K151594 · Ellipse Technologies, Incorporated · KCT · Jan 25, 2016 · General Hospital
Device Facts
Record ID
K151594
Device Name
Ellipse Supplemental Instrument Trays
Applicant
Ellipse Technologies, Incorporated
Product Code
KCT · General Hospital
Decision Date
Jan 25, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Ellipse Supplemental Instrument Trays are intended to organize, protect, and transport instruments during steam sterilization and subsequent storage. Sterility of the enclosed instruments is not maintained unless used in conjunction with an FDA-cleared sterilization wrap using the following cycle: Pre-vacuum Steam Sterilization parameters: 132°C for 4 minutes with 40 minutes dry time
Device Story
Ellipse Supplemental Instrument Trays consist of reusable, rigid, perforated anodized aluminum containers designed to organize, protect, and transport surgical instruments during steam sterilization and storage. The system includes a lid, base tray, and an optional insert tray for the reamer model. Instruments are secured using medical-grade Class VI silicone holders and stainless steel hardware. The device is used in clinical settings by healthcare personnel to prepare instruments for sterilization. The trays are placed in an FDA-cleared sterilization wrap and processed via pre-vacuum steam sterilization (132°C for 4 minutes, 40-minute dry time). The device ensures instruments remain organized and protected during the sterilization cycle and subsequent storage, facilitating efficient clinical workflows and maintaining instrument integrity.
Clinical Evidence
Bench testing only. Sterilization validation confirmed a sterility assurance level (SAL) of 10^-6 per ISO 17665-1 and AAMI TIR 12. Dry time validation confirmed effectiveness at 40 minutes. Material integrity and functionality were verified after 100 sterilization cycles. Biocompatibility testing (MEM elution assay on L-939 mouse fibroblasts) confirmed non-cytotoxicity. Cleaning validation (manual and automated) met acceptance criteria for hemoglobin, protein, and TOC levels.
Technological Characteristics
Materials: Anodized aluminum, stainless steel, medical-grade Class VI silicone. Design: Rigid, perforated containers with silicone instrument holders. Sterilization: Pre-vacuum steam (132°C, 4 min). Connectivity: None. Software: None. Dimensions: 510mm x 250mm x 45mm (Instrument Tray) and 510mm x 250mm x 87mm (Reamer Tray).
Indications for Use
Indicated for organizing, protecting, and transporting reusable medical instruments during steam sterilization and subsequent storage. No specific patient population, age, or gender restrictions apply.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
ContainMed, Inc. VersaPod™ Instrument Trays (K071783)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 25, 2016
Ellipse Technologies, Incorporated Ms. Cora Sim Regulatory Affairs Specialist 101 Enterprise, Suite 100 Aliso Vieio. CA 92656
Re: K151594
Trade/Device Name: Ellipse Supplemental Instrument Trays Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization wrap Regulatory Class: II Product Code: KCT Dated: December 28, 2015 Received: December 29, 2015
Dear Ms. Sim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements
{1}------------------------------------------------
as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151594
Device Name Ellipse Supplemental Instrument Trays
#### Indications for Use (Describe)
The Ellipse Supplemental Instrument Trays are intended to organize, protect, and transport instruments during steam sterilization and subsequent storage. Sterility of the enclosed instruments is not maintained unless used in conjunction with an FDA-cleared sterilization wrap using the following cycle:
Pre-vacuum Steam Sterilization parameters: 132°C for 4 minutes with 40 minutes dry time
The the worst case validated sterilization load is:
-One (1) Supplemental Instruments (total weight: 10.0 Ibs.) and one (1) Supplemental Reamer Tray with instruments (total weight: 13.1 lbs.); in a non-stacked configuration.
The worst case loading configuration included the 12 instruments supplied with the Supplemental Instrument Tray with the following worst case lumen dimensions:
Supplemental Instrument Tray (Model SIT1-000)
| Instrument Model | Description | Dimension (Inner Diameter x Length) | Number of Lumens |
|------------------|-----------------------------|-------------------------------------|------------------|
| TCD1-000 | Teardrop Cannulated Driver | 0.210" x 4.58" | 1 |
| CED1-011 | 11mm Cannulated Entry Drill | 0.146" x 10.75" | 1 |
| GWC1-000 | Guide Wire Chuck | 0.250" x 5.0" | 1 |
| LQC1-000 | Large AO Quick Connect | 0.175" x 2.91" | 1 |
| DGA1-000 | Direct AO Depth Gauge | 0.394" x 10.43" | 1 |
The worst case loading configuration included the 18 instruments supplied with the Supplemental Reamer Tray with the following worst case lumen dimensions:
Supplemental Reamer Tray (Model SRT1-000):
| Instrument Model | Description | Dimension (Inner Diameter x Length) | Number of Lumen |
|------------------|-------------------------|-------------------------------------|-----------------|
| RMR1-110 | Flexible Reamer, 10.0mm | 0.125" x 18.4" | 1 |
| RMR1-135 | Flexible Reamer, 13.5mm | 0.125" x 18.4" | 1 |
| RMR1-140 | Flexible Reamer, 14.0mm | 0.125" x 18.4" | 1 |
| RMR1-145 | Flexible Reamer, 14.5mm | 0.125" x 18.4" | 1 |
| RMR1-150 | Flexible Reamer, 15.0mm | 0.125" x 18.4" | 1 |
Do not exceed a maximum load of 12 instruments (10.0 lbs.) in the Supplemental Instrument Tray. Do not exceed a maximum load of 18 instruments (13.1 lbs.) in the Supplemental Reamer Tray.
{3}------------------------------------------------
| List of Devices:<br>Model | Dimensions<br>(Length x Width x Depth) | Number of Instruments: | Validated Weight of tray with<br>instruments: |
|---------------------------|----------------------------------------|------------------------|-----------------------------------------------|
| SIT1-000 | 510mm x 250mm x 45mm | 12 | 10.0 lbs. |
| SRT1-000 | 510mm x 250mm x 87mm | 18 | 13.1 lbs.* |
*Includes insert tray with dimensions 497mm x 240mm x 52mm
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (pursuant to 21 CFR 807.92 (a)(1))
### Ellipse Supplemental Instrument Trays Premarket Notification Number: K151594 January 22, 2016
| 1. | Company: | Ellipse Technologies, Incorporated<br>101 Enterprise, Suite 100<br>Aliso Viejo, CA 92656<br>USA |
|----|-------------------------|----------------------------------------------------------------------------------------------------|
| | Contact: | Cora Sim<br>Regulatory Affairs Specialist<br>Phone: (949) 837-3600 ext. 221<br>Fax: (949) 837-3664 |
| 2. | Proprietary Trade Name: | Ellipse Supplemental Instrument Trays |
| 3. | Classification Name: | Sterilization Wrap Containers, Trays, Cassettes & Other<br>Accessories (21 CFR 880.6850) |
| 4. | Product Code: | KCT (Sterilization wrap) |
| 5. | Device Classification: | Class II |
6. Predicate Device: The Ellipse Supplemental Instrument Trays are substantially equivalent to the following 510(k) cleared device:
- ContainMed, Inc. VersaPod™ Instrument Trays (K071783)
7. Product Description: The Ellipse Supplemental Instrument Trays are comprised of a Supplemental Instrument Tray and a Supplemental Reamer Tray. These trays are designed to organize, protect, and transport instruments during steam sterilization and subsequent storage. The trays are composed of a lid and base tray with handles and are perforated for steam The Supplemental Instrument Tray has the following approximate external penetration. dimensions: 510 mm (length) x 250 mm (width) x 45 mm (depth), and is designed to hold 12 instruments. The Supplemental Reamer Tray has the following external dimensions: 510 mm (length) x 250 mm (width) x 87 mm (depth), and is designed to hold 18 flexible reamer
{5}------------------------------------------------
instruments. An insert tray with the following dimensions is used with the Supplemental Reamer Tray: 497 mm (length) x 240 mm (width) x 52 mm (depth). The Supplemental Instrument Trays are reusable and are used to store instruments prior to, during, and after the sterilization process. The trays are made of anodized aluminum and have medical grade, Class VI Silicone instrument holders with stainless steel mounting hardware which are used to secure, separate, and organize the instruments within the trays. The tray handles are also encased in medical grade, Class VI Silicone.
8. Indications for Use: The Ellipse Supplemental Instrument Trays are intended to organize, protect, and transport instruments during steam sterilization and subsequent storage. Sterility of the enclosed instruments is not maintained unless used in conjunction with an FDAcleared sterilization wrap using the following cycle:
Pre-vacuum Steam Sterilization parameters: 132℃ for 4 minutes with 40 minutes dry time
The worst case validated sterilization load is:
- -One (1) Supplemental Instrument Tray (total weight: 10.0 lbs.) and one (1) Supplemental Reamer Tray (total weight: 13.1 lbs.); in a non-stacked configuration
The worst case loading configuration included the 12 instruments supplied with the Supplemental Instrument Tray with the following worst case lumen dimensions:
| Instrument<br>Model | Description | Dimensions<br>(Inner Diameter x Length) | Number<br>of<br>Lumens |
|---------------------|-----------------------------|-----------------------------------------|------------------------|
| TCD1-000 | Teardrop Cannulated Driver | 0.210" x 4.58" | 1 |
| CED1-011 | 11mm Cannulated Entry Drill | 0.146" x 10.75" | 1 |
| GWC1-000 | Guide Wire Chuck | 0.250" x 5.0" | 1 |
| LQC1-000 | Large AO Quick Connect | 0.175" x 2.91" | 1 |
| DGA1-000 | Direct AO Depth Gauge | 0.394" x 10.43" | 1 |
### Supplemental Instrument Tray (Model SIT1-000)
{6}------------------------------------------------
The worst case loading configurations included the 18 instruments supplied with the Supplemental Reamer Tray with the following worst case lumen dimensions:
| Instrument<br>Model | Description | Dimensions<br>(Inner Diameter x Length) | Number<br>of<br>Lumens |
|---------------------|-------------------------|-----------------------------------------|------------------------|
| RMR1-110 | Flexible Reamer, 10.0mm | 0.125" x 18.4" | 1 |
| RMR1-135 | Flexible Reamer, 13.5mm | 0.125" x 18.4" | 1 |
| RMR1-140 | Flexible Reamer, 14.0mm | 0.125" x 18.4" | 1 |
| RMR1-145 | Flexible Reamer, 14.5mm | 0.125" x 18.4" | 1 |
| RMR1-150 | Flexible Reamer, 15.0mm | 0.125" x 18.4" | 1 |
#### Supplemental Reamer Tray (Model SRT1-000):
Do not exceed a maximum load of 12 instruments (10.0 lbs.) in the Supplemental Instrument Tray.
Do not exceed a maximum load of 18 instruments (13.1 lbs.) in the Supplemental Reamer Tray.
List of Devices:
| Model | Dimensions (Length x Width x Depth) | Number of<br>Instruments | Validated Weight of<br>tray with<br>instruments |
|----------|-------------------------------------|--------------------------|-------------------------------------------------|
| SIT1-000 | 510mm x 250mm x 45mm | 12 | 10.0 lbs. |
| SRT1-000 | 510mm x 250mm x 87mm | 18 | 13.1 lbs.* |
*Includes insert tray with dimensions 497mm x 240mm x 52mm.
9. Technological Characteristics: Substantial equivalence of the Ellipse Supplemental Instrument Trays to the predicate is based on similar intended use, design, principles of operation, and materials of composition. The Ellipse Supplement Trays and the predicate device have a similar intended use, the same design, and the same principles of operation. Specifically, the Ellipse Supplemental Instrument Trays and the predicate device are both intended to organize, protect, and transport instruments during steam sterilization and subsequent storage. The Ellipse Supplemental Instrument Trays have the same design features as that of the predicate. Both the Ellipse Supplemental Instrument Trays and the predicate are reusable, rigid
{7}------------------------------------------------
sterilization containers. Both the Ellipse Supplemental Instrument Trays and predicate device have similar dimensions and are made of the same materials.
| Manufacturer | Predicate ContainMed | Ellipse Technologies, Inc. |
|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K071783 | K151594 |
| Product Code | KCT | KCT |
| Intended Use | Perforated instrument trays to organize, protect, and transport medical instruments during pre-vacuum steam sterilization and subsequent storage. | The Ellipse Supplemental Instrument trays are designed to store various reusable instruments in order to organize, protect, and transport instruments during steam sterilization and subsequent storage. |
| Design | Anodized aluminum tray with securing lid and base tray with handles. An optional anodized aluminum insert tray is available for use with the 82mm deep base only. Anodized aluminum placard, and silicone rubber mat and instrument holding accessories (with stainless steel mounting hardware), are available to identify, organize, and protect the contents during use. | The lid, base tray, and insert tray are composed of anodized aluminum. An internal individual insert tray is utilized by the Supplemental Reamer Tray. Medical grade Class VI Silicone instrument holders and with stainless steel mounting hardware are used to secure, separate, and organize the instruments within the trays. |
| Dimensions | Base tray: 510mm (length) x 250mm (width) x 40mm, 60mm, or 82mm (depth) | Supplemental Instrument Tray:<br>Base tray: 510mm (length) x 250mm (width) x 45mm (depth)<br>Supplemental Reamer Tray:<br>Base tray: 510mm (length) x 250mm (width) x 87mm (depth)<br>Insert Tray:<br>497 mm (length) x 240 mm (width) x 52 mm (depth).* |
| Material | Anodized aluminum, Stainless Steel,<br>Silicone | Anodized aluminum, Stainless Steel,<br>Silicone |
| composition<br>Physical Properties | Evenly distributed perforated hole pattern. | Evenly distributed perforated hole pattern. |
| Sterilization<br>Method | Steam | Steam |
| Sterilant<br>Penetration | Sterilant (steam) penetration through perforations in tray | Sterilant (steam) penetration through perforations in tray. |
| Sterilization cycle | Pre-vacuum | Pre-vacuum |
| Exposure<br>Temperature | 132° C | 132° C |
| Exposure time | 4 minutes | 4 minutes |
| Dry Time | 45 minutes | 40 minutes |
| Air permeance<br>Diameter of each<br>perforation | Yes<br>0.197 inches | Yes<br>0.197 inches |
| Number of<br>perforations | Tray Lid: 916<br>Base Tray: 1,044<br>Insert Tray: 1,044 | The Supplemental Instrument Tray:<br>Tray Lid: 916<br>Base Tray: 871<br>The Supplemental Reamer Tray:<br>Tray Lid: 916<br>Base Tray: 956<br>Insert Tray: 948 |
| Vent to volume<br>ratios | 0.045 in -1 | Supplemental Instrument Tray:<br>0.192 in-1<br>Supplemental Reamer Tray:<br>0.147 in-1 |
| Material<br>compatibility with<br>sterilization process | Materials of construction are<br>compatible with steam sterilization. | The materials used in the Ellipse<br>Supplemental Instrument Trays were<br>exposed to 100 sterilization cycles and<br>no material degradation was observed.<br>All materials maintained their integrity<br>and remained fully functional.<br>Performance testing demonstrated that<br>the materials of construction are<br>compatible with steam sterilization. |
| Toxicological<br>properties | Materials of construction are<br>biocompatible. | Extracts of test article (polymeric<br>component) on L-939 mouse fibroblast<br>cells in MEM elution assay did not<br>display a cytotoxic response and is<br>considered non-cytotoxic.<br>Testing demonstrated that the materials<br>of construction are biocompatible. |
| Results of<br>transportation<br>studies | Not available. | Shipping validation was performed on<br>the Ellipse Supplemental Instrument<br>Trays. The package performance testing<br>results demonstrated all acceptance<br>criteria were met. |
| Summary of<br>cleaning validation<br>information | Manual cleaning with use of neutral<br>pH (pH 6.0-8.0) enzymatic detergent<br>recommended. | The Ellipse Supplemental Instrument<br>Trays have been validated to be cleaned<br>using both a manual and automated<br>process.<br>Hemoglobin, Micro BCA protein, and<br>TOC test results met all acceptance<br>criteria. |
| Limitations of<br>reprocessing | The materials used in the<br>ContainMed VersaPod™ Instrument<br>Trays may be sterilized for an<br>indefinite number of cycles. The life<br>of the system is limited only by<br>irreparable physical damage from<br>mishandling. The function and physical construction of the Supplemental Instrument Trays and their components are safe and effective after exposure to 100 sterilization cycles. The end of tray life is determined through the inspection of each tray after the required cleaning and sterilization cycles. The life of the trays are limited only by irreparable physical damage from mishandling. | |
| Summary of drying<br>time validation | Dry time of 45 minutes was validated<br>for use in the pre-vacuum steam<br>cycle of 132°C for 4 minutes. | The dry time validation was successful<br>in determining an effective dry time of<br>40 minutes for the validated pre-vacuum<br>steam sterilization cycle. Acceptance<br>criteria below were met during<br>validation of the dry time:<br><br>1) Autoclave data showing appropriate<br>dwell time at 132°C (+3°C) for each full<br>cycle<br><br>2) Confirmation of the absence of<br>moisture on the trays and sterilization<br>wrap<br><br>3) No more than 3% difference in<br>weight prior to and after sterilization |
### Device Comparison Table for Substantial Equivalence:
{8}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{9}------------------------------------------------
*For use with the Supplemental Reamer Tray
10. Performance Testing: Non-clinical performance testing was conducted to establish substantial equivalence of the Ellipse Supplemental Instrument Trays to the predicate device. Sterilization validation testing was conducted to confirm that a sterility assurance level (SAL) of 10° was achieved at the validated sterilization parameter in the pre-vacuum steam sterilization cycle according to ISO 17665-1 and AAMI TIR 12. Dry time validation was performed following full cycles. Performance testing was conducted to demonstrate that the function and physical construction of the Ellipse Supplement Trays and its components are maintained after exposure to 100 sterilization cycles.
Conclusion: Based on the indications for use, technological characteristics, non-clinical 11. performance data, and a comparison to the predicate device, the Ellipse Supplemental Instrument Trays have been shown to be substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.