The Disposable Multistage Dilation Balloon Catheter is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract.
Device Story
Disposable Multistage Dilation Balloon Catheter; used for endoscopic dilation of gastrointestinal strictures. Device passes through endoscope working channel (≥2.8mm); accepts 0.035in guidewire. Features multistage balloon capable of 3 distinct, progressively larger diameters via controlled radial expansion. Operated by physician (gastroenterologist) in clinical setting. Output is mechanical dilation of stricture; provides therapeutic benefit by relieving obstruction. Device is single-use; supplied sterile.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing only, including balloon burst pressure, fatigue testing, inflation/deflation testing, radiopacity, tensile strength, tip pull testing, and package/shelf-life integrity.
Technological Characteristics
Multistage dilation balloon catheter; 3-stage radial expansion. Compatible with 2.8mm endoscope channels and 0.035in guidewires. Materials, design, and configuration are substantially equivalent to predicate. Supplied sterile (EO). Shelf life: 36 months.
Indications for Use
Indicated for adult and adolescent patients requiring endoscopic dilation of gastrointestinal tract strictures.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 16, 2015
Micro-Tech (Nanjing) CO., Ltd. Becky Li Quality Manager No.10 Gaoke Third Rd., Nanjing National Hi-Tech Industrial Development Zone Nanjing, Jiangsu 210032 China
Re: K151671
> Trade/Device Name: Disposable Multistage Dilation Balloon Catheter Regulation Number: 21 CFR §876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE, KNQ Dated: September 6, 2015 Received: September 10, 2015
Dear Becky Li,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K151671
Device Name
Disposable Multistage Dilation Balloon Catheter
Indications for Use (Describe)
The Disposable Multistage Dilation Balloon Catheter is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract.
Type of Use (*Select one or both, as applicable*)
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>□ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for MICRO-TECH. The logo consists of the letters "MT" in a stylized, three-dimensional font, with the word "MICRO-TECH" to the right of the letters. Below the word "MICRO-TECH" are two Chinese characters. The logo is blue and white.
#### 510K Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K151671
### 1. Date of Preparation: 10/14/2015
#### 2. Sponsor Identification
Micro-Tech (Nanjing) Co., Ltd.
No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing 210032, Jiangsu Province, PRC Establishment Registration Number: 3004837686 Contact Person: Becky Li Position: Manager of Quality Department Tel: +86-25-58646378 Fax: +86-25-58744269
Email: In@micro-tech.com.cn
#### 3. Identification of Proposed Device
Product Name: Disposable Multistage Dilation Balloon Catheter Common Name: Disposable Multistage Dilation Balloon Catheter Requlatory Information Classification Name: 1) Catheter, Biliary, diagnostic, 2) Dilator, esophageal Classification: 2 Product Code: 1) FGE, 2) KNQ Regulation Number: 1) 876.5010, 2) 876.3565 Review Panel: Gastroenterology/Urology
#### 4. Identification of Predicate Device
510(k) Number: K112994 Trade Name: CRE Dilatation Balloon Common Name: CRE Wireguided Dilatation Balloon
#### 5. Indications for Use
The Disposable Multistage Dilation Balloon Catheter is indicated for use in
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for Micro-Tech. The logo consists of a blue, stylized "MT" symbol on the left, followed by the text "MICRO-TECH" in a sans-serif font. Below the text, there are two Chinese characters. The overall design is clean and modern.
adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract.
#### 6. Device Description
The Disposable Multistage Dilation Balloon Catheter is capable of 3 distinct and progressively larger size diameters via controlled radial expansion. Specific balloon sizes are printed on each package and hub label. The Disposable Multistage Dilation Balloon Catheter is designed to pass through a 2.8mm or greater working channel of an endoscope. It will also accept a 0.035 in (0.89 mm) guidewire through its guidewire lumen. This catheter comes packaged with a 0.035 in (0.89mm) guidewire preloaded in the guidewire lumen. The guidewire is about 20cm longer than the balloon catheter with the excess length extending from single lumen tube.
The guidewire locking device is attached to the guidewire hub of the catheter. The locking device will be packaged in locked.
#### 7. Comparison of Technological Characteristics
The Disposable Multistage Dilation Balloon Catheter incorporates substantially equivalent device materials, design, catheter configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the Boston Scientific predicate devices.
| ltem | Proposed Device | Comparison to |
|---------------------|--------------------------------------------------|-------------------|
| | Disposable Multistage Dilation Balloon Catheter | Predicate Devices |
| Product Code | 1) FGE, 2) KNQ | Similar |
| Regulation No. | 1) 876.5010, 2) 876.3565 | Similar |
| Class | 2 | Same |
| Supplied Sterile | Yes | Same |
| Balloon Diameter | 6-7-8, 8-9-10, 10-11-12, 12-13.5-15, 15-16.5-18, | Same |
| (mm) | 18-19-20 | |
| Balloon length (mm) | 30, 55, 80. | Similar |
| Rated pressure(atm) | 3-6-10, 3-5.5-9, 3-5-8,3-4.5-8,3-4.5-7,3-4.5-6 | Same |
| Working Length | 1800, 2300 | Similar |
| (mm ) | | |
#### Comparison to predicate Devices:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MICRO-TECH. The logo consists of the letters "MT" in a stylized, blue font on the left side of the image. To the right of the letters is the word "MICRO-TECH" in a similar blue font, with a line underneath it. Below the line are two Chinese characters.
510K Summary
| Item | Proposed Device<br>Disposable Multistage Dilation Balloon Catheter | Comparison to<br>Predicate Devices |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Indications for Use | The Disposable Multistage Dilation Balloon<br>Catheter is indicated for use in adult and<br>adolescent populations to endoscopically dilate<br>strictures of the gastrointestinal tract. | Similar |
| Configuration | Tip, balloon, marker band, handle junction, and<br>guidewire. | Similar |
| Single Use | Yes | Same |
| Packaging | Single-use EO sterilized pouch with one device<br>per pouch | Similar |
| Shelf Life | Three years (36 months) | Similar |
#### 8. Performance Data
Bench testing was performed to support a determination of substantial equivalence. The Disposable Multistage Dilation Balloon Catheter performs as well as the predicate and is substantially equivalent to the predicate devices. The following in-vitro performance tests were completed on the Disposable
#### Multistage Dilation Balloon Catheter:
| Balloon Diameter | Endoscopic Compatibility Testing |
|--------------------|-------------------------------------|
| Balloon length | Balloon Burst Pressure |
| Rated pressure | Balloon Fatigue Testing |
| Working Length | Radiopacity |
| Retraction testing | Balloon Inflation/Deflation Testing |
| Tensile Strength | Package integrity |
| Tip Pull Testing | Shelf Life Testing |
| | |
#### 9. Clinical Test Conclusion
No clinical study is included in this submission.
#### 10. Substantially Equivalent (SE) Conclusion
Based on the indications for use, technological characteristics, and safety and performance testing, the Disposable Multistage Dilation Balloon Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the Boston Scientific CRE Wireguided Dilatation Balloon 。
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.