                          | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10                                                        | Substantially<br>equivalent |         |  |
|                                                                         |         | Test method in accordance<br>with ASTM D5151-06<br>(Reapproved 2011)                                 | Tested in accordance<br>with ASTM D5151<br>(Reapproved 2011)<br>with acceptable results            |                             |         |  |
| Residual<br>Powder                                                      |         | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)                                                         | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)                                                       | Substantially<br>equivalent |         |  |
|                                                                         |         | below 2mg of residual<br>powder                                                                      | Results generated<br>values below 2mg of<br>residual powder                                        |                             |         |  |
| Materials used<br>to fabricate the<br>devices                           |         | Nitrile                                                                                              | Nitrile                                                                                            | Substantially<br>equivalent |         |  |
| Dusting or<br>Donning<br>Powder:                                        |         | PU                                                                                                   | PU-120C                                                                                            | Substantially<br>equivalent |         |  |
| Dusting or<br>Donning<br>Powder: name                                   |         | Surface Coating Agent                                                                                | Surface Coating Agent                                                                              | Substantially<br>equivalent |         |  |
| Compare<br>performance<br>data supporting<br>substantial<br>equivalence |         | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D6319-10<br>ASTM D6124-06<br>(Reapproved 2011)   | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D6319-10<br>ASTM D6124-06<br>(Reapproved 2011) | Substantially<br>equivalent |         |  |
| Single Patient<br>Use                                                   |         | Single Patient Use                                                                                   | Single Patient Use                                                                                 | Substantially<br>equivalent |         |  |
| Biocompatibility                                                        |         | Under the conditions of this<br>study, the test article was a<br>non- irritant or non-<br>sensitizer | Under the conditions of<br>this study, the test<br>article was a non-<br>irritant or non-          | Substantially<br>equivalent |         |  |

{7}------------------------------------------------

|                |                          | sensitizer.          |               |
|----------------|--------------------------|----------------------|---------------|
|                | SKIN IRRITATION          |                      |               |
|                | DERMAL and               | SKIN IRRITATION      |               |
|                | SENSITIZATION            | DERMAL<br>and        |               |
|                | STUDIES Meets ISO        | SENSITIZATION        |               |
|                | 10993-10:2002/Amd.1:2006 | STUDIES Meets ISO    |               |
|                |                          | 10993-10: Third      |               |
|                |                          | Edition 2010-08-01.  |               |
| Labeling for   | -Powder Free             | -Powder Free         | Substantially |
| the legally    | -Patient Examination     | -Patient Examination | equivalent    |
| marketed       | Glove                    | Glove                |               |
| device to      | -Single Use Only         | -Single Use Only     |               |
| which          | - Manufactured For:      | - Manufactured For:  |               |
| substantial    | - Lot                    | - Lot                |               |
| equivalence is | -Blue color              | -Blue color          |               |
| claimed.       | - Non sterile            | - Non sterile        |               |

#### 10.0 Conclusion:

It can be concluded that the Powder Free Nitrile Patient Examination Gloves, Blue Color is as safe, as effective, and performs as well as the predicate device, Nitrile Powder Free Patient Examination Gloves, Cloured (Blue), Tangshan Zhonghong Pulin Plastic Co., Ltd., K120970.

The conclusions drawn from the nonclinical tests demonstrate that the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K151680](https://fda-staging.innolitics.com/device/K151680)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).