On Call Express Pro Blood Glucose Monitoring System
Applicant
ACON Laboratories, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Sep 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The On Call Express Pro Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips, forearm and palm. The On Call Express Pro Blood Glucose Monitoring System is intended for multiple patient use by health care professionals in health care facilities as an aid to monitoring the effectiveness of diabetes control programs. The system should only be used with single-use, auto disabling lancing devices. The On Call Express Pro Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, nor intended for use on neonates. The On Call Express Pro Blood Glucose Test Strips are used with the On Call Express Pro Blood Glucose Meter in the quantitative measurement of glucose in capillary blood from the fingertips, forearm and palm. The On Call Express Pro Blood Glucose Control Solution is for use with the On Call® Express Pro Blood Glucose Meter and Strips as a quality control check to verify the accuracy of blood glucose test results.
Device Story
System measures glucose in capillary whole blood via amperometric detection; uses FAD-glucose oxidase reagent on test strips. Meter calculates concentration from electrical current; displays plasma-equivalent results. Designed for multiple-patient use in healthcare facilities; operated by healthcare professionals. Requires single-use, auto-disabling lancing devices. Includes data port for optional cable transfer to Windows-based PC using On Call Diabetes Management Software. Provides quantitative data to assist clinicians in monitoring diabetes control programs. Meter features 300-record memory, 4-second test time, and automatic shutoff. Cleaning/disinfection required between patients.
Clinical Evidence
No new clinical studies performed. Performance characteristics, including precision, linearity, analytical specificity, and method comparison, were established in predicate k132086. Device validated for multiple-patient use via disinfection efficacy studies demonstrating complete inactivation of HBV and robustness testing simulating 3 years of use (10,950 cycles).
Technological Characteristics
Amperometric glucose monitoring system. Materials: FAD-glucose oxidase reagent. Power: 3.0V CR2032 coin cell. Dimensions: 3.46" x 1.93" x 0.65". Connectivity: Serial data port (9600 baud). Calibration: Plasma-equivalent, non-coding. Operating range: 20-600 mg/dL; 30-55% hematocrit. Software: Embedded firmware for signal processing and data management.
Indications for Use
Indicated for quantitative glucose measurement in fresh capillary whole blood from fingertips, forearm, and palm in patients with diabetes to monitor control program effectiveness. For multiple-patient use by healthcare professionals in clinical settings. Contraindicated for neonates, patients with dehydration, hypotension, shock, critical illness, hyper-osmolar state, or those undergoing oxygen therapy. Not for diabetes diagnosis or screening.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
On Call Express Blood Glucose Monitoring System (k132086)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k151595
B. Purpose for Submission:
Modified Device – modifications were made to the materials of the meter, the name of the test system and the intended use (for use on multiple patients).
C. Measurand:
Capillary whole blood glucose from the finger, palm and forearm
D. Type of Test:
Quantitative Amperometric assay (Glucose Oxidase)
E. Applicant:
ACON Laboratories, Inc.
F. Proprietary and Established Names:
On Call Express Pro Blood Glucose Monitoring System
G. Regulatory Information:
| Regulation Section | Classification | Product Code | Panel |
| --- | --- | --- | --- |
| 21 CFR 862.1345 | Class II | CGA, Glucose Oxidase, | Clinical Chemistry (75) |
| 21 CFR 862.1345 | Class II | NBW, System, Test, Blood Glucose, Over the Counter | Clinical Chemistry (75) |
| 21 CFR 862.1660 | Class I (exempt) | JJX, Quality Control Material (assayed and unassayed) | Clinical Chemistry (75) |
H. Intended Use:
1. Intended use(s):
See indication(s) for use below.
2. Indication(s) for use:
The On Call Express Pro Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the
{1}
fingertips, forearm and palm. The On Call Express Pro Blood Glucose Monitoring System is intended for multiple patient use by health care professionals in health care facilities as an aid to monitoring the effectiveness of diabetes control programs. The system should only be used with single-use, auto disabling lancing devices.
The On Call Express Pro Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, nor intended for use on neonates.
The On Call Express Pro Blood Glucose Test Strips are used with the On Call Express Pro Blood Glucose Meter in the quantitative measurement of glucose in capillary blood from the fingertips, forearm and palm.
The On Call Express Pro Blood Glucose Control Solution is for use with the On Call® Express Pro Blood Glucose Meter and Strips as a quality control check to verify the accuracy of blood glucose test results.
3. Special conditions for use statement(s):
- For in vitro diagnostic use
- For multiple patient use
- Not intended for use on neonates
- Not for diagnosis of or screening of diabetes mellitus
- Not to be used for patients who are dehydrated, hypotensive, in shock critically ill or in a hyper-osmolar state
- Meter should be cleaned and disinfected after use on each patient
- System should only be used with single-use, auto-disabling lancing devices
- Alternative site testing (AST) testing should only be done during steady-state times (when glucose is not changing rapidly).
- AST should not be used to calibrate continuous glucose monitors (CGMs)
- AST should not be used for insulin dose calculations
- Not for persons undergoing Oxygen therapy
- For fresh capillary whole blood, not for use with serum or plasma samples.
4. Special instrument requirements:
On Call Express Pro glucose meter
I. Device Description:
The On Call Express Pro Blood Glucose Monitoring System consists of the On Call Express Pro Blood Glucose Meter, On Call Express Pro Blood Glucose Test Strips and 3 levels of On Call Express Pro Glucose Control Solutions (Control Solution 0, Control Solution 1 and Control Solution 2), Carrying Case and User's Manual and single-use, auto-disabling lancing devices (On Call Safety Lancets). Test Strips, Control Solutions and single-use, auto-disabling lancing devices are required but not provided with the Meter and must be purchased separately.
{2}
The On Call Express Pro Blood Glucose Monitoring System that is the subject of this submission uses identical technology, software, measurement algorithm, test strips, and quality control solutions as the predicate device, the On Call Express Blood Glucose Monitoring System, which was cleared in k132086. The only differences between this device and the predicate are the name, labeling (for multiple patient use by health care professionals) and the materials of the meter.
Test strips are packaged in vials of 50 and also available individually packaged. Each test strip contains the following reagent compositions: Glucose Oxidase (from Aspergillus niger).
Each box of control solutions (Control Solution 0, Control Solution1, Control Solution 2) contains one vial of each control solution level (2.0 mL per vial) of viscosity-adjusted, buffered aqueous control solutions that contain known concentrations of d-glucose. Control solutions were previously cleared under k132086 as the On Call Express Blood Glucose Control Solutions and are being renamed under this submission.
## J. Substantial Equivalence Information:
1. Predicate device name(s): On Call Express Blood Glucose Monitoring System
2. Predicate 510(k) number(s): k132086
3. Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate Device
On Call Express Glucose Monitoring System
(k132086) | Candidate Device
On Call Express Pro Blood Glucose Monitoring System
(k151595) |
| Intended Use/Indications for Use | It is intended to be used for quantitative measurement of glucose in fresh capillary whole blood from the fingertips, forearm and palm as an aid to monitor the effectiveness of diabetes control in people with diabetes. | It is intended to be used for quantitative measurement of glucose in fresh capillary whole blood from the fingertips, forearm and palm as an aid to monitor the effectiveness of diabetes control in people with diabetes. |
| Use Setting | At home for single patient use only | Multiple patient use in a clinical setting. |
{3}
| Detection method | Amperometry | Same |
| --- | --- | --- |
| Enzyme | Glucose Oxidase | Same |
| Calibration Coding | Non-Coding | Same |
| Memory | 300 records with time and date | Same |
| Test range | 20 - 600 mg/dL | Same |
| Sample type | Capillary whole blood | Same |
| Sample sites | Fingertip, forearm, palm | Same |
| Sample volume | 0.4 μL | Same |
| Sample test time | 4 seconds | Same |
| Hematocrit range | 30-55% | Same |
| Altitude Study | Up to 8,516 feet | Same |
| Glucose Units of Measurement | The meter is preset to mg/dL when sold in the United States. | Same |
| Operating Temperature | 41-113°F (5-45°C) | Same |
| Operating Relative Humidity | 10-90% (non-condensing) | Same |
| Power Source | One (1) CR 2032 3.0V coin cell battery | Same |
| Meter Size | 3.46” x 1.93” x 0.65” | Same |
| Meter Weight | Approximately 50 g (with battery installed) | Same |
| Similarities and Differences for Control Solutions | | |
| Item | Predicate Device On Call Express Glucose Control Solution (k132086) | Candidate Device On Call Express Pro Control Solutions (k151595) |
| Intended use/Indications for Use | To check that the glucose meter and test strips are working together properly | Same |
| Matrix | Viscosity-adjusted, aqueous liquid | Same |
| Number of levels | 3 levels (Control Solution 0, Control Solution 1 and Control Solution 2) | Same |
{4}
K. Standard/Guidance Document Referenced:
- EN 1113 7-1: 2006 Sterilization of health care products. Radiation. Requirements for development, validation and routine control of a sterilization process for medical devices
- EN 11137-2: 2007 Sterilization of health care products. Radiation. Part 2: Establishing the sterilization dose
- EN 11137-3: 2006 Sterilization of health care products. Radiation. Guidance on dosimetric aspects
- EN 556-1:2001/AC: 2006 Sterilization of medical devices. Requirements for medical devices to be designated "STERILE". Requirements for terminally sterilized medical devices
- EN 1173 7-1: 2006 Sterilization of medical devices. Microbiological methods. Determination of a population of microorganisms on products
- EN 11737-2: 2000 Sterilization of medical devices. Microbiological methods. Tests of sterility performed in the validation of a sterilization process
- EN 11607-1: 2006 Packaging for terminally sterilized medical devices. Requirements for materials, sterile barrier systems and packaging systems
- EN 10993-5: 2009 Biological evaluation of medical devices. Tests for in vitro cytotoxicity
- EN 10993-10: 2009 Biological evaluation of medical devices. Tests for irritation and delayed-type hypersensitivity
- EN 61326-1: 2006 Class B Electrical Equipment for Measurement, Control and Laboratory Use- EMC Requirements
- EN 61326-2-6: 2006 Electrical Equipment for Measurement, Control and Laboratory Use. Particular requirements. In vitro diagnostic (IVD) medical equipment
- IEC/EN 61010-1: 2001 Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use. Part 1: General Requirements
- IEC/EN 61010-2-10: 2002 Safety requirements for electrical equipment for measurement, control and laboratory use. Particular requirements for in vitro diagnostic (IVD) medical equipment
- EN ISO 14971:2009 Medical devices - Application of Risk management to medical devices
- ISO 13640: 2002 Stability testing of in vitro diagnostic reagents
- EN 62366: 2008 Medical devices. Application of usability engineering to medical devices
- EN 62304: 2006 Medical device software. Software life-cycle processes
- CLSI EP-6A Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
- CLSI EP7-A2 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition
- ISO GP14-A: 1996 Labeling of Home-Use In Vitro Testing Products; Approved Guideline
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- General Principles of Software Validation; Final Guidance for Industry and FDA Staff: January 2002
{5}
- Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing' (Replaces #G87 -1 #8294) (blue book memo)(Text Only)
## L. Test Principle:
The test is based on electrochemical biosensor technology and the principle of capillary action. A blood sample is drawn into the test strip by capillary action, and reacts with reagents on the test strip. This reaction produces an electrical current which is proportional to the concentration of glucose in the samples. The electrical current is measured by the meter and is displayed to the user as a corresponding blood glucose level.
## M. Performance Characteristics:
Technological characteristics of this device are unchanged relative to the predicate device (k132086). Performance characteristics were established in k132086 and as described below.
### a. Precision/Reproducibility
Within-run and intermediate precision were established in k132086.
### b. Linearity/assay reportable range:
Linearity and reportable range were established in k132086.
### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
#### Traceability
Traceability is as described in k132086.
#### Stability and Expected Values
Value assignment:
Value assignment for the On Call Express Blood Glucose Control Solutions is as described in k132086.
#### Test Strip Stability:
Accelerated and real time stability studies were conducted to assess test strip shelf-life and open vial stability. The study protocols and acceptance criteria were reviewed and found to be adequate to support claims that test strips in vials and individually wrapped in foil pouches can be stored at 41-86 °F (5-30 °C) and 10-90% relative humidity and that test strips in vials are stable for 6 months after first opening when stored at 41-86 °F (5-30 °C) and 10-90% relative humidity.
{6}
Control Solution Stability:
Accelerated and real time stability studies were conducted to assess control solution open and closed vial stability. The study protocols and acceptance criteria were reviewed and found to be adequate to support claims that unopened control solutions have a 24 month shelf life and are stable for 6 months after first opening when stored at 41-86°F (5-30°C).
d. Detection limit:
The reportable range is 20 to 600 mg/dL based on linearity studies above (section M.1.b.).
e. Analytical specificity:
Analytical specificity was established in k132086. The labeling states:
Interference might occur when the values of the limiting concentrations of these compounds are greater than those listed below:
- Ascorbic acid > 3 mg/dL
- Acetaminophen > 20 mg/dL
- Bilirubin > 50 mg/dL
- Uric Acid > 23.5 mg/dL
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device
Performance of the On Call Express Pro Blood Glucose Monitoring System for use by trained professionals testing capillary samples from the fingertip, forearm, and palm was established in k132086.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
{7}
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
**User performance study:**
Performance of the On Call Express Pro Blood Glucose Monitoring System for lay-users self-testing of capillary samples from the fingertip, forearm, and palm was established in k132086.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
In the labeling the sponsor presents expected blood glucose levels for people without diabetes as:
| Time | Range (mg/dL) | Range (mmol/L) |
| --- | --- | --- |
| Fasting and Before Meals | 70-100 | 3.9-5.6 |
| 2 Hours After Meal | Less than 140 | Less than 7.8 |
These ranges were cited from the ADA Standards of Medical Care in Diabetes 2015.
N. Instrument Name:
On Call Express Pro Blood Glucose Meter
O. System Descriptions:
1. Modes of Operation:
Each test strip is single use and requires a sample volume of 0.4 µL.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X ☐ or No ☐
{8}
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes __________ or No ☐ x
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ x or No __________
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
The glucose test is intended to be used with capillary whole blood from the fingertip, forearm, or palm. The whole blood sample is applied directly to the test strip by capillary action.
5. Calibration:
There is no calibration required for the On Call Express blood glucose meter by the user. The meter is automatically coded by calibration information embedded in the test strip.
6. Quality Control:
Glucose control solutions at 3 different concentrations are available to be run with this device. The user marks control solution tests by selecting an option from the menu on the meter. Control solution tests are indicated on the display with a pound (#) sign, this is described in the labeling, and these test results are excluded from calculated average results. Recommendations on when to test the control materials are provided in the labeling. An acceptable range for each control level is printed on the test strip vial label and on the bottom of the test strip box. The user is cautioned not to use the meter if the control result falls outside these ranges.
{9}
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Sample volume study:
A minimal sample volume of 0.4 µL for the On Call Express Pro Blood Glucose Monitoring System and validation of the error code for insufficient sample volume was established in k132086.
2. Altitude Study:
Use of the On Call Express Pro Blood Glucose Monitoring System at up to 8,516 ft above sea level was established in k132086.
3. Hematocrit Study:
Use of the On Call Express Pro Blood Glucose Monitoring System with samples with a hematocrit range of 30–55% was established in k132086.
4. Test System operating conditions:
Use of the On Call Express Pro Blood Glucose Monitoring System at temperatures from 5-45°C and relative humidity conditions ranging from 10--90% was established in k132086.
5. Readability Assessment:
The readability of the labeling (user guides, test strip package insert and control solution package insert) was evaluated using a Flesch-Kincaid analysis and was found to be at the 8th grade level.
6. EMC Testing: As established in k132086.
7. Software documentation:
The software documentation was reviewed and found to be acceptable. The firm provided documentation to support the device was designed, developed and is under good software lifecycle processes.
8. Infection Control Studies:
The device system is intended for multiple-patient use. Disinfection efficacy studies were performed on the materials comprising the meter in two previous submissions (k122110 and k111371) by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, DisCide Ultra Disinfecting Towelettes (EPA Reg. No. 10492-4). Robustness studies were also performed by the sponsor using the On Call Express Pro Blood Glucose Monitoring System and demonstrating that there was no change in performance of the system or external materials of the meter after 10,950 cleaning and disinfection cycles, using DisCide Ultra Wipes Disinfecting Towelettes, to simulate 3 years of
{10}
multiple patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
**Q. Proposed Labeling:**
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
**R. Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.