K151656 · Heartway Medical Products Co., Ltd. · INI · Nov 25, 2015 · Physical Medicine
Device Facts
Record ID
K151656
Device Name
HEARTWAY Power Mobility Scooter
Applicant
Heartway Medical Products Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Nov 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Device Story
Battery-powered mobility scooter; four pneumatic wheels; swivel seat; turning tiller column; upper panel control; main frame; bumpers; anti-tipper. Operated by user via tiller-mounted controls for steering, speed, and lighting. Drive system uses two rear pneumatic wheels; steering via two front pneumatic wheels. Features independent suspension; 300 lbs weight capacity; 3.6 mph max speed. Used in various environments (avoiding sand, wet/icy surfaces, escalators, steep inclines). Provides mobility to patients with restricted sitting ability; improves independence. Healthcare provider/patient uses device for daily navigation; output is physical movement of the scooter.
Clinical Evidence
Bench testing only. Device tested against ISO 7176 series (stability, brakes, energy consumption, dimensions, speed, seating, strength, climatic, obstacle-climbing, test dummies, friction, power/control systems, information disclosure, ignition resistance) and ANSI/RESNA WC-2 (EMC). Biocompatibility testing performed on seat, hand grip, and seat belt PVC materials per ISO 10993-1, -5, and -10.
Indicated for persons restricted to a sitting position requiring mobility assistance.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
Predicate Devices
HEARTWAY Lightweight Power Mobility Scooter, S34 (K101142)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2015
HEARTWAY Medical Products Co., Ltd. No.6 Road 25, Taichung Industrial Park Taichung City, TW 40850 % Ke-Min Jen Official Correspondent No.6 Road 25, Taichung Industrial Park, Taichung City, 40850 TW
Re: K151656
Trade/Device Name: HEARTWAY Power Mobility Scooter, Cutie S17 Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: October 19, 2015 Received: October 28, 2015
Dear Dr. Ke-Min Jen,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# William J. Heetderks -S
for
Carlos L. Peña. PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151656
#### Device Name
Heartway Power Mobility Scooter, Cutie S17
#### Indications for Use (Describe)
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a business card for Heartway Medical Products Co., LTD. The card includes the company's logo, which features a stylized butterfly in purple and blue. The card also lists the company's address, phone number, fax number, website, and email address. The card also includes the TÜV CERT and ISO-9001 CERTIFICATED logos.
#### 5. 510(k) SUMMARY of the Safety and Effectiveness
Submitter's Name: HEARTWAY Medical Products Co., Ltd.
No.6, Road 25, Taichung Industrial Park,
Taichung, 40850, Taiwan, ROC
Date summary prepared: November 24, 2015
510(k) Number: K151656
Proprietary Name: HEARTWAY Power Mobility Scooter, Cutie S17
Common or Usual Name: POWERED SCOOTER
Classification Name Motorized three-wheeled vehicle
(21 CFR 890.3800, Class II, )
Product Code: INI
Mr. TIEN-HSING YANG (Email: yhead0722@hotmail.com) Company contact person:
Official Correspondent: Dr. KE-MIN JEN (email: ceirs.jen@msa.hinet.net)
TEL: 886-3-5208829, FAX: 886-3-5209783
Predicate Device: HEARTWAY Medical Products Co., Ltd.
> HEARTWAY Lightweight Power Mobility Scooter, S34 (K101142).
- Indications for Use: The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo and contact information for Heartway Medical Products Co., LTD. The logo features a stylized butterfly design in purple and blue. The text includes the company name, address, phone numbers (including sales and fax), website, and email address. Certifications from TUV and ISO-9001 are also displayed.
## Description of the device:
Cutie S17 power scooter is battery powered and configured with four pneumatic wheels, a swivel seat, a turning tiller column, an upper panel control, a main frame, a rear bumper, a front bumper, and an anti-tipper.
S17 power scooter is operated by two pneumatic rear wheels as the drive wheels and two pneumatic front wheels as the steering wheels, using the upper panel control to control the turning tiller column mechanism to control the front wheel as the power scooter steering direction mechanism. Upper panel control is able to control driving forward, driving backward, speed control and lighting function. The main frame is equipped with a rear bumper and a front bumper to allow the scooter to sustain an impact without damage to the power scooter safety system.
Turning tiller column is equipped with the head light, back mirrors, a swivel seat with the flipping arm-rests, and a lever to move the seat backward and forward. S17 power scooter maximum weight capacity: 300 lbs (135 kgs), S17 power scooter maximum speed: 3.6 mph (5.8 kph), and main frame is equipped with independent suspension. The following surfaces are recommended not to operate on:
- Sand surface
- Wet or icy surface
- Road maintenance hole metal cover ●
- Avoid going up multiple steps.
- Avoid using escalators. Use the elevator.
- Too steep incline over 10 degrees.
- Ground clearance to battery: 3.73"/100 mm
- Curb climbing ability: 3.73"/100 mm
{5}------------------------------------------------
Image /page/5/Picture/0 description: This image is a company logo for Heartway Medical Products Co., LTD. The logo includes a purple butterfly-like design on the left side of the company name. Below the company name is the company's address, phone number, fax number, website, and email address. On the right side of the logo are the TUV CERT and ISO-9001 CERTIFICATED logos.
## Performance Testing:
- (1) ANSI/ RESNA WC-2, Section 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and motorized scooters, 2009. (Exclusion of sections 4.2.3.1, 4.2.4, 4.2.5)
- ISO 7176-1 Wheelchairs Part 1: Determination of static stability, 1999. (2) (Exclusion of sections 9.4, 9.5, 10.4, 10.5, 11, 12.2)
- ISO 7176-2 Wheelchairs Part 2: Determination of dynamic stability of electric (3) wheelchairs, 2001. (Exclusion of sections 8.5, 8.6, 9.3, 9.4, 9.5, 10.3, 10.4, 10.5)
- ISO 7176-3 Wheelchairs Part 3: Determination of effectiveness of brakes, (4) 2012. (Exclusion of sections 7.3, 7.4, 7.5)
- ISO 7176-4 Wheelchairs Part 4: Energy consumption of electric wheelchairs (5) for determination of theoretical distance range, 2008.
- ISO 7176-5 Wheelchairs Part 5: Determination of overall dimensions, mass (6) and maneuvering space, 2008.
- ISO 7176-6 Wheelchairs Part 6: Determination of maximum speed, (7) acceleration and deceleration of electric wheelchairs, 2001.
- ISO 7176-7 Wheelchairs Part 7: Determination of seating dimensions -(8) Definitions and measuring method, 1998.
- (9) ISO 7176-8 Wheelchairs - Part 8: Static, impact and fatigue strength for manual wheelchairs, 1998.
- (10) ISO 7176-9 Wheelchairs Part 9: Climatic tests for electric wheelchairs, 2009.
- (11) ISO 7176-10 Wheelchairs Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs, 2008.
{6}------------------------------------------------
Image /page/6/Picture/0 description: This image shows the logo for Heartway Medical Products Co., LTD. The logo includes a purple butterfly design on the left side of the image. To the right of the butterfly is the company name and contact information, including the address, phone number, website, and email address. The image also includes the TÜV CERT and ISO-9001 CERTIFICATED logos.
- (12) ISO 7176-11 Wheelchairs Wheelchairs Part 11: Test dummies, 2012.
- (13) ISO 7176-13 Wheelchairs Part 13: Determination of coefficient of friction of test surfaces, 1989.
- (14) ISO 7176-14 Power and control system for electric wheelchairs, 2008.
- (15) ISO 7176-15 Wheelchairs Part 15: Requirements for information disclosure, documentation and labelling, 1996.
- (16) ISO 7176-16 Requirements and test methods for resistance to ignition of upholstered parts, 2012.
## Biocompatibility information
Patient contacting materials: Seat PVC material, Hand grip PVC material, Seat belt PVC material
We tested these patient contacting materials for compliance to the standards ISO 10993-1:2009, ISO 10993-5:2009 and ISO 10993-10:2010. We provided Standard Data Report for 510(k) Forms (FDA Form 3654)).
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with a purple flower-like shape on a blue square. The flower has four petals and is positioned above the word "HEARTWAX" in bold, black letters. A thin, yellow line extends from the left side of the word upwards, ending in a small circle.
# HEARTWAY MEDICAL PRODUCTS CO.,LTD.
NO.6, ROAD 25, TAICHUNG INDUSTRIAL PARK, TAICHUNG, TAIWAN R.O.C. 408 TEL:886-4-23580357 (Sales) - 23583232 (Rep) FAX:886-4-23590786 Web : www.heartway.com.tw 60
E-mail : sales@heartway.com.tw
# TÜV-ISO-9001
CERTIFICATED P
## COMPARISON TABLE
# Similarities
| Comparison<br>items | Predicate device | Subject device | Safety and effectiveness<br>of subject device<br>compared to the<br>predicate device |
|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Manufacturer | HEARTWAY Medical Products Co., Ltd. | | Same manufacturer |
| Proprietary name | Lightweight Power<br>Mobility Scooter Series | Power Mobility<br>Scooter Series | Heavier weight design |
| Model | S34 | Cutie S17 | Different models |
| 510K Number | K101142 | K151656 | New listing |
| Common or<br>Usual name | Powered Scooter | Powered Scooter | Same common name |
| Intended use | The device is intended<br>for medical purposes to<br>provide mobility to<br>persons restricted to a<br>sitting position. | The device is intended<br>for medical purposes to<br>provide mobility to<br>persons restricted to a<br>sitting position. | Same intended use |
| Footplates | ABS | ABS | Same material |
| Incline | 10 degrees | 10 degrees | Same incline angle |
| Back upholstery | Fabric | Fabric | Same upholstery |
| Armrest types | Flip-backward | Flip-backward | Same armrest |
| Wheel Lock | Push-to-Lock | Push-to-Lock | Same wheel lock |
| Patient<br>contacting<br>material | Seat PVC material<br>Hand grip PVC material<br>Safety belt PVC material | Seat PVC material<br>Hand grip PVC material<br>Safety belt PVC<br>material | Same material |
| Biocompatibility | ISO 10993-1:2009<br>ISO 10993-5:2009<br>ISO 10993-10:2009 | ISO 10993-1:2009<br>ISO 10993-5:2009<br>ISO 10993-10:2010 | Same biocompatibility<br>evaluation |
| Comparison<br>items | Predicate device | Subject device | Safety and effectiveness<br>of subject device<br>compared to the<br>predicate device |
| Electronic<br>controller | Penny & Giles<br>S-Drive | Dynamic Rhino2<br>120 Amp | Different controllers |
| Frame<br>Type<br>Material | Fixed<br>Carbon Steel Pipe | Fixed<br>Aluminum Alloy pipe | Same type<br>Different material |
| Dimensions<br>Overall length<br>Overall width<br>Overall height | 39.0" (995 mm )<br>19.3" (490 mm)<br>37.0" (940 mm) | 51.9" (1320mm)<br>26.3" (670mm )<br>45.6" (1160mm) | Larger dimensions |
| Seat dimension<br>Seat width<br>Seat height | 22.5" (570 mm)<br>16" (408 mm) | 18.1" (460 mm)<br>18.1" (460 mm) | Smaller width<br>Larger height |
| Weight limit | 220 lbs (100 kgs) | 300 lbs (135 kgs) | Heavier user |
| Suspension | Cross brace | Full | |
| Rear wheels | 8" x 3"<br>solid type x 2 | 11" x 4"<br>pneumatic type x 2 | Larger size &<br>different type |
| Front Casters | 8" x 3"<br>solid x 2 | 11"x3.5"<br>pneumatic x2 | Larger size &<br>different type |
| Turning Radius | 37.4" (950 mm ) | 40.1" (1020 mm ) | Larger turning radius |
| Ground<br>clearance | 2.8" (70 mm ) | 3.93" (100 mm) | Larger ground<br>clearance |
| Kerb climbing<br>ability | 2.0" (50 mm ) | 3.93" (100 mm) | Larger kerb climbing<br>ability |
| HEARTWAY | | | |
| Scooter weight | w/ batteries: 96 lbs<br>(43.6 kgs)<br><br>w/o batteries: 81 lbs<br>(36.6 kgs) | w/ batteries: 247 lbs<br>(110 kgs)<br><br>w/o batteries: 190 lbs<br>(85 kgs) | Heavier scooter<br>weight |
| Maximum speed | 4.8 mph (7.8 kph) | 3.6 mph (5.8 kph) | Smaller max<br>speed |
| Recharger type<br>Model | 24VDC<br>(UL E241359 certified)<br><br>4C24020A | 24VDC<br>(E201162 certified)<br><br>4C24050A | Different models<br>with UL<br>certificates |
| Motor | 3A, 24V, 270W | 10A, 24V, 700W | Larger power<br>motor |
| Batteries<br>Quantity | Two | Two | Same quantity |
| Type | 12Ah 12VDC | 50Ah 12VDC | Larger capacity |
| Range per full<br>charging | 6~9 miles (10~15 km) | 21 miles (34 km) | Larger range |
| Warranty | 3 years: Main frame<br>1 years: Controller / gear<br>motor / batteries w/o<br>exhaustive and wear<br>parts | 2 years: Main frame<br>1.5 years: Controller<br>1 year gear motor<br>w/o exhaustive and<br>wear parts | Shorter period<br>warranty |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo for an organization called HEARTWAY. The logo features a purple butterfly-like shape against a blue square background. The word "HEARTWAY" is written in bold, black letters below the square. A thin yellow line is also visible to the left of the logo.
# HEARTWAY MEDICAL PRODUCTS CO.,LTD.
TUV-
CERTIFICATED
P
6
00
NO.6, ROAD 25, TAICHUNG INDUSTRIAL PARK, TAICHUNG, TAIWAN R.O.C. 408 TEL:886-4-23580357 (Sales) · 23583232 (Rep) FAX:886-4-23590786 Web : www.heartway.com.tw 100
E-mail : sales@heartway.com.tw
## Differences
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo with a purple flower-like design on a blue square. The flower has four petals and is centered on the square. Below the square, the word "HEARTWAY" is written in bold, black letters. The logo appears to be for a company or organization called Heartway.
# HEARTWAY MEDICAL PRODUCTS CO.,LTD.
NO.6, ROAD 25, TAICHUNG INDUSTRIAL PARK, TAICHUNG, TAIWAN R.O.C. 408 TEL:886-4-23580357 (Sales) - 23583232 (Rep) FAX:886-4-23590786 Web : www.heartway.com.tw 100
Image /page/9/Picture/3 description: The image shows two logos. The first logo is the "TÜV CERT" logo, which is a blue circle with the letters "TÜV" inside. The second logo is the "ISO-9001 CERTIFICATED" logo, which is a blue rounded rectangle with the text "ISO-9001 CERTIFICATED" inside. The logos are both blue and white.
6
00
E-mail : sales@heartway.com.tw
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a logo with the word "HEARTWAY" in bold, black letters. Above the word is a blue square with a stylized purple flower or butterfly shape inside. To the left of the word and square is a thin, vertical yellow line with a small yellow circle at the top.
Image /page/10/Picture/2 description: The image shows the words "COMPARISON DISCUSSION" in bold, uppercase letters. The words are underlined with a thick black line. The text is centered and takes up most of the image.
The intended uses for two devices are the same. Incline capability, footplates, armrest type, wheel lock and the biocompatibility evaluation are the same. The back upholstery material is also the same and passed the resistance ignition test in accordance with ISO 7176-16.
Mainframes of two devices are fixed type, and frame materials of the two devices are different but they both passed the strength and fatigue tests. The weight limit of HEARTWAY Lightweight Power Mobility Scooter, S34 is 220 lbs (100kgs) and the subject device HEARTWAY Power Mobility Scooter Cutie S17 is 300 lbs (135 kgs). In order to load a heavier user, the subject device must be equipped with larger seat height and larger front and rear wheels. In order to have a longer cruise range, the motor power and the battery capability must be larger. The larger seat height can thus load a larger battery volume. The larger front and rear wheels lead to larger turning radius, ground clearance, kerb climbing ability. Finally, larger seat height, larger front and rear wheels, bigger batteries add to a heavier weight of the power scooter. The pneumatic tires and independent full suspension for the subject device can absorb more impact than the solid tires and cross brace suspension for the predicate device, thus leading to more comfort. Under normal driving and maintenance conditions, pneumatic tires have the same safety and effectiveness concerns as the solid tires.
In a word, the subject device has a higher seat due to a heavier user load design and thus possesses a higher center of gravity, which leads to less stable capabilities and more tip over hazard when moving fast. So, the maximum speed of the subject device is limited to smaller value. It is 3.6 mph (5.8 kph) for subject device and 4.8 mph (7.8 kph) for predicate device. This limitation can bring more safety and reduce tip-over hazard. This higher seat difference is equalized by limiting the maximum speed.
The electronic controllers are different, but the subject device met the requirements of the software validation and passed the relevant performance tests of ISO 7176-14:2008. Thus, the subject device doesn't raise any safety and effectiveness concerns.
The chargers are different type. The charger for the subject device is UL-certified. Thus, the subject device doesn't raise any safety and effectiveness concerns.
{11}------------------------------------------------
The larger power capability of battery for the subject device leads to a larger travel range per full charged battery than the predicate device. This difference is related to the safety aspect, but brings more convenience.
The warranty periods are shorter. Even the periods are shorter, the users are required to follow the instructions for use in the owner's manual to drive the scooter. The differences in warranty periods may lead to more consuming money paid by the users, but not related to the safety and effectiveness aspects.
At last, subject devices pass the performance tests according to Safety standards of ISO 7176 series and the EMC test of ANSI / RESNA WC-2, Section 21:2009. The overall performances of the subject devices are ensured.
### CONCLUSIONS
The subject device, HEARTWAY Power Mobility Scooter Cutie S17, is as safe and effective as, and functions in a manner equivalent to the predicate device, HEARTWAY Lightweight Power Mobility Scooter S34. The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed device identified in the submission. Thus, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.