St.Jude Medical MediGuide Enabled Livewire Steerable Electrophysiology Catheter, Livewire Electrophysiology Catheter, Response Electrophysiology Catheter with Lumen, Response Electrophysiology Catheter and SJM Epicardial Catheter System, Supreme Electrophysiology Catheter
K151622 · St Jude Medical · DRF · Jul 15, 2015 · Cardiovascular
Device Facts
Record ID
K151622
Device Name
St.Jude Medical MediGuide Enabled Livewire Steerable Electrophysiology Catheter, Livewire Electrophysiology Catheter, Response Electrophysiology Catheter with Lumen, Response Electrophysiology Catheter and SJM Epicardial Catheter System, Supreme Electrophysiology Catheter
Applicant
St Jude Medical
Product Code
DRF · Cardiovascular
Decision Date
Jul 15, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
The SJM™ Livewire™ Diagnostic Catheter, MediGuide Enabled™ can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites. The Livewire™ Diagnostic Catheter, MediGuide Enabled™ is compatible with MediGuide™ Technology to enable real-time tip positioning and navigation. The MediGuide™ Technology is indicated for use as an adjunct to fluoroscopy.
Device Story
Electrophysiology catheters used for cardiac arrhythmia evaluation; endocardial/intravascular sites. MediGuide-enabled versions include passive sensor at tip for real-time 3D positioning/navigation; tracked by gMPS system; superimposed on 2D X-ray fluoroscopy. Catheters transmit electrical signals to ECG recorders or deliver pacing signals from external pacemakers. Used by physicians in clinical settings. New manufacturing process for outer insulation alters surface morphology; design verification confirms safety/effectiveness.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993; design verification testing performed per ISO 10555. Results confirmed that catheters with modified surface morphology met all product performance specifications and maintained acceptable biocompatibility profiles.
Technological Characteristics
Diagnostic electrophysiology catheters; polyurethane shaft/insulation; platinum electrodes. MediGuide-enabled models include passive sensor for electromagnetic tracking. Dimensions/form factors vary by model (steerable, fixed curve, lumen/no-lumen). Connectivity: passive sensor interfaces with gMPS system. Sterilization: not specified. Standards: ISO 10993 (biocompatibility), ISO 10555 (catheter performance).
Indications for Use
Indicated for patients requiring evaluation of cardiac arrhythmias via endocardial and intravascular sites. Compatible with MediGuide technology for real-time tip positioning and navigation as an adjunct to fluoroscopy.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 15, 2015
St. Jude Medical Maria Obreztchikova Sr. Regulatory Affairs Specialist One St. Jude Medical Drive St. Paul, MN 55117
Re: K151622
Trade/Device Name: MediGuide Enabled Livewire Steerable Electrophysiology Catheter, Livewire Steerable Electrophysiology Catheter, Response Electrophysiology Catheter with Lumen, Response Electrophysiology Catheter, SJM Epicardial Catheter System. Supreme Electrophysiology Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: DRF Dated: June 15, 2015
Received: June 16, 2015
Dear Maria Obreztchikova:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized symbol. The symbol consists of three abstract human profiles facing right, with flowing lines beneath them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known):
Device Name: MediGuide Enabled Livewire Steerable Electrophysiology Catheter
Indications for Use:
- The SJMTM Livewire™ Diagnostic Catheter, MediGuide Enabled™ can be used in the ● evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.
- The Livewire™ Diagnostic Catheter, MediGuide Enabled™ is compatible with ● MediGuide™ Technology to enable real-time tip positioning and navigation.
- The MediGuide™ Technology is indicated for use as an adjunct to fluoroscopy. ●
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{3}------------------------------------------------
510(k) Number (if known): _ _ _ _ _ _ _ _ _ K151622
Device Name: Livewire™ Steerable Electrophysiology Catheter
Indications for Use:
The Livewire™ Steerable Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{4}------------------------------------------------
510(k) Number (if known): _
Device Name: Response™ Electrophysiology Catheter with Lumen
Indications for Use:
St Jude Medical (SJM) Electrophysiology Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{5}------------------------------------------------
510(k) Number (if known): _
Device Name: Response™ Electrophysiology Catheter
Indications for Use:
St Jude Medical (SJM) Electrophysiology Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{6}------------------------------------------------
510(k) Number (if known): _
Device Name: SJMTM Epicardial Catheter System
Indications for Use:
St Jude Medical (SJM) Epicardial Catheter System is intended for electrophysiology studies involving the epicardial surface of the heart.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _____________ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{7}------------------------------------------------
510(k) Number (if known): _
Device Name: Supreme™ Electrophysiology Catheter
Indications for Use:
St Jude Medical (SJM) Electrophysiology Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{8}------------------------------------------------
| 510(k) Summary | |
|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K151622 |
| Submitter Information: | |
| Date Prepared: | 15 June 2015 |
| Submitter Name &<br>Address: | St. Jude Medical<br>14901 DeVeau Place<br>Minnetonka, MN 55345<br>Establishment Registration Number: 3005188751 |
| Contact Person: | Maria Obreztchikova<br>Senior Regulatory Affairs Specialist<br>Phone (651) 756-4424<br>Fax (651) 756-3298<br>mobreztchikova@sjm.com |
| Device Information: | |
| Trade Name: | MediGuide Enabled Livewire Steerable Electrophysiology Catheter |
| Common Name: | Diagnostic Electrophysiology Catheter |
| Classification Name: | Electrode recording catheter or electrode recording probe |
| Class: | Class II, 21 CFR 870.1220, Product Code DRF |
| Predicate Device: | MediGuide Enabled Livewire Steerable Electrophysiology Catheter<br>(K101955, K150631) |
| Device Description: | The MediGuide Enabled Livewire Steerable Electrophysiology<br>Catheter is designed to allow electrophysiological mapping at<br>various endocardial and intravascular sites including the inferior<br>vena cava, superior vena cava, pulmonary veins and coronary sinus.<br><br>The use of the MediGuide Enabled Livewire diagnostic mapping<br>catheter in conjunction with the MediGuide gMPS system allows<br>real-time tip positioning and navigation during routine diagnostic<br>evaluation.<br><br>When connected to the ECG, the MediGuide Enabled Livewire<br>Catheter will transmit routine electrical signals while the passive<br>sensor at the tip of the catheter is tracked by gMPS and is<br>superimposed on any 2D X-ray image in real time (“Live”)<br>fluoroscopy mode (the actual 3D position of the catheter tip is<br>projected on the real time 2D image). |
| Intended Use:<br>(Indications for Use) | • The SJM™ Livewire™ Diagnostic Catheter, MediGuide<br>Enabled™ can be used in the evaluation of a variety of cardiac<br>arrhythmias from endocardial and intravascular sites.<br>• The Livewire™ Diagnostic Catheter, MediGuide Enabled™ is |
| compatible with MediGuide™ Technology to enable real-time<br>tip positioning and navigation. | |
| • The MediGuide™ Technology is indicated for use as an adjunct<br>to fluoroscopy. | |
| Comparison to<br>Predicate Devices | The MediGuide Enabled Livewire Steerable Electrophysiology<br>Catheter that is the subject of this application remains substantially<br>equivalent to the predicate device. There have been no changes to<br>the device materials, packaging, sterilization or labeling. The<br>proposed catheter includes a new outer insulation manufacturing<br>process that results in a different surface morphology than the<br>predicate catheter. The biocompatibility and design verification<br>(DV) testing demonstrated that the surface morphology modification<br>to the outer layer of the catheter proximal shaft does not adversely<br>affect the device safety and effectiveness. |
| Summary on Non-<br>Clinical Testing | Biocompatibility testing was performed according to ISO 10993.<br>The catheters with the modified surface morphology of the proximal<br>shaft demonstrated acceptable biocompatibility profile.<br>DV testing was performed according to ISO 10555. The catheters<br>with the modified surface morphology of the proximal shaft met<br>product performance specifications. |
| Statement of<br>Equivalence | The MediGuide Enabled Livewire Steerable Electrophysiology<br>Catheter that is the subject of this application has the same<br>indications for use and technological characteristics as the predicate<br>device. Based on this and the data provided in this pre-market<br>notification, the subject device remains substantially equivalent to<br>the predicate device. |
| Device Information: | |
| Trade Name: | Livewire™ Steerable Electrophysiology Catheter |
| Common Name: | Diagnostic Electrophysiology Catheter |
| Classification Name: | Electrode recording catheter or electrode recording probe |
| Class: | Class II, 21 CFR 870.1220, Product Code DRF |
| Predicate Device: | Livewire™ Steerable Electrophysiology Catheter (K913940,<br>K022380, K102721, K150631) |
| Device Description: | The SJM Livewire™ Steerable Electrophysiology Catheter is a<br>flexible electrode catheter constructed of a polyurethane<br>insulation/shaft and incorporates platinum electrodes. The active tip<br>may be manipulated by a remote means located at the proximal end<br>of the catheter. |
| Intended Use:<br>(Indications for Use) | The Livewire™ Steerable Electrophysiology Catheter can be used in<br>the evaluation of a variety of cardiac arrhythmias from endocardial<br>and intravascular sites. |
| Comparison to<br>Predicate Devices | The Livewire Steerable Electrophysiology Catheter that is the<br>subject of this application remains substantially equivalent to the<br>predicate device. There have been no changes to the device<br>materials, packaging, sterilization or labeling. The proposed catheter<br>includes a new outer insulation manufacturing process that results in<br>a different surface morphology than the predicate catheter. The<br>biocompatibility and design verification (DV) testing demonstrated<br>that the surface morphology modification to the outer layer of the<br>catheter proximal shaft does not adversely affect the device safety<br>and effectiveness. |
| Summary on Non-<br>Clinical Testing | Biocompatibility testing was performed according to ISO 10993.<br>The catheters with the modified surface morphology of the proximal<br>shaft demonstrated acceptable biocompatibility profile.<br>DV testing was performed according to ISO 10555. The catheters<br>with the modified surface morphology of the proximal shaft met<br>product performance specifications. |
| Statement of<br>Equivalence | The Livewire Steerable Electrophysiology Catheter that is the<br>subject of this application has the same indications for use and<br>technological characteristics as the predicate device. Based on this<br>and the data provided in this pre-market notification, the subject<br>device remains substantially equivalent to the predicate device. |
| Device Information: | |
| Trade Name: | Response™ Electrophysiology Catheter with Lumen |
| Common Name: | Diagnostic Electrophysiology Catheter |
| Classification Name: | Electrode recording catheter or electrode recording probe |
| Class: | Class II, 21 CFR 870.1220, Product Code DRF |
| Predicate Device: | Response™ Electrophysiology Catheter with Lumen (K914278,<br>K942379, K120544) |
| Device Description: | The Response™ Electrophysiology Catheter with Lumen is a sterile,<br>single use, electrophysiological diagnostic catheter with a fixed<br>distal curve and multiple electrodes. The electrodes detect cardiac<br>electrical signals that are transmitted to an electrocardiogram<br>recorder for diagnosis or deliver the electrical signal from an<br>external pacemaker to the heart for electrical pacing purposes. The<br>catheter body is a continuous tube with a central lumen for fluid<br>infusion through a 3-way valve. |
| Intended Use:<br>(Indications for Use) | St Jude Medical (SJM) Electrophysiology Catheters can be used in<br>the evaluation of a variety of cardiac arrhythmias from endocardial<br>and intravascular sites. |
| Comparison to<br>Predicate Devices | The Response™ Electrophysiology Catheter with Lumen that is the<br>subject of this application remains substantially equivalent to the |
| | predicate device. There have been no changes to the device<br>materials, packaging, sterilization or labeling. The proposed catheter<br>includes a new outer insulation manufacturing process that results in<br>a different surface morphology than the predicate catheter. The<br>biocompatibility and design verification (DV) testing demonstrated<br>that the surface morphology modification to the outer layer of the<br>catheter proximal shaft does not adversely affect the device safety<br>and effectiveness. |
| Summary on Non-<br>Clinical Testing | Biocompatibility testing was performed according to ISO 10993.<br>The catheters with the modified surface morphology of the proximal<br>shaft demonstrated acceptable biocompatibility profile.<br>DV testing was performed according to ISO 10555. The catheters<br>with the modified surface morphology of the proximal shaft met<br>product performance specifications. |
| Statement of<br>Equivalence | The Response™ Electrophysiology Catheter with Lumen that is the<br>subject of this application has the same indications for use and<br>technological characteristics as the predicate device. Based on this<br>and the data provided in this pre-market notification, the subject<br>device remains substantially equivalent to the predicate device. |
| Device Information: | |
| Trade Name: | Response ™ Electrophysiology Catheter |
| Common Name: | Diagnostic Electrophysiology Catheter |
| Classification Name: | Electrode recording catheter or electrode recording probe |
| Class: | Class II, 21 CFR 870.1220, Product Code DRF |
| Predicate Device: | Response™ Electrophysiology Catheter (K894500, K002976) |
| Device Description: | The Response™ Electrophysiology Catheter is a sterile, single use,<br>electrophysiological diagnostic catheter. The catheter has fixed<br>distal curve available in various styles, no lumen, and varying<br>electrode numbers and spacing. The electrodes detect cardiac<br>electrical signals that are transmitted to an electrocardiogram<br>recorder for diagnosis or deliver the electrical signal from an<br>external pacemaker to the heart for electrical pacing purposes. |
| Intended Use:<br>(Indications for Use) | St Jude Medical (SJM) Electrophysiology Catheters can be used in<br>the evaluation of a variety of cardiac arrhythmias from endocardial<br>and intravascular sites. |
| Comparison to<br>Predicate Devices | The Response™ Electrophysiology Catheter that is the subject of<br>this application remains substantially equivalent to the predicate<br>device. There have been no changes to the device materials,<br>packaging, sterilization or labeling. The proposed catheter includes a<br>new outer insulation manufacturing process that results in a different<br>surface morphology than the predicate catheter The |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
| Summary on Non-Clinical Testing | biocompatibility and design verification (DV) testing demonstrated that the surface morphology modification to the catheter outer shaft does not adversely affect the device safety and effectiveness.<br>Biocompatibility testing was performed according to ISO 10993. The catheters with the modified surface morphology of the proximal shaft demonstrated acceptable biocompatibility profile.<br>DV testing was performed according to ISO 10555. The catheters with the modified surface morphology of the proximal shaft met product performance specifications. |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Statement of Equivalence | The Response™ Electrophysiology Catheter with Lumen that is the subject of this application has the same indications for use and technological characteristics as the predicate device. Based on this and the data provided in this pre-market notification, the subject device remains substantially equivalent to the predicate device. |
| Device Information: | |
| Trade Name: | SJM™ Epicardial Catheter System |
| Common Name: | Epicardial Catheter System |
| Classification Name: | Electrode recording catheter or electrode recording probe |
| Class: | Class II, 21 CFR 870.1220, Product Code DRF |
| Predicate Device: | SJM™ Epicardial Catheter System (K081803)<br>Response™ Electrophysiology Catheter (K894500) |
| Device Description: | The SJM™ Epicardial Catheter System consists of the following SJM components:<br>• 17 Gauge Tuohy Needle, 11.4 cm<br>• Response™ Electrophysiology Catheter, 6F<br>• Agilis™ NxT Steerable Introducer, 40 cm (includes steerable sheath, dilator, and guidewire)<br><br>The Response™ Electrophysiology Catheter, 6F, is designed to provide pacing and recording capabilities during electrophysiology studies involving the epicardial surface of the heart. The distal tip of the catheter contains four platinum electrodes. It is curved to facilitate greater reach when advanced out of the steerable introducer. |
| Intended Use:<br>(Indications for Use) | The SJM™ Epicardial Catheter System is intended for electrophysiology studies involving the epicardial surface of the heart. |
| Comparison to<br>Predicate Devices | The Epicardial Catheter System that is the subject of this application remains substantially equivalent to the predicate device. There have been no changes to the device materials, packaging, sterilization or |
{13}------------------------------------------------
| Summary on Non-Clinical Testing | Biocompatibility testing was performed according to ISO 10993.<br>The catheters with the modified surface morphology of the proximal<br>shaft demonstrated acceptable biocompatibility profile.<br>DV testing was performed according to ISO 10555. The catheters<br>with the modified surface morphology of the proximal shaft met<br>product performance specifications.<br>labeling. The proposed catheter includes a new outer insulation<br>manufacturing process that results in a different surface morphology<br>than the predicate catheter. The biocompatibility and design<br>verification (DV) testing demonstrated that the surface morphology<br>modification to the Response catheter outer shaft does not adversely<br>affect the device safety and effectiveness. |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Statement of<br>Equivalence | The SJM™ Epicardial Catheter System that is the subject of this<br>application has the same indications for use and technological<br>characteristics as the predicate device. Based on this and the data<br>provided in this pre-market notification, the subject device remains<br>substantially equivalent to the predicate device. |
| Device Information | |
| Trade Name: | Supreme™ Electrophysiology Catheter |
| Common Name: | Diagnostic Electrophysiology Catheter |
| Classification Name: | Electrode recording catheter or electrode recording probe |
| Class: | Class II, 21 CFR 870.1220, Product Code DRF |
| Predicate Device: | Supreme™ Electrophysiology Catheter (K894500, K002976) |
| Device Description: | The Supreme™ Electrophysiology Catheter is a sterile, single use,<br>electrophysiological diagnostic catheter. The catheter has fixed<br>distal curve available in various styles, no lumen, and varying<br>electrode numbers and spacing. The electrodes detect cardiac<br>electrical signals that are transmitted to an electrocardiogram<br>recorder for diagnosis or deliver the electrical signal from an<br>external pacemaker to the heart for electrical pacing purposes. |
| Intended Use:<br>(Indications for Use) | St Jude Medical (SJM) Electrophysiology Catheters can be used in<br>the evaluation of a variety of cardiac arrhythmias from endocardial<br>and intravascular sites. |
| Comparison to<br>Predicate Devices | The Supreme™ Electrophysiology Catheter that is the subject of this<br>application remains substantially equivalent to the predicate device.<br>There have been no changes to the device materials, packaging,<br>sterilization or labeling. The proposed catheter includes a new outer<br>insulation manufacturing process that results in a different surface<br>morphology than the predicate catheter. The biocompatibility and<br>design verification (DV) testing demonstrated that the surface<br>morphology modification to the catheter outer shaft does not |
{14}------------------------------------------------
| Summary on Non-Clinical Testing | Biocompatibility testing was performed according to ISO 10993. The catheters with the modified surface morphology of the proximal shaft demonstrated acceptable biocompatibility profile. DV testing was performed according to ISO 10555. The catheters with the modified surface morphology of the proximal shaft met product performance specifications. adversely affect the device safety and effectiveness. |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Statement of Equivalence | The Supreme™ Electrophysiology Catheter that is the subject of this application has the same indications for use and technological characteristics as the predicate device. Based on this and the data provided in this pre-market notification, the subject device remains substantially equivalent to the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.