← Product Code [GCJ](/productcode/GCJ) · K151609

# ZeroWire Duo Wireless HD Video Transfer System G2 or ZeroWire G2 (K151609)

_Nds Surgical Imaging, LLC · GCJ · Nov 6, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151609

## Device Facts

- **Applicant:** Nds Surgical Imaging, LLC
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Nov 6, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The NDS ZeroWire G2 is a paired wireless video communication transmitter and receiver, intended for delivery of video signals from a source such as an endoscopy camera/processor, or other video source over a radio-frequency link to a ZeroWire Receiver for display of images during endoscopic and general surgical procedures. The ZeroWire G2 wireless video system is a non-sterile reusable device not intended for use in the sterile field.

## Device Story

ZeroWire G2 is a paired wireless video transmitter and receiver system; transmits video signals from endoscopy cameras or other sources to a display via 60 GHz radio-frequency link. Utilizes directional antenna and beam forming to minimize interference; proprietary memory-enabled pairing system facilitates setup. Used in operating rooms and surgical environments; operated by surgical staff. Eliminates video cables to enhance clinical efficiency and safety. Output is a high-definition video feed for real-time visualization during minimally invasive surgery and interventional procedures; assists clinicians in surgical decision-making by providing reliable, high-quality imaging.

## Clinical Evidence

No clinical data required per 510(k) guidance for this device type. Manufacturer provided test results from a clinical environment to demonstrate substantial equivalence to the predicate device.

## Technological Characteristics

Wireless video transfer system; 60 GHz frequency spectrum; directional antenna; beam forming technology; proprietary memory-enabled pairing. Complies with IEC 60601-1, ANSI/AAMI ES60601-1, IEC 60601-1-2, and FCC Part 15. Non-sterile, reusable.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- WIS1000 ([K112621](/device/K112621.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are arranged in a staggered formation, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

NDS Surgical Imaging, LLC Mr. Jim Leng Regulatory Engineer 5750 Hellyer Avenue San Jose, California 95138

November 6, 2015

Re: K151609 Trade/Device Name: ZeroWire Duo Wireless HD Video Transfer System G2 or ZeroWire G2 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: October 26, 2015 Received: October 28, 2015

Dear Mr. Leng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice

{1}------------------------------------------------

requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joshua C. Nipper -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K151609

## Device Name

ZeroWire Duo Wireless HD Video Transfer System G2 or ZeroWire G2

Indications for Use (Describe)

The NDS ZeroWire G2 is a paired wireless video communication transmitter and receiver, intended for delivery of video signals from a source such as an endoscopy camera/processor, or other video source over a radio-frequency link to a ZeroWire Receiver for display of images during endoscopic and general surgical procedures. The ZeroWire G2 wireless video system is a non-sterile reusable device not intended for use in the sterile field.

| <span style="font-size:10pt;">☒ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
  
Food and Drug Administration
  
Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for NDS Surgical Imaging. The logo consists of a grid of blue squares on the left, followed by the letters "NDS" in large, blue font. Below the letters, the words "SURGICAL・IMAGING" are written in a smaller, gray font.

## 510(K) SUMMARY

| A. Manufacturer:            | NDS Surgical Imaging, LLC<br>5750 Hellyer Avenue<br>San Jose, CA 95138<br>USA                                                                                                                                                                                                                                                |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B. Submitted By:            | Jim Leng<br>Regulatory Engineer/NDS Surgical Imaging, LLC                                                                                                                                                                                                                                                                    |
| B1, Address:                | NDS Surgical Imaging, LLC<br>5750 Hellyer Avenue<br>San Jose, CA 95138<br>USA                                                                                                                                                                                                                                                |
| C. Date of Preparation:     | Updated October 22, 2015                                                                                                                                                                                                                                                                                                     |
| D. Contact Information:     | Tel: 408-776-0085<br>Fax: 408-776-9878                                                                                                                                                                                                                                                                                       |
| E. Classification:          | Endoscope and Accessories                                                                                                                                                                                                                                                                                                    |
| F. Common Name:             | Wireless Device                                                                                                                                                                                                                                                                                                              |
| G. Proprietary Name:        | ZeroWire G2 Duo Wireless HD Video Transfer System or<br>ZeroWire G2                                                                                                                                                                                                                                                          |
| H. Classification number:   | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                              |
| I Product Code:             | GCJ                                                                                                                                                                                                                                                                                                                          |
| J. Substantial Equivalence: | Predicate device K112621 model WIS1000                                                                                                                                                                                                                                                                                       |
| K. Device Description:      | Model Wu-vwx-yz, where u = R or T; v=P or M; w=1 to 4;<br>x= 1 to 6; y= 0 to 2; z=1 to 3, the ZeroWire G2 wireless<br>device is designed as a wireless transmitter and receiver<br>pair which allows delivery of a video signal over a radio<br>frequency link to video destination such as Radiance and<br>EndoVue devices. |
| L. Indications for Use:     | The NDS ZeroWire G2 is a paired wireless video<br>communication transmitter and receiver, intended for<br>delivery of video signals from a source such as an<br>endoscopy camera/processor, or other video source over a                                                                                                     |

Image /page/3/Picture/5 description: The image contains two logos. The first logo is for Radiance, which includes a blue horizontal line, a circular symbol with an arrow, and the word "RADIANCE" in black, with the letters "HD" in blue above it. The second logo is for Dome, with the word "DOME" in black and a blue square above it.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for NDS Surgical Imaging. The logo features a grid of blue squares on the left, followed by the letters "NDS" in large, bold blue font. Below the letters, the words "SURGICAL - IMAGING" are written in a smaller, gray font.

radio-frequency link to a ZeroWire Receiver for display of images during endoscopic and general surgical procedures. The ZeroWire G2 wireless video system is a non-sterile reusable device not intended for use in the sterile field.

| M. Technological Characteristics: | ZeroWire G2 devices are the most advance medical grade<br>wireless video transfer solution for minimally invasive<br>surgery and interventional procedures. By utilizing<br>directional antenna and beam forming in the 60 GHz<br>+frequency spectrum, ZeroWire G2 provides a robust<br>wireless video link minimize the interference from other<br>devices. It enhances clinical efficiency and safety in the OF<br>by eliminating the need for a video cable. The proprietary<br>memory-enabled pairing system makes installation quick<br>and easy. ZeroWire G2 technology provides the highest<br>quality of service and is specifically designed for the video<br>transmission challenges of the surgical environment. |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| N. Performance:                   | Based upon our testing, the ZeroWire G2 meets and<br>exceeds IEC 60601-1, ANSI/AAMI ES60601-1, IEC<br>60601-1-2 and FCC part 15 which establish its safe design<br>and operation. The device had gone through its design<br>validation and clinical study to further demonstrate its<br>safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                           |
| O. Clinical Information:          | Clinical data is not needed for this type of wireless device<br>by 510(k) submission guidance document. However, this<br>submission provides test results in clinical environment to<br>demonstrate the device is substantial equivalent to the<br>predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| P. Conclusion:                    | Based upon validations results from both design and<br>clinical, ZeroWire G2 device demonstrates performance,<br>safety, and effectiveness that are equivalent to the predicate<br>device – Predicate device K112621 model WIS1000.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

Image /page/4/Picture/5 description: The image contains two logos. The first logo is for "RADIANCE" with the letters in bold, black font and a light blue underline. The second logo is a blue square with the letters "DOL" in bold, white font.

---

**Source:** [https://fda-staging.innolitics.com/device/K151609](https://fda-staging.innolitics.com/device/K151609)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
